Section 5 I&c Chap.10 H2s Gas Detector And Monitoring System (rev.1).pdf

  • Uploaded by: Richard Alexander
  • 0
  • 0
  • October 2019
  • PDF

This document was uploaded by user and they confirmed that they have the permission to share it. If you are author or own the copyright of this book, please report to us by using this DMCA report form. Report DMCA


Overview

Download & View Section 5 I&c Chap.10 H2s Gas Detector And Monitoring System (rev.1).pdf as PDF for free.

More details

  • Words: 51,821
  • Pages: 180
Human Technology & Future

INSTRUCTION MANUAL MODEL : GTC-200A

GTC200A POWER

GTC200A COMMON

GTC200A CHANNEL

VOLT

GTC200A CHANNEL

%LEL 100

GTC200A CHANNEL

%LEL 100

90

%LEL 100

90

90

MAIN POWER 80

80

70

70

60

BATT POWER

BATT GOOD

COM-PC

BZ. STOP

GTC200A

70

60

60

COM-CH

BATT TROUBLE

BATT TEST

80

BZ. STOP RESET

GTC200A

50

50

50

40

40

40

30

30

30

20

20

20

10

10

10

0

0

TEST

RESET

GTC200A

0

TEST

RESET

GTC200A

TEST

RESET

GTC200A

GTC200A POWER

GTC200A COMMON

GTC200A CHANNEL

VOLT

%LEL 100 90

MAIN POWER 80 70 60

BATT POWER

COM-CH 50 40

BATT GOOD

COM-PC

30 20 10

BZ. STOP

BATT TROUBLE

BATT TEST

GTC200A

BZ. STOP RESET

GTC200A

0

TEST

RESET

GTC200A

For proper use, please read this manual carefully!

GTC-200A

We much appreciate your wise purchase of GASTRON products Gastron, a world-renowned company for manufacturing gas detectors and gas monitoring devices exclusively, has enjoyed a great reputation from numerous consumers and customers in and out of Korea for its superb quality and convenience in use. We, at Gastron, have always strived and will further endeavor to let our consumers gain an easy access to our products at all times. Furthermore, we are proud to let you know that Gastron has exerted ourselves to ceaselessly develop and research for gas detectors catering to customers' satisfaction. Our customers are now invited to solve all the problems and dissatisfaction related to gas detectors by using our Gastron products. Our customers are warranted to receive contentment with our Gastron gas detectors at all times and at any place. The following operating manual serves to explain an installation method, an operating method, a simple maintenance method and other useful guidelines regarding the toxic gas leakage detector "GTC-200A" . We trust that this manual will be of good use whenever there are any questions or inquiries during usage of the detector. You are encouraged to carefully read and make good use of the operating manual by safekeeping and referring to the same at your every need. As always, we once again appreciate your patronage of our products and should you have any inconveniences or problems on our products, kindly contact us at the following address during use to receive prompt attention and necessary measures.

Note • You are kindly requested to get your gas detector inspected and calibrated at least once within a three-month period using calibration gas according to the type of gas in order to ensure an exact operation of the gas alarm, and to receive a periodic inspection including one or more calibration inspections for each six-month period under the provision of relevant Industrial Safety and Sanitary Act promulgated by the Korea Labor Ministry. • No periodic inspections and calibrations may be a cause of erroneous operation of the gas detector due to aged sensor unit. • The disintegration or disassembly of the gas detector needs a person who has a technical expertise on the gas alarm. • You are kindly invited to call our technical department or write to our e-mail address or web site for inspection and calibration of gas detectors or alarms.

2

GTC-200A

Contents 1. Introduction - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4

2. Features - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -

4

3. Specifications - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 5

4. Part’s name and major functions - - - - - - - - - - - - - - - - - 7

5. Terminal wiring diagram - - - - - - - - - - - - - - - - - - - - - - 14

6. How to use - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 16

7. External Drawings and Dimensions - - - - - - - - - - - - - - - 26

3

GTC-200A

1. Introduction GTC-200A Series has a high performance A/D converter and micro-processor so that it has various functions built-in. GTC-200A Series is centralized, that is, composed of one common alarm unit and several multi-channel control units. Each multi-channel control unit is connected to its sensing part. GTC-200A is encapsulated in DIN type case. There are two product types - Panel mount type and Wall mount type. GTC-200A has FND digital display (PV value) function and 3 colors LED bar graphic display (PV & alarm set value) function, it also has three instant alarm (1 st H/L, 2nd H/L, and 3rd H/L) and trouble alarm.

2. Features GTC-200A Series produces audible (buzzer) and visual (alarm LED and bar graphic LED flashing) warnings in case of instant alarm and trouble alarm. If an alarm occurs, this product can hold a Max. PV value holding. GTC-200A series enables remote control for alarm cancellation. Our product can carry out synchronous control functions because it has an output signal of the alarm (SPDT contact). GTC-200A series offers two outputs; Isolation type RS-485 signal (Option) is provided for a monitoring system from common unit, the other is a measurement output (4-20mA, DC) from channel control unit. Since GTC-200A series is composed of the state of art parts, it has stability and credibility. It can be expanded as large as its maximum (up to 64 channels)

4

GTC-200A

3. Specifications 3-1. Power Unit (Option) Power Unit 1

Input power

DC. 24V

2

Output power

DC. 24V (Aux. power : DC. 27V / 300mA)

3

Output power display

FND Digital Display

4

Main power display

Green LED

5

Aux. power display

Green LED

6

Aux. power monitoring

Red LED

7

Aux. power test

Aux. power test S/W

8

Aux. power

Ni-Cd Battery 24V / 600mA(under 6 channels)

3-2. Common Unit Common Alarm Unit 1

Input type

RS-485

2

Input check interval

100ms

3

Alarm type

Audible(Buzzer) & Visible (LED)

4

Alarm reset

Manual (Reset Switch)

5

Control input

Remote Reset– Buzzer stop/reset

6

Measuring output

Isolation RS-485 (Option) Buzzer - STDT Relay Dry contact Signal Output Fault Alarm- STDT Relay Dry contact Signal Output

7

Alarm output

1st Alarm - STDT Relay Dry contact Signal Output 2nd Alarm - STDT Relay Dry contact Signal Output 3rd Alarm - STDT Relay Dry contact Signal Output * Relay Dry contact capacity : AC125V 10A

8

Operating Power

DC 24V

5

GTC-200A

3-3. Channel Unit Channel Control Unit 1

Input type

4-20mA.DC / Full Scale

2

Measurement display

0.000 to 9999 Digital (settable by user)

3

Measurement Accuracy

FND Digital –bigger one between ±1% Full Scale or 1 Digit LED Bar --- bigger one between ±1% Full Scale or 1 Digit

4

Input check interval

100 ms

5

Alarm setting

3 stage Alarm (settable by user)

6

Alarm setting display

3 Color Bar Graphic (Green/Red/Yellow LED)

7

Alarm display

LED Bar Graphic

8

Alarm reset

Manual (Common Unit)

9

Self diagnosis

Test Switch & Reset Switch

10 Control In/out

RS-485

11

4-20mA.DC / Full Scale

Measurement Output

Fault Alarm - STDT Relay Dry contact Signal Output Alarm 1 STDT Relay Dry contact Signal Output 12 Alarm Output

Alarm 2 STDT Relay Dry contact Signal Output Alarm 3 STDT Relay Dry contact Signal Output * Relay Dry contact capacity : AC125V 10A

13 Operating Power

DC 24V

3-4. Wall Mount Type Wall Mount Type 1

2

Input Power

Applicable SMPS capacity by Channel quantities

AC 110V/220V 50/60Hz (default: AC220V 50/60Hz) Channel

Capacity

3(4) Channel

24V 1.5A (24V 1.5A)

5(6) Channel

24V 2.1A (24V 2.1A)

7(8) Channel

24V 2.1A (24V 3.5A)

9(10) Channel

24V 3.5A (24V 3.5A)

**( ): quantities and capacity when there is no Power Unit

6

GTC-200A

4. Part’s names and major functions 4-1. Configuration of Power unit 1

6

1

5

2

14

15 GTC200A POWER

VOLT

7

3

MAIN POWER

16

8 BATT POWER

9 BATT GOOD

10 BATT TROUBLE

BATT TEST

17 11 4

ON OFF

POWER

18

12 13

No. 1 2 3 4 5 6 7 8 9

DESCRIPTION FRONT COVER CASE MAIN BODY CASE ACRYLIC FRONT SUB COVER MAIN BODY FIXED BRACKET MAIN/BAT. POWER DISPLAY MAIN POWER LED BAT. POWER LED BAT. POWER GOOD LED

No. 10 11 12 13 14 15 16 17 18

DESCRIPTION BAT. POWER TROUBLE LED BAT. POWER TEST KEY BAT. POWER ON/OFF SWITCH FRONT COVER SCREW TERMINAL PCB TERMINAL PCB SCREW DC INPUT CONNECTOR BATTERY CONNECTOT DC OUTPUT CONNECTOR

4-1-1. MAIN/BAT. POWER DISPLAY Display voltage of main and Aux. power. This is also one for Common Unit and Channel Unit.

4-1-2. MAIN POWER LED This LED is turned on when using AC Power, and turned off in case of using Battery

4-1-3. BAT. POWER LED This LED is turned on when using battery power

7

GTC-200A

4-1-4. BAT. POWER GOOD LED If battery voltage is above 18V when connecting with battery, this LED is turned on.

4-1-5. BAT. POWER TROUBLE LED If battery voltage is below 18V when connecting with battery, this LED is turned on. This LED is flickering if there is no connection with Battery.

4-1-6. BAT. POWER TEST KEY This S/W is for testing operation of Battery. When pressing this S/W, battery power is applied and operates gas detection system. Also voltage of battery is displayed on FND.

4-1-7. BAT POWER ON/OFF SWITCH This is battery power on/off S/W.

Note)When the products is released the batter switch should be off. After turning on main power the battery power should be turned on.

8

GTC-200A

4-2. Configuration of Common unit 1

1

6

5

2

18

19

18

19

3 7 12

20

13

NC NO ALARM3

8

COM-CH

9

COM

COM-PC

10

NC NO ALARM2

14 11 15

COM

BZ. STOP RESET

NC NO ALARM1

COM

BZ. STOP

COM NC NO TROUBLE

B A BUZZER

B A RESET

GTC200A COMMON

4 ON OFF

POWER

16 17

Panel Mount Type

No. 1 2 3 4 5 6 7 8 9 10

DESCRIPTION FRONT COVER CASE MAIN BODY CASE ACRYLIC FRONT SUB COVER MAIN BODY FIXED BRACKET BUZZER POWER LED TROUBLE LED ALARM-3 LED ALARM-2 LED

No. 11 12 13 14 15 16 17 18 19 20

Wall Mount Type

DESCRIPTION ALARM-1 LED COMMUNICATION LED (CHANNEL) COMMUNICATION LED (PC) BUZZER STOP LED BUZZER STOP/RESET KEY POWER ON/OFF SWITCH FRONT COVER SCREW TERMINAL PCB TERMINAL PCB SCREW CONNECTOR

4-2-1. Buzzer Buzzer sounds when trouble or alarm occurs from each channel unit.

4-2-2. Power LED When power is applied from Common Unit, power LED turns on.

4-2-3. Trouble LED If trouble occurs from each channel unit, Trouble LED turns on. Ex) In case of bad wiring with gas detector or other trouble

9

GTC-200A

4-2-4. 1st Alarm LED (Alarm 1 LED) Alarm 1 LED is on when first alarm occurs on each channel unit. If the value reaches at 1st alarm value during test of each channel unit, Alarm 1 LED is on.

4-2-5. 2nd Alarm LED (Alarm 2 LED) Alarm 2 LED is on when second alarm occurs on each channel unit. If the value reaches at 2nd alarm value during test of each channel unit, Alarm 2 LED is on.

4-3-6. 3rd Alarm LED (Alarm 3 LED) Alarm 3 LED is on when third alarm occurs on each channel unit. If the value reaches at 3rd alarm value during test of each channel unit, Alarm 3 LED is on.

4-2-7. Channel communication LED (COM-CH LED) If Common Unit normally communicates with whole Channel Units set on common unit, this COM-CH LED is turning on continuously. If there is some channel units having no communication with common unit, this COM-CH LED is flickering. (If number of channel unit is set as under 1 on Common Unit, the LED is flickering with 0.5 second interval)

4-2-8. PC communication LED (COM-PC LED) (Option) If Common Unit normally receives communication data from PC, this COM-PC LED is off after turned on once. (If number of common unit address is set as under 1 on Common Unit, the LED is flickering with 0.5 second interval.)

4-2-9. Buzzer Stop LED (BZ-STOP LED) Buzzer stop s/w is used for stopping buzzer sound in case of alarm at each channel unit and buzzer stop LED is also on. In case of pushing buzzer stop s/w again on common or channel unit, buzzer stop LED is off. (This LED is activated only when set alarm type as “Hold” on Channel Unit)

4-2-10. Buzzer Stop & Reset Key This key is used for stop & reset of buzzer on alarm at each Channel Unit. - Push once .... Buzzer sound stops and BZ-STOP LED is on. - Push twice .... BZ-STOP LED is off and performs Reset function.

4-2-11. Main Power ON / OFF Switch This is Power On/Off switch of Common Unit.

10

GTC-200A

4-3. Configuration of Channel unit 1

6

1

5

21

2

22

21

22

+ -

%LEL 100

전원 80

8

70

COM

10

NC NO ALARM2

경보3 30

11

20

시험

복귀

14 15

기능

18 ON OFF

NC NO ALARM1

12

COM

경보1

mA + SENSOR

0

COM

경보2

10

17

23

9 고장

50 40

16

NC NO ALARM3

점검

60

전원

-

13

COM NC NO TROUBLE

7

90

3

OUTPUT 4-20mA

GTC210A CHANNEL

4 19 20

Panel Mount Type

No. 1 2 3 4 5 6 7 8 9 10 11 12

DESCRIPTION FRONT COVER CASE MAIN BODY CASE ACRYLIC FRONT SUB COVER MAIN BODY FIXED BRACKET FND DISPLAY POWER LED STANDBY LED TROUBLE LED ALARM-3 LED ALARM-2 LED ALARM-1 LED

No. 13 14 15 16 17 18 19 20 21 22 23

Wall Mount Type

DESCRIPTION 3 COLOR BAR GRAPHIC LED RESET KEY DOWN KEY TEST KEY FUNCTION KEY UP KEY POWER ON/OFF SWITCH FRONT COVER SCREW TERMINAL PCB TERMINAL PCB SCREW CONNECTOR

4-3-1. Measurement density display (FND Digital Display) Continuously displays measurement value of corresponding detector connected with each channel unit, and displays user setting value with flickering when doing Test function.

4-3-2. Power LED This Power LED is on when power is applied to channel unit.

4-3-3. Stand-by LED This STD-BY LED is flickering when checking gas detector.

4-3-4. Trouble LED This trouble LED is flickering in case of trouble on Channel Unit and/or gas detector. Ex) * bad connection with gas detector or other trouble 11

GTC-200A

4-3-5. 1st Alarm LED (Alarm 1 LED) Alarm 1 LED is on when first alarm occurs on channel unit. If the value reaches at1st alarm value during test of channel unit, Alarm 1 LED is on

4-3-6. 2nd Alarm LED (Alarm 2 LED) Alarm 2 LED is on when second alarm occurs on channel unit. If the value reaches at 2nd alarm value during test of channel unit, Alarm 2 LED is on

4-3-7. 3rd Alarm LED (Alarm 3 LED) Alarm 3 LED is on when third alarm occurs on channel unit. If the value reaches at 3rd alarm value during test of channel unit, Alarm 3 LED is on.

4-3-8. 3 Color Bar Graphic LED 3 Color Bar Graphic LED displays the measured value and alarm setting value continuously like FND display - When the measured value is lower than 1st alarm, the LED color is green. - When the measured value is lower than 2nd alarm, the LED color is orange - When the measured value is lower than 3rd alarm, the LED color is red. If measured value is higher than alarm setting value, bar graphic is being held at the maximum value and is flickering with the alarm

4-3-9. Test Key If pressing Test Switch, channel unit goes into self- diagnosis mode. - Measurement display FND is flickering and user can check alarm function on channel unit and common unit by changing flickering measurement value with ▲ or ▼ key - By pressing reset key on each channel unit once, self diagnosis mode is cancelled.

4-3-10. Reset Key This key is used for canceling of alarm, self test and/or program setting of Channel Unit.

4-3-11. Func. Key Function Key is used for switching or selecting mode when setting of alarm value, alarm type, alarm dead band and/or alarm delay time.

4-3-12. ▲ ▼ Key (UP/DOWN KEY) After selecting each mode by Func Key, desired value will be chosen by “▲”, “▼” Key. Furthermore, the value change will be accelerated by keeping “▲”, “▼” Key press if user need to change a lot of numbers at one time.

12

GTC-200A

4-4. Configuration of Wall Mount Type panel 4

5

GTC200A POWER

6

GTC200A COMMON

10

GTC200A CHANNEL

VOLT

GTC200A CHANNEL

%LEL 100

1

90

2

13

14

GTC200A CHANNEL

%LEL 100

11

%LEL 100

90

90

MAIN POWER 80

80

70

70

60

BATT POWER

BATT GOOD

50

No. 1 2 3 4 5 6 7 8 9

BZ. STOP RESET

GTC200A

8

50

40

40

30

30

20

7

60

50

40 30

20

10

BZ. STOP

GTC200A

70

60

COM-CH

COM-PC

BATT TROUBLE

BATT TEST

80

20

10

0

10

0

TEST

RESET

GTC200A

FRONT

0

TEST

RESET

GTC200A

9

TEST

RESET

GTC200A

3

12

DESCRIPTION

SIDE

15

No.

WALL MOUNT CASE FRONT WALL MOUNT CASE REAR WALL MOUNT CASE COVER POWER UNIT (Option) COMMON UNIT CHANNEL UNIT MOTHER BOARD FILTER PCB (Option) MOTHER BOARD COMMON PCB MOTHER BOARD CHANNEL PCB

16

Inner

17 18

DESCRIPTION

10 11 12 13 14 15

NAME PLATE DIN CASE FIXED BRACKET CASE COVER SCREW FRONT CASE FIXED HOOK WALL MOUNT FIXED HOLE MCCB (Molded Case Circuit breaket)

16

POWER SUPPLY

17 18

BATTERY (Option) BATTERY COVER (Option)

4-4-1. Power Unit (Option) This unit is only for Battery and supplied as Option item.

4-4-2. Mother Board Filter PCB (Option) This PCB is supplied when using Power Unit and protects circuit from electromagnetic wave, Noise and Surge.

4-4-3. Mother Board Common PCB Provides relay contact output terminals for buzzer, trouble and alarms.

4-4-4. Mother Board Channel PCB Provides terminals for relay contact output, 4~20mA output and connection with gas detector.

13

GTC-200A

5. Terminal wiring diagram 5-1. POWER CONNECTION 5-1-1. Panel Mount Type Channel Unit-#64

Channel Unit-…

Channel Unit-#2

Channel Unit-#1

Common Unit

Power Unit

Back-up battery (battery connection)

0V

24V

AC 110/220V

*. Power Unit is supplied only when using Battery. In case of no Power Unit, DC power (24V) is directly connected from Power Supply to connector (CN6) of common unit. (Power Unit is an option item.)

5-1-2. Wall Mount Type

GTC200A POWER

Wall Mount Front Case fixing ring

GTC200A COMMON

GTC200A CHANNEL

%LEL 100

90

%LEL 100

90

90

MAIN POWER 80

80

70

80

70

60

70

60

60

COM-CH

BATT GOOD

COM-PC

BZ. STOP

BATT TROUBLE

BATT TEST

GTC200A

14

GTC200A CHANNEL

%LEL 100

BATT POWER

Release the fixing ring and open front case. And then connect electrical power to NFB, located at left side in case Operating voltage is AC110V or 220V and user can select voltage on order. (Default voltage is AC220V)

GTC200A CHANNEL

VOLT

BZ. STOP RESET

GTC200A

50

50

50

40

40

40

30

30

30

20

20

20

10

10

10

0

0

TEST

RESET

GTC200A

0

TEST

RESET

GTC200A

TEST

RESET

GTC200A

GTC-200A

5-2. Gas Detector connection diagram (3Wire type) ( Gas detector ) TS-1100Ex TS-2000Ex TS-2100 Series TS-4000 Series TS-4100 Series TS-5100 Series GTD-1000 Series GTD-2000 Series GTD-3000 Series GIR-3000 Series

1

+

2

-

3

NO

4

NC

OUTPUT 4-20mA

TROUBLE (고장)

5 6

NO

7

NC

ALARM3 ( 경보3 )

8 9

NO

10

NC

ALARM2 ( 경보2 )

11 12

NO

13

NC

ALARM1 ( 경보1 )

14 15

+

16

mA

17

-

1

+

2

-

3

NO

4

NC

SENSOR (탐지기)

SHILD EARTH

Note) Please use CVVS or CVVSB 1.5sq↑ Shield Cable!

5-3. Gas Detector connection diagram (2Wire type) OUTPUT 4-20mA

TROUBLE (고장)

5

( Gas detector ) TS-1100Tx TS-2000Tx TS-3000 Series TS-3100 Series

6

NO

7

NC

ALARM3 ( 경보3 )

8 9

NO

10

NC

ALARM2 ( 경보2 )

11 12

NO

13

NC

ALARM1 ( 경보1 )

14

SHILD EARTH

Note) Please use CVVS or CVVSB 1.5sq↑ Shield Cable!

15

15

+

16

mA

17

-

SENSOR (탐지기)

GTC-200A

6. How to use 6-1. How to operate Power Unit -. Voltage of main power is displayed on FND Digital Display. -. If battery test key was pressed, the power is changed to Aux. GTC200A POWER

power (battery) and voltage of battery is displayed on FND VOLT

MAIN POWER

Digital Display. -. If a back-up battery is not connected, battery trouble LED is blinking (0.5 second interval).

BATT POWER

BATT GOOD

-. If a back-up battery voltage is above 18V,the back-up battery power is normal LED is on. Otherwise, battery trouble LED

BATT TROUBLE

is on.

BATT TEST

Note) this unit is off when shipment. ON OFF

POWER

Please turn on Aux. power switch after turned on main power. Back-up battery switch

6-2. How to operate Common Unit 6-2-1. Buzzer alarm -. gives short intermittent sound when receiving trouble signal from Channel Unit -. gives long intermittent sound when receiving alarm signal from Channel Unit.

6-2-2. Trouble/alarm LED -.These LEDs are turned on when receiving trouble or alarm signal from Channel Unit Alarm LED is flickering during alarm time.

6-2-3. Communication LED -.COM-CH LED represents communication status between Common Unit and Channel Unit. This LED is turned on at first communication (0.1 sec) with each channel unit and turned off. If Common Unit can normally communicate with whole pre-set channel units, this LED is continuously turning on. (Channel Unit Access time: 0.1) -.COM-PC LED is turned on once and immediately turned off when Common Unit receives normally communication data from PC. -. If SW1 (channel unit number) and SW2 (common unit address) are set under 1 in common unit, COM-CH and COM-PC LEDs are flickering (0.5sec interval) 16

GTC-200A

6-2-4. Buzzer Stop LED (BZ-STOP LED ) -. This LED is on when buzzer stops by pressing buzzer stop key in case of alarm from Channel Unit (This LED is activated only when alarm type is set as Hold in Channel Unit) -. This LED is off when pressing buzzer stop key once again in Common or Channel Unit.

6-2-5. Buzzer Stop & Reset Key - In case of trouble or alarm on each channel unit, this key is used for stopping buzzer and resetting - Push once .... Buzzer sound stops and BZ-STOP LED is on. - Push twice .... BZ-STOP LED is off and performs Reset function.

6-2-6. RS485 MODBUS communication Data 6-2-6-1. RS-485 communication Format - Baudrate : 9600BPS - Stop bit : 1 Stop - Parity : Even parity 6-2-6-2. Gas density value (Analog input) No

Function name

Address

1

Channel-1 unit gas density value

30001

2

Channel-2 unit gas density value

30002

3

Channel-3 unit gas density value

30xxx

Other Channel address is increased by 1 for each channel unit.

6-2-6-3. Channel unit status Bit Data(Digital input contact reading Data) Channel Unit No

Channel-1 Unit

Function name

Address

Alarm1 status Data

10001

Alarm2 status Data

10002

Alarm3 status Data

10003

Trouble status Data

10004

Test status Data

10005

Spare Data

Channel-2 Unit

8Bit is assigned for each channel unit

10006~10008

Alarm1 status Data

10009

Alarm2 status Data

10010

Alarm3 status Data

10011

Trouble status Data

10012

Test status Data

10013

Spare Data Channel-n Unit

Other

10014~10016 10001+((n-1)*8) ~ 10001+(((n-1)*8)+8)

Status Data

17

8Bit is assigned for each channel unit

GTC-200A

6-3. How to operate channel unit 6-3-1. Check of wiring - Check wirings to operation power, gas detector, common unit and channel unit with reference to Terminal wiring diagram (Section 5). * Power supply is AC110/220V, 50/60Hz. (Please make sure to check power type before turning on power)

6-3-2. Power ON -.Turn power on after checking voltage of power supply. -.Check power LED turned on at common unit and channel units. -.Check “SELF” on FND at channel units.

6-3-3. Display of gas density -.After power on of Channel Unit, “SELF” on FND at channel unit is flickering about 15 seconds and the channel unit goes into measurement mode. If the unit or gas detector has trouble, trouble signal occurs at this time. -.Channel unit displays gas density received from gas detector as number type on FND digital display. -.Gas density is also displayed with 3 Color Bar Graphic LED st Under 1 Alarm density: Green color st nd Above 1 Alarm and under 2 Alarm density: Orange color nd Above 2 Alarm density: Red color.

-. If there is no connection with gas detector or trouble at gas detector, channel unit displays “Undr” with flickering (0.5 sec interval). -. Trouble LED Lamp is also turned on. -. Lowest round LED at 3 Color Bar Graphic LED is turned on with Red color. -. If current value received from gas detector is 10% higher than predetermined High scale value, “OuEr” is displayed with flickering (0.5 sec interval). -. Highest round LED at 3 Color Bar Graphic LED is turned on with Red color. -.If gas density is higher than alarm setting value, corresponding alarm is activated after Alarm holding time. -.Alarm LED Lamp is flickering (0.5 sec interval) during alarm holding time and turned on after Alarm holding time. -.Alarm Relay is ON after Alarm holding time. -.If Alarm Latch Type is “on” mode, Alarm status and gas density value is held at maximum value when alarm is activated. Gas density value held is not automatically reset when real gas density is decreased below alarm setting value and only can be reset by “Reset” key. -.If Alarm Latch Type is “off” mode, Alarm is reset according to change of measured gas density.

6-3-4. Program Data setting -. Push “▲”, “▼” at the same time for over 2 seconds in gas density display mode and goes into program data setting mode. (During program setting mode, automatically return to density display mode if there is no key input for 10 seconds.) -. Firstly, “dP-S” (decimal point setting) is displayed. -. Press “Func” key to go into decimal point setting mode. -. If pressing “RESET” key, returns to density display mode. 18

GTC-200A -. By pressing “▲” or “▼”, decimal point can be changed in 3 types. -. Press “FUNC” key when desired decimal point is displayed to set the point and go to next item -. If pressing “RESET” key, returns to density display mode.

-. High Scale (Maximum display value of gas density) is displayed as “H-SL” -. Press “Func” key to go into High scale setting mode -. If pressing “RESET” key, returns to density display mode. -. High Scale value was set when shipping from factory according to domestic law -.High Scale value can be changed according to measuring range with “▲” or “▼”(increase or decrease). -.If desired High scale value is displayed, press “FUNC” key to set the value as High Scale and go to next item. -. If pressing “RESET” key, returns to density display mode -. This scale is set as same as detection range of gas detector at the factory. Ex) when set Range as 100 4 mA.DC input ................... 0 Display 20 mA.DC input ................... 100 Display -.SAD (compensating error from gas detector) setting message is displayed as “SAd ”. -. Press “Func” key to go into SAD value setting mode. -. If pressing “RESET” key, returns to density display mode -.When pressing “▲” or “▼” key, SAD value is increased or decreased (in case of negative number, “-“ sign is added and displayed. -.If desired SAD value is displayed, press “FUNC” key to set the value as SAD and go to next item. -.If pressing “RESET” key, returns to density display mode ex) In case of setting SAD as 2 : if output error from gas detector is -2, the display will be -2. In this case, User should compensate 2 (SAD setting value) to display 0 on FND. -. Channel Number setting message (setting recognition number of each channel unit) is displayed as “CHno ”. -. Press “Func” key to go into Channel Number setting mode.. -. If pressing “RESET” key, returns to density display mode -.Common unit can check operation status of each channel unit with channel number and this mode is to input each number (address) of channel unit. When pressing “▲” or “▼” key, address no. of channel unit will be increased or decreased. -.If desired address number is displayed, press “FUNC” key to set the value as Address NO. and go to next item (Alarm Reset Type function). -.If pressing “RESET” key, returns to density display mode -. If Channel No. was not inputted, Channel Unit and Common Unit can not communicate with each other. -. Please make sure to assign different number to Channel units.

6-3-5. Alarm Data setting -.Push “Func” for over 2 seconds in gas density display mode and go into alarm data setting mode. -.(During alarm data setting mode, automatically return to density display 19

GTC-200A mode if there is no key input for 10 seconds.) -.Function to set Alarm Latch Type “LACH” is displayed. -.Press “Func” to go into Alarm Latch Type setting mode. -.If pressing “RESET” key, returns to density display mode -.This step is for changing Alarm Reset Type. “ on” and “ oFF” are changed in turns when pressing “▲”or “▼” key. -.If desired Alarm Latch Type is displayed, press “FUNC” key to set the value as Alarm Latch Type and go to next item. - .If pressing “RESET” key, returns to density display mode -. There are two kinds of Alarm Latch Type (“ on” and “ oFF”) OFF mode is to automatically reset Alarm and ON mode is to reset alarm manually. -. Alarm1 value setting function (“AL-1”) is displayed. -. Press “Func” key to go into Alarm1 value setting mode. -.If pressing “RESET” key, returns to density display mode -.This mode is to change Alarm1 setting value (up to High Scale value) Alarm 1 value is increased or decreased by pressing “▲” or “▼” key. -. If desired Alarm1 value is displayed, press “FUNC” to set the value as Alarm1 and go to next item. -.If pressing “RESET” key, returns to density display mode -.Alarm Level was set when shipment according to density designated on domestic law. -. This mode is to set moving direction of Alarm1 “1H ” or “1L ” is displayed in turns when pressing “▲” or ▼” key. -.“1H ” mode is to activate alarm when value is equal and bigger than alarm 1 setting value and “1L ” mode is to activate alarm when value is equal and smaller than alarm 1 setting value. -. If desired mode is displayed, press “FUNC” to set mode and go to next item. -.If pressing “RESET” key, returns to density display mode -.Alarm Type was set at shipment (combustible: 1H & 2H & 3H / Oxygen: 3H & 2H & 1L / Toxic: 1H & 2H & 3H Type). -.This mode is to set alarm 1 dead band value and the value is increased or decreased by pressing “▲” or ▼” key. -. When Alarm1 is in “1H ”, alarm 1 is activated above the value (Alarm value plus Dead band value) and is deactivated below the value ( Alarm value minus Dead band value). -. When Alarm1 is in “1L ”, alarm 1 is activated below the value (Alarm value minus Dead band value) and is deactivated above the value ( Alarm value plus Dead band value) -. If desired alarm1 dead band value is displayed, press “FUNC” to set the value and go to next item. -.If pressing “RESET” key, returns to density display mode -. This function is to give hysteresis value to remove frequent alarm on/off around alarm setting value (this value is zero when shipment). Ex) If alarm setting value is 20%LEL and Dead Band is 2%LEL, alarm is activated at 22%LEL and deactivated at 18%LEL. -.Alarm1 delay time setting function (“AL1t”) is displayed. -.Press ”Func” key to go into alarm1 delay time setting mode. -.If pressing “RESET” key, returns to density display mode -. This function is to prevent malfunction by external shock or noise and change of alarm1 delay time is increased or decreased by pressing “▲” or “▼” key -. If desired alarm1 delay time is displayed, press “FUNC” key to set alarm1 20

GTC-200A delay time and go to next item -.If pressing “RESET” key, returns to density display mode Ex) alarm setting value is 20% LEL and delay time is 5 seconds, alarm is only activated when gas density is above than alarm setting value for over 5 seconds and no alarm is activated when gas density fall under alarm setting value within 5 seconds -.Alarm1 relay output setting function (“A1rl”) is displayed. -. Press “Func” key to go into Alarm1 relay output setting mode. -.If pressing “RESET” key, returns to density display mode -. This mode is to change Alarm1 relay output and output type is changed in turns (“on” and “oFF”) when pressing “▲” or “▼” key. -. If desired Alarm1 relay output type is displayed, press “FUNC” to set the type and go t onext item. -.If pressing “RESET” key, returns to density display mode -. There are two kinds of Alarm1 relay output modes (“on” and “ oFF”) oFF mode is to deactivate relay output and oN mode is to activate output. -. Alarm1 Blink (“A1bL”) is displayed This function is to set that alarm1 relay output is continuously on/off (1 sec interval) during operation of buzzer -. Press “Func” key to go into Alarm1 Blink output setting mode. -. If pressing “RESET” key, returns to density display mode -. This mode is to change Alarm1 Blink output type and “on” and “off” is changed in turns when pressing “▲” or “▼” key. -. If desired Alarm1 Blink output type is displayed, press “FUNC” key to set the type and go to next item. -.If pressing “RESET” key, returns to density display mode -. Alarm1 Blink output types are “ on” and “ oFF”. oFF type is to deactivate Alarm1 Blink output and oN type is to activate Alarm1 Blink output.(this output is activated only when Alarm1 relay output type is “on”) -.Alarm2 value setting function (“AL-2”) is displayed. -. Press ”Func” key and go into Alarm2 setting mode. -.If pressing “RESET” key, returns to density display mode -. This function is to change Alarm2 setting value (up to High Scale value) Alarm 2 value is increased or decreased by pressing “▲” or “▼” key -. If desired Alarm2 value is displayed, press “FUNC” to set the value as Alarm1 and go to next item -.If pressing “RESET” key, returns to density display mode -. Alarm Level was set when shipment according to density designated on domestic law -. This mode is to set moving direction of Alarm2 “2H ” or “2L ” is displayed in turns when pressing “▲” or ▼” key. -.“2H ” mode is to activate alarm when value is equal and bigger than alarm 2 setting value and “2L ” mode is to activate alarm when value is equal and smaller than alarm 2 setting value. -. If desired mode is displayed, press “FUNC” to set mode and go to next item. -. If pressing “RESET” key, returns to density display mode -. Alarm Type was set at shipment (combustible: 1H & 2H & 3H / Oxygen: 3H & 2H & 1L / Toxic: 1H & 2H & 3H Type). -. This mode is to set alarm2 dead band value and the value is increased or decreased by pressing “▲” or ▼” key. -. When Alarm2 is in “2H ”, alarm2 is activated above the value (Alarm value 21

GTC-200A plus Dead band value) and is deactivated below the value ( Alarm value minus Dead band value). -. When Alarm2 is in “2L”, alarm2 is activated below the value (Alarm value minus Dead band value) and is deactivated above the value ( Alarm value plus Dead band value) -. If desired alarm2 dead band value is displayed, press “FUNC” to set the value and go to next item. -.If pressing “RESET” key, returns to density display mode -. This function is to give hysteresis value to remove frequent alarm on/off around alarm setting value (this value is zero when shipment). Ex) If alarm setting value is 20%LEL and Dead Band is 2%LEL, alarm is activated at 22%LEL and deactivated at 18%LEL. -. Alarm2 delay time setting function (“AL2t”) is displayed. -. Press ”Func” key to go into alarm2 delay time setting mode -.If pressing “RESET” key, returns to density display mode -. This function is to prevent malfunction by external shock or noise and change of alarm2 delay time is increased or decreased by pressing “▲” or “▼” key -. If desired alarm2 delay time is displayed, press “FUNC” key to set alarm2 delay time and go to next item -.If pressing “RESET” key, returns to density display mode Ex) alarm setting value is 20% LEL and delay time is 5 seconds, alarm is only activated when gas density is above than alarm setting value for over 5 seconds and no alarm is activated when gas density fall under alarm setting value within 5 seconds . -. Alarm2 relay output setting function (“A2Ct”) is displayed. -. Press “Func” key to go into Alarm2 relay output setting mode. -. If pressing “RESET” key, returns to density display mode -. This mode is to change Alarm2 relay output and output type is changed in turns (“on” and “oFF”) when pressing “▲” or “▼” key. -. If desired Alarm2 relay output type is displayed, press “FUNC” to set the type and go to next item. -.If pressing “RESET” key, returns to density display mode -. There are two kinds of Alarm2 relay output modes (“on” and “ oFF”) oFF mode is to deactivate relay output and oN mode is to activate output.. -. Alarm2 Blink (“A2bL”) is displayed This function is to set that alarm1 relay output is continuously on/off (1 sec interval) during operation of buzzer -. Press “Func” key to go into Alarm2 Blink output setting mode. -.If pressing “RESET” key, returns to density display mode -. This mode is to change Alarm2 Blink output type and “on” and “off” is changed in turns when pressing “▲” or “▼” key. -. If desired Alarm2 Blink output type is displayed, press “FUNC” key to set the type and go to next item. -.If pressing “RESET” key, returns to density display mode -. Alarm2 Blink output types are “ on” and “ oFF”. oFF type is to deactivate Alarm2 Blink output and oN type is to activate Alarm2 Blink output.(this output is activated only when Alarm2 relay output type is “on”) -. Alarm3 value setting function (“AL-3”) is displayed. -. Press ”Func” key and go into Alarm3 settingmode. -.If pressing “RESET” key, returns to density display mode 22

GTC-200A

-. This function is to change Alarm3 setting value (up to High Scale value) Alarm 3 value is increased or decreased by pressing “▲” or “▼” key -. If desired Alarm3 value is displayed, press “FUNC” to set the value as Alarm1 and go to next item -.If pressing “RESET” key, returns to density display mode -. Alarm Level was set when shipment according to density designated on domestic law -. This mode is to set moving direction of Alarm3 “3H ” or “3L ” is displayed in turns when pressing “▲” or ▼” key. -.“3H ” mode is to activate alarm when value is equal and bigger than alarm3 setting value and “3L ” mode is to activate alarm when value is equal and smaller than alarm 3setting value. -. If desired mode is displayed, press “FUNC” to set mode and go to next item. -. If pressing “RESET” key, returns to density display mode -. Alarm Type was set at shipment (combustible: 1H & 2H & 3H / Oxygen: 3H & 2H & 1L / Toxic: 1H & 2H & 3H Type). -. This mode is to set alarm3 dead band value and the value is increased or decreased by pressing “▲” or ▼” key. -. When Alarm3 is in “3H ”, alarm3 is activated above the value (Alarm value plus Dead band value) and is deactivated below the value ( Alarm value minus Dead band value). -. When Alarm3 is in “3L”, alarm3 is activated below the value (Alarm value minus Dead band value) and is deactivated above the value ( Alarm value plus Dead band value) -. If desired alarm3 dead band value is displayed, press “FUNC” to set the value and go to next item. -.If pressing “RESET” key, returns to density display mode -. This function is to give hysteresis value to remove frequent alarm on/off around alarm setting value (this value is zero when shipment). Ex) If alarm setting value is 20%LEL and Dead Band is 2%LEL, alarm is activated at 22%LEL and deactivated at 18%LEL. -. Alarm3 delay time setting function (“AL3t”) is displayed. -. Press ”Func” key to go into alarm3 delay time setting mode -.If pressing “RESET” key, returns to density display mode -. This function is to prevent malfunction by external shock or noise and change of alarm3 delay time is increased or decreased by pressing “▲” or “▼” key - If desired alarm3 delay time is displayed, press “FUNC” key to set alarm3 delay time and go to next item -.If pressing “RESET” key, returns to density display mode Ex) alarm setting value is 20% LEL and delay time is 5 seconds, alarm is only activated when gas density is above than alarm setting value for over 5 seconds and no alarm is activated when gas density fall under alarm setting value within 5 seconds -. Alarm3 relay output setting function (“A3rL”) is displayed. -. Press “Func” key to go into Alarm3 relay output setting mode. -.If pressing “RESET” key, returns to density display mode -. This mode is to change Alarm3 relay output and output type is changed in turns (“on” and “oFF”) when pressing “▲” or “▼” key. -. If desired Alarm3 relay output type is displayed, press “FUNC” to set the type and go to next item. -.If pressing “RESET” key, returns to density display mode -. There are two kinds of Alarm3 relay output modes (“on” and “ oFF”) 23

GTC-200A oFF mode is to deactivate relay output and oN mode is to activate output. -. Alarm2 Blink (“A3bL”) is displayed This function is to set that alarm1 relay output is continuously on/off (1 sec interval) during operation of buzzer - Press “Func” key to go into Alarm3 Blink output setting mode. -.If pressing “RESET” key, returns to density display mode -. This mode is to change Alarm3 Blink output type and “on” and “off” is changed in turns when pressing “▲” or “▼” key. -. If desired Alarm3 Blink output type is displayed, press “FUNC” key to set the type and go to next item. -.If pressing “RESET” key, returns to density display mode -. Alarm3 Blink output types are “ on” and “ oFF”. oFF type is to deactivate Alarm3 Blink output and oN type is to activate Alarm3 Blink output.(this output is activated only when Alarm3 relay output type is “on”) -. The completion message “End” is displayed for 2 seconds and return to density display mode.

6-3-6. Maintenance setting -. After power “ON”, press “Test” and “Reset” keys at the same time when “SELF” is flickering and go into maintenance setting mode. -. Please make sure to press “Reset” key in maintenance mode, in order to return to density display mode. -. This mode is to select functions to set current input and output values, and “-In-” or “-oUt” are displayed when pressing “▲” or ▼” key. -. If desired mode is displayed, press “func” to go into selected mode. -.“-In-” mode is to set current input value and “-oUt” mode is to set current output value -.If pressing “RESET” key, returns to density display mode -. This mode is to select “-In-” and to set 20mA (input current) and message is shown as “IA20”. -. Press ”Func” key to go into number display mode (input current). -.If pressing “RESET” key, returns to density display mode -. This mode is to display current (20mA) inputted to (mA) terminal. -. If displayed value is stable and press ”Func” key to set the value and go to next item. -.If pressing “RESET” key, returns to density display mode -. This mode is to select “-Out” and to set 4mA, and the message is shown as “A-04”. -. Connect Ampere meter to 4~20mA Output terminal. Press “▲” or ▼” key to show 4mA in Ampere meter and push “Func” key to set output current and go to next item. -.If pressing “RESET” key, returns to density display mode -.This mode is to set 20mA output current and message is shown as “A-20”. -. Connect Ampere meter to 4~20mA Output terminal. Press “▲” or ▼” key to show 20mA in Ampere meter and push “Func” key to set output current and go to next item -.If pressing “RESET” key, returns to density display mode 24

GTC-200A -. The completion message “End” is displayed for 2 seconds and return to density display mode.

6-3-7. Test function display -. Press “Test” key over 2 seconds in gas density display mode and go into Test mode. -. (During Test mode, automatically return to density display mode if there is no key input for 30 minutes.) -. Gas density display is flickering during Test mode and this mode is to perform test without gas injection to sensor of gas detector. User can set desired gas density with “▲” or ▼” key and alarm is normally activated with inputted density by user. -. Press ”Func” key over 2 seconds and go into FND/LED/Bar LED Test mode. -.If pressing “RESET” key, returns to density display mode -. This mode is for FND/LED/Bar LED Test and 4 digits number and 8888 are displayed in turns on FND (0.5 second interval) 4 digits number represents PC address of common unit (SW1) and channel number (SW2). -.3 Color Bar Graphic LED shows green, orange and red color in turns and each function LED is also flickering (0.5 second interval) -. Press “Func” to go into next mode. -.If pressing “RESET” key, returns to density display mode -. This mode is to check channel unit normally communicated with Common unit and normal channel unit number is displayed in turns (0.5 second interval) -. If some channel number is missing, this channel units can not communicate with Common unit -. Press “Func” key and returns to gas density test mode. -.If pressing “RESET” key, returns to density display mode

6-3-8. Stand-by mode -. Press “▼(Standby)” key over 2 seconds in gas density display status and go into Stand-by mode. -. Please make sure to press “▼(Standby)” key on channel unit during stand-by mode, in order to return to density display mode. -.During Stand-by mode, STD-BY LED is flickering and every function is normal except trouble/alarm relay output of Common unit & Channel unit -. Press “▼(Standby)” key to return to gas density display mode.

25

GTC-200A

7. External Drawings and Dimensions 7-1. Dimension for each unit 7-1-1. Power Unit

34

119.3±1.0 104±1.0

12.7

36

97

GTC200A POWER

VOLT

MAIN POWER

160

144

BATT GOOD

136

BATT POWER

BATT TROUBLE

BATT TEST

GTC200A

7-1-2. Common Unit

34

119.3±1.0 104±1.0

12.7

36

97

GTC200A COMMON

BZ. STOP

BZ. STOP RESET

GTC200A

26

160

144

COM-PC

136

COM-CH

GTC-200A

7-1-3. Channel Unit

34

119.3±1.0 104±1.0

36

12.7

97

GTC200A CHANNEL

%LEL 100 90 80 70 60

136

144

20

160

50 40 30

10 0

TEST

RESET

GTC200A

7-2. Panel Mount Type Dimension and Panel cut-out Size A

GTC200A POWER

GTC200A COMMON

GTC200A CHANNEL

VOLT

GTC200A CHANNEL

%LEL

Minimum space of cut-out is 30mm in case of above 2 stages panel

GTC200A CHANNEL

%LEL

%LEL

100

100

100

90

90

90

MAIN POWER 80 70 60

144

BATT POWER

BATT GOOD

50

COM-PC

BZ. STOP

BATT TROUBLE

BATT TEST

GTC200A

BZ. STOP RESET

GTC200A

50

50

40

40

40

30

30

30

20

20

20

10

10

10

0

0

TEST

RESET

GTC200A

Min≥30

80 70 60

B

137

80 70 60

COM-CH

Panel cut-out

0

TEST

RESET

GTC200A

TEST

RESET

GTC200A

CHANNEL A(mm) B(mm) CHANNEL A(mm) B(mm) 1 Channel 108 107 6 Channel 288 287 2 Channel 144 143 7 Channel 324 323 3 Channel 180 179 8 Channel 360 359 4 Channel 216 215 9 Channel 396 395 5 Channel 252 251 10 Channel 432 431 -. The power unit is option item. If the power unit is used the width of power unit (36mm) shall be subtracted. Ex) If an user needs 5 channel units without the power the size of A is 216mm (252mm-36mm) and the size of B is 215mm (251mm-36mm). 27

GTC-200A

7-3. Wall Mount Type Dimension 7-3-1. The type including the power unit 145

B

35

A

GTC200A POWER

GTC200A COMMON

GTC200A CHANNEL

VOLT

GTC200A CHANNEL

%LEL

GTC200A CHANNEL

%LEL

100

100

90

90

%LEL 100 90

MAIN POWER 80

BATT POWER

80 70

70

60

60

COM-CH 50

50

40

BATT GOOD

80

70 60

COM-PC

20 10

0

TEST

GTC200A

0

RESET

TEST

GTC200A

RESET

TEST

GTC200A

RESET

GTC200A

290

345

GTC200A

30

20 10

BZ. STOP RESET

BATT TEST

40

30

20 10 0

BATT TROUBLE

50

40

30

BZ. STOP

20

10.5

Ø5 Ø10

CHANNEL A(mm) B(mm) CHANNEL A(mm) B(mm) 3 Channel 217 147 7 Channel 361 291 5 Channel 289 219 9 Channel 433 363 -. Above 4 types are standard model and the others size should be special ordered. .

7-3-2. The type excluding the power unit 145

B

35

A

GTC200A COMMON

GTC200A CHANNEL

GTC200A CHANNEL

%LEL

GTC200A CHANNEL

%LEL 100

100

90

90

90

80

80

GTC200A CHANNEL

%LEL

100

%LEL 100 90

80

80

70

70

70

70

60

60

60

60

COM-CH 50

50

40

COM-PC

BZ. STOP

BZ. STOP RESET

50

40

30

40

30

30

20

20

20

10

10

10

0

0

TEST

RESET

GTC200A

0

TEST

RESET

GTC200A

0

TEST

RESET

GTC200A

TEST

RESET

GTC200A

290

345

GTC200A

50

40

30 20 10

10.5

20

Ø5 Ø10

CHANNEL A(mm) B(mm) CHANNEL A(mm) B(mm) 4 Channel 217 147 8 Channel 361 291 6 Channel 289 219 10 Channel 433 363 -. Above 4 types are standard model and the others size should be special ordered. .

28

GTC-200A

The products and instruction manual can be revised for improvement performance and easy use without prior notice.

18-8, Dogeumdanji1-gil, Sanglog-gu, Ansan-city, Gyeonggi-do, Korea Tel : 82.31.490.0800

Fax : 82.31.490.0801

http://www.gastron.com

29

[email protected]

Design, development, production, installation, and servicing of Gas Detector P/N: GIM12073102

GTD2000-Tx Instruction Manual Revision: 1

Please read this manual carefully for proper use of the device

Copyright ⒞GASTRON, Co., LTD. All rights reserved.

GTD2000-Tx Instruction Manual

Thank you for purchasing the product of us Gastron Co., Ltd. Gastron has been recognized from many consumers for its top quality and ease of use as a professional company of Gas detectors & Gas Monitors. We are constantly researching and striving to help consumers to find with required products nearby and to develop consumer satisfied Gas detectors. From now on, Gastron detectors will resolve all the troubles you have and satisfy your needs. This manual contains method for the installation, operation method, and simple maintenance methods on GTD2000-Tx Gas detector. Read it carefully and keep it for further reference in case you have questions.

If there are any problems with our products, please contact us at the address below.  Address: 18-8 Dogeumdanji 1gil(Palgogi-dong), Sanglok-gu, Ansan-city, Gyeonggi-do, Korea,  Tel : 82-31-490-0800  Fax : 82-31-490-0801  URL : www.gastron.com  e-mail : [email protected]

Note • Inspection and calibration are recommended at least once every 3 months to ensure accurate operation of the Gas detector using calibration gas depending on the type of gas. • Negligence of periodic inspection and calibration may cause malfunction of the device due to the aging of the Sensor. • Only qualified personnel with professional skills on Gas detector may disassemble the unit if necessary. • For further information on Gas detector inspection and correction, please contact us at our technical department, e-mail or web site.

PAGE 2 / 44

GTD2000-Tx Instruction Manual

This page intentionally left blank

PAGE 3 / 44

GTD2000-Tx Instruction Manual

TABLE OF CONTENTS 1. Overview ............................................................................................................................................................................... 6 2. Structure ............................................................................................................................................................................... 6 3. Specification ........................................................................................................................................................................ 7 4. Name of Components and Main features ............................................................................................................ 8 4.1. Components ........................................................................................................................................................... 8 4.2. Description of components............................................................................................................................. 9 5. Terminal wiring diagram ............................................................................................................................................. 10 5.1. Separation of Detector Body and Cover ................................................................................................. 10 5.2. Main PCB configuration .................................................................................................................................. 11 5.2.1. Main PCB configuration ..................................................................................................................... 11 5.2.2. HART Board configuration ................................................................................................................ 12 5.3. Main PCB terminal description and wiring method .......................................................................... 14 5.3.1. 4~20mA Source drive type wiring method .............................................................................. 15 5.3.2. 4~20mA Sink drive type wiring method .................................................................................... 16 5.3.3. Connection method with our main Control Unit ................................................................... 17 6. Standard Type outside view and Dimensions ................................................................................................... 18 7. Warning light type: Outside view and Dimensions ........................................................................................ 20 8. Raincover type: Outside view and Dimensions................................................................................................. 21 9. Menu Configuration Table ......................................................................................................................................... 22 10. Detector activation Flow and KEY operation ................................................................................................... 25 10.1. Sensor activation Flow .................................................................................................................................. 25 10.2. Sensor KEY configuration and description .......................................................................................... 25 11. Initial status and Menu description ..................................................................................................................... 26 11.1. Initial operation status (Power On) ......................................................................................................... 26 11.2. Gas measuring status (Measuring Mode) configuration ............................................................... 26 11.3. PROGRAMMABLE MODE Setting ............................................................................................................. 27 11.4. CALIBRATION MODE Operation............................................................................................................... 28 11.4.1. Zero Calibration .................................................................................................................................. 28 11.4.2. Span Calibration .................................................................................................................................. 29 11.5. ALARM mode setting .................................................................................................................................... 30 11.6. SENSOR DATA MODE configuration ...................................................................................................... 32 PAGE 4 / 44

GTD2000-Tx Instruction Manual 11.7. Maintenance Mode setting......................................................................................................................... 33 11.8. Device Mode setting...................................................................................................................................... 35 11.9. Version mode configuration ...................................................................................................................... 36 11.10. Test mode configuration ........................................................................................................................... 37 11.11. Inspection mode use................................................................................................................................... 38 11.12. Data initialization .......................................................................................................................................... 38 11.13. Correction data initialization ................................................................................................................... 38 12. Troubleshooting............................................................................................................................................................ 39 13. Caution before installation....................................................................................................................................... 40 13.1. Selection of installation location (Occupational Safety and Health Law) .............................. 40 13.2. Selection of installation location (High Pressure Gas Safety Management Regulations) ............................................................................................................................................................................................ 40 13.3. Precaution before installation.................................................................................................................... 41 14. Ordering Information ................................................................................................................................................. 42 15. Revision history ............................................................................................................................................................. 44

PAGE 5 / 44

GTD2000-Tx Instruction Manual

1. Overview GTD2000-Tx toxic gas detector was developed in order to prevent accident by detecting gas leaking in industrial fields such as various toxic gases that are generated in plants manufacturing or using toxic gas, gas storage, or in the manufacturing process. GTD2000-Tx toxic gas leak detector is installed in areas of risk of leakage; constantly and continuously detects gas leak; displays the measured value in the liquid crystal display (LCD) installed in the detector; and provides DC4-20mA standard output signal. In addition, the Cable Connection Length of DC4 ~ 20mA standard output signal between sensor element and receiver extends up to 2,500m (When Cable CVVS or CVVSB 1.5sq or equivalent is used). This GTD2000-Tx toxic gas detector should be used at an elevation of less than 1000M.

2. Structure The body of GTD2000-Tx is made of aluminum alloy, and completely explosion-proof (Ex d IIC T6). The product may be installed anywhere risky of combustible gas leak and explosion, displaying the gas leak status in the installation site on the built-in liquid crystal display (LCD) within the sensing unit. The internal structure is composed of a liquid crystal display unit displaying the measurement; a terminal unit sending the measured value (DC4-20mA) to outside; and a PCB Board. A magnet bar is installed outside, so maintenance work is convenient because calibration is possible from outside using this magnet-bar.

PAGE 6 / 44

GTD2000-Tx Instruction Manual

3. Specification I T

E M S

SPECIFICATION

Measuring Type

Diffusion

Measuring Value Display

Local Digital LCD or OLED Display

Means Value Display

Back light, 2-line/8-Characters LCD or OLED

Approval

Ex d IIC T6

Detectible Gas

Toxic Gas

Measuring Method

Electro-Chemical Cell Heated-semiconductor Cell

Measuring Range

Refer to appendix.

Response Time

90% of full Scale in less than 45 sec

Accuracy

± 3% / Full Scale

Zero Drift

Less than 2% full Scale

Operating Temp.

-20 to 60℃

Operating Humidity

5 to 99% RH (Non-condensing) Measuring Output: 4 – 20mA.DC / Full Scale

Signal Output

Test Output: 3mA Calibration Output: 3mA Fault Output: 2mA

HART® Interface

HART REV7 / Optional Board (Note 1)

Calibration Work

Magnetic interface to configure Alarm

Cable Connection Length

Max. 2,500m : 4-20mA Signal

Power Supply

18 - 31V DC (24V DC / 60mA) With alarm device (24V DC / 110mA)

Conduit Connection

1/2” or 3/4”PF, NPT(Standard : 3/4” PF)

Signal Cable Connection

(CVVS or CVVSB 1.5sq↑×3 wires) + Shield HART® Interface Board

Option

GTL-100( Explosion-proof Warning Light ) Rain Cover

Dimensions

136(W) × 166(H) × 110 (D) mm

Weight

App. 1.5kg [Table 1. GTD2000-Tx Specification]

Note 1) For HART, refer to GTD2000 HART® Field Device Specification

PAGE 7 / 44

GTD2000-Tx Instruction Manual

4. Name of Components and Main features 4.1. Components

1

3

16

12

4

2

7

8 15

9

15 PR

ESS

DE MO

Tx

SERIAL No.

E I N KO R

EA

DATE

GA

13

UP

D MA

13

S D E T E C T OR

RESET

L I FT

L : GTD - 2 0 0 0

6 GA

5

S TR

PR

TD ON C O . , L P

ESS

L I FT

10

.

U

11

14

[Figure 1. GTD2000-Tx Components]

No.

Name

No.

Name

1

Detector housing body

9

Reset Switch

2

Detector housing cover

10

↑(up) Switch

3

AMP(main) PCB

11

↓(Down) Switch

4

Display parts

12

External Earth Ground( Min 1.5sq Cable)

5

Power/Signal Terminal

13

Mount Holes(ø7)

6

Sensor Terminal

14

Sensor

7

Power LED

15

Conduit Connection

8

Function Switch

16

Internal Ground

[Table 2. Components Reference Table]

PAGE 8 / 44

GTD2000-Tx Instruction Manual

4.2. Description of components No.

Name

1.

Detector Housing Body

2.

Detector Housing Cover

Description Protects Sensor and built-in PCB Board from external environment and shock. Is assembled with Detector Housing Body, and has a circular glass top through which LCD displayed Measuring Value can be seen. Amplifies the tiny Output generated by Sensor Element, converts it

3.

AMP PCB

to 4-20mA.DC by converting a standard Output, and transmits the Data to the Display unit. Displays the Data from AMP / Terminal (Transmitter) PCB to LCD

4.

Display PCB

display (Measuring Value), and indicates the power status with the Power Lamp. CN9 is composed of DC18V ~ 31V power supply and a DC 4 ~

5.

Power/Signal Terminal

20mA standard Output Connection terminal (VISO, +24V, mA, GND).

6.

Sensor Terminal

CN8 is Sensor Connection Terminal.

7.

Power LED

This Lamp is ON when Power is supplied. In Measuring Mode, touching this Switch for 2 sec with Magnet-

8.

Function Switch

Bar will enter to Function Setting Mode; or will store changed Data in Function Setting Mode. Touching this Switch once with Magnet-Bar will cancel the

9.

Reset Switch

Parameter setting, or return to the previous condition. (Every touch will return the Mode one previous step). Touching this Switch once with Magnet-Bar will convert one step

10.

↑(UP) Switch

or increase the displayed value in Mode conversion or number change. Touching this Switch once with Magnet-Bar will convert one step

11.

↓(DOWN) Switch

or decrease the displayed value in Mode conversion or number change.

12. 13. 14.

External Earth Ground

The outside of the Detector should be grounded to protect the circuit from external Noise or strong electric field.

Mount Hole

This Hole is used to secure the Gas Detector on exterior walls and

(ø7×2ea)

other installation places.

Sensor

Actually detects the gas leak, which is converted to electric signal and transmitted AMP PCB. This is provided for the inlet for power supply and Measuring

15.

Conduit Connection

Output signal. Cable connector has 3/4 ", 1/2" PF or NPT (Default spec. is PF 3/4".)

PAGE 9 / 44

GTD2000-Tx Instruction Manual

16.

The inside of the Detector should be grounded to protect the

Internal Ground

circuit from external Noise or strong electric field. [Table 3. Components details table]

5. Terminal wiring diagram

Warning

Never install, uncover, or manipulate the Detector other than authorized

personnel or installation/repair service person from Gastron, or serious loss of life and property damage such as fire or explosion may occur. In addition, check around for explosive Gas or flammable substances, followed by turning OFF before any work.

5.1. Separation of Detector Body and Cover The Gas Detector Cover can be removed by turning the Cover fastening Slotted Set Screw (M4 x 1ea) three to four rotations anti-clockwise using hex wrench (M2), followed by turning the Cover counterclockwise by hand. After the separation of Cover, the LCD indicator appears. Slotted Set Screw (M4)

[Figure 2. Slotted Set Screw]

After the Cover removal, remove the Display Part as follows.

PAGE 10 / 44

GTD2000-Tx Instruction Manual



Click the left and right retainer rings on the front of the LCD display inwards at the same time.

② While holding, pull the Display Part forward to separate from the Gas detector Body.

③ With the Display Part removed, the Main

PCB

is

shown

under

the

Detector Body.

[Figure 3. Display Part removal ]

5.2. Main PCB configuration 5.2.1. Main PCB configuration With the Display Part removed, the Main PCB terminal arrangement is shown as the following Figure.

[Figure 4. Main PCB terminal arrangement]

PAGE 11 / 44

GTD2000-Tx Instruction Manual No

Name

Description

1

CN9

2

J1

3

CN1

Display LCD Connector

4

CN8

Sensor Connector

5

CN6

Program download Connector

6

D1

7

CN2,CN3,CN5

8

CN4

Power & Output Signal Terminal 4~20mA Source / Sink selection jumper ( ON: Source Type, OFF: Sink Type )

Status LED ( blinks every second in normal operation ) HART Option Board Connector Warning Light (GTL-100) Interface Connector [Table 4. Main PCB main Parts Description]

5.2.2. HART Board configuration HART Board consists of Option Board, and is connected using CN5, CN6, CN7 terminals of Main PCB and the HART Board Screw at the top left.

[Figure 5. HART Board constituting Main PCB]

No

Name

Description

PAGE 12 / 44

GTD2000-Tx Instruction Manual 1

HART Board Screw

2

Blank Hole

3

HART Board

Constructed using 3Ø Screw Reserve Hole Option Board for HART Interface

[Table 5. HART Board main Parts description]

PAGE 13 / 44

GTD2000-Tx Instruction Manual

5.3. Main PCB terminal description and wiring method If you remove the Display Part, there is the Terminal Block under the Main PCB as shown in the following Figure 6. The Terminal Block can be removed from Main PCB by holding and pulling upward by hand. Unscrew the 5 terminal set screws above the separated Terminal Block CN9 (VIS, +, mA, -, ETH) Connector counter-clockwise with a Θ screwdriver; connect DC18-24V power to +, -; connect Signal Cable to mA; tighten the terminal set screws clockwise to keep the terminals in place; and insert it like before the removal.

[Figure 6. CN9 Terminal structure]

Description No

PCB Silk

Pin Name

4~20mA Source Drive

4~20mA Sink Drive

(J1 Jumper ON)

(J1 Jumper OFF)

N.C

4~20mA Sink In(+)

1

VISO

VIS

2

+24V

+

3

mA

mA

4

GND

-

GND / POWER (-)

5

ETH

ET

EARTH

+24V / POWER (+) 4~20mA Source Out

4~20mA Sink Out(-)

[Table 6. CN9 connector description]

Note1) Be sure to use CVVS or CVVSB 1.5sq↑ Shield Cable before Terminal construction.

PAGE 14 / 44

GTD2000-Tx Instruction Manual Note2) Fasten Terminals based on +24V of 2Pin to connect the 4Pin Terminal of existing conventional GTD2000-Tx.

5.3.1. 4~20mA Source drive type wiring method

Connect 4-20mA Signal Terminal of PLC to ‘mA’ of GTD2000-Tx. GND Terminal is used in common with the power. Turn J1 Jumper ON. ※

HART Communicator can be used only in models utilizing HART Option Board.

[Figure 7. Analog 4~20mA Source Configuration]

PAGE 15 / 44

GTD2000-Tx Instruction Manual

5.3.2. 4~20mA Sink drive type wiring method

Connect 4-20mA Sink Output (+) Terminal of PLC to VISO Terminal; and (-) Terminal to ‘mA’ Terminal. Turn J1 Jumper OFF. ※

HART Communicator can be used only in models utilizing HART Option Board.

[Figure 8. Analog 4~20mA Sink Configuration]

PAGE 16 / 44

GTD2000-Tx Instruction Manual

5.3.3. Connection method with our main Control Unit Connect CN9 ( VISO , + , mA , - , ET ) Connection Terminal of the Gas Detector and the Control Unit with reference to the Figure below. (See product manual for each Control Unit.)

GAS DETECTOR VISO POWER /mA

+

9

24V

mA

10

mA

-

11

-

ET

감지기 (SENSOR)

CONTROL UNIT

SHIELD EARTH

(GTC-100A Series Control unit)

GAS DETECTOR

VISO

POWER /mA

+

15

+

mA

16

mA

-

17

-

ET

SENSOR (탐지기)

CONTROL UNIT

SHIELD EARTH

(GTC-200A/210A Series Control unit)

GAS DETECTOR CN2

VISO

+ POWER /mA

+24V

mA

S-IN(mA)

ET

GND

CONTROL UNIT

SHIELD EARTH

(GTC-510A/520A Series Control unit)

PAGE 17 / 44

GTD2000-Tx Instruction Manual

6. Standard Type outside view and Dimensions

PAGE 18 / 44

GTD2000-Tx Instruction Manual

42

63

136 120 110

FUNC

RESET

SERIAL No.

DATE

35 25

110

113

176

2-Ø7

[Figure 9. GTD2000-Tx outside view]

PAGE 19 / 44

GTD2000-Tx Instruction Manual

7. Warning light type: Outside view and Dimensions 183.8 136 120 110

한국산업기술시험원 인증필

203

규격 및 형식: GTL-100 방 폭 형 식: II 2G Ex dIIC T6 Gb IP65 사 용 온 도: -30°C ~ +70°C 사 용 전 원: 24VDC / 80mA Max. 인 증 번 호: 12-KB2BO-XXXX

FUNC

RESET

SERIAL No.

DATE

25

49

110

113

176

2-Ø7

[Figure 10. GTD2000-Tx Warning Light type outside view]

PAGE 20 / 44

GTD2000-Tx Instruction Manual

8. Raincover type: Outside view and Dimensions 177.41

FUNC

RESET

DATE

217.5

244.75

SERIAL No.

[Figure 11. GTD2000-Tx Raincover type outside view]

PAGE 21 / 44

GTD2000-Tx Instruction Manual

9. Menu Configuration Table Level1

Level2

DEFAULT

NAME

PARAMETER

GAS TYPE (Gas Type)

[DEFIN./USER]

DEFIN.

GAS SEL (Gas Select)

See Table 8

COMB.

PROGRAM

UNIT SEL (Unit Select)

%/%LEL/PPM/PPB

%LEL

MODE

D-POINT (Decimal Point)

0.100/1.00/10.0/100

100

HIGH SCL (High Scale)

1~10000

100

PASSWORD

00~99

00

ZERO CAL [NO,YES]

[NO]

ZERO GAS

CALIBRA.

[

[ZERO] (Calibration Zero)

0]

ZERO >>>> [SUCCESS / FAIL] CAL. DATA [ 0/ FAIL]

CALIBRA.

SPAN CAL

MODE

[NO , YES]

(Calibration Mode)

SPAN GAS [

CALIBRA.

0]

SPAN SET

[SPAN] (Calibration Span)

[NO]

[ 50/ FAIL] SPAN

50

>>>>

[SUCCESS / FAIL ] CAL. DATA [

ALARM MODE

SENSOR MODE

0]

ALM USED (Alarm Used)

[OFF / ON]

OFF

AL LATCH (Alarm Latch)

[OFF / ON]

OFF

A1 LEVEL (Alarm-1 Level)

[1 ~ 9999]

20

A1 TYPE (Alarm-1 Type)

[INC / DEC]

INC

A1 DBAND (Alarm-1 Dead Band)

[0.0 ~ 10.0%F]

1.0

A1 TIME (Alarm-1 Time)

[0~ 60]SEC

1

A2 LEVEL (Alarm-2 Level)

[1 ~ 9999]

40

A2 TYPE (Alarm-2 Type)

[INC / DEC]

INC

A2 DBAND (Alarm-2 Dead Band)

[0.0 ~ 10.0%F]

1.0

A2 TIME (Alarm-2 Time)

[0~ 60]SEC

1

SEN. OUT (Sensor Output)

[ X.X mV ]

-

SEN. S/T (Sensor sensitivity)

[ X.X ]

-

MZ / MS (Manual Zero/Span)

X.X / XX.X

1.0 / -100.0

AZ / AS (Auto Zero / Auto Span)

X.X / XX.X

1.0 / -100.0

PAGE 22 / 44

GTD2000-Tx Instruction Manual

Level1

Level2

DEFAULT

NAME

PARAMETER

ZR / SR (Zero Rate / Span Rate)

XX.X % / XX.X %

49.9% / 60.0%

SEN GAIN (Sensor Gain)

[X.X] x G

1.0

SEN BIAS (Sensor BIAS)

[ X.XX V ]

-

TEMP. (Temperature)

[ XX ‘C ]

-

CURRENT(Sensor Current)

[ XXX mA]

-

24V VIN(GTD2000 Input Power)

[ XX.X V ]

-

CRO-SENS. (Cross Sensitivity)

[1.00] X G , 0.01~5.0 Adj

1.00

Z- SKIP (Zero Skip)

[0~10%F]

0.0

ODT (Operation Delay Time)

[0 ~ 60 SEC]

0

AUTO – Z (Auto Zero)

[ON / OFF]

ON

AZ MIN. (AutoZero Minimum)

[1.0~4.0] %F

2.0

BASE – Z (Base Zero)

[ON / OFF]

ON

MAINTEN.

BZ MAX. (BaseZero Maximum)

[0.5~2.0] %F

2.0

MODE

SP. HOLD (Span Hold)

[ON / OFF]

ON

(Maintenance Mode)

UNDER EN (Under Enable)

[ON / OFF]

OFF

SEN CHK. (Sensor Check)

[ON / OFF]

OFF

SEN-DIR (Sensor Direction)

[INC / DEC]

DEC

ENG MOD. (Engineer Mode)

[OFF / ON]

OFF

W/L TYPE (Warning Light Type)

[STEADY / BLINK.]

STEADY

M.-LEVEL (Maintenance Level)

[0 ~ Full Scale]

0

EMC T/O (Emergency Time Out)

[ON / OFF]

OFF

HART B/D (Hart Board)

[CHECK / EMPTY]

-

PADD / mA (Polling Address)

X / (ON/OFF)

-

DEV-CODE (Device Code)

[0xE1C3]

0xE1C3

FIX CUR (Fix Current)

[DISABLE / XX.XXmA]

-

SERIA.NO (Serial Number)

*XXXXXXXX

-

TAG

GTD-XXXX

GTD-0001

LONG TAG

GTD-XXXX-LT

GTD-0001-LT

DESCRIP. (Descriptor)

[ GASTRON GTD2000 ]

[ GASTRO >

MESSAGE

[ COMBUSTIBLE ]

[ COMBUS >

F/W REV (Soft Ware)

[REV 6]

-

E/X H/W (Hard Ware)

[REV 2]

-

HART DEV (Hart Device)

[REV 1]

-

HART REV (Hart Revision)

[REV 7]

-

mA OUT

[ON / OFF]

OFF

TEST

[ 0~Full scale ]

0

SENSOR MODE

DEVICE MODE

VERSION MODE

TEST MODE

[Table 7. Menu Configuration Table]

PAGE 23 / 44

GTD2000-Tx Instruction Manual ※

Gas name Reference Table No

Gas name

No

Gas name

No

Gas name

00

AN

32

DMDS

64

POCl3

01

AsH3

33

EDA

65

SI2H6

02

B2H6

34

EDC

66

SiCl4

03

BCl3

35

EtsH

67

SiF4

04

BF3

36

F2

68

SiH2.

05

C2H2

37

GeH4

69

SiH4

06

C2H4

38

H2

70

SO2

07

C2H4O

39

H2S

71

SO3

08

C2H6

40

H2Se

72

TBM

09

C3H6

41

HBr

73

TCS

10

C3H6O

42

HC

74

TEOS

11

C3H8

43

HCHO

75

THC

12

C4H8O

44

HCl

76

THF

13

C5H10

45

HCN

77

THT

14

C5H12

46

HF

78

TMP

15

C6H14

47

LNG

79

TOLU

16

C6H6

48

LPG

80

WF6

17

C6H6O

49

MCS

81

OTHER

18

C7H8

50

MeCHO

19

C7H8O

51

MEK

20

CH3CL

52

N2H4

21

CH3OH

53

NF3

22

CH4

54

NG

23

CiF3

55

NH3

24

CL2

56

NMP

25

CLF3

57

NO

26

CO

58

NO2

27

CO2

59

O2

28

COCL2

60

O3

29

COMB.

61

PCL3

30

DCS

62

PH3

31

DMC

63

PhoH

[Table 8. Gas name table]

PAGE 24 / 44

GTD2000-Tx Instruction Manual

10. Detector activation Flow and KEY operation 10.1. Sensor activation Flow - Timeout of Level1 and Level2 is 10 seconds, and 1 hour in the Calibration and Test Mode of Level2.

[Figure 12. Sensor workflow]

10.2. Sensor KEY configuration and description Item

Name

Description Sensor Mode entry function (more than 2 seconds of touch with Magnet-

FUNC

Function Key

bar in Measuring Mode). Entry to the next step of Level2 and storage of setting value.



RESET

Reset Key

Moving back to the previous step before the entered LEVEL



Up Key

Change to the next step after LEVEL1, and plus change of Level2 setting.



Down Key

Change to previous step before LEVEL1; minus change of Level2 setting.

Sensor Power ON followed by simultaneous input of Reset Key and Function Key will result in Factory Set in internal setting.

PAGE 25 / 44

GTD2000-Tx Instruction Manual

11. Initial status and Menu description 11.1. Initial operation status (Power On) After wiring of and power supplying to Power Terminal on MAIN PCB board, the following information will be displayed on the LCD display. About 30 minutes of stabilization time is required about 30 minutes’ stabilization time is required; normal operation begins after full stabilization.

GTD-2000 [ REV 6]

SELF TEST [ >>>>>>>>]

After the Power ON, the model name and product firmware Revision



number is displayed on LCD (OLED).

Self test runs for one minute, and the '>' character in second row



indicates the progress.

11.2. Gas measuring status (Measuring Mode) configuration [ [

%LEL] 0.0]



Operation is as follows in Normal state.



In the first row, Measuring GAS NAME and Measuring GAS Unit are displayed alternately every second; in the second row, current Measuring Value is displayed. Touching ‘Func’ Key with Magnet-bar for 2 seconds in current screen will change to



[ [

Setting Mode.

COMB.] 0.0]



During HART communications, '*' character is displayed on the left side of the 1st row.



When ENG. Mode is ON, '<' character is marked on the left side of the 2nd row. The operation is as follows when 1st or 2nd Alarm has occurred (ALARM EN item of



[ [ 0

%LEL] 30] ]

Maintenacne Mode must be ON to enable the operation). The 1st row is operated the same as in Normal mode; in the 2nd row, ALARM



messages and Gas Measuring Value are displayed alternately every 1 second. When GTL100 Explosion-proof Warning Light is installed, the red LED and Buzzer



[ %LEL] [ALARM1]

[

%LEL] [ OVER ]

[

%LEL] [ UNDER ]

[FAULT1] SEN EMPT

are blinking every second in the 1st Alarm; no flashing occurrs in the 2nd Alarm. Alarm condition continues unless it was released using Reset key if the Latch is on



during Alarm function.

If the entered Gas Measuring Valueis more than 10% higher than set High



Scale, the word “OVER” is displayed blinking every 1 second. At this time 4~20mA is operated as 21.6mA.



 If the entered Gas Measuring Valueis less than 10% lower, the word “UNDER” is displayed blinking every 1 second; ※ 

4 ~ 20mA is operated below 2mA.

This feature is operated only when UNDER button is ON. If there is any problem with the device, the Fault number and message will be displayed.



At this time, the 4 ~ 20mA is operated below 2mA.



The left Mode is displayed when Fault1 sensor is not equipped.

PAGE 26 / 44

GTD2000-Tx Instruction Manual

11.3. PROGRAMMABLE MODE Setting Touching the "RESET" switch in Program Mode screen will revert to Measuring Condition; and touching "RESET" switch in each Program Setting screen will revert to “PROGRAM MODE”. 

PASSWORD [**]

PROGRAM MODE

Touching "FUNC" switch with Magnet-bar for 2 seconds in

uring

Condition will enter the Password mode. 

Touch the "FUNC" switch after setting password using "↑" or "↓" switch.



It will enter Program Mode if the password is right.



The Mode will change in the specified order on every touching of "↑" or "↓" switch. (PROGRAM -> CALIABRA. -> ALARM -> SENSOR -> MAINTEN. > DEVICE -> VERSION -> TEST)



GAS TYPE [ DEFIN. ]

It is Gas Type Setting Mode: Gas Type is changed every time “↑” or “↓” switch is touched (DEFIN. / USER).



DEFIN. Shorts for Define, and is selected to use set Gas Type. USER is selected for the user to set the gas Type directly.



Touching “FUNC” switch when desired Gas Type is displayed will define the Gas Type and enter the next Program Mode.

GAS SEL [ COMB. ]



It is Gas name Setting Mode: Gas name is changed every time “↑” or “↓” switch is touched (DEFIN. / USER).



When Gas Type is set to DEFIN., Gases listed in Table 8 may be selected for use; when Gas Type is set to USER, the gas name is defined using 5

USER GAS

character keys. Numbers, uppercase alphabetic letters, space, and dot can be

[■SER ]

used. If the location moves to input letter, black box curser is displayed like the Figure.

UNIT SEL [ %LEL ]



Gas Measuring Unit Mode. Gas measuring unit is changed with touching of "↑" or "↓" switch each time (% /% LEL / PPM / PPB).



Touching "FUNC" switch will set the displayed Gas Measuring Unit desired, and will enter the next Program item.

D-POINT [ 100]



Decimal point setting Mode. The decimal point is changed on touching of "↑" or "↓" switch each time (0.100/1.00/10.0/100).



The desired and displayed decimal point is defined with the touch of "FUNC" switch; and the next Program item will be entered.

HIGH SCL [ 100]



High Scale setting Mode to be displayed in Full Range. The Scale value is increased or decreased on touching of “↑” or “↓” switch (1 ~ 10000).



Touching “FUNC” switch will define the displayed and desired High scale and enter the next Program.

PASSWORD [00]



Password setting Mode. Password is checked when entering Parameter Program Mode or Maintenance Mode.



"↑" switch or "↓" switch is used to set the password; touching "FUNC" switch will define the password and enter the next Program Item

PAGE 27 / 44

GTD2000-Tx Instruction Manual

11.4. CALIBRATION MODE Operation Due to the nature of the Gas detector, stabilization time of at least 30 minutes is required after power supply; management criteria may vary depending on the field condition.

11.4.1. Zero Calibration 

PASSWORD [**]

CALIBRA. MODE

Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.



Select “CALIBRA. MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “CALIBRA. MODE” is displayed will enter Calibration Mode.



CALIBRA. [ZERO]

Touching "FUNC" switch with Magnet-bar for 2 seconds in



Measuring Condition will return on touching “RESET” switch. Touching “FUNC” switch when “CALIBRA. MODE” is displayed will select Calibration Mode.



Select [ZERO] by touching "↑" or "↓" switch. Now, touch “FUNC” switch to enter Zero Calibration mode.

ZERO CAL [ NO]

ZERO GAS [ 0]

ZERO >> [SUCESS] ZERO >> [ FAIL ]

CAL.DATA [ 0]



Select [YES] by touching "↑" or "↓" switch. Now, touch “FUNC” switch to carry out Zero Calibration.



Inject clean air or 100% nitrogen gas into the sensor using a calibration device at 1000mL/min for a minute; touch “FUNC" Switch when Measuring Value was stabilized to carry out Zero calibration automatically.



After a successful Zero calibration, "ZERO SUCCESS" will be displayed on LCD Display for 2 seconds, and the mode will switch to "CALIBRATION DATA" Mode.



When Zero Calibration does not succeed, "ZERO FAIL" is displayed for two seconds, and it will switch to "CALIBRATION DATA" Mode.



ZERO FAIL happens when sensor input exceeds 70% of ADC input range.



It’s the Mode displaying Measuring Value after Calibration and the function to check if the calibration was performed normally.



If Failed, FAIL Text and current Measuring Value will be displayed at 1-second intervals.



"CALIBRATION MODE" returns on touching “RESET" Switch.

PAGE 28 / 44

GTD2000-Tx Instruction Manual

11.4.2. Span Calibration ※ Entering Calibration Mode is same as Zero Calibration.

CALIBRA. MODE

CALIBRA. [SPAN]

SPAN CAL [ NO]

SPAN GAS [ 50] [ [

50] 50]

SPAN SET [ 50] SPAN SET LOW FAIL SPAN SET HIGH FAIL



Select “CALIBRA. MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “CALIBRA. MODE” is displayed will enter Calibration Mode.



Measuring Condition will return on touching “RESET” switch.



Select [SPAN] by touching "↑" or "↓" switch. Now, touch “FUNC” switch to enter Span Calibration mode.



Select [YES] by touching "↑" or "↓" switch. Now, touch “FUNC” switch to carry out Span Calibration.



Inject standard gas into the Sensor using a calibration device at 500mL/min for 90 sec; touch “FUNC" Switch when Measuring Value was stabilized to enter the next Mode.



When HOLD function of Maintenance Mode is ON, the maximum value of the current SPAN gas is HOLD, and the HOLD value is displayed in the 1 st row.



In the 2nd row is the current Measuring Value displayed.



It’s the Mode to set the standard gas value, which is set by touching "↑" Switch or "↓" Switch if there’s no Fail message.



Fail message is displayed as follows when injected standard gas value is not normal. Fail message and Span set value are displayed alternately.



"LOW FAIL" occurs when the difference of injected Gas value and ZERO Calibration value is less than 1%.



"HIGH FAIL" occurs when injected Gas value is greater than 95% of the entire ADC input range.



"RNG FAIL" occurs when injected Gas value is operating as currently set SPAN value and exceeds more than 95% of the ADC's input range. The FAIL message is lifted when the normal operation resumes through SPAN value adjustment.

SPAN >> [SUCESS] SPAN >> [ FAIL ]

CAL.DATA [ 0]



SPAN calibration is performed after setting standard gas value and touching “FUNC" Switch. If the automatic SPAN calibration is successful, "SPAN SUCCESS” is displayed on LCD Display for 2 sec, and the mode is converted to “CAL DATA” Mode.



When Zero Calibration does not succeed, "SPAN FAIL" is displayed for two seconds, and it will switch to "CAL DATA" Mode.



It’s the Mode displaying Measuring Value after Calibration and the function to check if the calibration was performed normally.



"CALIBRATION MODE" returns on touching “RESET" Switch.

PAGE 29 / 44

GTD2000-Tx Instruction Manual

11.5. ALARM mode setting 

PASSWORD [**]

ALARM MODE

Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.



Select “ALARM MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “ALARM MODE” is displayed will enter Alarm setting Mode.



ALM USED [ OFF]

AL LATCH [OFF]

Touching "FUNC" switch with Magnet-bar for 2 seconds in

Measuring Condition will return on touching “RESET” switch.



An item that turns ON / OFF Alarm Mode setting.



Changes ON / OFF status by touching "↑" Switch or "↓" Switch. When it’s ON, Alarm function and Alarm Mode can be set.



When it’s ON only, the GTL100 Explosion-proof Warning Light can be used.



"FUNC" Switch shall be touched to enter into the next mode.



It’s a Mode setting Reset method after Alarm1 action; touching the ↑ "Switch or" ↓ "Switch toggles” ON "and" OFF ".



"OFF" setting automatically resets Alarm; "ON" setting resets Alarm only when Reset Switch is ON.



When the desired MODE is played, it is set by touching "FUNC" Switch; and the next Alarm setting item is entered.

A1 LEVEL [ 20]



Alarm1 level setting Mode. The Alarm1 level is increased or decreased on touching of “↑” or “↓” switch (1 ~ 10000).



Touching “FUNC” switch will define the Alarm1 level when desired Alarm1 level is displayed.

A1 TYPE [INC]



This MODE sets the direction of Alarm1 Mode; touching the "↑" Switch or "↓" Switch toggles the display of "INC" or "DEC".



"INC" Mode is operating when Alarm value is greater than or equal to Alarm setting value; "DEC" Mode is operating when Alarm value is less than

A1 TYPE [DEC]

or equal to Alarm setting value; 

The desired Mode is set by touching "FUNC" Switch when the desired Mode is displayed; and the next item is entered.

A1 DBAND [ 1.0]%F



The Mode sets Dead band where Alarm1 operates; the value is set using "↑" or "↓" key.



Alarm1 is operated at Alarm1 level plus Dead band value or greater; Alarm1 is released at Alarm1 level minus Dead band value or less.



Touching the "FUNC" Switch will set Alarm1 value when desired Dead band is displayed. The next item will be entered.

PAGE 30 / 44

GTD2000-Tx Instruction Manual

A1 TIME 1 SEC



It’s the function to prevent the transient malfunction of Alarm1 due to external shock and noise. Time can be set within the range of 0 ~ 60sec.



Alarm1 delay is increased or decreased by 1 sec at every push of "↑" or "↓" key.



When the desired Alarm1 delay time is displayed, press the "FUNC" key to set the Alarm1 delay time and to enter next item.



Example) At alarm set value=20%, and LEL / Delay time=5 Sec, an alarm occurs when the Measuring Value exists 5 sec more than the alarm setting value based on 20% LEL. No alarm will occur less than 5 sec.

A2 LEVEL [ 40]



Alarm2 level setting Mode. The Alarm2 level is increased or decreased on touching of “↑” or “↓” switch.



Touching “FUNC” switch will define the Alarm2 level when desired Alarm2 level is displayed.

A2 TYPE [INC]



This MODE sets the direction of Alarm2 Mode; touching the "↑" Switch or "↓" Switch toggles the display of "INC" or "DEC".



"INC" Mode is operating when Alarm value is greater than or equal to Alarm setting value; "DEC" Mode is operating when Alarm value is less than

A2 TYPE [DEC]

or equal to Alarm setting value; 

The desired Mode is set by touching "FUNC" Switch when the desired Mode is displayed; and the next item is entered.

A2 DBAND [ 1.0]%F



The Mode sets Dead band where Alarm2 operates; the value is set using "↑" or "↓" key.



Alarm2 is operated at Alarm2 level plus Dead band value or greater; Alarm2 is released at Alarm2 level minus Dead band value or less.



Touching the "FUNC" Switch will set Alarm2 value when desired Dead band is displayed. The next item will be entered.

A2 TIME 1 SEC



It’s the function to prevent the transient malfunction of Alarm2 due to external shock and noise. Time can be set within the range of 0 ~ 60sec.



Alarm2 delay is increased or decreased by 1 sec at every push of "↑" or "↓" key.



When the desired Alarm2 delay time is displayed, press the "FUNC" key to set the Alarm2 delay time and to enter next item.



Example) At alarm set value : 20%, and LEL / Delay time : 5 Sec, an alarm occurs when the Measuring Value exists 5 sec more than the alarm setting value based on 20% LEL. No alarm will occur less than 5 sec.

PAGE 31 / 44

GTD2000-Tx Instruction Manual

11.6. SENSOR DATA MODE configuration It’s the Mode displaying current Sensor value and Calibration condition. Setting is not possible. 

PASSWORD [**]

SENSOR MODE

Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.



Select “SENSOR MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “SENSOR MODE” is displayed will enter Alarm setting Mode.

 SEN. OUT 10.0mV

SEN. S/T 0.1mV

MZ 1.0 MS -100.0

AZ AS

ZR SR

1.0 -100.0

49.9% 60.0%

SEN GAIN [ 1.0 ] x G

SEN BIAS [ 3.18V ]

TEMPER [ 25’C ]

CURRENT [176mA]

24V VIN [24.0V]

Touching "FUNC" switch with Magnet-bar for 2 seconds in

Measuring Condition will return on touching “RESET” switch.



Displays sensor voltage value being currently measured.



Touching "FUNC" Switch will enter the next item.



It displays the difference of current Sensor Measuring Value and ZERO Calibrated Measuring Value.



The unit of the displayed value is mV.



Touching "FUNC" Switch will enter the next item.



It displays the normal Calibration ZERO and SPAN Measuring Value.



The unit of the displayed value is mV.



Touching "FUNC" Switch will enter the next item.



It displays the automatic Calibration ZERO and SPAN Measuring Value.



The corresponding values are automatically updated in Regular Calibration.



The unit of the displayed value is mV.



Touching "FUNC" Switch will enter the next item.



This Mode displays the percentage variation of current Calibration voltage and Sensor ADC maximum voltage.



Touching "FUNC" Switch will enter the next item.



Output Mode of Sensor Output values multiplied by the corresponding setting value.



The range of Setting is from 0.1 to 5.0.



This Mode displays the sensor bias applied to the sensor.



Touching "FUNC" Switch will enter the next item.



This Mode displays the temperature measured in the Sensor.



Touching "FUNC" Switch will enter the next item.



This Mode measures the current consumption in the Sensor.



Touching "FUNC" Switch will enter the next item.



It measures the Sensor input 24V power supply.



SENSOR MODE will return on touching of “FUNC” Switch.

PAGE 32 / 44

GTD2000-Tx Instruction Manual

11.7. Maintenance Mode setting ※

Only authorized personnel are allowed to perform these settings. 

PASSWORD [**] MAINTEN. MODE

CRO-SEN. [1.00]xG Z-SKIP [ 0.0]%F

Touching "FUNC" switch with Magnet-bar for 2 seconds in Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.



Select “Maintenance MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “MAINTEN. MODE” is displayed will enter Maintenance setting Mode.



Measuring Condition will return on touching “RESET” switch.



It sets the Cross sensitivity (0.1 ~ 5.0) of the sensor.



Value setting is done by touching "↑" Switch or "↓" Switch by 0.01 units.



Touching "FUNC" Switch will enter the next item.



It sets the Zero region sensitivity of the sensor.



Value setting is done by touching "↑" Switch or "↓" Switch by 1 unit.



The gas value less than corresponding setting is considered as 0; the setting is possible up to 10% of the High Scale.

ODT [ 0]SEC

AUTO-Z [ ON]



Touching "FUNC" Switch will enter the next item.



It sets the Measuring Value delay depending on the set time.



It sets whether to use Auto Zero function.



Touching the “↑ “or “↓ “Switch toggles” ON "and" OFF ". Auto zero function operates if it is ON (default is ON).



If Auto Zero is ON, automatic Zero Calibration is performed when the value is maintained over 10 minutes less than 0.5% of variation within the values from Auto Zero minimum to 5% of the High Scale.

AZ MIN. [2.0]%F BASE-Z [ ON]



It sets the minimum value of Auto Zero.



This Mode sets the percentage of High Scale in the range of 1.0%~4.0%.



Touching "FUNC" Switch will enter the next item.



It sets whether to use Base Zero function.



Touching the “↑ “or “↓ “Switch toggles” ON "and" OFF ". Base zero function operates if it is ON (default is ON).



If Base Zero is ON, automatic Zero Calibration is performed when the value is maintained over 10 minutes from 0.02% of the High Scale to Base Zero maximum.



Touching "FUNC" Switch will enter the next item.

PAGE 33 / 44

GTD2000-Tx Instruction Manual

BZ MAX. [2.0]%F

SP. HOLD [ ON]



It sets the minimum value of Base Zero.



This Mode sets the percentage of High Scale in the range of 0.5%~2.0%.



Touching "FUNC" Switch will enter the next item.



It sets whether to use the function to HOLD the maximum value during SPAN Calibration.



Touching the “↑ “or “↓ “Switch toggles” ON "and" OFF ". Measuring Value HOLD function operates during Span Calibration if it is ON.

UNDER EN [OFF]



Touching "FUNC" Switch will enter the next item.



It turns ON/OFF the Under Mode setting.



Touching the “↑ “or “↓ “Switch toggles” ON "and" OFF ". Under Mode is set when less than -10% of gas is measured if it is ON.

SEN CHK. [OFF]



If it is OFF, values under -10% will be processed as 0.



Touching "FUNC" Switch will enter the next item.



It sets whether to perform Combustible sensor check.



Touching the “↑ “or “↓ “Switch toggles” ON "and" OFF ". Whether the sensor is installed is automatically checked if it is ON. If it does not detect a sensor, Fault will be indicated.

S-DIR [INC] S-DIR [DEC] ENG MOD. [OFF]



If it is OFF, sensor check is disabled.



Touching "FUNC" Switch will enter the next item.



This item sets the direction of Sensor operation.



Touching the "↑" Switch or "↓" Switch toggles the display of "INC" or "DEC". On INC, the Sensor operates to the direction of voltage/current increase; on DEC, to the direction of decrease.



Touching "FUNC" Switch will enter the next item.



This item sets the display function of negative values of Gas Measuring Value.



Touching the "↑" Switch or "↓" Switch toggles the ON/OFF status, If it is ON, negative value of Gas Measuring Value is displayed; OVER and UNDER Mode is not indicated; leftmost character in the 2 nd row is expressed as ‘<’.

W / L TYPE [STEADY ]



MAINTENANCE MODE returns on touch of “FUNC” Switch.



This item is for setting the lighting method of the Warning Light in Normal condition.



Green light is on with “STEADY”; green light will flash at 1 second interval with “BLINK.”

M.- LEVEL [ 0 ] EMC T/O [ OFF ]



This item is for setting the current Output in Test Mode.



This item is for setting whether to set time in Test Mode.



If it is ON, Test Mode is activated for 30 minutes only; if it is OFF, it is operated without restriction.

PAGE 34 / 44

GTD2000-Tx Instruction Manual

11.8. Device Mode setting ※

Only authorized personnel are allowed to perform these settings. 

PASSWORD [**]

Touching "FUNC" switch with Magnet-bar for 2 seconds in Measuring Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.

DEVICE. MODE

HART B/D [ CHECK]

PADD / mA 0 / EN.

DEV-CODE [0xE1C3]

FIX CUR DISABLE

SERIA.NO *XXXXXXX

TAG GTD-0001

LONG TAG [GTD-00>

DESCRIP. [GASTRO>

MESSAGE [GASTRO>



Select “DEVICE MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “DEVICE MODE” is displayed will enter DEVICE setting Mode.



Measuring Condition will return on touching “RESET” switch.



This item is to check the connection status of HART BOARD.



CHECK is displayed if connected; EMPTY if not.



This item is to check the Polling Address of HART device.



This item is to check the Unique ID of the HART device.



Sensor ID cannot be changed since it is a unique ID of the product itself.



This item is to check the Fixed Current Mode of the HART device.



It is the Mode to check the Serial Number of the product.



This item is to check the Tag of the HART device.



Modification is not possible in the Sensor.



This Mode is to check the Long Tag of the HART device.



It is possible to check by shifting the string using "↑" Switch or "↓" Switch.



Modification is not possible in the Sensor.



This Mode is to check the Descriptor of the HART device.



Modification is not possible in the Sensor.



This Mode is to check the Message of the HART device.



Modification is not possible in the Sensor.

PAGE 35 / 44

GTD2000-Tx Instruction Manual

11.9. Version mode configuration This Mode is to display the important revision information of the interior of the equipment. 

PASSWORD [**]

Touching "FUNC" switch with Magnet-bar for 2 seconds in Measuring Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.

VERSION MODE

FW REV [REV 6]

EX H/W [REV 2]

HART DEV [REV 1]

HART REV [REV 7]



Select “VERSION MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “VERSION MODE” is displayed will enter VERSION setting Mode.



Measuring Condition will return on touching “RESET” switch.



It indicates the current version of the F/W.



It indicates the version of the H/W.



It indicates the version of the employed HART Device.



It indicates the version of the employed HART Protocol.



VERSION MODE will return on touching “FUNC” switch.

PAGE 36 / 44

GTD2000-Tx Instruction Manual

11.10. Test mode configuration 

PASSWORD [**]

TEST MODE

mA OUT [OFF]

[ TEST ] [ 0]

Touching "FUNC" switch with Magnet-bar for 2 seconds in Condition will enter the Password mode.



Touch the "FUNC" switch after setting password using "↑" or "↓" switch.



Select “TEST MODE” by touching "↑" or "↓" switch.



Touching “FUNC” switch when “TEST MODE” is displayed will enter TEST setting Mode.



Measuring Condition will return on touching “RESET” switch.



This item sets whether to include mA Output during the TEST.



Touching the “↑” or “↓” Switch will toggle the ON/OFF status; if it is ON, mA will be output by the setting value during TEST.



This item performs the TEST.



Test Gas value is set by touching “↑” or “↓” Switch; if the mA Output is ON, the mA Output is displayed coupled with Test Gas value.



TEST MODE will return on touching “FUNC” switch.

PAGE 37 / 44

GTD2000-Tx Instruction Manual

11.11. Inspection mode use This Mode is used for the inspector to identify the Sensor status and Fault details without affecting the equipment operated in emergency. Only authorized personnel are allowed to use this Mode.  [ #

%LEL] 0 ]

Press and hold the "↓" Switch for at least 3 seconds while Means value display gas concentrations in the display. Press and hold the "↓" Switch for at least 3 seconds to exit again.



Entering Inspection Mode, '#' mark will flash at the first place in the second row.



You can enter M/L item from Maintenance Mode to set Output value.

11.12. Data initialization Only authorized personnel are allowed to conduct this Mode because it will initialize all the values to the data set in the factory before shipping.

ALL - INIT [ YES ]



Turn the power ON while holding "FUNC" Key and "UP" Key.



When "ALL - INIT" is displayed on the Display window, select "YES" to carry out the data initialization.

11.13. Correction data initialization Only authorized personnel is allowed to conduct the initialization because this Mode initializes to the value of Calibration data set in the factory before shipping. This Mode is used for the inspector to initialize only the Calibration value among setting values.

CAL - INIT [ YES ]



Turn the power ON while holding "FUNC" Key and "UP" Key.



When "CAL - INIT" is displayed on the Display window, select "YES" to carry out the normal initialization of Calibration data.

PAGE 38 / 44

GTD2000-Tx Instruction Manual

12. Troubleshooting Fault code / Output

Description & Condition

Recovery

Message FAULT1 “SEN EMPT” FAULT2 “SEN HIGH” FAULT3 “SEN LOW” FAULT4 “EROM ERR” FAULT5 “+24V LOW” FAULT6 “ADC FAIL” FAULT7 “H/W REV”

Occurs

when

Combustible

sensor

module

is

connected.

not

sensor module connect fault

When Combustible sensor Output is over the maximum sensor module fault or ADC value.

transmitter Board ADC fault

When Combustible sensor Output is under the minimum sensor module fault or ADC value.

transmitter Board ADC fault

transmitter EEPROM Checksum fault

transmitter Board EEPROM fault Power supply input voltage

Occurs when 24V main power input is less than 10V.

confirm & transmitter ADC fault

transmitter ADC fault

transmitter ADC fault Check

H/W version error

resistance

PAGE 39 / 44

transmitter

ADC

GTD2000-Tx Instruction Manual

13. Caution before installation 13.1. Selection of installation location (Occupational Safety and Health Law) The Gas leak detection alarm system shall be installed in such place as follows. : 1) Near chemical accessory equipment installed inside/outside of a building and susceptible of gas leak such as compressors, valves, reactors, and piping connections, etc. dealing with

combustible

and toxic materials. 2) Locations risky of remaining gas near manufacturing equipments with ignition source like heaters. 3) Around connections of filling equipments of combustible and toxic substances. 4) Substations, distribution panel rooms, control rooms, etc. near explosion-proof area. 5) Other special gas-friendly places.

13.2. Selection of installation location (High Pressure Gas Safety Management Regulations) The Gas detector of the Gas leak detection alarm system shall be installed close to risky area of gas leakage. However, if the direct gas leaks are not expected, but gas residence is vulnerable, it should be installed in such places as follows.

1)

A gas leak detection alarm outside of a building shall be installed in a risky place of gas residence considering the wind direction, wind speed, and the gravity of the gas.

2)

A gas leak detection alarm inside of a building shall be installed in the lower part of the building if the gas is lighter than the air, and upper part or near the vent of the building.

3)

The alarm of the Gas leak detection alarm system shall be installed near Gas detector or in places where workers usually are.

PAGE 40 / 44

GTD2000-Tx Instruction Manual

13.3. Precaution before installation Rainwater shall be avoided because it can be an electrical hindrance, and accessibility should be considered for periodic maintenance before installation. Vibration or shock shall be avoided since it may affect the output value, and the sensor shall face the direction of gravity when installed. 

This device has high pressure explosion-proof structure; belongs to GROUP II targeting gas and steam from general workplaces and chemical plants; and can be used in hazardous places of ZONE 1 (ONE) –class 1 and ZONE 2 (TWO) –class 2.



Allowable temperature belongs to 85 ℃ or lower, which corresponds to T6.



The ambient temperature shall be in the range of -20 ℃ ~ 60 ℃.



Installation elevation: less than 1,000 M above sea level



Relative Humidity: 5-99%



Installation place: indoors or outdoors



Explosive ignition temperature of the gases or vapors used: Ex d IIC T6



The wire conduit shall be sealed to prevent the gas moving or the explosion flame propagation through the conduit under 45cm when the explosion-proof cable gland is used at the cable inlet or when metal conduit is used in wiring works.



At least 5 screw threads must be used for connection of this device and the conduit.



Other standards should be met in this work such as:

[Standards on the selection, installation

and maintenance of wiring for workplace explosion-proof structural electrical mechanism]. 

Only qualified materials shall be used in cable entry including CABLE GLAND and SEALING FITTING; and used in the closure of unused incoming part.

[Figure 13. Pressure packing type ]

[Figure 14. Y Sealing Compound ]

PAGE 41 / 44

GTD2000-Tx Instruction Manual

14. Ordering Information GAS NAME Acetic Acid Ammonia Antimony Pentachloride Arsetic Tafluoride Arsetic Pentafluoride Arsenic Tfichloride Arsenic Pentachloride Arsine Boron Trichloride Boron Tribromide Boron Trifluoride Bromine Chlorine Carbon Tetrachloride Carbon Monoxide Chlorine Tetrafluoride Diborane Dichlorosilane DIsilane Ethylene Oxide Fluorine Germane Germanium Tetrachloride Hydrazine Hydrogen Hydrogen Bromide Hydrogen Chloride Hydrogen Cyanide Hydrogen Fluoride Hydrogen Iodine Hydrogen Selenide Hydrogen Sulfide Iodine Isopropyl Alcohol(IPA) Molybdenum Fluoride Nitric Acid Nitrogen Monoxide Nitrogen Dioxide Nitrogen Trifluoride Nitrogen Tetraoxide Oxygen Ozone

CH3COOH NH3 SbCℓ5 AsF3 AsF5 AsCl3 AsCl5 AsH3 BCℓ3 BBr3 BF3 Br2 Cℓ2 CCℓ4 CO CℓF3 B2F6 SiH2Cℓ2 Si2H6 C2H4O F2 GeH4 GeCl4 N2H4 H2 HBr HCℓ HCN HF Hi H2Se H2S I2 CH3CHOHCH3 MoF6 HNO3 NO NO2 NF3 N2O4 O2 O3

PAGE 42 / 44

MESURING RANGE

TLV-TWA

0 ~ 30 ppm 0 ~ 75 ppm 0 ~ 15 ppm 0 ~ 9 ppm 0 ~ 9 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0 ~ 0.3 ppm 0 ~ 15 ppm 0 ~ 9 ppm 0 ~ 9 ppm 0 ~ 1 ppm 0 ~ 3 ppm 0 ~ 30 ppm 0 ~ 150 ppm 0 ~ 1 ppm 0 ~ 0.3 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0~30ppm 0 ~ 3 ppm 0 ~ 2 ppm 0 ~ 15 ppm 0 ~ 10 ppm 0 ~ 2000 ppm 0 ~ 9 ppm 0 ~ 15 ppm 0 ~ 30 ppm 0 ~ 9 ppm 0 ~ 5 ppm 0 ~ 0.2 ppm 0 ~ 30 ppm 0 ~ 1 ppm 0 ~ 2000 ppm 0 ~ 9 ppm 0 ~ 20 ppm 0 ~ 100 ppm 0 ~ 15 ppm 0 ~ 30 ppm 0 ~ 15 ppm 0 ~ 25% Volume 0 ~ 1 ppm

10 ppm 25 ppm 5 ppm 3 ppm 3 ppm 5 ppm 5 ppm 0.05 ppm 5 ppm 3 ppm 3 ppm 0.1 ppm 1 ppm 5 ppm 25 ppm 0.1 ppm 0.1 ppm 5 ppm 5 ppm 1ppm 1 ppm 0.2 ppm 5 ppm 0.01 ppm LEL=4%VOL 3 ppm 5 ppm 10 ppm 3 ppm 2 ppm 0.05 ppm 10 ppm 0.1 ppm 400 ppm 3 ppm 2 ppm 25 ppm 3 ppm 10 ppm 3 ppm 0.1 ppm

GTD2000-Tx Instruction Manual

MESURING RANGE

GAS NAME

TLV-TWA

Phosgene

COCℓ2

0 ~ 0.3 ppm

0.1 ppm

Phosphine Phosphorus Oxychloride PhosPhorus Pentafluoride Phosphorus Trichloride Silane Silicon Tetrachloride Silicon Tetrafluoride Sulfur Dioxide Sulfur Tetrafluoride Sulfur Hexafluoride Tantalum Fluoride Tetraethyl Orthosilicate Tin Tetrachloride Titanium Fluoride Titanium Tetrachloride Trichlorosilane Trimetoxy Phosphate Tungsten Hexafluoride

PH3 POCℓ3 PF5 PCℓ3 SiH4 SiCℓ4 SiF4 SO2 SF4 SF6 TaF5 TEOS SnCℓ4 TiF4 TiCℓ4 SiHCℓ3 P(OCH3)3 WF6

0 ~ 1 ppm 0 ~ 15 ppm 0 ~ 9 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0 ~ 9 ppm 0 ~ 10 ppm 0 ~ 9 ppm 0 ~ 2000 ppm 0 ~ 9 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0 ~ 9 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0 ~ 15 ppm 0 ~ 9 ppm

0.3 ppm 5 ppm 3 ppm 5 ppm 5 ppm 5 ppm 3 ppm 2 ppm 3 ppm

PAGE 43 / 44

3 ppm 10 ppm 5 ppm 3 ppm 5 ppm 5 ppm 2 ppm 3 ppm

GTD2000-Tx Instruction Manual

15. Revision history Version 0 1

Contents

Date

* Initial revision of the Manual

JUL. 31 , 2012

* Spec. revised(Added power consumption),

JAN 23, 2013

Revised company address.

END

This product and instruction manual are subject to change without prior notice for the improvement of product performance and ease of use.

Gastron Co.,Ltd. Address: 18-8 Dogeumdanji 1gil(Palgogi-dong), Sanglok-gu, Ansan-city, Gyeonggi-do, Korea, Tel : 031-490-0800 http://www.gastron.com

PAGE 44 / 44

Fax : 031-490-0801 [email protected]

Multi-gas Monitor Product Manual Set-up Operation Service

Part Number: 17152357-1 Version 11

1

Ventis™ MX4 Product Manual

Table of Contents COPYRIGHT NOTICE ................................................................................................................................................... 3 WARNINGS AND CAUTIONARY STATEMENTS ........................................................................................................ 3 General ................................................................................................................................................................... 3 Personnel ............................................................................................................................................................... 3 Hazardous Conditions, Poisons, and Contaminants .............................................................................................. 3 General Usage ....................................................................................................................................................... 4 Agency-issued Conditions of Use and Warnings .................................................................................................... 4 Recommended Practices ....................................................................................................................................... 5 VENTIS MX4™ RESOURCES....................................................................................................................................... 6 VENTIS MX4 CAPABILITIES ........................................................................................................................................ 6  

UNPACKING THE MONITOR ....................................................................................................................................... 7 Contents.................................................................................................................................................................. 7 Reporting a Problem .............................................................................................................................................. 7 MONITOR OVERVIEW ................................................................................................................................................. 8 Hardware Features and Functions .......................................................................................................................... 8 Display Screen ........................................................................................................................................................ 9 Alarms ................................................................................................................................................................... 11 MONITOR SET-UP ...................................................................................................................................................... 13 Battery Properties and Monitor Compatibility ....................................................................................................... 13 Charging the Lithium-ion Battery Packs ............................................................................................................... 14 Instruction ...................................................................................................................................................... 14 Power-on and –off................................................................................................................................................. 15 Configuration......................................................................................................................................................... 16 Introduction .................................................................................................................................................... 16 Instructions .................................................................................................................................................... 16 Configuration Process.................................................................................................................................... 17 MONITOR USE AND SERVICE .................................................................................................................................. 24 Zero, Calibration, and Bump Testing .................................................................................................................... 24 Procedures .................................................................................................................................................... 24 Recommendations ......................................................................................................................................... 25 General Information .............................................................................................................................................. 25 Instructions ................................................................................................................................................... 26 Recommended Practices for In-field Air Sampling ............................................................................................... 32 Cleaning ............................................................................................................................................................... 32 Service .................................................................................................................................................................. 33 Battery Packs ................................................................................................................................................. 33 Monitor Conversion ....................................................................................................................................... 36 Sensor, Sensor Water Barrier, LCD, and Vibrating Motor Replacement........................................................ 38 Pump Module ................................................................................................................................................. 40 Ventis MX4 Three-Dimensional View Diagrams ............................................................................................ 41 PRODUCTS, SPECIFICATIONS, AND CERTIFICATIONS ........................................................................................ 44 Ventis MX4 Accessories and Parts ....................................................................................................................... 44 Monitor Specifications ........................................................................................................................................... 46 Operating Conditions ............................................................................................................................................ 46 Storage Conditions ............................................................................................................................................... 47 Sensor Specifications ........................................................................................................................................... 47 Toxic Gas Sensor Cross-Sensitivity Table ............................................................................................................ 47 LEL and LEL Correlation Factors for Combustible Gases .................................................................................... 48 Certifications ......................................................................................................................................................... 49 WARRANTY ................................................................................................................................................................ 50 Limitation of Liability .............................................................................................................................................. 50 INDUSTRIAL SCIENTIFIC CORPORATION GLOBAL LOCATIONS..................................................... BACK COVER

© 2014Industrial Scientific Corporation

2

Ventis™ MX4 Product Manual

►Copyright

Notice

Ventis MX4™ and Ventis™ are trademarks of Industrial Scientific Corporation. All trademarks and registered trademarks are the property of their respective owners. These help materials or any part thereof may not, without the written consent of Industrial Scientific Corporation, be copied, reprinted, or reproduced in any material form including but not limited to photocopying, transcribing, transmitting, or storing it in any medium or translating it into any language, in any form or by any means, be it digitally, electronic, mechanical, xerographic, optical, magnetic, or otherwise. The information contained in this document is proprietary and confidential and all copyright, trademarks, trade names, patents, and other intellectual property rights in the documentation are the exclusive property of Industrial Scientific Corporation unless otherwise specified. The information (including but not limited to data, drawings, specification, documentation, software listings, source or object code) shall not at any time be disclosed directly or indirectly to any third party without prior written consent. The information contained herein is believed to be accurate and reliable. Industrial Scientific Corporation accepts no responsibility for its use by any means or in any way whatsoever. Industrial Scientific Corporation shall not be liable for any expenses, costs by damage that may result from the use of the information contained within this document. Although every effort is made to ensure accuracy, the specifications of this product and the content herein are subject to change without notice.

►Warnings and Cautionary Statements General IMPORTANT Failure to perform certain procedures or note certain conditions may impair the performance of this product. For maximum safety and optimal performance, please read and understand the Ventis MX4 Product Manual available online at the Ventis MX4 Resource Center at www.indsci.com/ VentisMX4resources.

Personnel CAUTION: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the product manual completely before operating or servicing.

Hazardous Conditions, Poisons, and Contaminants WARNING: Servicing the unit, replacing or charging battery packs, or using the communications port must only be done in an area known to be nonhazardous. Not for use in oxygen-enriched atmospheres. WARNING: Power-off the monitor before servicing the unit or replacing the battery. WARNING: Substitution of components may impair intrinsic safety and may cause an unsafe condition. CAUTION: High off-scale readings may indicate explosive gas concentration(s). CAUTION: Any rapid up-scale reading followed by a declining or erratic reading may indicate gas concentration(s) beyond the upper scale limit which may be hazardous. Silicone compound vapors or other known contaminants may affect the combustible gas sensor and cause readings of combustible gas to be lower than actual gas concentrations. If the monitor has been used in an area where silicone vapors were present, always calibrate the monitor before next use to ensure accurate measurements.

3

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

General Usage Oxygen-deficient atmospheres may cause combustible gas readings to be lower than actual concentrations. Oxygen-enriched atmospheres may cause combustible gas readings to be higher than actual concentrations. Sudden changes in atmospheric pressure may cause temporary fluctuations in the oxygen reading. Verify the calibration of the combustible gas sensor after any incident where the combustible gas content has caused the monitor to display an over-range condition. Sensor openings, water barriers, and the pump inlet must be kept clean. Obstruction of the sensor openings or pump inlet and/or contamination of the water barriers may cause readings to be lower than actual gas concentrations. To avoid the potential of liquid being pulled into the sample tubing and pump assembly, it is recommended that Industrial Scientific filter (P/N 17027152) be used on the sample tubing when drawing samples using the aspirated monitor. WARNING: INSERT THE ALKALINE BATTERIES WITH THE CORRECT POSITIVE “+” AND NEGATIVE “-“ ORIENTATION. WARNING: The Ventis MX4 is only approved for use with AAA battery types Energizer EN92 and Duracell MN2400. Do NOT mix battery types. WARNING: The use of leather cases can produce inaccurate readings with diffusion (non-aspirated) gas detection instruments for specific monitoring applications. Leather cases should be used ONLY as carrying cases, and NOT for continuous monitoring, with diffusion instruments configured to measure gases other than O2, CO, CO2, H2S, and combustible gases (LEL/CH4).

Agency-issued Conditions of Use and Warnings Ensure all part-use restrictions (e.g., battery) meet any agency-mandated conditions of use. Ensure all instrument-configurable settings (e.g., always-on setting) meet any agency-mandated conditions of use. When using instrument-compatible Industrial Scientific docking stations, maintain mandated settings through the software (e.g., iNet Control or Accessory Software) or by manually configuring the instrument settings after docking. The Ventis MX4 is CSA certified according to the Canadian Electrical Code for use in Class I, Division 1 and Class I, Zone 1 Hazardous Locations within an ambient temperature range of Tamb: -20°C to +50°C. CSA has assessed only the %LEL combustible gas detection portion of this instrument for performance according to CSA Standard C22.2 No. 152. This is applicable only when the monitor is used in the diffusion mode and has been calibrated to 50% LEL CH4, and when the monitor is used in the aspirated mode with an extended range lithium-ion battery and has been calibrated to 50% LEL CH4. CAUTION: CSA C22.2 No. 152 requires before each day’s usage, sensitivity must be tested on a known concentration of pentane or methane equivalent to 25% or 50% of full scale concentration. Accuracy must be within -0% to +20% of actual concentration. Accuracy may be corrected by referring to the zero/calibration section of the Product Manual. The equipment complies with the standards IEC 60079-29-1 and EN 60079-29-1 for methane, propane, and hexane with the following exception: as for the methane (mine) detector, the battery run time was verified to be seven (7) hours rather than the eight (8) hours recommended by the standards, respectively. MED-certified instruments may be used only when configured and maintained to disallow power-off when the unit is in alarm. The Mine Safety and Health Administration (MSHA) has approved the Ventis MX4 as a Permissible Multi-Gas Monitor with the following warnings: 

MSHA approved for use with either the P/N 17134453-X2, 3.7 volt, lithium-ion battery or P/N 17148313-2, 3.7 volt, lithium-ion extended battery pack assembly only. The battery pack is not user-replaceable. The monitor battery and the lithium battery on the main PCB are technician replaceable only. Charge battery pack with an ISC battery charger designed for use with this monitor in fresh air locations only.

© 2014Industrial Scientific Corporation

4

Ventis™ MX4 Product Manual

  

The monitor is to be calibrated according to the procedures in the instruction manual only. The aspirated version of the Ventis MX4 is only approved for use with the extended battery pack. The monitor must display methane in the percent-by-volume mode (0-5%) for compliance determinations required by 30 CFR Part 75, subpart D.

SANS 1515-certified units may be used only as follows:  Diffusion applications  Configured and maintained to disallow power-off when the unit is in alarm  The Methane alarms are set as follows: low alarm = 1%vol and high alarm =1.4%vol  With approved Lithium-ion battery packs (see Ventis MX4 Accessories and Parts in this manual).

Recommended Practices Industrial Scientific Corporation recommends the monitor be fully charged (when equipped with a rechargeable battery pack), configured, and calibrated before first time use. If the lithium-ion battery is deeply discharged, it can take up to an hour for the instrument display to indicate that the battery is charging. Monitors used infrequently should be fully charged every four months. No part of the unit should be covered by any garment, part of a garment, or other item that would restrict the flow of air to the sensors or impair the operator's access to the audible, visual, or vibration alarms. Industrial Scientific Corporation recommends a full monitor calibration be performed monthly (at a minimum), using a certified concentration(s) of Industrial Scientific calibration gas(es) to help ensure monitor accuracy. Industrial Scientific Corporation recommends the monitor be zeroed and bump tested before each use with a certified concentration(s) of Industrial Scientific calibration gas(es). Battery contacts are exposed on battery packs when they are removed from the monitor. Do not touch the battery contacts and do not stack battery packs on top of one another. When reassembling the instrument or installing a battery pack, maintain ingress protection by tightening each fastener to its stated torque value (see the “Ventis MX4 Monitor three-Dimensional Diagram” and its key in this manual). Contact your service representative immediately if you suspect that the Ventis MX4 is working abnormally. Industrial Scientific recommends the “2 & 2 Sampling Rule” when sampling with a motorized pump and tubing, one should allow for 2 minutes plus 2 seconds per foot of tubing used, prior to noting the monitor readings. This allows time for the gas to reach the instrument and for the sensors to adequately react to any gases present. Industrial Scientific recommends that clear urethane tubing, part number 17065970, be used with the pumped versions of the Ventis MX4 when sampling for the following gases: Nitrogen Dioxide (NO2) and Sulfur Dioxide (SO2).

5

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

►Ventis

MX4 Resources

The Ventis MX4 Product Manual is the primary resource, within a full suite of learning tools, developed for the monitor user. Its step-by-step “walk through” format covers everything from unpacking to set-up, operation, and service. All Ventis MX4 users should read and understand the Product Manual prior to unpacking or using the monitor. A companion to the manual, the Ventis MX4 Reference Guide ships with the monitor. It serves to announce all warnings and cautionary statements relevant to general monitor use. The guide also features process charts that provide an overview of four fundamental tasks: operation/start-up, configuration, calibration, and functional “bump” testing. These charts are tools for the user who is both familiar with the manual and proficient in the performance of the given task. Ventis MX4 product-specific resources are part of the organization’s broader training line-up, featuring online training modules and face-to-face classroom programs for technicians, operators, first responders, trainers, and distributors. Courses combine theory with hands-on learning, and can be tailored to the customer’s unique requirements and gas monitoring applications. The organization’s customer and technical support call centers provide product and order information, how-to product assistance, and guidance for in-depth technical applications. Its service centers offer comprehensive factory repair and maintenance services. Industrial Scientific Corporation provides a full suite of resources to aid customers in the competent and safe use of its products and services. With 19 manufacturing, support, and service centers and hundreds of distributors worldwide, Industrial Scientific serves the globe’s gas detection needs.

►Ventis MX4 Capabilities The Ventis MX4 is a portable multi-gas monitor. Offered as a diffusion monitor, it detects and measures gas(es) present in open space. To enable monitor use within confined space locations, the Ventis MX4 is also offered as an aspirated monitor. A pump module and battery accessories enable the conversion of either monitor for dedicated use in either confined or open spaces. Based on the customer’s monitor order, up to four sensors are factory installed enabling the monitor to continuously and simultaneously detect and measure the presence of up to four specific gases. Sensor Category

Number available per Gases Monitored monitor

Oxygen

1

O2 (Oxygen) only

Combustible

1

Monitor can be configured for sensor to measure ONE of the following:  LEL (Pentane)  LEL (Methane)  CH4 (0%-5%)

Toxic

2

Each sensor detects and measures only ONE of the following:  CO (Carbon Monoxide)  CO/H2 Low (Carbon Monoxide with low H2 interference)  H2S (Hydrogen Sulfide)  NO2 (Nitrogen Dioxide)  SO2 (Sulfur Dioxide)

Equipped with a multi-mode (audible, visual, and vibration) and multi-level alarm system, the Ventis MX4 monitor is capable of notifying its user of potentially hazardous gas concentrations. The monitor performs continuous datalogging at 10 second intervals. It can store approximately 90 days of data for a four-sensor configuration. Its date- and time-stamped event log records and stores data for the following: 60 alarm events, 30 error events, and 250 manually performed calibrations or bump tests. The memory, when full, overwrites the oldest data as the newest readings and events are logged.

© 2014Industrial Scientific Corporation

6

Ventis™ MX4 Product Manual

The Ventis MX4 monitor functions as an independent device to monitor the environment for hazardous gas concentrations. It is also part of the Ventis System including products that charge, calibrate, bump test, datalink, protect, and otherwise enable or enhance use of the monitor and its data. (Datalink refers to capabilities that enable access to the download and use of monitor datalogs, reports, and other information.) For a complete list of the Ventis MX4 system of products, please refer to the manual section, Products and Parts.

►Unpacking the Monitor CONTENTS The monitor box contains the following items including, when ordered, those marked optional. Each item ordered should be accounted for in the unpacking process. Quantity

Item

Notes

1 as ordered

Ventis MX4 Portable Multi-gas Monitor

The monitor type is indicated on the box label. Options:  Ventis MX4 Diffusion  Ventis MX4 Aspirated  Ventis MX4 Aspirated with Conversion Kit

1

Ventis MX4 Reference Guide

A companion to the Ventis MX4 Product Manual.

1 installed as ordered

Battery Pack

One of three battery types is factory installed as indicated on the box label. Options:  Rechargeable Lithium-ion  Rechargeable Extended Range Lithium-ion  Alkaline

1 as ordered

Charger

Universal power cord. AC charger products include interchangeable plugs (US, UK, EU, and AUS).

0 or 1

Calibration Cup

Diffusion – 1 included Aspirated – 0 included

1

Calibration and Bump Test Tubing

Diffusion – two feet of clear tubing

0 or 1

In-field Sampling Tubing

Diffusion – 0 included Aspirated – Ten feet of black tubing

1

Final Inspection & Test Report

Contains the following factory set* information:  Monitor Set-up Date  Monitor Part Number (P/N)  Monitor Serial Number (S/N)  For Each Sensor*:  P/N  S/N  Type  Location  Alarm level values  Span gas values  Span reserve values *Some factory set sensor values subject to user changes.

1

Warranty Card

--

REPORTING A PROBLEM After unpacking, if any item is missing or appears to have been damaged, contact a local distributor of Industrial Scientific products or Industrial Scientific Corporation (for contact information, please see the manual’s last page).

7

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

►Monitor Overview HARDWARE FEATURES AND FUNCTIONS The monitor’s case top (front of monitor) has two main sections. As shown below, the upper section contains the sensor ports. The lower section houses the user interface features, a LCD display screen and two buttons. Each feature’s general functions are noted below. As shown, the diffusion and aspirated monitors differ with respect to the location of the air intake mechanism and visual alarm indicators. The instrument may be used in any orientation when clipped to the user or when used with a compatible carrying case. Normal instrument orientation for measuring gas concentration is hand held with sensors and display facing the operator.

Number

Feature

Functions

1

Visual alarm indicator

Signals an alarm or warning; frequency varies by alarm level. Also used as a confidence indicator.

2

Pump inlet (aspirated) Sensor ports (diffusion)

Air intake; calibration and bump test gas intake.

3

LCD display

User interface; backlight flashes when monitor is in system, high, or low alarm states.

4

Audible alarm ports

On when monitor is in system, high, or low alarm states; frequency and tone vary by alarm level. Also used for warnings and as confidence indicator.

5

On/Off/Mode button

Used to power-on and power-off. Also used to bypass a process/step or advance to a next screen in both gas monitoring and configuration modes. Sets values in configuration mode.

6

Enter button

Used to begin a process/step in a process. Edits values in configuration mode.

7

IrDA interface

Indicates infrared light data exchange in-progress.

8

Charging contacts

Battery charging.

© 2014Industrial Scientific Corporation

8

Ventis™ MX4 Product Manual

DISPLAY SCREEN The Ventis MX4 Boot-up Screen, as shown below, serves to introduce all icons and the alpha-numeric items (e.g., 8.8.8) that can appear on the display when the monitor is in use, docked, or charging. Each display item is stationary, communicates unique information, and appears only when relevant to the task being performed. A sample Gas Monitoring Screen is also shown below, next to the boot-up screen. This illustrates how the icons and the alpha-numeric characters work together to communicate several points of information to the monitor user.

Boot-Up Screen All possible screen images.

Gas Monitoring Screen Sample screen in gas monitoring mode. NOTE: Display screens featured throughout this manual include the “pump” icon. Similar in appearance to a fan, it indicates an aspirated monitor is in use. For a diffusion monitor, the pump icon does not appear on the display.

It is helpful to view the boot-up screen in sections. The top and bottom rows each contain icons. The main function of the middle section, in gas monitoring mode, is to communicate gas concentration readings. Definitions for all icons, gas name abbreviations, gas measurement units, and other indicators are provided below. Where applicable, display variations are noted. Top Row Icons

Definition Status: indicates no monitor or sensor faults. Warning: indicates monitor or sensor fault. Zero: communicates zero status (e.g., zero results, zero in-progress, etc.). Gas Cylinder: communicates calibration related information (calibration due, calibration apply gas, etc.). Clock: indicates a process is in-progress. Calendar: communicates overdue warnings for service items (calibration, bump testing, etc.). Alarm: indicates an alarm causing condition. Low level audio alarm is on. High level audio alarm is on. Peak: displayed when peak detection values are viewed.

Alpha-numeric display values

Definition Carbon Monoxide (CO) Methane (CH4) Sulfur Dioxide (SO2)

9

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Lower Explosive Limit. Display variations: “LEL” (English) “LIE” (French) “UEG” (German) Oxygen (O2) Nitrogen Dioxide (NO2) Hydrogen Sulfide (H2S)

COL

CO H2/Low Percentage Volume: O2 and CH4 measurement unit Percentage unit for combustible gases; display variations: “% LEL” (English) “% LIE” (French) “% UEG” (German) Parts Per Million: H2S, CO, SO2 and NO2 measurement unit. Over-range: for any sensor in over-range, indicates the measured gas concentration is greater than the measurement range of the sensor. Display variations: “Or” (English and German) “Sup” (French) Negative Over-range: for any sensor in negative over-range indicates the measured gas concentration is less than the negative measurement range of the sensor. Display variations: “-Or” (English and German) “InF” (French)

Bottom Row Icons

Definition Battery level indicator; display variations: 0 black bars = low battery warning 1 black bar < 33% charge remaining 2 black bars = 34% - 66% charge remaining 3 black bars = 67% – 100% charge remaining Security Code: indicates code is set or to be entered. Pump: shown anytime an aspirated monitor is in use. Indicates IrDA communication is in-progress. Short Term Exposure Limit: communicates STEL values. Display variations: “STEL” ( English and German) “VLE” (French) Time Weighted Average: communicates TWA values. Display variations: “TWA” (English and German) “VME” (French)

© 2014Industrial Scientific Corporation

10

Ventis™ MX4 Product Manual

ALARMS NOTICE  All monitor alarms and warnings should be taken seriously and responded to as stated in company safety standards.  Once initiated, an alarm will remain on while the alarm condition is present. For gas-related alarms, once the detected gas concentration changes, the alarm indicators will change to reflect any new condition such as low-alarm gas, high-alarm gas, over-range gas, or no gas alarm.  When the latch alarm feature is enabled and the monitor goes into alarm, it will remain in alarm until the alarm condition no longer exists and the monitor user presses the ENTER button for one second. This applies only to gas-related alarms. It is practical for the monitor user to be aware of the possible alarms prior to monitor set-up and use. The Ventis MX4 has four alarm and warning levels. A “system level” alarm generates the highest frequency tone and highest level visual and vibration signals. It is used to indicate such events as a pump or sensor failure. The “high” or “low” level audio alarms, in combination with visual and vibration indicators, turn on when gas concentration readings are overrange, high, or low. The lowest level indicator is a warning with beep patterns to indicate service needs (e.g., low battery or calibration due). The beep is also used as a confidence indicator when enabled. Alarm types and their alarm generating conditions are described below.

Display

Description An over-range condition occurs when the gas concentration value sensed is above the sensor’s measuring range. After any over-range alarm, the monitor should be calibrated.

Over-range Alarm Screen The “Or” message indicates which sensor(s) is reading an overrange condition(s). All other sensors show their current gas concentration readings on a numeric display (left) or gas names on a text display (right). The high level alarms turn on and the alarm icon displays.

NOTES: The O2 and toxic sensor values normally reset when the gas sensed reaches an acceptable range.

If the LEL reads over-range, the alarm latches and the LEL sensor is automatically turned off. Press the enter button to turn on the LEL sensor. This will turn off the alarm indicators. After a warm-up period of approximately 30 seconds, an LEL reading will display. If the new reading is an over-range or other alarm condition, the alarm indicators will turn on.

A negative over-range condition occurs when the gas concentration value sensed is less than the sensor’s measuring range. After any negative over-range alarm, the monitor should be calibrated. Negative Over-range Alarm Screen The “-Or” message indicates which sensor is reading a negative over-range condition. All other sensors display their current gas concentration readings*. The high level alarms turn on and the alarm icon displays. A high alarm condition occurs when the concentration of gas sensed reaches a level greater than the monitor’s high alarm value setting for a sensor(s).

High Alarm Screen A flashing gas concentration value* indicates which sensor(s) reading(s) is the cause for alarm. The high level alarms turn on and

11

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

the up arrow icon displays. A low alarm condition occurs when the concentration of gas sensed reaches the monitor’s low alarm value setting for a sensor(s).

Low Alarm Screen A flashing gas concentration value* indicates which sensor(s) reading(s) is the cause for alarm. The low level alarms turn on and the down arrow icon displays. A TWA alarm occurs when the calculated time weighted average reaches the monitor’s hazardous value for the set time frame.

TWA Alarm Screen A flashing gas concentration value* indicates which sensor(s) reading(s) is the cause for alarm. The low level alarms turn on and the TWA icon flashes. The STEL alarm occurs when the short term exposure value exceeds the acceptable limit.

STEL Alarm Screen A flashing gas concentration value* indicates which sensor(s) reading(s) is the cause for alarm. The low level alarms turn on and the STEL icon flashes. Alarm occurs when the monitor registers no sensors installed.

No Sensor Installed Screen The system level alarms turn on and the error icon displays. Alarm occurs when any installed sensor’s data-related operations fail and the sensor is not operational.

Sensor Data Fail Screen A flashing “F” indicates which sensor is the cause for alarm. The audio alarm turns on and the error icon displays. Error codes 4XX to 5XX (404 shown here) indicate the monitor has detected a malfunction. The unit is not operational and and should be examined by a qualified technician or reported to Industrial Scientific for service or repair information. Critical Error Screen

© 2014Industrial Scientific Corporation

12

Ventis™ MX4 Product Manual

Alarm occurs when, if attached, the pump is not operating correctly. While in alarm, every ten seconds the monitor attempts to restart the pump. If unsuccessful, the monitor remains in alarm. Pump Fault Alarm The system level alarms turn on and the error icon displays.

Note: The nominal flow rate is >200 cc/m (.2 LPM). A pump fault alarm will occur when the flow is less than 200 cc/m +0, -25%. Alarm occurs when the monitor’s battery reaches a low level of charge or is nearing its end of life.

Low Battery Warning Screen A beep sounds every 30 seconds and the empty battery icon flashes. Alarm occurs when one or more sensors are due for a bump test. If the monitor settings permit, an in-field bump test may be performed in an area known to be nonhazardous. Bump Overdue Screen A “b” indicates which sensor(s) is overdue for bump testing. Two beeps sound every 30 seconds and the calendar and alarm icons display. Alarm occurs when one or more sensors are due for calibration. If the monitor settings permit, an in-field calibration can be performed in an area known to be nonhazardous. Calibration Due Alarm Screen The gas value flashes for each sensor overdue for calibration. Three beeps sound every 30 seconds and the calendar and alarm icons display. The gas cylinder icon flashes. * The numeric mode display shows gas concentration values; the text mode display shows gas type names in place of gas values.

►Monitor Set-up Preparing the monitor for first time use is a “3-C” process: charge (if equipped with a lithium-ion battery pack), configure, and calibrate. This manual section covers charging and configuration for set-up purposes and can be consulted for ongoing instruction thereafter. Immediately following this section, calibration is covered in the manual section, Use and Service.

BATTERY PROPERTIES AND MONITOR COMPATIBILITY Based on the customer order, the Ventis MX4 comes equipped with one of three factory installed batteries: rechargeable Lithium-ion (Li-ion), rechargeable Extended Range Lithium-ion (extended range Li-ion), or replaceable alkaline. The factory installed battery pack type is stated on the label affixed to the monitor box. Basic battery properties and acceptable monitor/battery combinations are shown below.

13

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Properties and Compatibility

Battery Pack Rechargeable Li-ion

Rechargeable Extended Range Li-ion

Replaceable Alkaline

No

Yes

Yes

Ventis MX4 diffusion monitor compatible

Yes

Yes

Yes

Battery lifetime

300 charge cycles

300 charge cycles

--

Battery re-charge time

3-5 hours

3-7.5 hours

N/A

Ventis MX4 aspirated monitor compatible

Nominal run time (when fully charged and operating at room temperature) For the aspirated monitor

--

12 hours

4 hours

For the diffusion monitor

12 hours

20 hours

8 hours

NOTE: When a lithium-ion battery becomes deeply discharged and the instrument is docked, it can take up to an hour for the instrument display to indicate that the battery is charging.

CHARGING THE LITHIUM-ION BATTERY PACKS The lithium-ion battery packs are charged at the factory. As some or all of the charge may deplete before the monitor arrives or is unpacked, it is recommended that the monitor be fully charged before first time use. The lithium-ion equipped Ventis MX4 can be charged with any of the products listed below.         

DS2 Docking Station™ for Ventis MX4 V-Cal™ Calibration Station V-Cal™ 6-Unit Calibration Station Single Unit Charger Single Unit Charger/Datalink 6-Unit Charger Single Unit Automotive Charger, 12 VDC Single Unit Truck-Mount Charger, 12 VDC, with Cigarette Adapter Single Unit Truck-Mount Charger, 12 VDC, Hard Wired NOTE: The above products are all equipped with a yellow LED “presence” indicator. This LED confirms that the monitor is properly seated in the cradle such that the monitor can charge; however, it is NOT intended to be used as a charging indicator. This LED indicator may go out intermittently during normal charging functions and will not light if the unit is fully charged when placed in the cradle. Always refer to the monitor display’s battery level indicator to confirm the battery charge level.

The Single Unit Charger is generally shipped with the monitor. Equipped with a movable partition, which fits in each of two dedicated slots, it charges the diffusion and aspirated monitors with their compatible lithium-ion battery packs. Place the partition in the back slot for a diffusion or aspirated monitor with an Extended Range Li-ion battery pack. Place the partition in the front slot to charge a diffusion monitor with a Li-ion battery pack.

Instructions NOTICES  Charge the monitor in an area known to be nonhazardous.  When using the charger and adjusting its partition, take care NOT to touch the battery contacts located at the front of the cradle bottom.  

The single unit charger has a universal power cord; change the plug insert, if needed, and plug into the appropriate outlet. To properly adjust the partition, if needed, complete and observe the following. o Lift up to remove from slot. o Push down to place in the desired slot. o When partition is inserted correctly, a click sounds. Recommended Practice: To prevent the loss of the partition, it should always reside in the cradle in one of its two dedicated slots. Choose the most used slot. Do not place the partition in the forward most compartment of the charger where the battery contacts are located.



To properly place the monitor in the charger, complete or observe the following. o The monitor’s display side faces the user.

© 2014Industrial Scientific Corporation

14

Ventis™ MX4 Product Manual

o o

The charging contacts on the monitor bottom meet the contact pins inside the charger’s cradle. Refer to the monitor’s battery icon to confirm the battery charge level.  If the battery is less than fully charged, the monitor displays the battery icon (flashing empty to full, repeatedly).  If the battery is fully charged, the monitor displays a full battery icon.

POWER-ON AND -OFF To power-on the Ventis MX4, press ON/OFF/MODE and hold for three to five seconds. During the first ten to15 seconds the monitor is on, its firmware completes internal tests and the user sees or hears what is described and shown below. Following this initialization phase, a countdown screen displays. During this 20-second countdown, the monitor user can enter configuration mode to manually adjust monitor settings. Display and Options

Instructions No user action required.

Visual Test Screen Displays for up to five seconds as the monitor completes a sensor and alarm check. Visual, vibration, and audio alarms turn on briefly, then off. Be sure the pump inlet is not blocked.

Pump Set-up Screen Displays for five to seven seconds for an aspirated monitor. The monitor checks for the presence of a pump. If present, the pump is started and, if needed, adjusted for optimum flow. No user action required.

Software Version Screen The Software Version Screen message displays for five seconds. No user action required.

Calibration Days Screen When the up arrow (▲) is featured, the number of days displayed for each sensor indicates when the next calibration is due. When the down arrow (▼) is featured, the number of days displayed indicates when the last calibration occurred.

15

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

To enter gas monitoring mode: allow the countdown to complete and advance to the Gas Monitoring Screen. Proceed to the manual section, Monitor Use and Service. Countdown Screen Displays the 20 second countdown, one second at a time, from 20 to one. Options : Enter gas monitoring mode Enter configuration mode

To enter configuration mode: simultaneously press ON/OFF/MODE and ENTER, hold for three seconds, and release.

Press ON/OFF/MODE, hold for the full five second countdown to zero, and release to power-off the monitor.

Power-Off Screen The screen displays a five-second countdown accompanied by five beeps and LED flashes.

CONFIGURATION Introduction Before first time use of the monitor, its settings should be reviewed and, if needed, be adjusted. Qualified safety personnel should complete the following tasks.    

Review the monitor settings for compliance with company policy and any applicable regulations, laws, and observed guidelines as issued by regulatory agencies and government or industry groups. Determine which settings, if any, require adjustment. Make the adjustments or supervise other qualified personnel in the process. When using instrument-compatible Industrial Scientific docking station software (e.g., iNet, Docking Station Server Administrative Console [DSSAC], or Accessory Software), maintain the mandated settings through the software or by manually configuring the instrument settings after docking.

Monitor settings should be reviewed regularly and adjusted as needed. The following settings are adjustable or “configurable" for the Ventis MX4. LEL Type Calibration Mode Setting Low Alarm Settings High Alarm Settings TWA Alarm Settings TWA Interval Settings STEL Alarm Setting Calibration Gas Settings Clock Settings Date Settings

Display Mode Setting Confidence Indicator (on/off) Confidence Indicator (type) Bump Test In-field Bump Test Due Warning Bump Test Time Set-point Bump Test Percentage Bump Test Response Time Alarm Latch Set Zero In-field

Calibration In-field Calibration Due Alarm Calibration Due Set-point Security Code Language Selection Always-on Setting Shutdown In Alarm Setting Alarm on Dock Setting

The Ventis MX4 can be configured manually as instructed below. Any changes made take effect immediately upon exiting the configuration mode. Instructions NOTICES  The configuration mode should be accessed only by safety personnel authorized to change monitor settings based on company policy.  Read ALL requirements and instructions outlined below, including the screen-by-screen process description, before beginning the configuration process.

© 2014Industrial Scientific Corporation

16

Ventis™ MX4 Product Manual The configuration mode can be entered during the 20-second countdown of the power-on process. During the countdown, simultaneously press ON/OFF/MODE and ENTER, hold for three seconds, and release to enter configuration mode. (While in the configuration mode, the same button presses cause the monitor to exit configuration). Each configuration screen times out after 30 seconds and the monitor enters gas monitoring mode. To re-enter the configuration mode, power-off the monitor, then power-on and repeat the entry process. Throughout the configuration process, the main functions of the two buttons are as follows.  The ENTER button is used to edit values. It is also used, where noted, to begin a process or a step in a process.  The ON/OFF/MODE button is used to set the value. Where noted, it is also used to bypass a process or step in a process, or to advance to the next configuration screen. The first screen to display in configuration mode depends on three things:  security code setting,  the presence or absence of the China MA feature,  and the presence or absence of an LEL sensor. If the security code setting is 000, the security feature is disabled and the Enter Security Code Screen does NOT appear. If the security code is NOT 000, the security feature is enabled and the monitor displays the Enter Security Code Screen. The monitor next checks for the presence of a China MA mining feature. If this feature is operational, the monitor displays the Zero Initiate Screen. If the China MA mining feature is NOT operational, the monitor then checks for an installed LEL sensor. If installed, the monitor displays the LEL Type Screen. If no LEL sensor is installed, the monitor displays the Zero Initiate Screen. Configuration Process Display and Options

Instructions Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace to reach the valid security code. Press ON/OFF/MODE to enter configuration mode and arrive at the next applicable screen.

Enter Security Code Screen The presence of this screen indicates an enabled security feature. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Zero Initiate Screen.

LEL Type Set Screen Options LEL CH4

17

NOTE: If the LEL type is changed, the sensor goes into calibration fail mode. A full calibration is required before the monitor can be used and is accessible from the next screen in the configuration process, the Zero Initiate Screen. For complete calibration instructions, proceed to the manual section, Zero, Calibration, and Bump Testing.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Press ON/OFF/MODE to bypass the zero and calibration processes and advance to one of two screens.

Zero Initiate Screen Options Bypass zero and calibration process. Begin zero and calibration process.

If the installed sensor set includes H2S and NO2, OR, SO2 and NO2, the monitor is pre-set for standard calibration mode and the Low Alarm Set-point Screen displays. For all others installed sensor combinations, the Calibration Mode Selection Screen displays. Press ENTER to begin the zero and calibration process. Proceed to the manual section, Zero, Calibration, and Bump Testing The quick calibration option sets the monitor to calibrate all four sensors simultaneously. The standard calibration option sets the monitor to calibrate each sensor independently.

Calibration Mode Selection Options 0 = Standard Calibration 1 = Quick Calibration

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Low Alarm Set Screen.

NOTE: The user can edit the values for four alarm types in configuration mode. The monitor presents these options in the order shown below. 1. Low alarm 2. High alarm 3. TWA (if toxic sensors installed) 4. STEL (if toxic sensors installed) For each alarm type (e.g., low alarm), the user can edit the alarm settings for each installed sensor, one sensor at a time. The order in which the sensors are subject to change is as follows. 1. Toxic sensor 1 2. LEL sensor 3. Toxic sensor 2 4. O2 sensor

Low Alarm Set-point Screen Displays the existing low alarm value for each installed sensor. If any one of the sensors is NOT installed, its position on the display is blank.

Press ON/OFF/MODE to bypass the low alarm value set process and advance to the High Alarm Set-point Screen. Press ENTER to begin the low alarm value set process. On the display, the first sensor subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next sensor subject to change flashes. Continue to use the ENTER and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s low alarm value. After the alarm value is set for each installed sensor, press ON/OFF/MODE to advance to the High Alarm Setpoint Screen.

© 2014 Industrial Scientific Corporation

18

Ventis™ MX4 Product Manual

High Alarm Set-point Screen Displays the existing high alarm value for each installed sensor. If any one of the sensors is not installed, its position on the display is blank.

Press ON/OFF/MODE to bypass high alarm value set process and advance to one of two screens as noted below. Press ENTER to begin the high alarm value set process. On the display, the first sensor subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next sensor subject to change flashes. Continue to use the ENTER and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s high alarm value. After the alarm value is set for each installed sensor, press ON/OFF/MODE and advance to one of two screens. If at least one toxic sensor is installed, the TWA Alarm Set Screen displays. If NO toxic sensors are installed, the Calibration Gas Set Screen displays.

TWA Alarm Set-point Screen Displays the existing TWA values for the toxic sensors installed. No other sensor readings appear.

Press ON/OFF/MODE to bypass the TWA alarm value set process and advance to the TWA Interval Set-point Screen. Press ENTER to begin the TWA alarm value set process. On the display, the first sensor subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next sensor subject to change flashes. Continue to use the ENTER and ON/OFF/MODE buttons, respectively, to edit and set each alarm value. After the alarm value is set for each installed sensor, press ON/OFF/MODE to advance to the TWA Interval Set Screen. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the STEL Alarm Set-point Screen.

TWA Interval Set-point Screen Displays the existing TWA interval. The value can be set from one to 40 hours, in increments of one.

STEL Alarm Set-point Screen Displays the existing STEL values for the toxic sensors installed. No other sensor readings appear.

Press ON/OFF/MODE to bypass the STEL alarm value set process and advance to the Calibration Gas Set Screen. Press ENTER to begin the STEL alarm value set process. On the display, the first sensor subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next sensor subject to change flashes. Continue to use the ENTER and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s STEL alarm value. After the alarm value is set for each installed sensor, press ON/OFF/MODE to advance to the Calibration Gas Set Screen.

19

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Calibration Gas Set Screen Displays the existing calibration gas value for each installed sensor. If any one of the sensors is not installed, its position on the display is blank.

Clock Set Screen Displays the existing time values using a 24-hour time format.

Press ON/OFF/MODE to bypass the calibration gas set process and advance to the Clock Set Screen. Press ENTER to begin the calibration gas value set process. On the display, the first sensor subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next sensor subject to change flashes. Continue to use the ENTER and ON/OFF/MODE buttons, respectively, to edit and set each sensor’s calibration gas value. After calibration gas value is set for each installed sensor, press ON/OFF/MODE to advance to the Clock Set Screen. Press ON/OFF/MODE to bypass the clock set process and advance to the Date Set Screen. Press ENTER to begin the clock set process. On the display, the first time value subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next value subject to change flashes. Use the ENTER and ON/OFF/MODE buttons, respectively, to edit the value. After all values are set, press ON/OFF/MODE and advance to the Date Set Screen.

Date Set Screen Displays the existing date. The value displayed on the far left is the month and to its right the day. The year is displayed beneath the day.

Press ON/OFF/MODE to bypass the date set process and advance to the Display Mode Set Screen. Press ENTER to begin the date set process On the display, the first date value subject to change flashes. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value. The next date value subject to change flashes. Continue to use the ENTER and ON/OFF/MODE buttons, respectively, to edit and set each value. After all values are set, press ON/OFF/MODE and advance to the Display Mode Set Screen. The display mode selected determines whether the monitor user will see a numeric or text display (including alarm displays) when the monitor is in the gas monitoring mode.

Display Mode Set Screen Options 0 = Numeric Mode 1 = Text Mode

© 2014 Industrial Scientific Corporation

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Confidence Indicator Set Screen.

20

Ventis™ MX4 Product Manual

With an enabled confidence indicator, the monitor will emit a signal, every 90 seconds in gas monitoring mode, to inform the user it is operational.

Confidence Indicator On-Off Screen Options 0 = Disable/off 1 = Enable/on

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to one of two screens. If the confidence indicator is enabled, the Confidence Indicator Type Set Screen displays. If the confidence indicator is disabled, the Bump Test Infield Option Screen displays. Sets the type of signal that will be emitted by an enabled confidence indicator. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Bump Test In-field Option Screen.

Confidence Indicator Type Set Screen Options 1 = audible chirp 2 = LED flash 3 = combination audible chirp and LED flash When enabled, permits all monitor users to bump test the monitor from the gas monitoring mode.

Bump Test In-field Option Screen Options 0 = Disable/off 1 = Enable/on

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to one of two screens. If Bump Test In-field is enabled, the Bump Due Warning Option Screen displays. If the Bump Test In-field is disabled, the Alarm Latch Set Screen displays. When enabled, the monitor will sound two beeps every 30 seconds and its display icons will indicate a bump test is due. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Bump Test Time Set-point Screen.

Bump Due Warning Option Screen Options 0 = Disable/off 1 = Enable/on Sets the elapsed time allowed between bump tests. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value and advance to the Bump Test Percentage Requirement Screen. Bump Test Time Set-point Screen Value range: .5 days to 7.0 days Value increment: .5 days

21

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Sets the percentage of calibration gas the monitor expects to be exposed to.

Bump Test Percentage Requirement Screen Value range: 50% to 99% Value increment: one percent

Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value and advance to the Bump Test Response Time Screen.

Sets the bump test response time period. Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value and advance to the Latch Alarm Set Screen. Bump Test Response Time Screen Value range: 30 to 300 seconds Value increment: five seconds When enabled, if the monitor goes into any gas-related alarm, it will remain in alarm until after the gas concentration is less than (or more than for oxygen) the alarm set point, and the monitor user presses the ENTER button for one second. Latch Alarm Set Screen Options 0 = Normal mode 1 = Latching mode

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Zero In-field Screen.

When enabled, all monitor users are permitted to zero the monitor from the gas monitoring mode.

Zero In-field Screen Options 0 = Disable/off 1 = Enable/on

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to one of two screens. If Zero In-field is enabled, the Calibration In-field Option screen displays. If Zero In-field is disabled, the Calibration Due Alarm screen displays. When enabled, all monitor users are permitted to calibrate the monitor from the gas monitoring mode. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Calibration Due Alarm Option.

Calibration In-field Option Screen Options 0 = Disable/off 1 = Enable/on When enabled, the monitor will activate the calibration due alarm, in gas monitoring mode, when any sensor is due for calibration. A flashing gas cylinder and gas type will appear on the display and three beeps will sound every 30 seconds. Calibration Due Alarm Option Screen Options 0 = Disable/off 1 = Enable/on

© 2014 Industrial Scientific Corporation

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Calibration Due Set-point screen.

22

Ventis™ MX4 Product Manual

Sets the elapsed time allowed between calibrations. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Calibration Days Set Screen. Calibration Due Set-point Screen Value range: one to 365 days Value increment: one day Sets how the Calibration Days Screen will display in operation mode.

Calibration Days Set Screen Options 0 = display days since last calibration 1 = display days until next calibration

NOTE: The up arrow ( ▲) will be featured on-screen when the unit is set to display the number of days before a sensor’s next calibration is due. The down arrow (▼) will be featured when the unit is set to display the number of days since the last calibration was performed. A value will be displayed for each installed sensor. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the Security Code Set Screen A security code value of 000 permits all monitor users to enter configuration mode and gain access to change the monitor’s settings. A value other than 000 will restrict access to the configuration mode; it will also restrict access to the shutdown process for an instrument that is configured for “always on”.

Security Code Set Screen Valid values: 000 to 999. Increment value: one

Press ENTER to edit the value, if needed; press repeatedly or hold down to speed the increment pace. Press ON/OFF/MODE to set the value and advance to the Language Selection Screen. Allows the choice of display languages as applied to select screens. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and return to the LEL Type Set Screen.

Language Selection Screen Options E = English F = French d = German When enabled, the shutdown process is security-code protected only if the security code is not equal to 000. If the code is set to anything other than 000, the user will be prompted to enter the unit's security code to complete the shutdown process. Always-on Set Screen Options 0 = Disable/off 1 = Enable/on

23

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the next configuration mode screen.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Disallow or allow operator-activated shutdown when the unit is in alarm.

Shutdown In Alarm Screen Options 0 = Disallows shutdown 1 = Allows shutdown

Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the next configuration mode screen.

Disable or enable alarm indicators when the unit is docked. Press ENTER to edit the value, if needed. Press ON/OFF/MODE to set the value and advance to the next configuration mode screen. Alarm on Dock Screen Options 0 = Disable/off 1 = Enable/on

►Monitor Use and Service Proper monitor use and service includes everything from bump testing and calibration to keeping the monitor clean, proper air sampling, and the replacement of parts and components. Beginning with calibration and bump testing, the following sections provide information and instruction on all use and service tasks.

ZERO, CALIBRATION, AND BUMP TESTING Gas detection instruments are potentially life-saving devices. When completed regularly, the procedures defined below help to maintain proper instrument functionality and enhance operator safety. Procedures Configuration. The configuration process allows qualified personnel to review and adjust a unit's settings. Bump Test (or "functional test"). Bump testing checks for sensor and alarm functionality. The installed sensors are briefly exposed to expected concentrations of calibration gases that are greater than the sensors’ low alarm set points. When one or more sensors “pass” the test, they are “functional” and the unit will alarm. Each sensor’s “pass” or “fail” result is indicated on the unit’s display. Note: a bump test does not measure for sensor accuracy (see “Calibration”).

Zero. Zeroing sets each installed sensor to recognize the ambient air as clean air. If the ambient air is not truly clean air, any gasses that are present and relevant to the installed sensor types will be measured and displayed as zero. Readings will be inaccurate until the unit is correctly zeroed in truly fresh air or with a zero air cylinder. Calibration. All sensors gradually degrade over time. This diminishes a sensor's ability to measure gas concentrations accurately; however, regular calibrations adjust the instrument to compensate for this decline in sensitivity. During calibration, the installed sensors are exposed to expected concentrations of calibration gases and, when needed, the instrument will self-adjust to ensure the accurate measurement and display of gas concentration values. Note: when a sensor has degraded beyond an acceptable level, no further adjustment is possible and the sensor will no longer pass calibration.

Peak Readings. The instrument stores the highest detected gas readings, the "peak readings" or "peaks". Bump testing and calibration will often register new peak readings. Therefore, the clearing of the peak readings should follow each calibration. The instrument operator may also wish to clear the peak readings after a bump test, before a change in location, or after an alarm is addressed and cleared.

© 2014 Industrial Scientific Corporation

24

Ventis™ MX4 Product Manual Note: The peak readings and the data log readings are stored independently of one another; therefore, clearing the peak reading does not affect the data log. Powering the instrument off or changing its battery does not affect the peak reading. These checks and balances help promote operator safety, and serve to contain the peak readings in a "black-box" manner. In the event of a gas-related incident, this black-box record can be useful to the safety team or a prospective investigator.

Recommendations Industrial Scientific Corporation (ISC) minimum frequency recommendations for each procedure are summarized in the table below. These recommendations are based on field data, safe work procedures, industry best practices, and regulatory standards to help ensure worker safety. Industrial Scientific is not responsible for setting safety practices and policies. These policies may be affected by the directives and recommendations of regulatory groups, environmental conditions, operating conditions, instrument use patterns and exposure to gas, and other factors. Procedure

ISC Recommended minimum frequency

Configuration

Before first use and as needed thereafter.

Calibrationa

Before first use and monthly thereafter.

Bump testb

Prior to each day’s use.

aBetween

regular calibrations, ISC also recommends a calibration be performed immediately following each of these incidences: the unit falls, is dropped, or experiences another significant impact; is exposed to water; fails a bump test; or has been repeatedly exposed to an over-range (positive or negative) gas concentration. A calibration is also recommended after the installation of a new (or replacement) sensor. bIf conditions do not permit daily testing, bump tests may be done less frequently based on company safety policy. Note: The use of calibration gases not provided by ISC may void product warranties and limit potential liability claims.

General information The zero, calibration, and bump testing tasks are in-field enabled or in-field disabled in the configuration process. This permits or denies access to these functions from the gas monitoring mode. When any of these options is enabled, it is accessible to all monitor users. In gas monitoring mode, a series of presses on the ON/OFF/MODE button gives the user access to the following screens and processes in the order shown.       

Gas Monitoring Screen Days Since Calibration Zero Initiate (if in-field enabled) o Calibration Apply Gas Screen (if in-field enabled) Bump Test Initiate (if in-field enabled) Peak Readings TWA Readings STEL Readings

The monitor is capable of performing two types of calibration, and this option is set in configuration mode. The calibration type selected also determines the monitor’s bump test type. With a “quick” calibration, the monitor is set to calibrate and bump test all installed sensors simultaneously. With a “standard” calibration setting, these tasks are completed independently for each installed sensor in the order shown below. 1. Oxygen sensor* 2. Toxic sensor 1 3. LEL sensor 4. Toxic sensor 2 *If set to the default value of 20.9% or 21%, the Oxygen sensor calibrates during the zero process and toxic 1 is the first to calibrate in the calibration process.

The Ventis MX4 monitor can be calibrated with any of the accessories listed.  Calibration cup and/or tubing shipped with the monitor (see instructions below)  V-Cal Calibration Station (consult the calibration station manual for instruction)  DS2 Docking Station for Ventis (consult the docking station manual for instruction)

25

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual Instructions Calibration and Bump Testing with Calibration Cup and/or Tubing Read all instructions before beginning: notices, supply check-list, gas cylinder preparation, and the complete screenby-screen walk-through of the zero, calibrate, and bump test processes. Each process is presented in the order in which it is accessible from gas monitoring mode. NOTICES  Industrial Scientific recommends that full monitor calibration be performed, using a known certified concentration(s) of Industrial Scientific calibration gas(es), to prepare the monitor for first time use, and monthly (at a minimum) thereafter, to help ensure monitor accuracy.  Industrial Scientific also recommends that each monitor be zeroed and bump tested before each use with a known certified concentration(s) of Industrial Scientific calibration gas(es).  Read ALL requirements and instructions outlined below, including the screen-by-screen process description, before beginning the zero, calibration, or bump testing processes.  Only qualified personnel should zero, calibrate, or bump test a monitor.  Zero, calibration, and bump testing functions should be performed in a fresh air environment known to be nonhazardous.  After calibration or bump testing, or after terminating either process, stop the flow of gas. Supplies Item

Monitor/Regulator** Aspirated monitor with Demand Flow Regulator**

Aspirated monitor with Positive Flow Regulator**

Diffusion monitor with Positive Flow Regulator**

Calibration cup*

No

No

Yes

Calibration tubing 2 feet in length*

Yes

No

Yes

Calibration tubing 2 feet in length with integrated “t” fitting

No

Yes

No

Calibration gas cylinder

Yes

Yes

Yes

*Shipped with monitor. **Industrial Scientific recommends 1) the use of positive flow regulators with a flow rate of .5 LPM, and 2) the diffusion monitor be calibrated or bump tested using a positive flow regulator, NOT a demand flow regulator.

Diffusion monitor with positive flow regulator.

© 2014 Industrial Scientific Corporation

Aspirated monitor with demand flow regulator.

26

Ventis™ MX4 Product Manual Prepare the gas cylinder for use  According to the supply chart above, attach the correct regulator to the gas cylinder and turn clockwise to tighten.  Next, choose instruction A., B., or C. based on the monitor/regulator combination in use. A.

Aspirated with demand flow regulator Attach either end of the tubing to the cylinder’s nipple. DO NOT ATTACH THE OTHER END OF THE TUBING TO THE MONITOR BEFORE REACHING THE “APPLY GAS SCREEN”. Completing the connection of the tubing will cause gas to flow. If gas is applied before reaching the appropriate screen, the monitor will go into alarm and a failure will be logged.

B.

Aspirated with positive flow regulator The calibration tubing with the t-fitting (not included) has two different sized openings, a narrow opening at one end and a wider opening at the other end. Attach the wider opening to the nipple on the cylinder’s regulator. Attach the smaller opening to the pump inlet. DO NOT APPLY THE GAS BEFORE REACHING THE “APPLY GAS SCREEN”. If gas is applied before that point, the monitor will go into alarm and a failure will be logged.

C. Diffusion with positive flow regulator Attach either end of the tubing to the cylinder’s nipple. Attach the other end of the tubing to the calibration cup’s nipple. DO NOT ATTACH THE CALIBRATION CUP TO THE MONITOR OR APPLY THE GAS BEFORE REACHING THE “APPLY GAS SCREEN”. If gas is applied before that point, the monitor will go into alarm and a failure will be logged. Zero and Quick Calibration Process Display and Options

Instructions Press ON/OFF/MODE to advance to the Days since Calibration Screen.

Gas Monitoring Screen Numeric mode display (left) Text mode display (right)

Displays the gas concentration readings (or gas names in text mode) for all installed sensors. If a sensor is NOT installed, its position on the LCD is blank. Press ON/OFF/MODE to advance to one of three screens. If zero in-field is enabled, the user advances to the Zero Initiate Screen. Days Since Calibration Screen Displays the number of days since the last successful calibration for each installed sensor. Each value can be different.

If zero in-field is disabled and bump test in-field is enabled, the user advances to the Bump Test Initiate Screen. If zero in-field and bump test in-field are both disabled, the user advances to the Peak Readings Screen.

NOTE: When zero, calibration, and bump test are ALL in-field enabled, and the user has entered zero from the gas monitoring mode, the monitor expects to be calibrated following a successful zero. If the desired task, after zero, is bump testing (or clearing the peaks) and NOT calibration, follow the instructions at the Calibration Apply Gas Screen to terminate calibration.

27

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Press ENTER to begin the zero process and advance to the Zero In-process Screen. Press ON/OFF/MODE to bypass zero and calibration and advance to one of two screens.

Zero Initiate Screen Options: Enter Zero Bypass Zero

If bump test in-field is enabled, the user advances to the Bump Test Initiate Screen. If bump test in-field is disabled, the user advances to the Peak Readings Screen. Allow the zero process to complete and advance to the Zero Results (Pass or Fail) Screen.

Zero In-process Screen Each sensor’s numerical value becomes zero except O2. An updated O2 span value displays. The clock icon flashes and the zero icon displays.

After the zero process, press ON/OFF/MODE to bypass calibration of the installed toxic and combustible sensors. The user returns to the mode from which the calibration process was entered (configuration or gas monitoring). NOTE: during the zero process, the O2 sensor is calibrated (when set to default gas volume of 20.9%)

NOTE: The “span reserve” of a sensor measures its sensitivity. The displayed span value divided by the calibration gas value equals the span reserve percentage. A span reserve percentage of greater than 70% indicates a “good” sensor; 50%-70% indicates “marginal” sensitivity. When the span reserve percentage is less than 50%, the sensor will not pass calibration.

Within Ten Seconds Press ENTER to repeat the zero process. Press ON/OFF/MODE to advance to one of two screens. If zero was entered from… Zero Results (Pass) Screen The check mark displays to indicate a successful zero and a short beep sounds. Each sensor’s numerical values display at zero except O2. Options: Repeat zero Begin calibration Enter gas monitoring mode

…configuration mode, the user advances to the Calibration Apply Gas Screen. … gas monitoring mode and the calibration in-field option is enabled, the user advances to the Calibration Apply Gas Screen. …gas monitoring mode and the calibration in-field option is disabled, the user advances to the Gas Monitoring Screen in the gas monitoring mode. If neither ENTER nor ON/OFF/MODE is pressed, within ten seconds, the user advances to the Gas Monitoring Screen in the gas monitoring mode. Press ON/OFF/MODE (or wait ten seconds) to return to the Zero Initiate Screen and repeat the zero process.

Zero Results (Fail) Screen Displays an “F” or “P”, respectively, for each failed or passed sensor. For O2, if the sensor passed its calibration, the sensor reading displays.

© 2014 Industrial Scientific Corporation

28

Ventis™ MX4 Product Manual

To Terminate Press ON/OFF/MODE while the gas cylinder icon flashes to terminate the quick calibration process (or to skip a sensor’s calibration in standard calibration) and return to the gas monitoring mode. Calibration Apply Gas Screen* The gas cylinder icon flashes. Each sensor’s display shows the calibration gas concentration to be applied. (The O2 display is blank as the sensor was calibrated during zeroing.) The monitor waits up to five minutes to successfully sense the gas. Calibration In-progress Screen* If gas is sensed, the gas values for the LEL and toxic sensors increase and the O2 value decreases. If gas is NOT sensed, a failed calibration registers and the Calibration Failed Screen displays.

To Calibrate From the already prepared gas cylinder, start the flow of gas as noted below for the monitor/regulator combination in use. For an aspirated monitor with a demand flow regulator, complete the tubing connection from the regulator to the pump inlet. For an aspirated monitor with a positive flow regulator, turn (counterclockwise) the regulator’s knob. For a diffusion monitor with a positive flow regulator:  Place the calibration cup over the upper portion of the monitor’s case top (front of monitor). To attach properly, complete or observe the following. o The cup fully covers the sensor ports. o The monitor’s display and buttons are NOT covered. o The cup’s side arms fit securely in the grooves on the sides of the monitor. o The Ventis MX4 name on the calibration cup is upright and readable. o The cup’s nipple points up and away from the monitor.  Turn (counterclockwise) the regulator’s

knob. STOP THE FLOW OF GAS. After calibration, or if calibration is terminated at anytime during the process, stop the flow of gas as follows. For an aspirated monitor with a demand flow regulator, disconnect the tubing from the pump inlet. For an aspirated or diffusion monitor with a positive flow regulator, turn (clockwise) the regulator’s knob.

Sensor Results Screen* Pass (top) or Fail (bottom) Screen The display alternately shows a “P” for pass (or “F” for fail) and the final span value reading for each sensor. A check mark displays and a single beep sounds.

All Sensors Pass The user returns to the mode from which the calibration process was entered (configuration or gas monitoring). Sensor Fail If one or more sensors fail calibration, the Calibration Fail Screen displays and a system level alarm turns on.

NOTE: The “span reserve” of a sensor measures its sensitivity. The displayed span value divided by the calibration gas value equals the span reserve percentage. A span reserve percentage of greater than 70% indicates a “good” sensor; 50%-70% indicates “marginal” sensitivity. When the span reserve percentage is less than 50%, the sensor will not pass calibration.

29

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Any failed sensor stays in alarm until it passes a calibration or is replaced. Press ON/OFF/MODE to repeat calibration.

Calibration Failed Screen* Gas readings display for all successfully calibrated sensors and an “F” displays for any failed sensors. A system level alarm turns on. The warning icon and gas cylinder display to indicate a sensor calibration failure. * During the standard calibration or standard bump test process, a series of apply gas, in-progress, and results screens show for EACH sensor as it is calibrated or tested.

Bump Test Process Display and Options

Instructions Press ON/OFF/MODE to bypass the bump test process and advance to the Peak Readings Screen. Press ENTER to begin the bump test process.

Bump Test Initiate Screen Options Begin process Bypass process To Terminate Press ON/OFF/MODE while the gas cylinder icon flashes to terminate the quick bump test process (or to skip a sensor’s testing in standard bump testing). The user returns to the Gas Monitoring Screen. Bump Test Apply Gas Screen Displays the bump test gas concentrations the monitor is expecting to receive. The monitor waits up to five minutes to successfully sense the gas.

To Bump Test From the already prepared gas cylinder, start the flow of gas as noted below for the monitor/regulator combination in use.

If gas is sensed, the user advances to the Bump Test In-progress Screen.

For an aspirated monitor with a demand flow regulator, complete the tubing connection from the regulator to the pump inlet.

If gas is NOT sensed, a failed bump test occurs and the user advances to the Bump Test Results Screen displays.

For an aspirated monitor with a positive flow regulator, turn (counterclockwise) the regulator’s knob. For a diffusion monitor with a positive flow regulator:  Place the calibration cup over the upper portion of the monitor’s case top (front of monitor). To attach properly, complete or observe the following. o The cup fully covers the sensor ports. o The monitor’s display and buttons are not covered. o The cup’s side arms fit securely in the grooves on the sides of the monitor. o The Ventis MX4 name on the calibration

© 2014 Industrial Scientific Corporation

30

Ventis™ MX4 Product Manual

cup is upright and readable. The cup’s nipple points up and away from the monitor. Turn (counterclockwise) the regulator’s knob.

o 

As the bump test progresses, observe the display activity (left). After the bump test, the Bump Test Results Screen displays. Bump Test In-progress Screen Displays when gas is sensed within five minutes. The clock icon flashes to indicate the test is in-progress. The sensor reading(s) display. The LEL and toxic sensor readings increase and the O2 reading decreases.

STOP THE FLOW OF GAS. After bump testing, or if bump testing is terminated at anytime during the process, stop the flow of gas from the cylinder as follows. For an aspirated monitor with a demand flow regulator, disconnect the tubing from the pump inlet. For an aspirated or diffusion monitor with a positive flow regulator, turn (clockwise) the regulator’s knob.

No User Action Required After a passed bump test, the monitor goes into gas monitoring mode.

Bump Test Results (Pass) Screen The above displays an all sensor pass result. If one or more sensors fail, the “F” shows in place of the “P”. The pass/fail screen (left) and the final sensor reading screen (right) display alternately three times. A single beep sounds to indicate the bump test is completed. After a failed bump test, the Bump Test Fail Screen displays and a low level audio alarm turns on. The monitor should be fully calibrated after a failed bump test. Bump Test Results (Fail) Screen The “bF” displays under each gas type to indicate a bump test failure. The system level alarm turns on and the gas cylinder icon flashes.

NOTE: After a full calibration, the O2 sensor must pass a bump test to clear the bump test fail status.

Press and release ENTER to clear the peak values, if desired. Press ON/OFF/MODE to advance to one of two screens. If toxic sensors are installed, the user advances to the TWA Readings Screen. Peak Readings Screen Displays the peak icon and peak gas concentrations for each installed sensor since the last time the peak readings were cleared. (For O2, the lowest reading is shown.)

31

If no toxic sensors are installed, the user advances to the Gas Monitoring Screen.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Press and release ENTER to clear the displayed TWA reading(s), if desired. Press ON/OFF/MODE to advance to the STEL Readings Screen.

TWA Readings Screen Displays the TWA (time weighted average) icon and calculated readings for each toxic sensor installed; all other sensor values are blank. Press and release ENTER to clear the reading(s), if desired. Press ON/OFF/MODE to advance to the Gas Monitoring Screen.

STEL Readings Screen Displays the STEL (short term exposure limit) icon and STEL values for each toxic sensor installed; all other sensor values are blank. The STEL value is the running average over the last 15 minutes. * During the standard calibration process, a series of apply gas, in-progress, and results screens show for EACH sensor as it is calibrated or tested.

RECOMMENDED PRACTICES FOR IN-FIELD AIR SAMPLING Diffusion monitor When worn, the diffusion monitor should be fastened securely and attached to ensure the sensor portals are exposed to the air. The monitor should be in full view. No part of the monitor should be covered by any garment or part of a garment. Aspirated monitor Before using an aspirated monitor, force a pump fault to ensure the pump and air-sample accessories (sample tubing, probe, or sample tubing and probe) are working as intended: 1 Power on the monitor. 2 Attach the air-sample accessory(ies) to the pump inlet. 3 Using a finger, cover the opening at the other end of the air-sample tubing (or probe) to block the flow of air. This should result in a pump fault alarm. 4 After a pump fault occurs, remove the finger from the tubing (or probe). If the unit resumes normal operation, the pump and air-sampling accessories are working as intended. Note: If no pump fault occurs or if the unit stays in pump fault, check for debris at these locations: the pump nipple, inside the inlet barrel, at the water barrier, and inside the pump cradle; attempt to clear any blockage. Check the airsample accessories, too, for debris and for possible crack or leaks. The Ventis MX4 aspirated monitor is rated to sustain a continuous sample draw for up to 100 feet (30.48 m) with 0.125 inch (0.3175 cm) inside diameter sample tubing. In confined space, an air sample should be taken in four foot (1.2192 m) intervals. With each sample, the minimum time required to successfully draw air and read the gas concentrations should include two minutes plus two seconds for every 12 inches (30.48 cm) of tubing.

CLEANING    

NEVER use solvents or cleaning solutions of any type. When necessary, wipe the outside of the Ventis MX4 with a soft, clean cloth. Make sure the sensor diffusion membrane, inside and out, is free of debris; wipe gently with a cloth or brush that is soft, clean, and dry. Make sure the aspirated monitor’s inlet is free of debris.

© 2014 Industrial Scientific Corporation

32

Ventis™ MX4 Product Manual

SERVICE Instructions are provided for battery service; monitor conversion (diffusion to aspirated and vice versa); sensor, sensor barrier and LCD service; and pump assembly service. Refer to the three-dimensional view diagrams to identify the parts referenced in the instruction sets, and for screw torque values. Read all instructions before beginning any monitor service. NOTICES  Before beginning any service tasks, power-off the monitor.  Only qualified staff should perform monitor service and should take the following precautions.  Take care not to touch battery contacts on the monitor or the battery itself.  Perform work in a clean air environment that is known to be nonhazardous.  Perform work on a nonconductive work surface.  Wear grounding straps. BATTERY PACKS NOTICES  WARNING: INSERT THE ALKALINE BATTERIES WITH THE CORRECT POSITIVE “ + ” AND NEGATIVE “-“ ORIENTATION. FAILURE TO FOLLOW PROPER BATTERY ORIENTATION WILL RESULT IN DAMAGE TO THE MONITOR.  WARNING: The Ventis MX4 is only approved for use with AAA battery types Energizer EN92 and Duracell MN2400. Do NOT mix battery types. Battery service instruction sets are provided below for each allowable monitor/battery pack combination. Please choose, read, and then follow the appropriate instruction set. Aspirated Monitor Battery Replacement For an aspirated monitor, two of the three Ventis MX4 battery packs can be used. The Extended Range Li-ion battery is replaced as a single part. The Alkaline battery unit consists of batteries and a pack for the batteries. ASPIRATED MONITOR BATTERY REPLACEMENT. Removing the Battery Pack. 1

Power-off the monitor.

2

Loosen the four captive screws on the lower portion of the pump module bottom (back of the module).

3

Loosen the single captive screw on the pump door.

4

Slide the pump door down; lift to reveal and access the monitor.

5

Lift and remove the monitor from the pump module; set aside the monitor.

6

Lift and remove the battery pack from inside the lower portion of the pump module.

Replacing the Battery Pack. If inserting the Alkaline battery pack, follow steps 1-5. If inserting the Extended Range Li-ion battery pack, follow step 5 ONLY.

33

1

To prepare the alkaline battery pack, unlatch and lift the hinged portion of the battery pack.

2

If needed, remove and set aside the spent batteries from within the pack.

3

Insert new batteries so that their negative contacts are in contact with the springs inside the pack. DO NOT MIX BATTERY TYPES.

4

Close the latch. A click sounds.

5

To properly place the Extended Range Li-ion (or Alkaline) battery pack inside the pump module, complete or observe the following:  the battery circuitry is exposed and faces the user;  a small groove at the back of the battery bottom fits over the small rib at the back of the pump module bottom.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

Reassembling the Aspirated Monitor. 1

Re-place the monitor inside the pump module. The monitor is display side up and its logo readable. Its lower exposed bottom portion covers the battery.

2

Tighten the four screws on the pump module bottom to secure the module to the monitor.

3

Lower and close the pump door; slide up to click in place.

4

Tighten the pump door screw to secure the door in place.

5

Dispose of any spent batteries according to company policy.

Diffusion Monitor Battery Replacement or Changeover The diffusion monitor can be used with all three Ventis MX4 battery packs.  The Li-ion battery kit is a single part consisting of the monitor’s lower case bottom and the battery. It is removed from and attached to the diffusion monitor as a single item.  The Extended Range Li-ion battery pack and its cover are two distinct items.  The Alkaline battery pack consists of three items: the batteries, the battery pack, and a battery pack cover (the same cover that holds the Extended Range Li-ion battery). NOTE: When an Extended Range Li-ion (or Alkaline) battery pack is attached to a diffusion monitor, the monitor’s lower portion becomes deeper than its upper portion. The Conversion Kit’s suspender clip components are used to make the upper and lower portions depthcompatible. The Conversion Kit provides a battery cover (battery not included), a suspender clip and spacer, and a screw and washer.

ATTACHING THE LI-ION BATTERY TO A DIFFUSION MONITOR. Battery Replacement. 1

Power-off the monitor.

2

Loosen the four captive screws on the lower portion of the case bottom (or the battery cover if the Extended Range Li-ion or Alkaline battery pack is attached).

3

Lift and remove the Li-ion kit (or Extended Range Li-ion battery Alkaline battery pack and cover); set aside.

4

To properly attach the new Li-ion battery kit to the monitor, align its exterior charging contacts with the monitor’s bottom.

5

Tighten the four captive screws to secure the battery pack to the monitor.

6

Dispose of any spent batteries according to company policy, or properly store any battery packs that have a remaining life.

To changeover from an Extended Range Li-ion (or Alkaline) battery pack to a Li-ion battery continue with the following suspender clip instruction sets. Removal of the Suspender Clip and Spacer. 1

On the monitor’s case bottom, open the suspender clip.

2

Using a Phillips head screwdriver, remove the screw and washer that attach the suspender clip and spacer to the monitor.

3

Set aside and store the removed items for future use with the Extended Range Li-ion (or Alkaline) battery pack on a diffusion monitor.

Attachment of the Suspender Clip. 1

Open the suspender clip.

2

Place the washer on the screw (P/N 17139262).

3

Insert the screw with washer through the clip’s center hole and into the screw hole on the monitor.

4

Tighten the screw to secure the suspender clip to the monitor.

© 2014 Industrial Scientific Corporation

34

Ventis™ MX4 Product Manual

ATTACHING THE EXTENDED RANGE LI-ION (OR ALKALINE) BATTERY PACK TO A DIFFUSION MONITOR. Battery Removal. 1

Power-off the monitor.

2

Loosen the four captive screws on the lower portion of the Case Bottom (back of the monitor.

3

Lift the battery unit to remove; set it aside.

Battery Attachment. If inserting the Alkaline battery, follow steps 1-7. If inserting the Extended Range Li-ion battery, follow steps 5-7 ONLY. 1

To prepare the Alkaline battery pack, unlatch and lift the hinged lid of the battery housing.

2

If needed, remove and set aside the spent batteries from within the housing.

3

Insert new batteries so that their negative contacts are in contact with the springs inside the housing. DO NOT MIX BATTERY TYPES.

4

Close and latch the lid. A click sounds.

5

To properly place the Extended Range (or Alkaline) battery pack inside the battery cover ensure:  the battery circuitry is exposed and faces the user;  a small groove at the back of the battery bottom fits over the small rib at the back of the pump module bottom.

6

To properly attach the cover (and its enclosed battery) to the monitor, ensure the battery contacts are showing at the bottom of the monitor. NOTE: each of the above battery packs easily fits within the battery cover. If the battery does not easily insert, stop to ensure proper placement as noted above.

7

Tighten the four captive screws to secure the battery cover to the monitor.

To changeover from a Li-ion battery kit to an Extended Range Li-ion (or Alkaline) battery pack, continue with the following suspender clip instruction sets. Removal of the Suspender Clip. 1

Open the suspender clip.

2

Using a Phillips head screwdriver, remove the screw and washer that attach the clip to the monitor

3

Set aside and store the removed clip, washer, and screw for future use with a diffusion monitor and Li-ion battery pack.

Attachment of the Suspender Clip and Spacer (from the Conversion Kit).

35

1

Place the monitor face down.

2

The spacer has one flat end. Place the spacer on the monitor’s case back so its flat end meets top edge of the battery cover; the hole in the spacer aligns with the screw hole on the monitor.

3

Place the washer on the screw.

4

Insert the screw with washer through the clip’s hole and through the center hole on the clip back.

5

Place the clip on top of the spacer, inserting its screw into the spacer’s hole. Be sure the clip’s ring is at the top of the monitor.

6

Grasp the clip and spacer with one hand. With the other hand, tighten the screw with the Phillips screwdriver.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual MONITOR CONVERSION To convert a diffusion monitor to an aspirated monitor, only the Extended Range Li-ion or Alkaline battery packs are approved for use. To convert an aspirated monitor to a diffusion monitor, a suspender clip, washer, and screw are recommended for use with the Li-ion battery pack. If attaching the Extended Range Li-ion or Alkaline battery packs, the conversion kit is required. NOTE: When an Extended Range Li-ion (or Alkaline) battery pack is attached to a diffusion monitor, the monitor’s lower portion becomes deeper than its upper portion. The Conversion Kit’s suspender clip components are used to make the upper and lower portions depthcompatible. The Conversion Kit provides a battery cover (battery not included), a suspender clip and spacer, and a screw and washer.

CONVERTING A DIFFUSION MONITOR TO AN ASPIRATED MONITOR. Removing the Suspender Clip and Battery Pack. 1

Power-off the monitor.

2

On the monitor’s case bottom (back of the monitor), open the suspender clip.

3

Using a Phillips head screwdriver, remove the screw and washer that attach the suspender clip (and spacer, if applicable) to the monitor. Set aside and store any removed, unused items for future use with a diffusion monitor.

4

Loosen the four captive screws on the lower portion of the case bottom. Lift and remove the battery pack (and cover, if applicable). Set aside and store any removed, unused items for future use with a diffusion monitor.

Assembling the Aspirated Monitor (including battery insertion). If inserting the Alkaline battery pack, follow steps 1-10. If inserting the Extended Range Li-ion battery pack, follow steps 5-10 ONLY. 1

To prepare the Alkaline battery pack, unlatch and lift the hinged portion of the battery pack.

2

If needed, remove and set aside any spent batteries from within the pack.

3

Insert new batteries so that their negative contacts are in contact with the springs inside the pack. DO NOT MIX BATTERY TYPES.

4

Close the latch. A click sounds.

5

To properly place the Extended Range Li-ion (or Alkaline) battery pack inside the pump module, complete or observe the following:  the battery circuitry is exposed and faces the user;  a small groove at the back of the battery bottom fits over the small rib at the back of the pump module bottom.

6

Loosen the captive screw on the front of the pump module. Slide the module’s door down and lift to open.

7

Place the monitor inside the pump module. The monitor is display side up and its logo readable. Its lower exposed bottom portion covers the battery.

8

Tighten the four screws on the pump module bottom to secure the module to the monitor.

9

Close the pump door; slide up to click in place.

10

Tighten the pump door screw.

CONVERTING AN ASPIRATED MONITOR TO A DIFFUSION MONITOR. Removing the Pump Module. 1

Power-off the monitor.

2

Loosen the four captive screws on the lower portion of the pump module bottom (back of the module).

3

Loosen the single captive screw on the pump module door.

4

Slide the pump door down; lift to reveal and access the monitor.

5

Lift and remove the monitor from the pump module; set aside and store for future use.

© 2014 Industrial Scientific Corporation

36

Ventis™ MX4 Product Manual

Choose OPTION 1 or OPTION 2 below depending on the battery pack to be attached. OPTION 1: Attaching the Li-ion Battery Kit and its Compatible Suspender Clip Components. 1

To properly place the Li-ion battery kit, align its contacts with the monitor’s contacts, at the monitor bottom.

2

To secure the battery kit to the monitor, tighten the four captive screws on the kit bottom.

3

Open the suspender clip.

4

Place the washer on the screw.

5

Insert the screw with washer through the clip’s center hole and into the screw hole on the monitor. Be sure the clip’s ring is at the top of the monitor.

6

Tighten the screw.

OPTION 2: Attaching the Extended Range Li-ion (or Alkaline) Battery Pack and its Compatible Suspender Clip Components. If inserting the Alkaline battery pack, follow steps 1-14. If inserting the Extended Range Li-ion battery pack, follow steps 5-14 ONLY. 1

To prepare the Alkaline battery pack, unlatch and lift the hinged portion of the battery pack.

2

If needed, remove and set aside any spent batteries from within the pack.

3

Insert new batteries so that their negative contacts are in contact with the springs inside the pack. DO NOT MIX BATTERY TYPES.

4

Close the latch. A click sounds.

5

To properly place the Extended Range Li-ion (or Alkaline) battery pack inside the battery cover, ensure:  the battery circuitry is exposed and faces the user;  a small groove at the back of the battery bottom fits over the small rib at the back of the pump module bottom. NOTE: each of these two battery packs easily fits within the battery case. If the battery does not easily insert, stop to ensure proper placement as noted above.

37

6

To properly attach the cover containing the Extended Range Li-ion or Alkaline battery pack, ensure the battery contacts are showing at the bottom of the monitor.

7

To secure the battery cover (and its enclosed battery pack) to the monitor, tighten the four captive screws on the battery cover.

8

Place the monitor face down.

9

The spacer for the suspender clip has one flat end. Place the spacer on the monitor’s case back so its flat end meets top edge of the battery cover; the hole in the spacer aligns with the screw hole on the monitor.

10

Open the suspender clip.

11

Place the washer on the screw.

12

Insert the screw with washer through the clip’s hole and through the center hole on the clip back.

13

Place the clip on top of the spacer, inserting its screw into the spacer’s hole and ensuring the clip’s ring is at the top of the case back.

14

Grasp the clip and spacer with one hand. With the other hand, tighten the screw with the Phillips screwdriver.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual SENSOR, SENSOR BARRIER, LCD, AND VIBRATING MOTOR REPLACEMENT Service instruction sets are provided below for each monitor type. Please choose, read, and then follow the appropriate instruction set . Within each set of instructions, follow those relevant to the desired task(s) and note the following.  The monitor has a two-part circuit board assembly, the main board and a smaller sensor board. They are attached to one another with a connecter at the center of the sensor board.  The sensor barrier can be replaced as an assembly that fits in the monitor’s case top, or the full case top can be replaced. NOTE: When a sensor is replaced, it is recommended that the sensor barrier/case top also be replaced. After reassembling the 

monitor, a full calibration should be completed.

The LCD is removed and attached as a single component. ASPIRATED Disassembling the Monitor. 1

Power-off the monitor.

2

Loosen the four captive screws on the lower portion of the pump case module bottom (back of the module).

3

Loosen the single captive screw on the pump case module top.

4

Slide the case door down; lift the hinged door to reveal and access the monitor.

5

Lift and remove the monitor from the pump module; set aside the module.

6

Place the monitor display side down.

7

Loosen the two captive screws on the upper portion of the case bottom.

8

Lift to separate the monitor case top from the monitor case bottom to reveal the circuit board assembly.

9

Remove the circuit board assembly and set aside the monitor case top and bottom.

10

Separate the main circuit board from the sensor board.

Replacing the LCD (if needed). 1

Grasp the sides of the LCD and lift straight up to remove from the main circuit board.

2

To properly place the new LCD, align the pins on the LCD with their receptacles on the main circuit board.

3

Gently press straight down and into place.

Replacing the Sensor(s) (if needed). 1

Identify the sensor to be removed.

2

Gently lift and remove the sensor.

3

To add the new sensor, align its pins or connector(s), with the respective receptacles on the sensor board.

4

Press down. A slight click indicates the sensor is securely in place.

Reassembling the Circuit Board Assembly. 1

Re-attach the main circuit board to the sensor board, aligning their connectors.

2

Press. A slight click indicates the boards are securely attached.

Replacing the Sensor Barrier or Case Top (if needed). To replace the sensor barrier on the inside of the case top, follow steps 1-5 below. To replace the entire case top, skip to the instruction set, “Reassembling the Monitor”. 1

Observe the placement of the existing sensor barrier. Note that each cut-out is shaped to match the sensor it protects.

2

Lift and remove the sensor barrier and gasket from inside the monitor case top. Ensure the entire case top is free of adhesive; gently scrape, if needed. Wipe with a clean, dry, soft cloth or brush.

3

Lift the backing from the new sensor barrier assembly to reveal the adhesive.

4

Carefully position the new barrier. Each shaped opening matches the shape of the sensor it protects. Press to attach to the inside of the case top.

© 2014 Industrial Scientific Corporation

38

Ventis™ MX4 Product Manual

Replacing the Vibrating Motor (if needed). 1

Place the monitor’s case top face down.

2

Lift the vibrating motor from its partition. The partition has two sections divided by a ridge. Discard the used motor.

3

To properly place the new vibrating motor, its contact pins face the user and align with the left edge of the partition. (The motor’s movable component fits within the small section of the partition.)

4

Press into place.

Reassembling the Monitor. 1

Re-place the board assembly into the monitor’s case bottom. The LCD faces the user.

2

Re-place the monitor’s case top (or place its new case top).

3

Tighten the two captive screws on the upper portion of the monitor case bottom.

4

Re-place the monitor inside the pump module. The monitor is display side up and its logo readable. Its lower exposed bottom portion covers the battery.

5

Tighten the four screws on the pump module bottom to secure the module to the monitor.

6

Close the pump module door; slide up to click in place.

7

Tighten the pump door screw to secure.

8

Dispose of the used sensor(s) according to company policy.

9

Perform a full calibration following the addition or replacement of any sensor, or the replacement of the sensor water barrier or monitor case top.

DIFFUSION Disassembling the Monitor. 1

Power-off the monitor.

2

Loosen the four captive screws on the lower portion of the case bottom (back of the monitor) to remove the battery pack. Set aside the battery kit (or pack and cover if applicable).

3

Loosen the two captive screws on the upper portion of the case bottom.

4

Lift to separate the case top from the case bottom.

5

Remove the circuit board assembly.

6

Separate the main circuit board from the sensor board.

Replacing the LCD (if needed). 1

Grasp the sides of the LCD and lift straight up to remove.

2

To properly position the new LCD, align the pins on the LCD with their receptacles on the main circuit board.

3

Gently press straight down and into place.

Replacing the Sensor(s) (if needed). 1

Identify the sensor to be removed.

2

Gently lift and remove the sensor.

3

To add the new sensor, align its pins or connector(s), with the respective receptacles on the board.

4

Press down. A slight click indicates the sensor is securely in place.

Replacing the Sensor Barrier (if needed). To replace the sensor barrier assembly, follow instruction 1-5 below. To replace the entire case top, including its sensor barrier, skip to the next instruction set, “Reassembling the circuit board assembly and monitor.” 1

39

Observe the placement of the existing water barrier. Note that each cut-out is shaped to match the sensor it protects.

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

2

Lift and remove the sensor barrier and gasket from the inside the monitor case top. Ensure the entire case top is free of adhesive; gently scrape, if needed. Wipe with a clean, dry, soft cloth or brush.

3

Lift the backing from the new sensor barrier to reveal the adhesive.

4

Carefully position the new barrier. Each shaped opening matches the shape of the sensor it protects. Press to attach to the inside of the case top.

Replacing the Vibrating Motor (if needed). 1

Place the monitor’s case top face down.

2

Lift the vibrating motor from its partition. The partition has two sections divided by a ridge. Discard the used motor.

3

To properly place the new vibrating motor, its contact pins face the user and align with the left edge of the partition. (The motor’s movable component fits within the small section of the partition.)

4

Press into place.

Reassembling the Circuit Board Assembly and Monitor. 1

Re-attach the main board to the sensor board, aligning their connectors.

2

Press. A slight click indicates the boards are securely attached.

3

Re-place the circuit board assembly into the monitor’s case bottom.

4

Re-place the monitor’s case top (or its new case top).

5

Tighten the two captive screws on the upper portion of the case bottom.

6

Re-place the battery pack and tighten the four captive screws on the case bottom.

7

Dispose of the used sensor(s) according to company policy.

8

Perform a full calibration following the addition or replacement of any sensor, or the replacement of the sensor water barrier or the monitor case top.

PUMP MODULE There are two field replaceable parts on the aspirated case: the water barrier and case top door. As needed, choose, read, and follow the appropriate instruction set(s). pump module Pump Inlet Filter Cap and/or Water Barrier Replacement (if needed). 1

Power off the monitor.

2

To remove the pump inlet filter cap, turn counter clockwise; set aside.

3

Remove the water barrier from inside the inlet.

4

To properly place the new water barrier, its bottom will come into contact with the pump inlet surface on which the barrier sits. The barrier’s bottom side has the smaller diameter opening in the black ring.

5

Re-place the pump inlet filter cap (or new cap); turn clockwise to tighten.

Pump Door Removal and Replacement (if needed). 1

Power off the monitor.

2

Loosen the single captive screw on the pump module door.

3

Slide the door down and lift.

4

The upper portion of the door has pegs on either side. The pegs fit into grooves where the door meets the module; they flex slightly for easy door removal. Angle the door so that one peg moves to the bottom of its groove and the other moves the top of its groove. Lift to unhinge the pegs and remove the door.

5

To properly place the new door, secure one of the pegs into its groove on the pump module.

6

Angle the door and press the other peg into its groove.

7

Close the door and slide up into place.

8

Tighten the single captive screw on the pump door.

© 2014 Industrial Scientific Corporation

40

Ventis™ MX4 Product Manual

VENTIS MX4 MONITOR THREE-DIMENSIONAL DIAGRAM

41

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

KEY FOR VENTIS MX4 MONITOR THREE-DIMENSIONAL DIAGRAM Number

Part Number (P/N)

1

17152380-X

Description Ventis MX4 Diffusion Case Top Assembly (includes items 2 and 3) X = Case Color, where: 0 = Black, 1 = Orange

2

17152429

Sensor Barrier Assembly

3

17145285

Vibrating Motor

4

17150772

Ventis MX4 LCD Assembly

5

17134495

Ventis MX4 Sensor, Combustible Gas (LEL/CH4)

6

17134461

Ventis MX4 Sensor, Oxygen (O2)

7

17134487

Ventis MX4 Sensor, Carbon Monoxide (CO)

7

17155564

Ventis MX4 Sensor, Carbon Monoxide/Low Hydrogen Interference (CO/H2 Low)

8

17134479

Ventis MX4 Sensor, Hydrogen Sulfide (H2S)

9

17134503

Ventis MX4 Sensor, Nitrogen Dioxide (NO2)

10

17143595

Ventis MX4 Sensor, Sulfur Dioxide (SO2)

11

17147281

Captive Case Screw, Torx (torque value: 55 oz-in or .39 N.m +/- 10%)

11

17147273

Captive Case Screw, Phillips (torque value: 55 oz-in or .39

12

17152506

Suspender Clip Spacer

13

17120528

Suspender Clip

14

17153137

Locking Washer

15

17152507

N.m +/- 10%)

Screw, Phillips (for use with items 12, 13 and 14) (torque value: 115 oz-in or .81 N.m +/- 10%)

16

17139262

17

17134453-XY

Screw, Phillips (torque value: 115

oz-in or .81 N.m +/- 10%)

Lithium-ion Battery Kit X = Battery Cover Color, where: 0 = Black, 1 = Orange (captive screw torque value: 55 oz-in or .39 N.m +/- 10%) Y=1: UL, CSA, ATEX, IECEx, GOST-R, GOST-K, KOSHA, MED, and SANS approvals Y=2: MSHA Y=3: China Ex Y=4: ANZEx Y=5: INMETRO

18

17148313-Y

Extended Range Lithium-ion Battery Pack Y=1: UL, CSA, ATEX, IECEx, INMETRO, GOST-R, GOST-K, KOSHA, MED, and SANS approvals Y=2: MSHA Y=3: China Ex Y=4: ANZEx

19

17150608

Alkaline Battery Pack UL, CSA, ATEX, IECEx, ANZEx, China MA, China KA, and INMETRO approvals (CSA, China MA, and China KA approved for diffusion monitor only.)

20

17151184-XY

Cover, Extended Range Lithium-ion or Alkaline X = Cover Color, where: 0 = Black, 1 = Orange (captive screw torque value: 55 oz-in or .39 n.m . +/- 10%) Y=1: UL, CSA, ATEX, IECEx, GOST-R, GOST-K, and KOSHA approvals Y=2: MSHA (Extended Range Lithium-ion only) Y=3: China Ex Y=4: ANZEx Y=5: INMETRO Y=C: China KA

* Item is not user replaceable. The Ventis MX4 monitor must be sent to an authorized ISC Service Center for this item to be replaced.

© 2014 Industrial Scientific Corporation

42

Ventis™ MX4 Product Manual VENTIS MX4 PUMP MODULE THREE-DIMENSIONAL DIAGRAM

KEY FOR VENTIS MX4 PUMP MODULE THREE-DIMENSIONAL DIAGRAM Number

Part Number (P/N)

1

17151150-X0

Description Ventis MX4 Pump Door Assembly X = Pump Door Assembly Color, where: 0 = Black, 1 = Orange (captive screw torque value: 55 oz. in. or .39 n.m . +/- 10%)

2

17129909

Pump Inlet Filter Cap

2

17141581

Pump Inlet-Filter Cap for use 6’ extendable probe

2

17141599

Filter Cap, 1/8 NPT Female

3

17152395

Water Barrier

4

17151275

Screw, Torx (torque value: 55 oz-in or .39 N.m +/- 10%)

5

17052558

Screw (torque value: 25 oz-in or .17 N.m +/- 10%)

6

17148313-Y

Extended Range Lithium-ion Battery Pack Y = Approval where: 1 = UL, CSA, ATEX, IECEx, INMETRO, GOST-R, GOST-K, and KOSHA 2 = MSHA 3 = China Ex 4 = ANZEx

7

43

17150608

Alkaline Battery Pack: UL, ATEX, IECEx, ANZEx, and INMETRO approvals (CSA, China KA and China MA approved for diffusion monitor only.)

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

8

17151002

Captive Case Screw, Phillips (torque value: 55 oz-in or .39 N.m +/- 10%)

8

17151028

Captive Case Screw, Torx (torque value: 55 oz-in or .39 N.m +/- 10%)

9

17151010

Captive Case Screw, Phillips (torque value: 55 oz-in or .39 N.m +/- 10%)

9

17151036

Captive Case Screw, Torx (torque value: 55 oz-in or .39 N.m +/- 10%)

* Item is not user replaceable. The Ventis MX4 Pump Module must be sent to an authorized ISC Service Center for this item to be replaced.

►Products, Specifications, and Certifications VENTIS MX4 ACCESSORIES AND PARTS Docking and Calibration Stations 18108630-0BC

DS2 Docking Station™ for Ventis MX4 B = Quantity of iGas® Readers C = Power Cord, where 0 = US, 1 = UK, 2 = EU, 3 = AUS, 4 = ITA, 5 = DEN, 6 = SWZ

18108631-AB

V•Cal ™ Calibration Station A = Instrument Type, where 0 = Diffusion, 1 = Aspirated B = Power Cord, where 0 = US, 1 = UK, 2 = EU, 3 = AUS, 4 = ITA, 5 = DEN, 6 = SWZ

Printer 18107763

Serial data dot matrix printer for V•Cal™ – 120 VAC powered

Chargers 18108191

Single-Unit Charger

18108209

Single-Unit Charger/Datalink

18108651

Single-Unit Automotive Charger, 12VDC

18108652

Single-Unit Truck-Mount Charger, 12VDC, with Cigarette Adapter

18108653

Single-Unit Truck-Mount Charger, 12VDC, Hard Wired

18108650-A

6-Unit Charger

Pumps 18108830 VTSP-ABCD

Ventis MX4 Hand Pump (manual bulb) Ventis MX4 Pump Module* A = Battery Type, where 0 = No Battery, 2 = Li-ion Extended Range Battery Pack, 3 = Alkaline Battery Pack B = Color, where: 0 = Black, 1 = Orange C = Approvals, where 1 = UL and CSA; 2 = ATEX, and IECEx,; 3=MSHA; 4= ANZEx; 5 = China EX; 7 = GOST-R and GOST-K; 8 = KOSHA; 9 = INMETRO; and A = MED D = Assembly Guide language, where: 1 = English, 2 = French, 3 = Spanish, 4 = German, 5 = Italian, 6 = Dutch, 7 = Portuguese, 9 = Russian, A = Polish, B = Czech, C = Chinese, D = Danish, E = Norwegian, F = Finnish, G = Swedish

*Ventis MX4 Pump Module replacement parts 17151150-X0

Ventis MX4 Pump Door Assembly X = Pump Door Assembly Color, where: 0 = Black, 1 = Orange

17129909

Pump Inlet Filter Cap

17152395

Water Barrier

17151275

Screw, Torx

17052558

Screw

17148313-Y

Extended Range Lithium-ion Battery Pack, Y = Approvals where: 1 = UL, CSA, ATEX, IECEx, INMETRO, GOST-R, GOST-K, and KOSHA 2 = MSHA 3 = China Ex 4 = ANZEx

17151002

Captive Case Screw, Phillips

17151028

Captive Case Screw, Torx

© 2014 Industrial Scientific Corporation

44

Ventis™ MX4 Product Manual

17151010

Captive Case Screw, Phillips

17151036

Captive Case Screw, Torx

17116096

Calibration Tubing Assembly with Integrated “T” Fitting (For use when calibrating a monitor with pump using a positive flow regulator)

Carrying Cases 18108175

Ventis MX4 Diffusion Soft Carrying Case, Lithium-ion Battery (black)

18109150

Ventis MX4 Diffusion Soft Carrying Case, Lithium-ion Battery (orange)

18108183

Ventis MX4 Diffusion Soft Carrying Case, Extended Range Lithium-ion Battery or Alkaline (black)

18109151

Ventis MX4 Diffusion Soft Carrying Case, Extended Range Lithium-ion Battery or Alkaline (orange)

18108813

Ventis MX4 Diffusion Hard Carrying Case with Display, Lithium-ion Battery

18108814

Ventis MX4 Diffusion Hard Carrying Case with Display, Extended Range Lithium-ion Battery or Alkaline

18108815

Ventis MX4 Diffusion Hard Carrying Case without Display, Lithium-ion Battery

18108816

Ventis MX4 Diffusion Hard Carrying Case without Display, Extended Range Lithium-ion Battery or Alkaline

18108810

Ventis MX4 Aspirated Soft Carrying Case (black)

18109152

Ventis MX4 Aspirated Soft Carrying Case (orange)

18108811

Ventis MX4 Aspirated Hard Carrying Case with Display

18108812

Ventis MX4 Aspirated Hard Carrying Case without Display

Clip Assembly 17120528

Suspender Clip (for diffusion monitor with Lithium-ion Battery Pack)

17139262

Clip screw and washer (for diffusion monitor with Lithium-ion Battery Pack)

Sensors, Sensor Barrier, Vibrating Motor, LCD, and Calibration Cup 17134495

Ventis MX4 Sensor, Combustible Gas (LEL/CH4)

17134461

Ventis MX4 Sensor, Oxygen (O2)

17134487

Ventis MX4 Sensor, Carbon Monoxide (CO)

17155564

Ventis MX4 Sensor, Carbon Monoxide/Low Hydrogen Interference (CO/H2 Low)

17134479

Ventis MX4 Sensor, Hydrogen Sulfide (H2S)

17134503

Ventis MX4 Sensor, Nitrogen Dioxide (NO2)

17143595

Ventis MX4 Sensor, Sulfur Dioxide (SO2)

17152380-X

Diffusion Case Top Assembly (includes Sensor Barrier Assembly) X = Case Top Color, where: 0 = Black, 1 = Orange

17152429

Sensor Barrier Assembly

17145285

Vibrating Motor

17150772

Ventis MX4 LCD Assembly

17156189

Ventis Calibration Cup Assembly, calibration cup with tubing

17152455

Ventis Calibration Cup

Battery Packs and Components 17134453-XY

Lithium-ion Battery Kit X = Battery Cover Color, where: 0 = Black, 1 = Orange (captive screw torque value: 55 oz-in or .39 N.m +/- 10%) Y = Approvals where: 1 = UL, CSA, ATEX, IECEx, GOST-R, GOST-K, KOSHA, MED, and SANS 2 = MSHA 3 = China Ex 4 = ANZEx

45

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

5 = INMETRO

17148313-Y

Extended Range Lithium-ion Battery Pack Y = Approvals where: 1 = UL, CSA, ATEX, IECEx, INMETRO, GOST-R, GOST-K, KOSHA, MED, and SANS 2 = MSHA 3 = China Ex 4 = ANZEx

17150608

Alkaline Battery Pack: UL, ATEX, IECEx, ANZEx, and INMETRO approvals (CSA, China KA and China MA approved for diffusion monitor only.)

Conversion Kit** for the diffusion monitor. NOTE: For use in converting an aspirated monitor with a rechargeable Extended Range Lithium-ion battery pack to a diffusion monitor with a rechargeable Extended Range Lithium-ion (or Alkaline) Battery Pack. (Batteries sold separately.) Kit parts listed below can be ordered separately as replacement parts.

**17151184-XY

Cover, Extended Range Lithium-ion or Alkaline X = Cover Color, where: 0 = Black, 1 = Orange (captive screw torque value: 55 oz-in or .39 n.m . +/- 10%) Y = Approvals where: 1 = UL, CSA, ATEX, IECEx, GOST-R, GOST-K, and KOSHA; MED (Lithium-ion only) 2 = MSHA (Extended Range Lithium-ion only) 3 = China Ex 4 = ANZEx 5 = INMETRO C = China KA

**17152506

Suspender Clip Spacer

**17120528

Suspender Clip

**17153137

Locking Washer

** 17152507

Screw, Phillips

MONITOR SPECIFICATIONS Item

Description

Display

Backlit Liquid Crystal Display (LCD)

Buttons

Two (ON/OFF/MODE and ENTER)

Monitor case

Polycarbonate with ESD protective rubber overmold

Alarms

Ultra-bright LEDs, loud audible alarm (95dB at 30 cm), and vibrating alarm

Size and Weight

Diffusion with Lithium-ion (typical)

Aspirated with Extended Range Lithium-ion (typical)

Size

103 mm x 58 mm x 30 mm (4.1” x 2.3” x 1.2”)

172 mm x 67 mm x 66 mm (6.8” x 2.6” x 2.6”)

Weight

182 g (6.4 oz)

380 g (13.4 oz)

OPERATING CONDITIONS Warm-up time

40 seconds (includes stabilization time)

Temperature range

-20 °C to +50 °C (-4 °F to +122 °F)

Humidity range

15–95% relative humidity (RH) noncondensing (during continuous operation)

Pressure range

1 atm ±20%

 

         

© 2014 Industrial Scientific Corporation

46

Ventis™ MX4 Product Manual

STORAGE CONDITIONS Temperature range

0−25 ºC (32−77 ºF)

Humidity range

40–70% relative humidity (RH) noncondensing

Pressure range

0.9–1.1 atm

Maximum time

Up to 6 months Note: Industrial Scientific recommends that infrequently used lithium-ion batteries be fully charged every four months.

  SENSOR SPECIFICATIONS Gas Name

Abbr.

Measuring Range

Resolution

Accuracy at Time and Response Temperature of Time Calibration* (typical) T50

Response Time (typical) T90

Oxygen

O2

0–30% vol

0.1% vol

± 0.5%

15

30

Carbon Monoxide with low H2 interference

CO/H2 Low

0–1000 ppm

1 ppm

± 5% (0-300 ppm) ± 15% (301-1000 ppm)

8

17

Carbon Monoxide

CO

0–1000 ppm

1 ppm

± 5%

15

50

Hydrogen Sulfide

H2 S

0–500 ppm

0.1 ppm

± 5%

15

30

Nitrogen Dioxide

NO2

0–150 ppm

0.1 ppm

± 10%

10

30

Sulfur Dioxide

SO2

0–150 ppm

0.1 ppm

± 10%

20

80

Combustible

LEL

0–100% LEL

1% LEL

± 5%

15

35

Methane

CH4

0–5% vol

0.01% vol

± 5%

15

35

*The accuracy specification for each sensor is ± the stated percentage or 1 unit of resolution, whichever is greater.

TOXIC GAS SENSOR CROSS-SENSITIVITY TABLE Target Gas

CO

Sensor

CO

CO (H2 Low)

H2S

SO2

NO2

Cl2

ClO2

HCN

HCl

PH3

NO

H2

NH3

100

100

1

1

0

0

0

0

0

0

0

20

0

H2S

5

5

100

1

-40

-3

-25

10

300

25

10

20

25

SO2

0

5

5

100

0

0

0



40



0

0

-40

NO2

-5

5

-25

-165

100

45



-70





30

0

-10

Cl2

-10

0

-20

-25

10

100

60

-20

6

-20

0

0

-50

ClO2











20

100













HCN

15



10

50

1

0

0

100

35

1

0

30

5

HCl

3



0

5

0

2

0

0

100

0

15

0

0

PH3













-100

425

300

100







NO

25

40

1

1

5





-5





100

30

0

H2

22

3

0.3

0.5

0

0

0

0

0

0

0

100

0

NH3

0

0

0

0

0

0

0

0

0

0

0

0

100

47

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual The Sensor Cross Sensitivity Table (above) reflects the percentage response provided by the sensor (top row) when exposed to a known concentration of the target gas (column 1). The numbers were measured under these environmental conditions: 20 °C (68 °F) , 50% RH and 1 atm. The specified cross-interference numbers apply to new sensors only, and may vary with time as well as from sensor to sensor. “—” means no data available. This table is given as a reference only and is subject to change.

LEL AND LEL CORRELATION FACTORS FOR COMBUSTIBLE GASES Sample gas*

LEL (% vol)

LEL correlation factors Calibration gas Butane

Hexane

Hydrogen

Methane

Pentane

Propane

Acetone

2.5%

1.00

0.70

1.70

1.70

0.90

1.10

Acetylene

2.5%

0.70

0.60

1.30

1.30

0.70

0.80

Benzene

1.2%

1.10

0.80

1.90

1.90

1.00

1.20

Butane

1.9%

1.00

0.58

1.78

1.67

0.83

1.03

Ethane

3.0%

0.80

0.60

1.30

1.30

0.70

0.80

Ethanol

3.3%

0.89

0.52

1.59

1.49

0.74

0.92

Ethylene

2.7%

0.80

0.60

1.40

1.30

0.70

0.90

Hexane

1.1%

1.71

1.00

3.04

2.86

1.42

1.77

Hydrogen

4.0%

0.56

0.33

1.00

0.94

0.47

0.58

Isopropanol

2.0%

1.10

0.90

2.00

1.90

1.00

1.20

Methane

5.0%

0.60

0.35

1.06

1.00

0.50

0.62

Methanol

6.0%

0.60

0.50

1.10

1.10

0.60

0.70

Nonane

0.8%

2.22

1.30

3.95

3.71

1.84

2.29

Pentane

1.4%

1.21

0.71

2.15

2.02

1.00

1.25

Propane

2.1%

0.97

0.57

1.72

1.62

0.80

1.00

Styrene

0.9%

1.30

1.00

2.20

2.20

1.10

1.40

Toluene

1.1%

1.53

0.89

2.71

2.55

1.26

1.57

Xylene

1.1%

1.50

1.10

2.60

2.50

1.30

1.60

JP-4











1.20



JP-5











0.90



JP-8











1.50



NOTE: The table above provides the LEL for select combustible gases*. It also provides correlation factors that help the safety technician and instrument operator determine the actual percentage LEL when the sample gas differs from the gas that was used to calibrate the unit. For example, if the unit reads 10% LEL in a pentane atmosphere, and was calibrated to methane, the actual percentage LEL is determined as follows: 1. Locate the table cell where the sample gas (pentane) intersects with the calibration gas (methane). 2. Multiply the cell's value (2.02) by the unit's LEL reading (10%) to calculate the actual concentration of 20.2% LEL. * The combustible gas list is not a comprehensive list of all combustible gases that can be detected by the Ventis MX4. For additional information about combustible gas detection and the Ventis MX4, contact the ISC Technical Service department.

© 2014 Industrial Scientific Corporation

48

Ventis™ MX4 Product Manual

CERTIFICATIONS Directive/Code

Certification Marking

ATEX

Ex ia IIC T4 Ga and Ex ia I Ma; Equipment Group and Category II 1G and I M1; IP66; IP67

ANZEx 

Ex ia s Zone 0 I/IIC T4; IP66; IP67

China Ex

Ex ia IIC T4 Ga

China CMC

Metrology Approval

China MA

Approved for Underground Mines; diffusion (without pump) alkaline version only

CSA

Class I, Division 1, Groups A B C D, T4; Ex d ia IIC T4 C22.2 No. 152 for %LEL reading only

GOST – R and GOST - K

PBExdiaI X and 1ExdiaIICT4 X

IECEx

Ex ia IIC T4 Ga; IP66; IP67

INMETRO

Ex ia IIC T4 Ga; IP66; IP67 -20°C ≤ Ta ≤ +50 °C

KOSHA

Ex d ia IIC T4

MED

Portable Multigas Detector; Category 2 (MED 2012/32/EU); Li-Ion only Charger/Docking Station accessories; category 1

MSHA

Permissible for Underground Mines; li-ion versions only

SANS-1515

SANS 1515-1 Type A; Ex ia I/IIC T4 IP66/67; Li-Ion only

UL

Class I, Division 1, Groups A B C D, T4; Zone 0, AEx ia IIC T4 Class II, Groups F G (Carbonaceous and Grain Dust); Class I, Zone 0, AEx ia IIC T4 IP66 ; IP67

Marking Requirements ATEX Markings

IECEx Markings

Industrial Scientific Corp. 15071 USA VENTIS MX4 DEMKO 10 ATEX 1006410 Ex ia IIC T4 Ga / Ex ia I Ma IP 66/67 o o -20 C ≤ Ta ≤ +50 C [Serial Number] [Month/Year of Production] Charging contact parameters: Um = 6.2V; Ii = 1.3A EN 60079-29-1; EN 50104

Industrial Scientific Corp. 15071 USA VENTIS MX4 IECEx UL10.0034 Ex ia IIC T4 Ga IP 66/67 o o -20 C ≤ Ta ≤ +50 C [Serial Number] [Month/Year of Production] Charging contact parameters: Um = 6.2V; Ii = 1.3A

Diffusion Version: Warning: Do not recharge or replace battery in hazardous locations. Replace only with P/N 17148313-1, P/N 17134453-X1, or P/N 17050608

Aspirated Version: Warning: Do not recharge or replace battery in hazardous locations. Replace only with P/N 17148313-1 or P/N 17050608

Li-Ion Battery Packs, P/N 17148313-1 or P/N 17134453-X1 Warning: Only charge instrument in non-hazardous locations. Charging contact parameters: Um = 6.2V; Ii = 1.3A; Do Not Connect in Hazardous Locations

Alkaline Battery Pack, P/N 17150608 Warning: Only replace batteries in non-hazardous locations; Only approved for use with three (3) AAA battery types Duracell MN2400 and Energizer EN92. Replace all batteries at the same time.

Warning: Read and understand manual before use. Warning: Read and understand manual before use.

49

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

►Warranty Industrial Scientific Corporation's Ventis MX4 portable gas monitors are warranted to be free from defects in material and workmanship for a period of two years after purchase. This warranty includes the sensors, the pump, and the lithium-ion battery pack as shipped with the Ventis MX4. Filters are warranted to be free from defects in material and workmanship for 18 months from date of shipment, or one year from date of first use, whichever occurs first, except where otherwise stated in writing in Industrial Scientific literature.

LIMITATION OF LIABILITY INDUSTRIAL SCIENTIFIC MAKES NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. SHOULD THE PRODUCT FAIL TO CONFORM TO THE ABOVE WARRANTY, BUYER’S ONLY REMEDY AND INDUSTRIAL SCIENTIFIC’S ONLY OBLIGATION SHALL BE, AT INDUSTRIAL SCIENTIFIC’S SOLE OPTION, REPLACEMENT OR REPAIR OF SUCH NON-CONFORMING GOODS OR REFUND OF THE ORIGINAL PURCHASE PRICE OF THE NON-CONFORMING GOODS. IN NO EVENT WILL INDUSTRIAL SCIENTIFIC BE LIABLE FOR ANY OTHER SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, INCLUDING LOSS OF PROFIT OR LOSS OF USE, ARISING OUT OF THE SALE, MANUFACTURE, OR USE OF ANY PRODUCTS SOLD HEREUNDER WHETHER SUCH CLAIM IS PLEADED IN CONTRACT OR IN TORT, INCLUDING STRICT LIABILITY IN TORT. It shall be an express condition to Industrial Scientific’s warranty that all products be carefully inspected for damage by Buyer upon receipt, be properly calibrated for Buyer’s particular use, and be used, repaired, and maintained in strict accordance with the instructions set forth in Industrial Scientific’s product literature. Repair or maintenance by non-qualified personnel will invalidate the warranty, as will the use of non-approved consumables or spare parts. As with any other sophisticated product, it is essential and a condition of Industrial Scientific’s warranty that all personnel using the products be fully acquainted with their use, capabilities, and limitations as set forth in the applicable product literature. Buyer acknowledges that it alone has determined the intended purpose and suitability of the goods purchased. It is expressly agreed by the parties that any technical or other advice given by Industrial Scientific with respect to the use of the goods or services is given without charge and at Buyer’s risk; therefore, Industrial Scientific assumes no obligations or liability for the advice given or results obtained.

© 2014 Industrial Scientific Corporation

50

Ventis™ MX4 Product Manual

51

© 2014 Industrial Scientific Corporation

Ventis™ MX4 Product Manual

►Ventis MX4 Resource Center Product documentation. Online training. And more! www.indsci.com/ventis

Global Locations Americas USA Corporate Headquarters 1001 Oakdale Road Oakdale, PA 15071-1500 USA Phone: +1 412-788-4353 1-800-DETECTS (338-3287) e-mail: [email protected] Canada Phone: +780-467-2423 e-mail: [email protected] Europe/Middle East/Africa France Phone: +00 800 WORKSAFE +33-157329261 e-mail: [email protected] Germany Phone: +49-69299571416 e-mail: [email protected] Czech Republic Phone: +420 234 622 222 e-mail: [email protected] United Arab Emirates Phone: +971 50 455 8518 e-mail: [email protected] England Phone: +44 12 80 70 61 14 e-mail: [email protected] Asia Pacific/China People's Republic of China Phone: +86 21 5899 3279 +86 400 820 2515 e-mail: [email protected] Singapore Phone: +65 6561 7377 e-mail: [email protected] Australia Phone: +03 96447777 e-mail: [email protected]

© 2014 Industrial Scientific Corporation

52

Product Manual The Essential Guide for Safety Teams and Instrument Operators Part Number: 17154993-1 Edition: 5 August 12, 2014

Industrial Scientific Corporation. Oakdale, PA USA Shanghai, China © 2012, 2013, 2014 Industrial Scientific Corporation All rights reserved. Published 2014 Revision 7

ii

Contents General Information .................................................................................................................................................................................................. 1  Introduction ........................................................................................................................................................................................................... 1  Certifications .................................................................................................................................................................................................... 1  Product Overview ................................................................................................................................................................................................. 2  Product Specifications .......................................................................................................................................................................................... 3  Recommended Practices .......................................................................................................................................................................................... 7  Introduction ........................................................................................................................................................................................................... 7  Procedures ........................................................................................................................................................................................................... 7  Recommendations................................................................................................................................................................................................ 8  First Use ............................................................................................................................................................................................................... 9  Wearing the Instrument ........................................................................................................................................................................................ 9  Instrument Basics ................................................................................................................................................................................................... 11  Unpacking the Instrument................................................................................................................................................................................... 11  Hardware Overview ............................................................................................................................................................................................ 12  Display Overview ................................................................................................................................................................................................ 12  Start-up and Shutdown ....................................................................................................................................................................................... 14  Instrument Preparation and Use ............................................................................................................................................................................. 17  Configuration ...................................................................................................................................................................................................... 17  Operation ............................................................................................................................................................................................................ 25  Zero, Calibration, and Bump Test....................................................................................................................................................................... 28  Alerts, Warnings, and Alarms ................................................................................................................................................................................. 31  Overview ............................................................................................................................................................................................................. 31  Event Causes and Recommended Actions ........................................................................................................................................................ 32  Service and Warranty ............................................................................................................................................................................................. 37  Service Instructions ............................................................................................................................................................................................ 37  Supplies.......................................................................................................................................................................................................... 37  Three-dimensional Diagrams ......................................................................................................................................................................... 38  Service Tasks ................................................................................................................................................................................................. 40  Warranty Policy .................................................................................................................................................................................................. 44  Limitation of Liability ....................................................................................................................................................................................... 44  Contact Information…………………………………………………………………………………………………………………………………Back Cover

iii

Tables and Figures Table 1.2 Warnings and cautionary statements ....................................................................................................................................................... 2  Table 1.3 Sensor-type options .................................................................................................................................................................................. 3  Table 1.4 Instrument specifications .......................................................................................................................................................................... 4  Table 1.5 Sensor specifications ................................................................................................................................................................................ 4  Table 1.6 Battery properties...................................................................................................................................................................................... 5  Figure 1. Probability of sensor failure by bump test frequency ................................................................................................................................. 9  Table 2.2 Securing the garment or belt clip .............................................................................................................................................................. 9  Table 3.1 Package contents ................................................................................................................................................................................... 11  Table 3.2. Hardware overview ................................................................................................................................................................................ 12  Table 3.3 Display screen indicators and abbreviations .......................................................................................................................................... 12  Table 3.4 Start-up and shutdown ............................................................................................................................................................................ 14  Table 4.1 Configuration instruction ......................................................................................................................................................................... 18  Table 4.2 Operation instruction ............................................................................................................................................................................... 26  Table 4.3 Zero, calibration, and bump test ............................................................................................................................................................. 28  Table 5.1 Overview ................................................................................................................................................................................................. 31  Table 5.2 Events ..................................................................................................................................................................................................... 31  Table 5.3 Recommended action by event type ...................................................................................................................................................... 32  Figure 6.1 Disassembled Tango TX1 ..................................................................................................................................................................... 38  Figure 6.2 Disassembled Tango TX1 case top assembly ...................................................................................................................................... 38  Table 6.1 Key for the Tango TX1 diagram .............................................................................................................................................................. 39  Table 6.2 Service tasks........................................................................................................................................................................................... 40  Table A.1. Sensor cross interference (percent response) ...................................................................................................................................... 45  Table A.2. ATEX and IECEx marking requirements ............................................................................................................................................... 45 

iv

1 General Information Introduction Certifications Product Overview Product Specifications

Introduction CERTIFICATIONS Each Tango TX1TM is certified by one or more certifying bodies (CBs). The approved uses for which a unit is certified appear on label(s) affixed to the instrument. When a new certification is received, it is not retroactive to any unit that does not bear the new marking on its label. Instrument certifications at the time of this document's publication are noted below (see Table 1.1). To determine for which uses a unit is certified, always refer to the unit's labels. CBs issue warnings and cautionary statements to notify the safety team and instrument operators of important information, or to restrict instrument use or service (see Table 1.2). Those items listed under the heading, "General", are issued by more than one CB or by Industrial Scientific Corporation (ISC); these apply to each unit regardless of its certifications. Additionally, those items listed under the heading of a specific CB apply to units that bear its markings. Table 1.1 Certifications Directive or CB

Area Classifications

Approved Temperature Range

CSA

Ex ia IIC T4; Class 1; Groups A, B, C, D; T4

-40 °C to +50 °C (-40 °F to +122 °F)

IECEx

Ex ia I Ma Ex ia IIC T4 Ga

-40 °C to +50 °C (-40 °F to +122 °F)

INMETRO

Ex ia IIC T4 Ga Ex ia I Ma

-40 °C ≤ T amb ≤ +50 °C (-40 °F ≤ T amb ≤ +122 °F)

UL (C-US)

Class I, Groups A, B, C, and D; Class II, Groups E, F, and G; T4; Exia Class I, Zone 0, AEx ia IIC T4

-40 °C to +50 °C (-40 °F to +122 °F)

Americas

General Information

Table 1.1 Certifications Directive or CB

Area Classifications

Approved Temperature Range

ATEX

Ex ia I Ma Ex ia IIC T4 Ga Equipment Groups and Categories: I M1 and II 1G

-40 °C to +50 °C (-40 °F to +122 °F)

GOST EAC

PO Ex ia I X 0 Ex ia IIC T4 X

-40 °C to +50 °C (-40 °F to +122 °F)

China Ex

Ex ia IIC T4 Ga

-20 °C to +50 °C (-4 °F to +122 °F)

KOSHA

Ex ia IIC T4

-20°C ≤ Ta ≤ +50°C

Europe and Russia

Asia and Pacific

Note: See the Appendix for ATEX and IECEx marking requirements.

Failure to perform certain procedures or note certain conditions may impair the performance of this product. For maximum safety and optimal performance, please read and follow the procedures and conditions listed below. Table 1.2 Warnings and cautionary statements For maximum safety and optimal performance, read and understand the manual before operating or servicing the unit. Failure to perform certain procedures or note certain conditions may impair the performance of this product. For safety reasons, this equipment must be operated and serviced by qualified personnel only. Substitution of components may impair intrinsic safety and may cause an unsafe condition. Do not replace battery in hazardous locations. Only certified for use with one Tadiran TL-5955 battery cell. Obstruction of sensor openings – due to dust, dirt, water, or another cause – can inhibit the unit’s ability to measure gas concentrations accurately. When this occurs, readings may appear lower than the actual gas concentration. Keep sensor openings clean, dry, and properly exposed to the ambient air. Obstructed, contaminated, or damaged sensor water barriers (or their gaskets) can inhibit the unit’s ability to measure gas concentrations accurately. When this occurs, readings may appear lower than the actual gas concentration. Replace the sensor water barriers and gaskets as needed (see “Service” for instructions). Service the unit, use its communications port, and change its battery cell only in nonhazardous locations. Not for use in oxygen-enriched atmospheres. Contact your service representative immediately if you suspect that the unit is working abnormally.

Product Overview The Tango TX1 is a portable, long-life, single-gas monitor (instrument) for personal protection. It is a diffusion instrument for use in detecting and measuring gas present in open space. Based on the customer’s order, two redundant sensors – sensors of the same type – are factory installed. Four sensor types are available (see Table 1.3).

2

Product Overview

Table 1.3 Sensor-type options Sensor Category Number of sensors available per instrument

Sensor type

Toxic

Carbon Monoxide (CO) only, Hydrogen Suflide (H2S) only, Nitrogen Dioxide (NO2 ) only, or Sulfur Dioxide ( SO2) only

Two of the same type

Note: See Table 1.5 for sensor specifications.

When two working sensors are installed, they measure gas simultaneously and independently and the Tango TX1 operates on DualSenseTM Technology from ISC. If only one sensor is installed or only one installed sensor is operational, the unit operates as a single-sensor or non-DualSense unit. If two working sensors are installed, the data log contains a data set for each sensor, plus a third data set. This third data set – labeled "VIRTUAL" or "3" – is created by the DualSense technology. Each VIRTUAL data point is an algorithm-calculated value that is based on sensor 1 and sensor 2 data. The VIRTUAL data are saved to and downloaded from the data log just as data are for the installed sensors. VIRTUAL gas readings are displayed to the instrument operator. The Tango TX1 multiplane sensor ports each sample air from three directions; this promotes continuous operation if one or two planes become obstructed. Tango TX1 measures gas at two second intervals, and continuously logs data every ten seconds. The data log can store approximately three months of data for a unit that is on 24 hours a day and has two installed, operational sensors. As the newest data are logged to memory, the oldest data are overwritten. The data log's date- and timestamped event log records and stores event data for 60 alarm events and 30 error events. It also stores the data for up to 250 manual calibrations and bump tests. The data log is downloaded when the unit is docked in a compatible docking station. The instrument has two modes: configuration and operation. When in configuration mode, a unit's settings can be manually edited. Entry to configuration mode can be password protected. When the instrument is on and is not in configuration mode, it is said to be in operation mode. The instrument features an always-on option that can be enabled or disabled from configuration mode. The Tango TX1 has a multisensory (audible, visual, and vibration), multilevel warning and alarm system. Warnings indicate a service need (calibration due) or an operating condition (confidence indicator). Alarms indicate potentially hazardous gas concentrations or system faults. Alarms can be set to latch. The instrument also features a country-oforigin option that automatically sets the values for the low-gas and high-gas alarm set points for each of five different countries or regions; each alarm set point value can also be manually edited. An optional AlarmAmpTM is available; when used, the audible alarm volume is increased by approximately 10 decibels (dB). The user interface consists of two buttons and an LCD (liquid crystal display). The buttons are used to power on and power off the instrument, navigate the operation and configuration loops, perform tasks, and access information. The unit can be set to display select information in English or French. The unit's suspender clip is intended for attachment to a garment; it is not intended for attachment to a belt or hard hat. An optional belt clip is available from ISC (see "Service"). The Tango TX1 is iNet ready and compatible with the Tango TX1 iNet DS (docking station).

Product Specifications Effective use of the Tango TX1 includes knowledge of the instrument's specifications and its sensor and battery specifications (see Tables 1.4 through 1.6). 3

General Information

Table 1.4 Instrument specifications Item

Description

Display

Segment LCD

Keypad buttons

Two buttons

Case materials

Case top: polycarbonate with a protective rubber over-mold Case bottom: conductive polycarbonate

Alarms

Three strobe-emitting visual alarm LEDs (two red; one blue) 100 dB audible alarm at a distance of 10 cm (3.94"), typical Vibration alarm

Dimensions

99 x 51 x 35 mm (3.9" x 2.0" x 1.4")

Weight

126 g (4.4 oz.), typical

Ingress protection

IP66 and IP67

Operating temperature range1

-20 °C to +50 °C (-4 °F to +122 °F)

Operating humidity range

15 to 95% relative humidity (RH) noncondensing (continuous)

1 Operating temperatures above 50 °C (122 °F) may cause reduced instrument accuracy. Operating temperatures below -20 °C (-4 °F) may cause reduced instrument accuracy and affect display and alarm performance.

Table 1.5 Sensor specifications Sensor properties

Accuracy

Response time

Sensor temperature range

Sensor RH range

At time and temperature of calibration

Over full sensor temperture and RH ranges

(% vol, % LEL, or ppm)

degrees (°)

(%)

(%)

(%)

Measurement range2

Measurement resolution2

Gas Names

Sensor technology1

Sensor Category

Abbreviation

(typical)

T50

T90

Seconds (s)

Toxic

4

Carbon Monoxide with H2 low interference

CO/ H2 Low

E

0-1000 ppm

1 ppm

-20 to +50 °C (-4 to +122 °F)

15-95

± 5% (0-300 ppm) ± 15% (3011000 ppm)

± 15%

9s

18 s

Carbon Monoxide

CO

E

0-1000 ppm

1 ppm

-40 to +50 °C (-40 to +122 °F)

15–95

±5

± 15

12 s

48 s

Hydrogen Sulfide

H2 S

E

0-200 ppm

0.1 ppm

-40 to +50 °C (-40 to +122 °F)

15–95

±5

± 15

7s

14 s

Nitrogen Dioxide

NO2

E

0-150 ppm

0.1 ppm

-40 to +50 °C (-40 to +122 °F)

15–95

± 10

± 15

10 s

30 s

Sulfur Dioxide

SO2

E

0 to 150 ppm

0.1 ppm

-20 to +50 °C (-4 to +122 °F)

15 – 95

± 10

± 15

20 s

80 s

Product Overview

1"E" stands for electrochemical. 2ppm stands for parts per million; vol for volume; and LEL for lower explosive limit. Note: See the Appendix for supplemental information about sensor types and gases.

Table 1.6 Battery properties Battery Pack

Properties

3.6 V Primary Lithium-thionyl chloride (Li-SOCl2), 1.5AH, 2/3AA

Replaceable* Nonrechargeable Run time of 36 months depending on operating conditions; the amount of time the unit is in alarm; and the enablement of unit's confidence indicator, bump test due alert, or calibration due alert.

* See "Service" for instructions. Some restrictions may apply (see "Introduction", Table 1.2 Warnings and cautionary statements).

5

6

2 Recommended Practices Introduction Procedures First Use Wearing the Instrument

Introduction Gas detection instruments are potentially life-saving devices. When completed regularly, the procedures defined below help to maintain proper instrument functionality and enhance operator safety.

Procedures Configuration. The configuration process allows qualified personnel to review and adjust a unit's settings. Self-test. The self-test verifies the functionality of the instrument’s memory operations, battery, and each alarm indicator (audible, visual, and vibration). Note: a self-test does not verify sensor functionality (see “Bump Test”) or instrument accuracy (see “Calibration”).

Bump Test (or "functional test"). Bump testing checks for sensor and alarm functionality. The installed sensors are briefly exposed to an expected concentration of calibration gas that is greater than the sensors’ low alarm set point. When one or more sensors “pass” the test, they are “functional” and the unit will alarm. Each sensor’s “pass” or “fail” result is indicated on the unit’s display. Note: a bump test does not measure for accuracy (see “Calibration”).

Zero. Zeroing sets each installed sensor to recognize the ambient air as clean air. If the ambient air is not truly clean air, any gases that are present and relevant to the installed sensor types will be measured and displayed as zero. Readings will be inaccurate until the unit is correctly zeroed in truly fresh air or with a zero air cylinder. Calibration. All sensors gradually degrade over time. This diminishes a sensor's ability to measure gas concentrations accurately; however, regular calibrations adjust the instrument to compensate for this decline in sensitivity. During calibration, the installed sensors are exposed to an expected concentration of calibration gas and, when needed, the instrument will self-adjust to ensure the accurate measurement and display of gas concentration values. Note: when a sensor has degraded beyond an acceptable level, no further adjustment is possible and the sensor will no longer pass calibration.

Recommended Practices

Peak Readings. The instrument stores the highest detected gas readings, the "peak readings" or "peaks". Bump testing and calibration will often register new peak readings. Therefore, the clearing of the peak readings should follow each calibration. The instrument operator may also wish to clear the peak readings after a bump test, before a change in location, or after an alarm is addressed and cleared. Note: The peak readings and the data log readings are stored independently of one another; therefore, clearing the peak reading does not affect the data log. Powering the instrument off or changing its battery does not affect the peak reading. These checks and balances help promote operator safety, and serve to contain the peak readings in a "black-box" manner. In the event of a gas-related incident, this black-box record can be useful to the safety team or a prospective investigator.

Recommendations Industrial Scientific Corporation (ISC) minimum frequency recommendations for each procedure are summarized in the table below. These recommendations are based on field data, safe work procedures, industry best practices, and regulatory standards to help ensure worker safety. Industrial Scientific is not responsible for setting safety practices and policies. ISC recommendations include a daily bump test for any ISC instrument that is not operating on DualSense technology. This includes the Tango TX1 when it is operating with only one working sensor. When two redundant, working sensors are installed in the Tango TX1, the instrument is operating on DualSense and the probability of sensor failure – compared with a single-sensor instrument – is diminished regardless of bump test frequency (see Figure 1). The frequency of bump testing for DualSense instruments, between monthly calibrations, is best determined by a company's safety policies. These policies may be affected by the directives and recommendations of regulatory groups, environmental conditions, operating conditions, instrument use patterns and exposure to gas, and other factors. Table 2.1 ISC recommended practices Procedure

Recommended minimum frequency

DualSense instruments Configuration

Before first use, when there is a change in the installed sensor type, and as otherwise needed.

Calibrationa

Before first use and monthly thereafter.

Bump testb

As desired between monthly calibrations.

Self-testc

Prior to each day’s use for a unit that is always on or is left on.

Non-DualSense instruments Configuration

Before first use and as needed thereafter.

Calibrationa

Before first use and monthly thereafter.

Bump

testb

Self-testc aBetween

Prior to each day’s use. As desired between daily bump tests.

regular calibrations, ISC also recommends a calibration be performed immediately following each of these incidences: the unit falls, is dropped, or experiences another significant impact; is exposed to water; fails a bump test; or has been repeatedly exposed to an over-range (positive or negative) gas concentration. A calibration is also recommended after the installation of a new (or replacement) sensor. bIf conditions do not permit daily bump testing, the procedure may be done less frequently based on company safety policy. cThe Tango TX1 self-test is performed automatically during the start-up process. It can be user-initiated from operation mode. Note: The use of calibration gases not provided by ISC may void product warranties and limit potential liability claims.

8

First Use

Average number of failures per instrument use-days

1 in 1 1 in 10 1 in 100 1 in 1,000 1 in 10,000 1 in 100,000 1 in 1,000,000 0

5

10

15

20

25

30

Bump test interval (days) Non-DualSense instruments DualSense instruments

Figure 1. Probability of sensor failure by bump test frequency

First Use To prepare the Tango TX1 for first use, qualified personnel should configure and calibrate the unit.

Wearing the Instrument Based on the U.S. Department of Labor's Occupational Safety and Health Administration (OSHA) definition of the breathing zone, it is recommended that the unit be worn within a 25.4 cm (10") radius of the nose and mouth. Refer to OSHA and to other agencies or groups as needed for additional information. ISC also recommends that the unit be worn within the instrument operator's sight line. The instrument operator may wear the unit with its factory-installed suspender clip or with the optional belt clip. The suspender clip is solely intended for attachment to a garment. The belt clip may be attached to a hard hat, belt, or garment. The clips should be securely fastened and attached in a manner that ensures the unit's sensor portals are fully exposed to the air. No part of the unit should be covered by any garment, part of a garment, or other item that would restrict the flow of air to the sensors or impair the operator's access to the audible, visual, or vibration alarms. Attach the desired clip as shown below. Table 2.2 Securing the garment or belt clip Suspender clip

Lift the clip cover.

Position the garment between the clip's upper and lower teeth. Press down on the clip cover to secure the clip in place. 9

Recommended Practices

Table 2.2 Securing the garment or belt clip Belt clip

Position the hard hat or garment between the clip and clip back.

10

Slide the clip to secure it in place. The instrument can be worn right side up or upside down.

3 Instrument Basics Unpacking the Instrument Hardware Overview Display Overview Start-up and Shutdown

Unpacking the Instrument The items that are shipped with the unit are listed below (see Table 3.1); each item should be accounted for in the unpacking process. Table 3.1 Package contents Quantity

Item

Notes

1 as ordered

Tango TX1

Part number 18109075

1

Suspender clip (installed)

Optional belt clip sold separately.

1

Calibration cup



1

Calibration and bump test tubing

60.96 cm (2 ') of urethane tubing; 4.762 mm (3/16 ”) ID

1

Reference Guide

Companion to the Tango TX1 Product Manual

1

Final Inspection & Test Report and Declaration of Conformity

The Final Inspection & Test Report includes these values: Unit setup date Unit part number Unit serial number For each sensor:  Part number  Serial number  Type  Alarm settings1  Span information1

1At

the time of shipment. Note: If any item is missing or appears to have been damaged, contact ISC (see "Contact Information") or a local distributor of ISC products.

Instrument Basics

Hardware Overview The instrument's main hardware components are identified below (see table 3.2). Table 3.2. Hardware overview Tango TX1 Visual alarm (or alert) indicator

Visual alarm (or alert) indicators

IrDA (infrared data exchange) window Sensor port 1 and dust filter

Sensor port 2 and dust filter

Case top Enter button

On-off-mode button LCD

Audible alarm (or alert) indicator and dust filter

Case bottom

Garment clip (closed)

Garment clip (open)

Display Overview The visual test screen shown below contains all the indicators that can appear on the display screen. Each indicator is stationary and appears only when relevant to the task being performed. For example, in the gas-monitoring screen shown below (numeric display), the following apply: the check mark indicates there are no sensor faults; the sensortype icon indicates that H2S sensors are installed; the numeric display shows a gas reading of 5.1 ppm. Table 3.3 Display screen indicators and abbreviations Display screens

Visual test screen

Gas-monitoring screen (numeric display)

Gas-monitoring screen (text display)

Status indicators only

Two sensors are installed and neither is in fault. Two sensors are installed and one is in fault; a sensor location icon also displays to indicate which sensor is in fault.

12

Display Overview

Table 3.3 Display screen indicators and abbreviations and

Only one sensor is installed and is not in fault. Two sensors are installed and both are in fault or one sensor is installed and in fault. The warning icon is also used in combination with other indicators to communicate a system alarm or an alert condition.

!

The unit is in configuration mode. Alarms indicators The alarm icon is used in combination with other indicators to communicate a variety of conditions. and ▲

High-level gas alarm.

and ▼

Low-level gas alarm.

and

STEL alarm.

and

TWA alarm.

and

Positive over-range gas alarm.

and

Negative over-range gas alarm. Low battery alarm.

Process and time-based indicators The zero icon is used in combination with other indicators to communicate sensor zero information. The bump test icon is used in combination with other indicators to communicate bump test information. The calibration icon is used in combination with other indicators to communicate calibration information. A process is in progress. In configuration mode, indicates a time-based setting (e.g. bump test response time). Used in combination with other indicators to communicate overdue warnings. In configuration mode, indicates a date-based setting (e.g., bump test interval). Gas name and unit-of-measure abbreviations Carbon Monoxide (CO) Sulfur Dioxide (SO2 or SO2) Nitrogen Dioxide (NO2 or NO2) Hydrogen Sulfide (H2S or H2S)

PPM

Parts per million is the unit of measure for CO, SO2, NO2, and H2S.

Other abbreviations Positive over-range: the detected gas concentration is greater than the upper limit measurement range of the sensor. Display variations: “Or” (English) and “Sup” (French). Negative over-range: the detected gas concentration is less than the lower limit measurement range of the sensor. Display variations: “-Or” (English) and “InF” (French). Short-term exposure limit. Display variations: “STEL” ( English) and “VLE” (French).

13

Instrument Basics

Table 3.3 Display screen indicators and abbreviations Time-weighted average. Display variations: “TWA” (English) and “VME” (French). Security code is set or to be entered. In configuration mode, indicates a feature may be operation-mode enabled or disabled. Peak reading.

Start-up and Shutdown The start-up and shutdown sequences are outlined below and feature reproductions of the display screens the instrument operator will see during these processes (see Table 3.4). Instructions accompany any display screen where the instrument operator must press a button to proceed. The instrument operator may be prompted to complete the time- and date-setting tasks during start-up. This may happen after a battery has been removed or changed. If prompted by the unit to do so, it is essential – for data log accuracy – that the time- and date-setting tasks be completed. The data log plays an important role in preserving operator safety and in the prospective investigation of an incident. The instrument operator may be prompted to enter a security code during shutdown. This will occur if the unit is configured for "always-on" and is security-code protected. Table 3.4 Start-up and shutdown Start-up. Press and hold for three seconds, then release to initiate the start-up sequence and power on the unit.  If all start-up diagnostics pass, the audio, visual, and vibration indicators turn on then off. Four start-up screens display followed by the gas-monitoring screen.  If any start-up diagnostic fails, an error message displays (see "Alarms and Warnings"). Start-up screens.

Visual test screen

Version display screen

Calibration date screen (last calibration date shown above)

Gas-monitoring screen. No-fault indicator Gas concentration

14

Installed sensor type Unit of measure

Countdown screen For qualified personnel only: During the 15-second countdown, press both buttons and hold for three seconds to enter configuration mode (see "Configuration").

Start-up and Shutdown

Table 3.4 Start-up and shutdown What to do if the time setting screen is activated.

Time setting

This display screen features the clock icon and the current time setting. The instrument's clock uses a 24-hour time format. Its values are edited in this order using these ranges*: Hours: 00 to 24 Minutes: 00 to 59 The first press activates the first value to be edited. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value and activates the next value to be edited. Continue to use the buttons, and , to edit and save the values, respectively. After all values are edited and saved, one press activates the next configuration-mode screen.

Date setting

This display screen features the calendar icon and the current date setting. The year is displayed in the lower left corner. In the main display, the first two digits represent the date and the second two digits represent the month. The values are edited in this order using these ranges: Year: 2012 to 9999 Day: 00 to 31 Month: 00 to 12 The first press activates the first value to be edited. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value and activates the next value to be edited. Continue to use the buttons, and , to edit and save the values, respectively. After all values are edited and saved, one press activates the gas-monitoring screen.

Shutdown. Press and hold for five seconds.

Countdown

After a five-second countdown: The instrument powers off if  the always-on feature is disabled or  the always-on feature is enabled and the security code is set to 000. If this screen is activated, the shutdown process is security-code protected. To complete shutdown, the user must enter the correct three-digit code. Value range: 000 to 999

Enter security code Increments the value by one; hold to speed the increment pace. Enters the value. If the value is correct, the unit powers off. If the value is incorrect, the gas-monitoring screen is activated. Dock the unit or store it for next use. *When editing a value, once the last value in the range is reached, the display starts again with the first value.

To prepare the instrument for first use, qualified personnel should proceed with the configuration process (see "Configuration"). To operate a field-ready unit, refer to "Operation". 15

16

4 Instrument Preparation and Use Configuration Operation Zero, Calibration, and Bump Testing

Configuration Read and understand all configuration instruction before configuring the unit. As noted in "Recommended Practices", the unit should be configured before first use, when there is a change in the installed sensor type (e.g., H2S sensors are replaced with CO sensors), and as needed. Only qualified personnel should access the configuration mode and adjust the unit's settings. Configuration mode can be accessed only during the start-up sequence (see “Start-up and Shutdown”). Review the unit's configured settings for compliance with company policy and any applicable regulations, laws, and guidelines as issued by regulatory agencies and government or industry groups. Determine which settings, if any, require adjustment. Choose alarm- and warning-related options that maximize safety within the air-sampling environment. When the unit is in configuration mode, the following apply: 

The tool icon ( ) displays in the lower right corner of each screen.



With successive short presses of the on-off-mode button ( ), the user can scroll through the configuration loop.



The enter button (



When editing a value, the enter button ( value.



When editing a value, once the last value in the range is reached, the display starts again with the first value.



When both buttons ( and ) are simultaneously pressed and held for three seconds, the unit leaves configuration mode; it enters operation mode and the gas-monitoring screen is activated.



Unless otherwise noted, when no button is pressed for 30 seconds, the unit enters operation mode and the gasmonitoring screen is activated.

) is used to start the editing process or start a task (e.g., zero). ) increments the value and the on-off-mode button ( ) saves the

Any changes made in configuration mode are automatically saved to the unit and take effect immediately. Upon next docking, settings are updated according to the unit's settings in iNet Control. The configuration-mode loop is outlined below (see Table 4.1). Instructions for button use accompany each configuration-mode display screen.

Instrument Preparation and Use

Table 4.1 Configuration instruction Screen Button

Description Button effect If this display screen is activated, the configuration mode is security-code protected. To enter configuration, the user must enter the correct three-digit code. If the security code is set to 000, entry to configuration mode is not security-code protected. The first configuration-mode screen is activated, the initiate-zero screen.

Enter security code Increments the value by one; hold to speed the increment pace. Saves the displayed value. Notes: If an incorrect code is entered, the unit does not enter configuration mode and the gas-monitoring screen is activated. If the security code is unknown, configuration mode can be accessed as follows: enter 412, then press both buttons simultaneously and hold briefly. This activates the next configuration mode screen. This also resets the security code to 000, leaving configuration mode unprotected. The security code can be reset in configuration mode at the security-code-setting screen. This screen's activation allows the technician to complete the zero and calibration processes from configuration mode.

Initiate zero Starts the zero process. Skips the zero process and activates the next configuration-mode screen.

Low gas alarm set point

See also country-of-origin setting. This display screen features the low alarm and sensor-type icons, and the alarm's current set point and unit of measure. The alarm set point can be edited based on the following: Value range = within the sensor measurement range Value increment = sensor measurement resolution See Table 1.5 for the measurement range and resolution for the installed sensor type. The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

High gas alarm set point

See also country-of-origin setting. This display screen features the high alarm and sensor-type icons, and the alarm's current set point and unit of measure. The alarm set point can be edited based on the following: Value range = within the sensor measurement range Value increment = sensor measurement resolution See Table 1.5 for the measurement range and resolution for the installed sensor type. The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

18

Configuration

Table 4.1 Configuration instruction Screen Button

TWA operation-mode setting

Description Button effect This display screen features the lock icon to indicate the technician can enable or disable the option for operation-mode access. When enabled, the instrument operator is permitted to view and clear the unit's TWA reading while the unit is in operation mode. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen. This display screen features the alarm and sensor-type icons, and the alarm's current set point and unit of measure. The alarm set point can be edited. Value increment = within the sensor measurement resolution See Table 1.5 for more information about the installed sensor type.

TWA alarm set point The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen. This display screen features the clock and TWA icons, and the current TWA time-base. The set point value can be edited based on the following: Value range: 01 to 40 hours Value increment: 1 hour TWA time-base setting The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

STEL operation-mode setting

This display screen features the lock icon to indicate the technician can enable or disable the option for operation-mode access. When enabled, the instrument operator is permitted to view and clear the unit's STEL reading while the unit is in operation mode. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

STEL alarm set point

This display screen features the alarm, STEL, and sensor-type icons, and the current set point. The set point can be edited. Value increment: sensor measurement resolution See Table 1.5 for more information about the installed sensor type. The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

19

Instrument Preparation and Use

Table 4.1 Configuration instruction Screen Button

Calibration gas setting

Description Button effect This display screen features the calibration cylinder and sensor-type icons, and the current calibration gas setting. This setting reflects the concentration of calibration gas that the instrument expects to read when calibrated; it should be edited to match the cylinder's gas concentration. Value range: within the sensor measurement range Value increment: sensor measurement resolution See Table 1.5 for the measurement range and resolution for each sensor type. Increments the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

Time setting

This display screen features the clock icon and the current time setting. The instrument's clock uses a 24-hour time format. Its settings are edited in this order using these values: Hours: 00 to 24 Minutes: 00 to 59 Value increment: 1 The first press activates the first value to be edited. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value and activates the next value to be edited. Continue to use the buttons, and , to edit and save the values, respectively. After all values are edited and saved, one press activates the next configuration-mode screen.

Date setting

This display screen features the calendar icon and the current date setting. The year is displayed in the lower left corner. In the main display, the first two digits represent the date and the second two digits represent the month. The settings are edited in this order using these values: Year: 2012 to 2099 Day: 00 to 31 Month: 00 to 12 The first press activates the first value to be edited. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value and activates the next value to be edited. Continue to use the buttons, and , to edit and save the values, respectively. After all values are edited and saved, one press activates the next configuration-mode screen.

Style setting

This display screen allows the technician to choose the display style for the gas-monitoring screen. A numeric display will feature the numeric gas reading and the sensor type icon. A text display will feature the sensor type in place of the numeric gas reading (see "Operation" for sample display styles). Values: 0 = Numeric display 1 = Text display Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

20

Configuration

Table 4.1 Configuration instruction Screen Button

Confidence indicator setting

Description Button effect This display screen features the alarm icon and check mark indicator. The technician can disable the indicator, or enable the indicator and choose the indicator type. When enabled, the unit will emit the selected signal every 90 seconds in operation mode. Values: 0 = disabled 1 = enabled for audible chirp 2 = enabled for blue LED flash Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

Operation-mode bump test setting

This display screen features the bump test icon. The lock icon indicates the technician can enable or disable this operation-mode feature. When enabled, the instrument operator is given access to bump test the unit from operation mode. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

Bump test due alert setting

This display screen features the alarm, bump test, calendar, and warning icons. The technician can disable the warning, or enable the warning and choose the warning type. When enabled, the alert screen will be activated and the unit will emit the selected indicator every 60 seconds to notify its user that a bump test is due; the instrument will continue to operate. Values: 0 = disabled 1 = enabled for audible chirp 2 = enabled for blue LED flash 3 = enabled for combination audible chirp and blue LED flash Increments the value. One press saves the displayed value and activates the next configuration-mode screen. This display screen features the bump test and calendar icons. The technician can set the interval at which the bump test due alert is to be activated. Value range: 0.5 to 30.0 days Value increment: 0.5 days

Bump test interval setting The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

21

Instrument Preparation and Use

Table 4.1 Configuration instruction Screen Button

Bump test percentage setting

Description Button effect This display screen features the bump test icon and the current setting. The technician can set the percentage of calibration gas to which the unit will respond. Value range: 50% to 95% Value increment: 1% See Table 1.5 for sensor information that can aid in the setting of bump test values. The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

Bump test response-time setting

This display screen features the bump test and clock icons, and the current setting in seconds. A sensor passes a bump test when it senses the specified percentage of calibration gas within the specified response time setting. Value range: 30 to 120 seconds Value increment: 1 second The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

Alarm latch setting

This display screen features the alarm icon and the current setting. The lock icon indicates the technician can enable or disable this operation-mode feature. When disabled, a unit in alarm will turn off its alarm when the gas reading is no longer at the alarm-producing concentration. When enabled, a unit in alarm will remain in alarm until it is manually reset. The instrument operator can reset a latched alarm from operation mode with a long press of the enter button ( ). This resets the alarm; it does not disable an enabled latch. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

Vibration alarm setting

This display screen features the alarm, check mark, and vibration motor icons. When enabled, the vibrating alarm will be activated when the unit is in alarm. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

22

Configuration

Table 4.1 Configuration instruction Screen Button

Description Button effect

Operation-mode zero setting

This display screen features the zero icon. The lock icon indicates the technician can enable or disable this operation-mode feature. When enabled, the instrument operator is given access to zero the unit from operation mode. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

Operation-mode calibration setting

This display screen features the calibration icon. The lock icon indicates the technician can enable or disable this operation-mode feature. When enabled, the instrument operator is given access to calibrate the unit from operation mode. Values: 0 = disabled 1 = enabled Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

Calibration due alert setting

This display screen features the alarm, calibration, calendar, and warning icons. The technician can disable the warning, or enable the warning and choose the warning type. When enabled, the alert screen will be activated and the unit will emit the selected indicator every 60 seconds to warn its user that a calibration is due; the instrument will continue to operate. Values: 0 = disabled 1 = enabled for audible chirp 2 = enabled for blue LED flash 3 = enabled for combination audible chirp and blue LED flash Increments the value. One press saves the displayed value and activates the next configuration-mode screen. This display screen features the calibration and calendar icons, and the current interval setting. The technician can set the interval at which the calibration due alert is to be activated. Value range: 1 to 365 days Value increment: 1 day

Calibration interval setting The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

23

Instrument Preparation and Use

Table 4.1 Configuration instruction Screen Button

Calibration date setting

Description Button effect This display screen features the calibration cylinder, calendar, warning, and arrow icons. The technician can choose whether the operation-mode calibration date screen will display the due date for the unit's next calibration or the date of the unit's last calibration. The up arrow (▲) will be featured on-screen when the unit is set to display the next the calibration due. The down arrow (▼) will be featured when the unit is set to display the last calibration date. Values: 0 = displays date of last calibration 1 = displays next calibration due date Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

Security code setting

This display screen features the lock icon and the current security code. The security code controls two things: access to a unit's configuration mode and the ability to power off a unit that is configured for always-on operation. If the security code is set at 000, entry to configuration mode is not security-code protected, and an always-on unit can be powered off without a security code. Any other value will enable the security code. Value range: 000 to 999 Value increment: 1 The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value; a second press activates the next configuration-mode screen.

Country-of-origin setting

This display screen features the country-of-origin icon and the current setting. This feature sets automatically the low- and high-gas alarm set points. The technician must choose one of these options: “DEF” = USA and default value “CAn” = Canada “EUr” = Europe “CR” = Czech Republic “AUS” = Australia The unit's settings are immediately updated to reflect that country's (or Europe's) low- and high-gas alarm set points for the installed sensor-type. When the country-of-origin selection does not contain a value for an alarm set point, the default value (DEF) is automatically substituted. The first press activates the value. Continued presses increment the value; hold to speed the increment pace. One press saves the displayed value and activates the next configuration-mode screen. Notes: Each alarm setting can be edited individually, in configuration mode, at that alarm's set point screen. Because low- and high-gas alarm set points can be edited both individually and through the country-of-origin option, it is important to understand override behavior. Example. The H2S low-gas alarm set point was edited – at its set point screen – to a value of 9 ppm. Afterwards, a country-of-origin selection was made where the H2S low-gas alarm set point value is 10 ppm. The last-entered setting overrides the first. Therefore, in this example, the value of the H2S low gas alarm set point is 10 ppm. Another aspect to alarm settings applies to replacement or new sensors. For example:  If the installed H2S sensor(s) is replaced with other H2S sensor(s), the last-entered low-gas alarm set point (a value of 10 ppm in the above example) will be applied to the newly installed sensor(s).  If the installed H2S sensors are replaced with a different sensor type (e.g., CO), the alarm settings will be read from the newly installed sensor(s).

24

Configuration

Table 4.1 Configuration instruction Screen Button

Description Button effect This display screen features the language icon and the current setting. The technician can choose from these options: “En” = English “F” = French

Language setting Increments the value. One press saves the displayed value and activates the next configuration-mode screen. This display screen features the battery icon. The lock icon indicates the technician can enable or disable this feature. When enabled, the entry of the unit's security code (if the security code is not 000) will be required to complete the shutdown process. Values: 0 = disabled 1 = enabled

Always-on setting

Increments the value. One press saves the displayed value and activates the next configuration-mode screen. Disallow or allow operator-activated shutdown when the unit is in alarm. Values: 0 = disallows shutdown 1 = allows shutdown Shutdown in alarm setting Increments the value. One press saves the displayed value and activates the next configuration-mode screen.

After the configuration process is completed and before the unit’s first use, calibrate the instrument (see “Zero, Calibration, and Bump Test”).

Operation In operation mode, the following apply:  With successive short presses of the on-off-mode button ( ), the instrument operator can scroll through the operation-mode loop.  The zero, calibration, and bump test processes can be completed only if these task settings are enabled for operation-mode access.  The peak reading can be viewed and cleared.  The TWA and STEL readings can be viewed and cleared only if these task settings are enabled for operationmode access.  In general, the buttons are used as follows: 

Press

to scroll through the operation-mode loop.



Press

to initiate a task or to clear a reading.

25

Instrument Preparation and Use



A long press on



When



Except where noted, when no button is pressed for 30 seconds, the gas-monitoring screen is activated.

and

will reset a latched alarm; it does not disable an enabled latch. are simultaneously pressed and held for three seconds, the unit will complete a self-test.

The operation-mode loop is outlined below (see Table 4.2). Instructions for button use accompany each display screen. Table 4.2 Operation instruction Screen Buttons

Screen description Button effects This display screen (numeric shown) features the check mark and sensor-type icons, and the current gas reading and unit of measure. The check mark indicates the unit is operational and there are no sensor faults.

Gas monitoring One short press turns on the backlight if the unit senses it is not in a well-lit environment. When the unit is in alarm, a long press will reset a latched alarm; the alarm will recur if the alarm-causing condition is still present. Activates the next enabled operation-mode screen.

Calibration date

This display screen features the calibration, calendar, and check mark icons, an up or down arrow, and a date value. When the up arrow (▲) is featured, the next calibration date is displayed. When the down arrow (▼) is featured, the last calibration date is displayed. Values: Date: XX (day) and XX (month) Year: XXXX No effect. Activates the next enabled operation-mode screen. This display screen is activated if the unit is enabled for operation-mode zeroing. It features the check mark and zero icons.

Initiate zero Starts the zero process (see "Zero, Calibration, and Bump Testing"). Activates the next enabled operation-mode screen. This display screen is activated if the unit is enabled for operation-mode bump testing. The screen features the check mark and bump test icons.

Initiate bump test Starts the bump test process (see "Zero, Calibration, and Bump Testing"). Activates the next enabled operation-mode screen.

26

Operation

This display screen features the check mark, peak, and sensor-type icons, and the most recent peak reading.

Peak reading Clears the peak reading. Activates the next enabled operation-mode screen. This display screen is activated if this feature is operation-mode enabled. The screen features the check mark, sensor-type, and TWA icons, and the current TWA reading.

TWA reading Clears the TWA reading. Activates the next enabled operation-mode screen. This display screen is activated if this feature is operation-mode enabled. The screen features the check mark, sensor-type, and STEL icons, and the current STEL reading.

STELreading Clears the STEL reading. Activates the next enabled operation-mode screen.

27

Instrument Preparation and Use

Zero, Calibration, and Bump Test Perform the zero, calibration, and bump testing tasks in an area known to be nonhazardous. Table 4.3 Zero, calibration, and bump test Supplies, preparation, and instruction Supplies Calibration cup (shipped with the unit) Calibration tubing (shipped with the unit) Calibration gas cylinder suitable for the installed sensors and the unit's calibration gas settings Positive flow regulator suitable for the calibration gas cylinder Preparation

Holding the regulator, turn the calibration gas cylinder in a clockwise direction to tighten.

28

Connect either end of the calibration tubing to the regulator's nipple.

Connect the other end of the tubing to the calibration cup. Proceed with the instruction set below for the desired task: zero and calibration or bump test.

Zero, Calibration, and Bump Test

Instruction Zero

Initiate zero

Zero in-progress

Zero results (pass)

Note: From anywhere in the operation-mode loop, press until the initiate-zero screen is activated. At the initiate-zero screen, press to start the zero process.

While the sensors are zeroed, the zero-in-progress screen is activated.

After the sensors are zeroed, the zero-results screen is activated and an audible alert is emitted. If the result for either sensor is an "F" for fail, press to reactivate the initiate-zero screen. Repeat the zero process. If the result for both sensors is a "P" for pass, press then to display the initiate-calibration screen. If calibration is not desired, wait approximately 30 seconds for the zero-results screen to deactivate; the gas-monitoring screen will be automatically activated.

Zero results (fail)

Calibration

Place the calibration cup over the case top; align its top groove with the small ridge at the top of the instrument. Press down to secure the cup in place; a click will sound. Visually inspect the calibration cup to ensure its edges along the top and sides align with the case top edges.

Calibration in progress To start the flow of gas, turn the regulator knob in a counterclockwise direction. While the sensors are calibrated, the calibration-in-progress screen displays the span reserve value. If desired, press to cancel the calibration.

Initiate calibration

Calibration apply gas

To start the calibration process, press . Both sensors will be calibrated simultaneously. To cancel the calibration, press .

Once the calibration is started, the apply-gas screen is activated; the expected type and concentration of calibration gas are displayed. This screen remains activated for up to 5 minutes as the unit awaits the application of calibration gas. To cancel the calibration, press .

Calibration results (pass)

Calibration results (fail)

If at least one sensor passes calibration, two results screens are alternately activated; one indicates the pass or fail result for each sensor and the other displays the span reserve value. If neither sensor passes calibration, the audible, visual, and vibrating alarms turn on. Two results screens are alternately activated; one indicates the fail results and the other displays the span reserve value. Note: With two installed, working sensors, the span reserve value is the algorithm calculation of the DualSense Technology.

29

Instrument Preparation and Use



Span reserve value If at least one sensor passes the calibration, the gas-monitoring screen will be automatically activated. Note: The span reserve value divided by the calibration gas concentration yields the span reserve percentage. A span reserve percentage of greater than 70% indicates a “good” sensor; 50%-70% indicates “marginal” sensitivity. When the span reserve percentage is less than 50%, the sensor will not pass calibration.

To stop the flow of gas, turn the regulator knob in a clockwise direction and tighten.

To remove the calibration cup, lift up from the cup's tabs. Set aside or store for future use.

Initiate bump test

Bump test apply gas

Note: From anywhere in the operation-mode loop, press until the initiate-bump-test screen is activated. Press to start the bump test process. Press to cancel the bump test.

Once the bump test is started, the apply-gas screen is activated; the expected type and concentration of calibration gas are displayed. This screen remains activated for up to 5 minutes as the unit awaits the application of calibration gas.

Bump test results (pass)

Bump test results (fail)

Bump testing

Place the calibration cup over the case top; align its top groove with the small ridge at the top of the instrument. Press down to secure the cup in place; a click will sound. Visually inspect the calibration cup to ensure its edges along the top and sides align with the case top edges.

Bump test in progress To start the flow of gas, turn the regulator knob in a counterclockwise direction.

If either or both sensors fail the bump test, the calibration due warning screen will be automatically activated. Calibrate the instrument. If both sensors pass the bump test, the gas-monitoring screen will be automatically activated. 

To stop the flow of gas, turn the regulator knob in a clockwise direction and tighten.

30

To remove the calibration cup, lift up from the cup's tabs. Set aside or store for future use.



5 Alerts, Warnings, and Alarms Overview An alert indicates an instrument operating condition; a warning, the unit is in need of service; and an alarm, the unit has detected a potentially hazardous gas concentration or has encountered a system fault. Each category of notification is listed in Table 5.1 along with its indicators (audible, visual, and vibration) and their patterns. Table 5.1 Overview Indicators Notification Category Alert1 Warning Low-battery alarm Low-level alarm High-level alarm System alarm 1Alert

Pattern

Audible

Visual (LEDs)

Vibration

No Yes2 Yes Yes Yes Yes

No Blue only2 Red only Red and blue Red only Red only

No No No Yes2 Yes2 Yes2

n/a Intermittent Intermittent Continuous sequence with pauses Continuous sequence Continuous sequence

notifications are indicated on the display screen only. enabled.

2When

Table 5.2 lists all event types that can activate an alert, warning, or alarm. Here, they are grouped by gas-related alarms, instrument-based alarms, instrument service warnings, and instrument status alerts. The notification category is indicated below for each event type. Some categories can indicate more than one type of event. For example, the low-level alarm can be activated by a TWA event or a low-alarm gas event, and the unit’s display screen differentiates between the two (see Table 5.3 for more information). Table 5.2 Events Event type Gas-related alarm events Low-alarm gas event TWA event High-alarm gas event STEL event Over-range gas event Instrument-based alarm events Low-battery event Dead battery indicator

Status Alert

Warning

Alarms Low-level

High-level

Low-battery

System











√ √



































√ √ √

─ ─

















√ √







Alerts, Warnings, and Alarms

Table 5.2 Events Event type

Status Alert

Low-level

High-level

Low-battery

System

System fault















√ √ √ √ √









































Instrument service warnings Bump test overdue Calibration overdue Single-sensor zero failure Single-sensor bump test failure Single- sensor calibration failure

─ ─ ─ ─

Warning

Alarms

Instrument status alerts One sensor installed*













Single-sensor data failure*













*Display screen indicators only.

Event Causes and Recommended Actions As noted earlier, more than one event type can activate the same notification category; the display screen icons, text, and data help the instrument operator identify the event. For example, the low-alarm gas event and the TWA event can activate the low-level alarm. The display screen for each event features the alarm icon ( ) and only one of these distinguishing icons: low-gas (▼) or TWA ( ).The display screen for every event type is shown in Table 5.3. Table 5.3 Recommended action by event type Event type

Possible causes

Status and recommended actions

The unit has no faults and needs no service. Both sensors are functional and the unit is operating on DualSense technology.

The unit is operational.

No events

Gas-monitoring screen ▲▼

!

indicate that the unit is in alarm.

Gas-related alarm events The detected gas concentration exceeds the unit’s low-alarm set point.

Low-alarm gas event2

32

Gas warning: Respond according to company safety policy and guidelines.

Event Causes and Recommended Actions

Table 5.3 Recommended action by event type Event type

Possible causes

Status and recommended actions

The cumulative gas readings have reached the unit’s setting for the time weighted average (TWA) exposure limit.

Leave the area. Respond according to company safety policy and guidelines.

The detected gas concentration exceeds the high-alarm set point.

Leave the area. Respond according to company safety policy and guidelines.

The cumulative gas readings have reached the unit’s setting for short-term exposure limit (STEL).

Leave the area. Respond according to company safety policy and guidelines.

The gas concentration is outside of the sensor's measuring range. Note: For a negative over-range condition, the "—OR" icon displays.

Leave the area. Respond according to company safety policy and guidelines.

TWA alarm event

High-alarm gas event2

STEL alarm event

Over-range alarm gas event2 (positive shown) Instrument-related alarm events

Low-battery alarm event

Dead battery indicator

System alarm event (sample screen)

From the time of initial display, there are fewer than 96 hours of remaining battery life. The battery-warning icon appears on the unit's display. The audible indicator and a red LED turn on and off every minute for the first 72 hours of the alarm condition. During the last 24 hours, these indicators continue but at a faster pace. The unit is operational.

Power warning: Respond according to company safety policy and guidelines. Qualified personnel can complete the battery replacement task (see “Service”).

After 96 hours, the battery has reached its end of life and the unit is not operational. The indicators noted above, along with the vibration indicator, turn on and off for approximately 10 minutes. The red LEDs will then turn on and off simultaneously for approximately 24 hours.

Leave the area. Do not use the unit. Qualified personnel can complete the battery replacement task (see “Service”).

A critical hardware or system fault has occurred. Note: The "483" value shown here represents a specific error code (both sensors in fault). The code number will vary based on the system alarm event that has occurred.

Leave the area. Do not use the unit. Respond according to company safety policy and guidelines. See a supervisor or contact ISC Technical Support (see "Contact Information"). For the sensor-related error codes (406, 408, 483, and 499) qualified personnel can complete the checks noted below and install or re-install the 33

Alerts, Warnings, and Alarms

Table 5.3 Recommended action by event type Event type

Possible causes

Status and recommended actions needed sensors (see “Service”).

406: illegal sensor position.

The instrument does not accept the installed sensor type.

Check the installed sensors for instrument compatibility.

408: no sensors found.

No sensors are installed or the installed sensors are not detected by the unit.

Check any installed sensor for proper installation.

483: both sensors in fault.

Neither sensor is operational.

Check each installed sensor for proper installation.

499: sensor-type mismatch.

The installed sensors are not of the same type.

Check to see that the installed sensors are of the same type.

! indicate that the unit is in need of service. Instrument service warnings The unit is overdue for bump testing. Depending on the unit's configured settings, some combination of visual and audible indicators may turn on every 60 seconds.

The unit is operational. Respond according to company safety policy and guidelines. Dock1 the unit.

The unit is overdue for calibration. Depending on the unit's configured settings, some combination of visual and audible indicators may turn on every 60 seconds.

The unit is operational. Respond according to company safety policy and guidelines. Dock1 the unit.

One sensor has failed zeroing. The warning and sensor-location icons indicate which sensor is in failure (sensor 2 or "SE 2" shown here). The zero icon is displayed to indicate the sensor must pass a zero.

Use Alert: The unit is operating as a single-sensor instrument; respond according to company safety policy and guidelines. Dock1 the unit. If the message persists, see a supervisor.

One sensor has failed bump testing. The warning and sensor-location icons indicate which sensor is in failure (sensor 2 or "SE 2" shown here). The bump test and calibration cylinder icons are displayed to indicate the sensor must pass a bump test or calibration.

Use Alert: The unit is operating as a single-sensor instrument; respond according to company safety policy and guidelines. Dock1 the unit

One sensor has failed calibration. The warning and sensor-location icons indicate which sensor is in failure (sensor 2 or "SE 2" shown here). The calibration cylinder icon is displayed to indicate the sensor must pass calibration.

Use Alert: The unit is operating as a single-sensor instrument; respond according to company safety policy and guidelines. Dock1 the unit If the message persists, see a supervisor.

Bump test overdue

Calibration overdue

Single-sensor zero failure

Single-sensor bump test failure

Single-sensor calibration failure

!

provide information about the unit’s operational status.

Instrument status alerts

34

Event Causes and Recommended Actions

Table 5.3 Recommended action by event type Event type

Possible causes

Status and recommended actions

Only one sensor is installed and operational (indicated by the check mark). The warning and sensor-location icon are displayed to indicate which sensor is missing (sensor 1 or "SE 1" shown here).

Use Alert: The unit is operating as a single-sensor instrument; respond according to company safety policy and guidelines.

A data-related function has failed for one of the sensors. The warning and sensor-location icons indicate which sensor is in failure (sensor 1 or "SE 1" shown here).

Use Alert: The unit is operating as a single-sensor instrument; respond according to company safety policy and guidelines.

One sensor installed

Single-sensor data failure 1 When an instrument is docked, the iNet DS and the DS2 docking stations will automatically perform scheduled tasks; otherwise, qualified personnel can manually complete any needed task (see "Zero, Calibration, and Bump Testing"). 2Once the detected gas concentration changes, the alarm indicators will change to reflect any new condition such as low-alarm gas, high-alarm gas, over-range gas, or no gas alarm. Note: For additional assistance in resolving any alert, warning, or alarm, see a supervisor or contact ISC Technical Support (see "Contact Information")

35

.

36

6 Service and Warranty Service Instructions Supplies Three-dimensional Diagrams Service Tasks Warranty Policy Limitation of Liability

Service Instructions Perform all service tasks on a nonconductive surface in a well-lit area that is known to be nonhazardous. Wear grounding straps to prevent electrostatic discharge (ESD) which can cause damage to the unit's electronics. When working with the adhesive-backed filters and gaskets:  Be careful not to pierce or tear these items.  When using tweezers, apply gentle pressure.  Once the adhesive touches a surface, any attempt to remove or reposition the item may cause it damage. When working with sensors and the case top's water barriers:  Do not touch the white membranes as this can contaminate these items.  Use care not to damage the membranes.  Use care not to separate the sensor from its membrane.

SUPPLIES Torx screwdriver (for case bottom screws) Philips screwdriver (for clip screw) Needle-nose tweezers (for barrier and filter service tasks)

Service and Warranty

THREE-DIMENSIONAL DIAGRAMS Refer to the three-dimensional diagrams for disassembled views of the instrument and its case top assembly. Use the diagram number to identify parts, part numbers, and field-replaceable items (see Table 6.1).

Figure 6.1 Disassembled Tango TX1

Figure 6.2 Disassembled Tango TX1 case top assembly

38

Service Instructions

Table 6.1 Key for the Tango TX1 diagram Diagram number

Part name

Field replaceable

Part number

Notes

--

Case top assembly The case top can be replaced as a single assembly that includes the parts labeled with diagram numbers: 1, 14, 15, 16, and 17. These components are also sold separately.

Yes

17153951

Assembly includes case top (17153952); sensor water barriers and gaskets (17154219 and 17154051, respectively); and sensor and audible alarm dust barriers (17154540 and 17154581, respectively).

1

Case top

Yes

17153952

15

Sensor water barrier

Yes

17154219

16

Sensor water barrier gasket

Yes

17154051

17

Sensor dust barrier

Yes

17154540

18

Audible alarm dust barrier

Yes

17154581

19

Vibration alarm motor

Yes

17127275

2

Sensors

Yes

Varies

CO

Yes

17155161

Parts 17154219 and 17154051 should be replaced at the same time. Kit 18109230 contains 10 barriers and 10 gaskets. Kit 18109218 contains 10 sensor dust barriers and 5 audible alarm dust barriers. Notes: The dust barriers are not water impenetrable. More frequent replacement service may be needed in harsh environments.

Includes two sensors and polycarbonate plates.

H2S

Yes

17155164

Includes two sensors and polycarbonate plates.

NO2

Yes

17155162

Includes two sensors and polycarbonate plates.

SO2

Yes

17155163

Includes two sensors and polycarbonate plates.

4

Battery

Yes

17154367

3 and 5

LCD

No*

17153786

7

Polycarbonate plate (not sold separately)

Yes



6 and 8

Board assembly

No*



11 and 12

Unit labels

No*



13

Suspender clip

Yes

17154484

See “Sensors” above.

not shown

Belt clip (optional)

Yes

17120908

not shown

Audio alarm amplifier (optional)

Yes

17154915

14

Screw (for use with installed suspender clip and optional belt clip)

Yes

17139262

Torque: 81 newton cm (115 ounce-force inch)

10

Case bottom screws

Yes

17154328

Torque: 85 newton cm (120 ounce-force inch)

9

Case bottom

No*

17153769

*For items that are not field-replaceable, contact ISC (see "Contact Information") or a local distributor of ISC products.

39

Service and Warranty

SERVICE TASKS Table 6.2 Service tasks Power off the unit before disassembling or performing any service task. Instrument disassembly 

Using a torx screwdriver, remove all four screws from the case bottom; set aside the screws.

Hold the case bottom near the upper screw holes; lift the case top slightly to separate the top and bottom.

Continue to lift the case top straight up to remove it and to avoid unintentionally loosening the sensors. If replacing just the case top, refer to the task below, "Instrument assembly". Otherwise, proceed with the desired service tasks below.

Sensor port and audible alarm dust barrier replacement (sensor port dust barrier shown)

Using a finger or needle-nose tweezers, peel off the dust barrier and discard.

Place the barrier packet on the work surface. Scrape lightly across the paper to the barrier's edge; gently lift to expose a portion of its adhesive back. Peel the barrier from the packet.

Guide the new barrier—adhesive side down—onto the case top. Using a finger, press and hold for five seconds to activate the adhesive.

Sensor water barrier assembly replacement Note. The sensor water-barrier assembly consists of two parts: the filter and its gasket. Replace both items at the same time.

Inside the case top, grip the gasket and underlying sensor filter with the needle-nose tweezers; peel to remove.

40

Remove any remnants of the adhesive, filter, or gasket. Clear away any dirt, dust, or debris.

Place the filter packet on the work surface. Using the tweezers, scrape lightly across the paper to the filter's edge; gently lift to expose a portion of the adhesive back. Grip the filter lightly with the tweezers; peel the filter from the packet.

Service Instructions

Table 6.2 Service tasks

Guide the new filter—adhesive side down—into the filter opening. For proper placement, take care to ensure the filter edge meets the inner edge of the filter opening.

Using a clean, soft cloth, press gently around the filter edge; hold for five seconds to activate the adhesive.

Place the gasket packet on the work surface. Using the tweezers, scrape lightly across the paper to the gasket ring's edge; gently lift to expose a portion of the adhesive back. Grip the gasket ring lightly with the tweezers; peel the gasket ring from the packet.



Guide the gasket—adhesive side down—into the filter opening, placing it on top of the filter. Ensure the gasket ring edge meets the outer edge of the filter opening and fully covers the white filter membrane.



Using a clean, soft cloth, press gently around the gasket edge; hold for five seconds to activate the adhesive.

Vibrating alarm motor replacement

Insert the tweezers between the case top and the motor. Pry up to remove.

Using the tweezers, gently grip the new motor. Place the new motor – contact side up – into the case top. Without touching the motor's contacts, use the tweezer handle to press the item into place.

Sensor replacement ─

Lift the sensor to remove it. Some sensors may have an adhesive backing holding them in place; use gentle pressure to lift and remove the sensors. Set aside for future use or dispose of according to company policy.

Examine the circuit board for the presence of a black polycarbonate plate in each sensor position. As shown above (left) neither sensor position has the polycarbonate plate. If both sensor positions are equipped with the plates as shown above (right), skip to the sensor placement steps below.

41

Service and Warranty

Table 6.2 Service tasks ─

The plates are adhered to a paper backing. Bend the paper backing to separate the plate. Lightly grip the plate near its top with tweezers; gently lift to completely remove from paper backing. Guide the plate – adhesive side down – for correct placement on the circuit board as shown above (right). Take care to ensure the plate edge aligns with the sensor outline on the board.

Using a clean, soft cloth, press gently into place.



For each sensor, use tweezers to remove the paper liner from the bottom of the sensor and expose the adhesive backing.

Position the sensor to align with its connector on the instrument board.

Secure the sensor in place by applying gentle pressure to the sides of the sensor case. Do not touch the sensor's white membrane. A slight connection impact can be felt when the sensor is secured into place.

Battery replacement 

Lift the battery up from its cradle. Dispose of according to company policy.

Orient the replacement battery so the positive and negative ends align with the "+" and "" cradle markings, respectively. Place the new battery into the cradle negative end first.

Press down on the battery to secure it in the cradle.

Note: Once started, if the battery replacement task is not completed within 60 minutes, the following will occur:  Any data will be lost that was not downloaded prior to the start of the task.  The unit's time and date settings will be erased. When these settings are erased, the user will be prompted, during the next start-up sequence, to enter the correct time and date. Note: It is essential—for data log accuracy—that these tasks be completed. The data log plays an important role in preserving operator safety, and in the investigation of any potential incident, it can be useful to the safety team or a prospective investigator.

42

Service Instructions

Table 6.2 Service tasks Clip removal and attachment (suspender clip shown) 

Lift the clip's cover.

To attach the clip cover, place the screw through the center hole of the clip back.

Use a Philips screwdriver to access the clip's screw. Turn counterclockwise to loosen and remove the screw and washer.

Lift the clip to remove it. Close the clip; store it for future use. 



Press to secure the case top and case bottom.

Using a torx screwdriver, insert and tighten each of the four screws into the case bottom.

Guide the screwdriver through the clip's hole and into the screw head. Turn clockwise to tighten. Note: Refer to Table 6.1 for torque values.

Instrument assembly

Hold the case bottom near the upper screw holes.

Lower the case top assembly onto the case bottom. Use a straight-down motion to guide the sensors into the sensor barrels.

Note: Refer to Table 6.1 for torque values.

43

Service and Warranty

Warranty Policy Industrial Scientific Corporation’s portable Tango TX1 gas-monitoring instrument is warranted to be free from defects in material and workmanship under normal and proper use and service for three years from the initial purchase date. The above warranty does not include the sensors, battery, or filters, but the sensors carry their own separate warranty. The factory-installed sensors are warranted to be free from defects in material and workmanship under normal and proper use and service as follows, except where otherwise stated in writing in Industrial Scientific literature accompanying the product: 

CO and H2S sensors are warranted for three years from the initial purchase date.



All other sensors are warranted for two years from the initial purchase date.

LIMITATION OF LIABILITY THE WARRANTY SET FORTH ABOVE IS STRICTLY LIMITED TO ITS TERMS AND IS IN LIEU OF ALL OTHER WARRANTIES, GUARANTEES, EXPRESS OR IMPLIED, ARISING BY OPERATION OF LAW, COURSE OF DEALING, USAGE OF TRADE OR OTHERWISE. INDUSTRIAL SCIENTIFIC MAKES NO OTHER WARRANTIES, EITHER EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE WARRANTIES OF MERCHANTABILITY OR FITNESS FOR PARTICULAR PURPOSE. Should the product fail to conform to the above warranty, buyer's only remedy and Industrial Scientific’s only obligation shall be, at Industrial Scientific’s sole option, replacement or repair of such non-conforming goods or refund of the original purchase price of the non-conforming goods. In no event will Industrial Scientific be liable for any other SPECIAL, INCIDENTAL OR CONSEQUENTIAL OR OTHER SIMILAR DAMAGES, including loss of profit or loss of use, arising out of the sale, manufacture or use of any products sold hereunder whether such claim is pleaded in contract or in tort, including strict liability in tort and whether Industrial Scientific has been advised of the potential for such damages. Industrial Scientific’s total liability hereunder from any cause whatsoever (except liability from personal injury caused by Industrial Scientific’s negligence), whether arising under contract, warranty, tort (including negligence), strict liability, products liability or any other theory of liability, will be limited to the lesser of Buyer’s actual damages or the price paid to Industrial Scientific for the Products that are the subject of Buyer’s claim. All claims against Industrial Scientific must be brought within one year after the cause of action arises, and Buyer expressly waives any longer statute of limitations. It shall be an express condition to Industrial Scientific's warranty that all products be carefully inspected for damage by Buyer upon receipt, be properly calibrated for Buyer's particular use, and be used, repaired, and maintained in strict accordance with the instructions set forth in Industrial Scientific's product literature. Repair or maintenance by non-qualified personnel will invalidate the warranty, as will the use of non-approved consumables or spare parts. As with any other sophisticated product, it is essential and a condition of Industrial Scientific's warranty that all personnel using the products be fully acquainted with their use, capabilities and limitations as set forth in the applicable product literature. Buyer acknowledges that it alone has determined the intended purpose and suitability of the goods purchased. It is expressly agreed by the parties that any technical or other advice given by Industrial Scientific with respect to the use of the goods or services is given without charge and at Buyer’s risk; therefore, Industrial Scientific assumes no obligations or liability for the advice given or results obtained.

44

Appendix Supplemental Information about Sensors and Gases Table A.1. Sensor cross interference (percent response) Sensor type Carbon Monoxide %

Hydrogen Sulfide %

Nitrogen Dioxide %

Sulfur Dioxide %

Carbon Monoxide

100.0

2.0

0.0

1.0

Hydrogen Sulfide

10.0

100.0

8.0

1.0

Sulfur Dioxide

0.0

10.0

0.0

100.0

Nitrogen Dioxide

20.0

20.0

100.0

100.0

Chlorine

10.0

20.0

90.0

25.0

Target gas

Chlorine Dioxide









Hydrogen Cyanide

15.0

10.0

1.0

50.0

Hydrogen Chloride

3.0

0.0

0.0

0.0

Phosphine









Nitric Oxide

10.0

1.0

0.0

1.0

Hydrogen

60.0

0.08

0.0

0.5

Ammonia

0.0

0.0

0.0

0.0

— No data available Note: This table is provided as a guide only and is subject to change. The data reflect the percentage response of the sensor type when exposed to a known concentration of a target gas. For example, when the carbon monoxide sensor is exposed to carbon monoxide in the air sample, the gas concentration reading accurately reflects the actual concentration of carbon monoxide, so the sensor's response is said to be 100%. When an air sample contains hydrogen, it triggers a carbon monoxide reading. The carbon monoxide sensor's response to hydrogen is approximately 60% meaning that exposure to 100 ppm hydrogen will produce a reading of approximately 60 ppm carbon monoxide.

Marking Requirements Table A.2. ATEX and IECEx marking requirements ATEX markings

IECEx markings

Industrial Scientific Corp. 15071 USA TANGO TX1 DEMKO 12 ATEX 1209126 Ex ia I Ma Ex ia IIC T4 Ga -40 oC ≤ Ta ≤ +50 oC [Serial Number] [Month/Year of Production]

Industrial Scientific Corp. 15071 USA TANGO TX1 IECEx UL12.0041 Ex ia I Ma Ex ia IIC T4 Ga -40 oC ≤ Ta ≤ +50 oC [Serial Number] [Month/Year of Production]

45

46

47

Contact Information Americas USA 1001 Oakdale Rd. Oakdale, PA 15071-1500 USA Phone: +1 412-788-4353 1-800-DETECTS (338-3287) e-mail: [email protected]

Canada

Phone: +780-467-2423 e-mail: [email protected]

Europe/Middle East/Africa France Phone: +00 800 WORKSAFE + 33-157329261 e-mail: [email protected]

Germany Phone: + 49-69299571416 e-mail: [email protected]

Czech Republic Phone: +420 234 622 222 e-mail: [email protected]

United Arab Emirates Phone: +971 50 455 8518 e-mail: [email protected]

England Phone: +44 12 80 70 61 14 e-mail: [email protected]

Asia Pacific/China People's Republic of China Phone: +86 21 5899 3279 +86 400 820 2515 e-mail: [email protected]

Singapore Phone: +65 6561 7377 e-mail: [email protected]

Australia Phone: +03 96447777 e-mail: [email protected]

S100D-LR / S125D-LR

2) S125DLR ▪ General Specification

OPERATING MANUAL

Model

LED Revolving Warning Light with Phone

Rated Voltage

Current

Weight

DC12V

0.395A

1.18Kg

DC24V

0.342A

1.18Kg

AC110V

0.092A

1.57Kg

AC220V

0.046A

1.57Kg

S125DLR

RPM

Color

Green-R

120-140/Min.

Amber-A Green-G Blue-B

Thank you for purchasing Q.Light product. Please read this manual carefully before installing this system for safe and long use of the product.

▪ External View Drawing

▪ Signal sound built-in product of LED revolving warning light which can transmit both visual signal and auditory signal at same time. ▪ Excellent in diffusing signal light by employing unique direct-diffusion system. ▪ Material : Lens-AS, Body-ABS ▪ RPM : 120-140/Min.(Controllable) ▪ Volume : Max. 105dB/1m ▪ Protection Class : IP54 ▪ Each model has 5 sound tones built-in. Select a suitable sound tone for the purpose of use Main Built-in Sound

CH

5 Warning Sounds

ch1 ch2 ch3 ch4 ch5

Fire A-ANG(Wail) Emergency WA-U(Yelp) Ambulance PI-PO(High-Low) Machinery Fault( PI-PI-PI-) Foam Release Alarm(PI~ Continuous)

105dB

WM 5 Melodies

ch1 ch2 ch3 ch4 ch5

Sweet Home For Elise Cuckoo’s Waltz Piano Sonata Turkish March

105dB

WA 5 Alarms

ch1 ch2 ch3 ch4 ch5

Beep(Intermittent) Door Chime Phone Ring Machinery Fault(PI.PI.PI.PI.) Tripping Bead Sound

103dB

ch1 Reproduction of Voice Recording ch2 3 Channels ch3

95dB

Type

1) S100DLR

WS

▪ General Specification Model

Rated Voltage

Current

Weight

DC12V

0.395A

1.17Kg

DC24V

0.342A

RPM

Green-R

1.17Kg 120-140/Min.

S100DLR AC110V

0.092A

1.56Kg

AC220V

0.046A

1.56Kg

Color

Amber-A Green-G Blue-B

▪ External View Drawing WV

3 Voice Signals

Volume

Sound Tone

Power line and signal line come out from the product. Please refer to below wiring diagram for correct connection of power supply and signal line. ▸ Power Line : UL1015-NO.16(1.25sq) x 2C

400mm

▪ In case of product using AC voltage : Connect the line without polarity(wire color).

▪ In case of product using DC voltage : Black line is negative(-).

wire disconnection or short circuit. ▪ Do not strongly pull the wire out from the product. It can cause a

▸ Signal Line : UL1007-NO.22(0.3sq) 400mm

disorder. ▪ In case of product using AC voltage, do not allow power supply either to common line or to signal line.

Situation Non- Lighting Partly

Feeble Lighting Low Revolving Speed

Double Sound Tone

Indoor

Outdoor

Main Cause

Making Sure Using Voltage

Wrong Selecting Voltage Selection Switch to 220V (In Case of AC Type)

Selecting Voltage Selection Switch to 110V

Making Sure Using Voltage

Short Circuit of Sound Making Sure Wiring Tone Signal Line Problem with Sound Contact Q.Light or Tone Selection Switch Their Agent Wrong Wiring

Making Sure Wiring

Problem with Horn

Contact Q.Light or Their Agent

Wrong Wiring

Making Sure Wiring

No Sound

Neither Lighting nor Revolving

Solution

LED is Partly Damaged Supplying Higher Voltage than Rated Voltage Supplying Voltage Lower Than Rated Voltage

No Power Supply

Making Sure Power Supply

Q-LIGHT CO., LTD. http://www.qlightkr.com E-mail : [email protected]

Seoul Office (Domestic Sales & A/S) Tel) 02-2679-6152 Ansung BD A-312, 410-13 Sindorim-Dong, Guro-Gu, Seoul

▪ This product was manufactured with the structure for indoor use. Installing in any direction is okay for indoor use. ▪ Outdoor use is limited at a place where no water or very little water percolates. Mounting also should be done only with vertical standing shape. ▪ Mount at the place where exists low vibration, low impact, low moisture and low dust as long as possible. ▪ Refer to below information and drill necessary mounting holes on profitable part of the product. (Make sure PCD from external view drawing.) ▪ Fix the product by using nuts(M5-3EA) originally equipped on mounting bolt on bottom of the product.

▪ If using voltage is seriously higher or lower than rated voltage, it can reduce the self-life of the product or can cause a problem on product operating. Make sure the using voltage does not exceed +/- 5% of the rated voltage. ▪ If lens or body of the product is contaminated by chemicals such as thinner, benzene, the product can be damaged, discolored or deformed. Please be careful. ▪ Do not give strong impact to the product. It can be a reason of

Overseas Marketing Team / Plant Tel) 055-328-4082 Fax) 055-328-4083 Plant #1 : 178-1 Woogye-Ri, Sangdong-Myun, Kimhae-Si, KyungNam, Korea Plant #2 : 417-1 Geumsa-Dong, Geumjung-Gu, Busan

Related Documents


More Documents from ""