Professional Ultrasonic Nebulizer NE-U17 For Hospital & Clinic

for Effective nebulization respirology therapy for Asthma, Bronchopulmonary and other Obstructive respiratory disorders.

Explanation of LCD display

Medication cup cover

Display of air volume Display of nebulation rate When the timer is used For continuous operation “ERROR” display.

Features

Remove the medication cup cover by roating it counter clockwise and than attach the inhalation hose and the mouthpiece set.

Medication cup holder Fill medication into the medication cup. Max. of 150 ml, Min. of 5 ml. Assemble the medication cup cover by turning it clockwise, the install it to main unit.

High Volume Ultrasonic for humidication or medication therapy Particle size 4,4 m Variable output up to 4 ml/min Very quiet operation Backlit display shows settings and timer Wide variety of accessories (optional) AC operation

created by joko sutantiyo [email protected] www.nautiza.com

About the Principle of Nebulization in the NE-U17

The NE-U17 transmits the energy of ultrasonic vibration from the vibrator located at the bottom of the water tank into water. Air

The ultrasonic vibration is transmitted to the medication in the medication cup through the water in the water tank. The ultrasonic vibration emits medication from the medication cup (like a fountain) and this dispersed as aerosol.

Medication Water

Air from the fan carries this nebulised medication out of the medication cup.

Vibrator

mask inhalation

Ultrasonic

Nebulizer

NE-U1 7

O

O

mouthpiece

N

FF

You can use an optimal inhalation mask instead of a mouthpiece

Technical Data Product name Model Power source Power consumption Fuse Ultrasonic frequency Nebulization rate Particle size

: : : : : : : :

Air volume Amount of cooling water Capacity of medication cup External dimensions Weight of the main unit Operating Conditions Accessories included

: : : : : : :

Do not bend the inhalation hose or block the nozzle

Do not use if the main unit higher than you head

The medication will be pooled in the hose and the nebulization rate may be lowered.

The medication may splash on your face or clothes.

Omron Ultrasonic Nebulizer NE-U17 230 VAC, 50Hz Approximately 80 VA T2A L250V Approximately 1.7MHz Approx. 0 to 3 ml adjustable 4.4 m MMAD, base on cascade impaction particle size measurement method (according to EN 13544-1 European nebulizer standard). Maximum of 17L/min Approximately 375 ml Approximately 150 ml (min. 5ml) Approximately 276 (W) x 243 (H) x 226 (D) Approximately 4.0 Kg 5 C to 40 C (42 F to 73,4 F), 30 to 85% RH Inhalation hose m (70 with a cuff), mouthpiece set, 2 medication cups, power cord, instruction manual (with warranty card)

*) May change depending on the type of medication Graha Mas Pemuda Blok AB 19 Jl. Pemuda, Jakarta Timur - 13220 Telp. (021) 4788 2599 (hunting), Fax. (021) 4788 2598 E-mail : [email protected] Website : www.indocore.co.id created by joko sutantiyo [email protected] www.nautiza.com

Distributed by:

IDC/JULY/2008

OMRON SERVICE CENTER

Related Documents

Eng
May 2020 59
Eng
May 2020 56
Eng
April 2020 50
Eng
November 2019 69
Eng
June 2020 32
Eng
November 2019 50

More Documents from ""

Hbf-306 Eng
April 2020 10
Ne-u17 Eng
May 2020 20
Sem-2 Eng
May 2020 12
Ia1 Eng
April 2020 10
Pedometer Hj-113 Eng
May 2020 21