General Specifications

PH400G 4-Wire Type pH Meter

GS 12B7C1-E

GENERAL The EXA PH intelligent pH conveter is a highly reliable advanced function pH measurement system designed for wide application in all industries such as for medium-scale waste water treatment system, and general-purpose pH measurement and control systems. The EXA PH intelligent pH converter features highly effective built-in functions for sensor diagnosis similar to FLXA21 2-wire liquid analyzer. These functions monitor sensor errors even while pH measurements are in progress, enabling more highly reliable pH measurement. The converter also offers an interactive system for setting the various parameters, facilitating effective use of its wealth of functions. The EXA PH intelligent pH converter is available not only in versions for general-purpose applications but also for high-purity water service.

High-Purity Water pH Measurements High-purity water today is widely used in numerous process plants in a variety of chemical processes and boilers, with heavy emphasis placed on pH measurement and monitoring for water quality control and water purification unit operational control. However, the use of general-purpose pH meters for pH measurements in high-purity water meets with the following problems. d Since measurement solutions are chemically pure, their pH values may vary owing to carbon dioxide in the atmosphere and / or impurities in measurement lines. d The pH electrode output is likely to be affected by streaming potentials, externally induced potentials, etc. and thus likely to produce measurement error because the liquid resistance between the electrodes in measurement solution is high owing to low solution conductivity. Since the pH sensors and their holders for high-purity water service have been designed considering countermeasures against the above problems, they can provide highly reliable pH measurements. Moreover, since the EXA PH intelligent pH converter is equipped with a reference temperature conversion function to compensate for pH variation owing to measured liquid temperature variation, any effect from sample liquid temperature fluctuation is slight, and devices such as sampling system coolers, etc. can thus be simplified.

FEATURES Microprocessor - Based Intelligent pH converter d Sensor failure can be monitored even during pH measurement. d Automatic calibration using standard buffer solutions can be initiated by merely a single touch of a key. Tables for three types of standard solution are incorporated. This allows for simple automatic calibration.

PH400G

d Sensor characteristic deterioration is automatically checked during automatic calibration using the standard solutions. The sensor replacement time that previously depended on intuition can now be decided automatically. d Daily operation can be performed without opening the case. Even in rainy weather, since the routine maintenance operation such as calibration using standard solutions can be performed without opening the case lid, there is no worry about insulation deterioration. d Ranges can be freely set in the field if the span is 1 pH or more. d A timer function for cleaning is incorporated. d Abundant built-in contact output function d Operating panel has excellent operability.

Ryton pH Sensor Body d The sensor body uses Ryton which compares with Teflon in corrosion resistance and heat resistance and has excellent mechanical properties. This pH sensor can cover most applications. d This pH sensor is easy to handle. As this is a composite pH sensor, calibration using standard solutions or maintenance can be easily performed. Glass electrode and junction can also be simply replaced.

Yokogawa Electric Corporation 2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 Japan Tel.: 81-422-52-5617 Fax.: 81-422-52-6792

GS 12B7C1-E ©Copyright Jan. 1991 6th Edition Nov. 2012

2

<> <>

SYSTEM CONFIGURATION For the pH sensors, see GS 12B07B02-E. For the holders or cleaning devices, see GS 12J05C02-00E. Fig. 1-a System Configuration (General Purpose) pH sensors d Xerolyt HA405/HA485

d KCL refillable type PH8ERP

Holder, Holder with Cleaning system d Suspension Type Holder HH350G

d Guide Pipe PH8HG



pH Converter d 4-wires pH Converter PH400G

<With Jet Cleaning> pH

NO

MODE

>

>

MEASURE AUT CAL MAN CAL TCMP DISP HOLD

YES

ENT

CONTACTS S1 S2 S3 FAIL

d for Chemical DPA405/DPA485

d Angled Floating Ball Holder PB350G

d Submersion Type Holder d KCL filling type PH8EFP

Float

d Vertical Floating Ball Holder

d Hydrofluoric Acid-resistive HF405

No Cleaning

Accessories

Ultrasonic Cleaning Jet Cleaning Brush Cleaning d Flow Type Holder PH8HF

Cleaning Devices d Ultrasonic Oscillator PUS400G

d Sensor Stand

No Cleaning

Ultrasonic Cleaning Jet Cleaning Brush Cleaning

Fig. 1-b System Configuration (For Pure Water) Holder d For pure water PH8HH d For pure water PH8EHP

Accessories

pH Converter d 4-wires pH Converter PH400G

PH8AX d Sensor Stand d Calibration Reagent and KCL Solution

pH

MEASURE AUT CAL MAN CAL TCMP DISP HOLD

YES

NO

MODE

>

>

pH sensor

ENT

CONTACTS S1 S2 S3 FAIL

F1-a.b.c.EPS

CAUTION Installation Location of Holders (Guide Pipe, Submersion Type, etc) The holder should be used in a place that is as vibration free as possible. Using the holder in a place where it is affected by vibration, may result in damage to the holder.

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

3

<> <>

SPECIFICATIONS EXA PH Series General Specifications Measured object: Hydrogen ion concentration (pH) in aqueous solutions Measuring principle: Glass electrode method Measuring range: 0 to 14 pH Measuring conditions: Solution temperature: See Table 1, for high-purity water, 0 to 508C Solution pressure: See Table 2, for high-purity water, atmospheric pressure (measured liquid outlet side open to atmosphere) Solution flow rate: Max. 2 m/s for submersion guidepipe type 3 to 11 l/min for flow-through type, for high-purity water, see Figure 2. Solution conductivity: Min. 50 mS/cm, for high-purity water, see Figure 1. Temperature compensation range: -10 to 1308C (for pH converter) Table 1. Process Temperature Range

(a) pH and Liquid Temperature Ranges for pH sensors Solution pH Ranges

Solution Temperature Ranges (8C)

2 to 12

-5 to 80

Guide-pipe

2 to 12

-5 to 80

Submersion *3 Flow-through

0 to 14

Holder Type

Guide-pipe *3 KCl refilable Solution type Flow-through KCl filling type

*2

Table 2. Process Pressure Range pH Sensor Holder Submersion

PH8ERP

PH8EFP

Atmospheric pressure (Submersion depth: Max. 3 m)

Guide-pipe Atmospheric pressure Suspension (Submersion depth: Max.3 m) Float Atmospheric pressure to 10 kPa when general purpose reserve Atmospheric tank used. Flow-through pressure Atmospheric pressure to 500 kPa to 50 kPa when medium-pressure reserve tank used. (Note 1)

For flow-through types, refer also to the solution temperature and pressure diagram of Holder GS 12J5C2-E. T03.EPS

*1

-5 to 105

*2 *1

T01.EPS

ml/min

pH Sensor

Notes: (*1): The solution temperature should preferably be below 808C when the pH value is in the range 11 to 14. Even if the solution is below 808C, sensor service life may be shortened in store solutions. In such cases the KCl filling type is recommended. (*2): Maximum temperature 508C when Hard PVC guidepipe is used. (*3): For flow-through types, refer also to the solution temperature and pressure graph (in notes following Model and Suffix code table for flow-through type holders). (*4): Solutions with normal pH ranges of 3 to 14 are recommended for stainless steel (SUS316).

700 600

(b) Solution Temperature Ranges for Holders Holder Type

Guide-pipe

Holder Material

Cleaner

Liquid Temperature (8C)

PV

None

-5 to 50

PP

None

-5 to 80

None

-5 to 100

Provided

-5 to 80

PP Submersion S3

*4

PP Flow-through *3

S3

*4

PV: Hard PVC PP: Polypropylene S3: Stainless steel (SUS 316)

None

-5 to 100

Provided

-5 to 80

None

-5 to 80

Provided

-5 to 80

None

-5 to 105

Provided

-5 to 80

Solution flow rate

500 400 300

Usable range

200 100 0 0

0.1

0.2

0.3

0.4

0.5

0.6

50

Solution conductivity (mS/cm) F02.EPS

Figure 2 Solution flow rate and solution conductivity of sensor and holder for high purity water

T02.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

4

<> <>

1. pH Sensors: PH8ERP, PH8EFP, PH8EHP

Table 4. Utilities (For Jet Cleaning or Brush Cleaning)

A compound sensor assembly consisting of glass electrode, reference electrode, temperature sensor, and ground tip.

Water jet

200 to 400 + Liquid pressure

5 to 20 l/min

Type: KCl filling and KCl refillable types

Water brush 100 to 250 + Liquid pressure

20 to 30 l/min

Measuring range: 0 to 14 pH for KCl refillable type; normal pH =2 to 12 pH) Refer to "Selection Criteria for pH Sensors and Holders" shown on the last page to select KCl filling or refillable type sensors. Temperature compensation sensor: Platinum RTD (Pt1000) Wetted part materials:

Air jet

200 to 400 + Liquid pressure 100 to 300 Nl/min

Air brush

150 to 250 + Liquid pressure 300 to 600 Nl/min

pH Sensor

Wetted parts materials

KCl refillable type

Ryton (PPS resin), glass, ceramics, titanium or Hastelloy C, fluorocarbon rubber and chlorinated polyethylen rubber (carbon sheath only when suspended in guide-pipe)

KCl filling type

Same as above. Plus heat resistant soft PVC (KCl tube only when suspended in guide-pipe) T04.EPS

Weight: KCl refillable type; Approx. 0.4 kg KCl filling type; Body: Approx. 0.4 kg General purpose: Approx. 0.3 kg Medium-pressure: Approx. 1 kg For high-purity water; Body: Approx. 0.4 kg Tank: Approx. 0.3 kg Tank:

Note: Although sensors other than the EXA PH pH sensor (Pt1000 temperature sensor) can also be used, there are some restrictions on function. Also, since reference performance values will differ, please consult your Yokogawa representative.

2. Holders 2.1 General-Purpose Holders: PH8HG, PH8HF, PH8HS Materials: Polypropylene, Stainless steel, PVC (guidepipe only) Cleaning: Ultrasonic, jet, or brush cleaning Cleaning Unit Wetted Materials: Ultrasonic unit: Stainless steel, PVC (guide-pipe only) Jet unit: Polypropylene Brush unit: Polypropylene, titanium (shaft), Ryton(bearings) Mounting: Mount on 50 A vertical or horizontal pipe Mounting hardware: One or two sets (submersion type) or one set (flow-through type) Weight: See Table 3.

Pressure (kPa)

Flow Rate

T06.EPS

2.2 Holder For High-purity Water: PH8HH Materials: Acrylic resin, SUS316, chloroprene rubber Process connections: Rc1/4 (inlet) and R1/2 (outlet), or 1/4 NPT (inlet) and 1/2 NPT (outlet) Mounting: Mount on 50 A vertical or horizontal pipe, or wall-mount (specify mounting hardware) Weight: Body: 1.7 kg Mounting hardware: 0.7 kg

3.

4-wires type pH Converters: PH400G

Measuring Range pH: - 2 to15 pH Temperature : -10 to 1308C Display Method: Digital display Display Range pH : -2 to15 pH Temperature : -10 to 1308C Transmission Signal Output 4 to 20 mA DC, isolated transmission output, Max. load 600 V 0 to 1 V DC, isolated transmission output, Minimum load 1 kV (However, the above current (4 to 20 mA) signal and the voltage (0 to 1 V) signal are not isolated from each other.)

pH or temperature can be freely set.

Transmission Signal Range pH: Freely adjustable to any desired range of 1 pH span or greater (set at 0 to14 pH when shipped). Temperature: Freely settable to any desired range of 508C span or greater (set to 0 to 1008C when shipped).

Table 3. Holder Weight

Materials Holder

Submersion Flow-through Guide-pipe

Polyethylen

Stainless steel (SUS316)

Approx. 0.5 to Approx. 1.5 2.2kg to 6kg Approx. 0.5 to Approx. 7 to 8.5kg 2kg Approx. 1kg

Hard PVC

Approx. 1.6kg T05.EPS

Note: Table does not include mounting hardware weights. Mounting hardware weight: For submersion type: Approx. 1 kg / set For flow-through type: Approx. 0.5 kg All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

5

<> <> Contact ON / OFF

Contact Output ?Contact output function that can be set Contact

ON/OFF, Proportional duty pulses Proportional frequency pulses *2

S2

OFF, low limit, high limit, HOLD

Ditto

OFF, low limit, high limit, HOLD, Cleaning timer, Hi - Hi limit, Lo - Lo limit

Ditto

S3

Failure

ON/OFF T07.EPS

output is to be used.

(*1) Proportional duty pulse Example of high limit alarm (conrol) output 10%

% duty cycle control 90% 100

50%

50%

t on

t off

50

10%

90%

0

Proportional range

pH

Duty cycle

% of output range

Time F02.EPS

(*2) Proportional frequency pulse Example of high limit alarm (conrol) output

% controller output 100

Maximum pulse frequency 0.3 s

Setpoint

pH

OFF

Open

OFF Open

ON Action

ON Closed OFF Open

OFF Closed

No pulses

% of output range

When closed, each indicator lamp is lit corresponding to each contact status. ?Contact Type: Relay contact output (dry contact) ?Contact Rating 250 V AC, 2 A, Max. 100 VA 220 V DC, 2 A, Max. 50 W Contact Input: Dry contact for manual start for automatic cleaning Ambient Temperature: -10 to 558C (hood may be fitting as option) Storage Temperature: -30 to 708C Ambient Humidity: 10 to 90% RH Construction: Watertight complying with JIS C0920 equivalent to NEMA 4 waterproof construction Material: Case; Aluminum alloy casting Cover and window; Polycarbonate Finish: Baked polyurethane resin coating Colors: Cover; Deep sea moss green (Munsell 0.6GY3.1/2.0) Case; Frosty white (Munsell 2.5Y8.4/1.2) Mounting: Pipe mounting, wall mounting or rack mounting Panel mounting Power Supply Voltage: 88 to 132 V AC, 50 / 60 Hz 176 to 264 V AC, 50/60 Hz Power Consumption: Approx. 8.5 VA Electrical Connection: pH sensor: Watertight plastic gland equivalent to JIS A8 attached to pH sensor side, f13.5 hole Others: Watertight plastic gland equivalent to JIS A15 (for cable OD. 9 to 12 mm) attached, f21 hole Weight: Body: Approx. 2.5 kg Mounting bracket: Approx. 0.4 kg Dimensions: 144 (W) x 144 (H) x 135 (D) mm

Note: Use a sensor with solution earth sensor because of differential amplifier (that uses two high-impedance amplifiers).

50% pulse frequency

Proportional range

Action

ON Closed

Function Specifications Input Impedance: 1012 V or more

50

0

*1

Power

T08.EPS

(*1) and (*2) are effective when the low high limit contact

Setpoint

ON

Operation (free selectable)

OFF, low limit, high limit, HOLD

FAIL

FAIL Power

Function Setting (free selectable)

S1

S1, S2, S3

Time F03.EPS

Settable parameters ? pH value setpoint ? Control range ? Pulse periode

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

Asymmetry Potential Adjustable Range: pH 7 6 2 pH Slope Adjusting Range: Adjustable in the range from 70 % to 110 % of the theoretical value. Automatic Temperature Compensation Range: -10 to 1308C (manual compensation is also available) Reference Temperature Conversion Coefficient (Reference Temperature 258C): On shipment: 0 Adjustable range: -1.00 to 1.00 pH / 108C Conversion to the reference temperature is used only for high-purity water or when the measured solution temperature coefficient is known.

GS 12B7C1-E

6th Edition 2012.11.10-00

6

<> <> Standard Performance (performance when combined with a pH sensor) Repeatability: 0.05 pH (sensor submerged 3 times in the same buffer solution) Response Time: 10 sec (90 % response, using pH sensor and buffer solution both at temperature equilibrium at 208C, with adequate agitation) Accuracy: 60.1 pH (using KCl filling type pH sensor or high-purity water pH sensor ) 60.15 pH (using KCl refillable pH sensor) Temperature Repeatability: 18C Operating Functions Display: 3 1/2 digit digital display (data display) Six digit alphanumerics (message or data display) Display function: pH value Temperature value mA output mV output Reference electrode impedance The e.m.f. slope Asymmetry potential mV (e.m.f.) 90% response time Error display (at error occurrence) Hold display (when holding) Manual temperature compensation display (in manual temperature compensation setting) Interactive message Key operation requesting display Functions that can be set or executed at operation level: One-touch calibration (Buffer selection is manual but indication stability check is automatic.) Measurement of sensor 90 % response time Manual calibration Selection of message area display contents Temperature coefficient setting Auto / manual temperature compensation selection Manual temperature setting Hold set / reset Functions that can be set or executed at setting level: pH value setting for contacts (Sl, S2, and S3) Output range setting (mA and mV outputs) Hold parameter setting: Hold provided / no selection Selection of holding either the value immediately before or the preset value Preset value setting (mA and mV outputs) Cleaning parameter setting: Auto / manual cleaning selection Manual cleaning start / stop selection Automatic cleaning ON / OFF selection Cleaning period: Setting range 0.1 to 36 hours Relaxation time: Setting range 0.1 to 10 min Cleaning time: setting range 0.1 to 10 min Functions that can be set or executed at service level: 8C / 8F selection Use / no use of conversion to reference temperature Temperature sensor selection Checking item setting (asymmetry potential, e.m.f. slope, reference electrode impedance)

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

Setting of reference electrode impedance high limit value Setting of response time high limit value One-point temperature calibration Sensor type selection (glass electrode or antimony electrode) Error reset pH display value selection (0.1 pH / 0.01 pH) Half value recovery time check ON / OFF Half value recovery time setting (0.1 to 10 min) Burn-up or burn-down ON/OFF Response stabilization judging parameter setting Auto-return (approx. 60 min) ON/OFF pH standard solution temperature characteristics setting Temperature sensor cable length correction Signal output selection 4 to 20mA DC

OFF, pH Temperature

0 to 1V DC

OFF, pH Temperature T09.EPS

Contact (S1, S2, or S3) output selection Setting of contact output delay time (0.1 to 20 sec) and hysteresis (0.01 pH to 0.2 pH) Setting of proportional duty pulse contact output period (5 to 100 sec) Setting of maximum frequency (50 to 120 pulses / min) at the proportional frequency pulse contact output Control range (0 to 10 pH) Details of failure detection by self-diagnosis function: Response time error during calibration (time until pH value is settled) Asymmetry potential failure e.m.f. slope failure Temperature range failure pH range failure Glass electrode impedance failure (measured solution must be (50 mS/cm or more and temperature 608C or less) Reference electrode impedance failure (measured solution must be 50 mS/cm or more) Half value recovery time failure 90% response time failure Calibrating solution temperature failure

GS 12B7C1-E

6th Edition 2012.11.10-00

7

<> <>

4. Ultrasonic Oscillator (Model PUS400G) Combination device: Holder with ultrasonic cleaner (PH8HS, PH8HF) Connection cable are provided with holders. Cleaning method: Continuous ultrasonic emission (Frequency sweep method) Oscillation frequency: Approx. 61 to 81 kHz Output voltage: Approx. 40 to 70V Power supply: 100/110/115/200/220/240 V AC610% 50/60 Hz Power consumption: Approx. 15 VA Ambient Temperature: -10 to 50 8C (hood may be fitting as option) Storage Temperature: -25 to 70 8C Construction: JIS C0920 Watertight (NEMA 4 equivalent waterproof construction) Material: Case: Aluminum alloy casting Window: Polycarbonate Mounting bracket: Stainless steel Finish: Baked polyurethane resin (Standard) Baked epoxy resin (Option) Color: Case; Frosty white (Munsell 2.5Y8.4/1.2 or equivalent) Cover; Deep sea-moss green (Munsell 0.6GY3.1/ 2.0 equivalent) Mounting: (2-inch) pipe mounting, wall or rack mounting or panel mounting Cable inlet: f22.7 x 2 Pg16 watertight plastic gland Cable / Terminal: For 7 to 12 mm, M4 Conduit adapter: Power Supply side only (Option) Connection: G 1/2 or 1/2 NPT Weight: Body; Approx. 2.5 kg Mounting bracket; Approx. 0.7 kg (Note) Output of ultrasonic oscillator changes with power supply voltage or connected cable length.

5. Solenoid Valve for Jet/Brush Cleaning PH8MV Pilot kick operated, 2-port valve. Open when energized. Fluid: Normal tap water, industrial water, or air Operating pressure: 0 to 1 MPa Forward (reverse) pressure resistance: 2 MPa Fluid temperature: Water; 5 to 608C , Air; -10 to 608C Cv: 4.5 Process connection: Rc 1/2 Power supply: 100/110/200/220 V AC, 50/60 Hz Construction: IP53 Material: Body: Bronze Sealing: Nitrile rubber Ambient temperature: Maximum 508C Electric connection: G 1/2 Weight: Approx. 0.9 kg

Cleaning water outlet flow rate: Max. 30 l/min, (which can be regulated by the bypass valve.) Connectable unit: Holder or sensor with the jet cleaning system Tank: With a ball tap (ball float valve for level control) Effective capacity of approx. 40 liters. Pump: Suction pump with capacitor-starting-type induction motor. Equipped with an electromagnetic actuator with overcorrect protection. Cleaning water inlet connection: G 1/2 Cleaning water outlet connections: G 1/2 (Optional Rc 1/2 or ANSI 1/2NPT adapters are available.) Electric connection: Plastic watertight gland equivalent to JIS A15. Applicable cable outside diameter: 9 to 12 mm Conduit adapter: Optional Ambient temperature: 5 to 508C Power supply: 100 or 200 V AC 610%, 50/60 Hz, single phase (Must be equal to the converter’s power supply.) Power consumption: 0.4 kW Construction: For installation outdoors (in a place safe from rain, wind, and direct sunlight) Color: Light gray (Munsell 2.8GY6.4/0.9 equivalent) Dimensions: Approx. 442 (W) x 804 (D) x 800 (H) mm Finish: Baked polyurethane Weight: Approx. 55 kg

7.

Terminal Box PH8TBG Used when pH converter is installed remotely from the pH sensor. Ambient temperature: -10 to 50 8C Construction: JIS waterproof Case material: Fiberglass reinforced polycarbonate resin Case color: Grayish green (Munsell 2.5GY5.0/1.0) Electrical connections: pH Converter side: f21 mm hole (With G1/2 plastic gland) pH sensor side: f13 mm hole (With G1/2 plastic gland) With Cable (Maximum length 20 m) Conduit adapter (optional)

8.

Accessories (Purchased Separately)

See Model and Suffix Codes.

6. Cleaning Pump/Tank Assembly PH8PU1 Pump/tank assembly for water jet cleaning. Allows tap water to be used for cleaning, using a level-controlled tank. Cleaning water: Normal tap water or industrial water (when isolation from other water lines is required); maximum pressure of 500 kPa Cleaning water outlet pressure: Max. 300 kPa, (which can be regulated by the bypass valve.)

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

8

<> <>

MODEL AND SUFFIX CODES When ordering, specify model and codes, product name and / or part numbers. Item to be specified.

1. pH Sensors For selection of KCl filling or refillable type pH sensors, refer to "Selection Criteria For pH Sensors and Holders" on page 24, and "Table of Corrosion-Resistant Materials" on page 25 of this manual. d KCl Refillable Type pH Sensor

1. pH Sensor

PH8ERP, PH8EFP

2. Holder

PH8HG, PH8HS, PH8HF or PH8HH

3. pH Converter

PH400G

Model

Cable Length -03

7. Terminal box

PH8PU1

Only for installation of pH converter PH8TBG at distance from pH sensor.

8. Accessories

3m 5m 7m 10m 15m 20m

-05 -07 -10 -15 -20

PH8MV

For jet or brush cleaning only

Specifications KCl Refillable Type pH Sensor

PH8ERP

4. Ultrasonic oscillator for ultrasonic cleaning PUS400G 5. Solenoid valve 6. Cleaning pump/tank

Option Code

Suffix Code

Titanium Hastelloy C

-TN -HC

Solution Ground Tip

PH8AX

Always -N

-N T10.EPS

pH Measuring System

d KCl Filling Type pH Sensor Model

Option Code

Suffix Code

PH8EFP -03

3m 5m 7m 10m 15m 20m

Cable Length -05 and KCl Tube -07 Length -10

-15 -20 Solution Ground Tip

Titanium Hastelloy C

-TN -HC -TT1

For general purpose (250ml solution inlet) For medium pressure, Big volume tank (With 500 ml tank) For maintenance (for TT1, TT3) For maintenance (for TT2)

KCl Reserve Tank -TT2 (*1)

-TT3 -TN1 -TN2

Always -N

-N pH Measuring System

Style Option

Specifications KCl Filling Type pH Sensor

Style

*A O-ring /PF Special glass sensor /HA Special junction /TF

Option

O-ring

Suffix Code

Model

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

Specifications

PH8EHP Cable Length -03

3m

-05

5m

-07

7m

-10

10m

-15

15m

-20

20m

Solution Ground Tip

Titanium

-TN

(*1)

For general purpose (250 ml solution inlet) Big volume tank (With 500 ml tank) For maintenance (for TT1)

-TT1 -TT3 -TN1 -N

Measuring System

Air Sup. Regulator (to be prepared separately) To pH sensor *3: Mark band is shown by mark and fork terminals are used. *4: Mark band is shown by numeral and pin terminals are used. *5: The tag which indicated the color, the sign, and the number is attached to the cable of a sensor. *6: Choose Daielperfrow when this is used in organic solvent, high alkali or high temperature solution. T12.eps *7: Choose when using in high alkali solution. *8: Choose when using in the heavily contaminated application. *9: Mark band is shown by numeral and M3 ring terminals are used.

Option Code

pH sensor for high purity water

KCl Reserve Tank

*1: A 50A pipe mounting bracket is supplied with TT1, TT2 and TT3. Only a supply tube, but no KCl solution, is supplied with TN1 and TN2. Since a KCl solution is not supplied with TT2 and TT3, arrange it from among accessories or auxiliary parts. *2: Prepare an air pressure regulator as shown in the diagram below when the medium-pressure reserve tank is used. To pH sensor, Regulator, (to be prepared separately)

Daielperfrow (*4)

/PF

d pH Sensor for High-Purity Water

Style A Daielperfrow (*6) Glass sensor for high alkali (*7) Teflon junction (*8)

For PH450G,PH202/TB (*5)

Style A

*A

*1: Mark band is shown by mark and fork terminals are used. *2: Mark band is shown by numeral and pin terminals are used. *3: The tag which indicated the color, the sign, and the number is attached to the cable of a sensor. *4: Choose Daielperfrow when this is used in organic solvent, T11.eps high alkali or high temperature solution. *5: Mark band is shown by numeral and M3 ring terminals are used.

For PH200/PH400 (*3) For PH202 (*4) For PH100 (*5) For PH450G,PH202/TB (*9)

-T -E -B -G

For PH200/PH400 (*1) For PH202 (*2) For PH100 (*3)

-T -E -B -G

Style

Always -N

-H

For PH200/PH400 (*2)

-E

For PH202 (*3)

-G

For PH450G,PH202/TB (*4)

*A

Style A

* 1: Since a KCl solution is not supplied with TT3, arrange it from among accessories or auxiliary parts. T13.eps * 2: Mark band is shown by mark and fork terminals are used. * 3: Mark band is shown by numeral and pin terminals are used. * 4: Mark band is shown by numeral and M3 ring terminals are used.

GS 12B7C1-E

6th Edition 2012.11.10-00

9

<> <>

2.

Holders

d Guide-pipe Type Model

Option Code

Suffix Code

PH8HG

Description Guide-pipe (*1)

Material

Polyvinylchloride (fluid temperature 508C or less)

-PV -PP

Style Code

Polypropylene (fluid temperature 808C or less) Style A

*A

Option, Mounting Bracket (*2)

/MS5

Stainless steel mounting bracket (1 set)

*1 : Pipe length: 2 m *2 : A set of 2-inch galvanized pipe mounting bracket is provided as standard. T14.EPS

d Submersion Type Suffix Code

Model

Option Code

PH8HS Material

-PP

Polypropylene

-S3

Stainless steel

Pipe length

pH Measuring System

Cleaning System

-10

1.0m

-15

1.5m

-20

2.0m

-25

2.5m

-30

3.0m

-T

Always -T

-NN

None

-S3

For ultrasonic cleaning (Transducer: SUS316) (*1)

-TN

For ultrasonic cleaning (Transducer: Titanium) (*2)

-HC

For ultrasonic cleaning (Transducer: Hastelloy C) (*3)

-JT

For jet cleaning. The solenoid valve must be specified separately.

-BR

For brush cleaning. The solenoid valve must be specified separately.

Cable Length for Ultrasonic Cleaning

Connector for Jet or Brush Cleaning

-NN

None

-C3

3m

-C5

5m

-C6

7m

-C7

10m

-C8

15m

-C9

20m

-JP

Rc1/2

-NP

1/2 NPT

*A

Style Code

Mounting Bracket Options

Description Submersion type holder

(*5)

Style A

/MS1

Mounting bracket : 1 set

/MS2

Mounting bracket : 2 sets

/MS3

Stainless steel mounting bracket : 1 set

/MS4

Stainless steel mounting bracket : 2 sets

Special Mounting

/F1

With flange (Without Cleaning System)

/F2

With flange (With Cleaning System)

O-ring

/PF

Daielper (*4)

*1: General purpose (Normal pH3 to 14) *2: For salt water *3: For acid (Normal pH0 to 4) *4: Choose Daielperfrow when this holder is used in organic solvent, high alkali or high temperature alkali. *5: The required number of mounting bracket sets depends on the installation location and flow rate. In general, one set is sufficient for pipe lengths of 1 meter, and otherwise two sets are required. T15.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

10

<> <> d Flow-Through Type Model

Suffix Code

Option Code

PH8HF Material

Description Flow-through type holder

-PP

Polypropylene

-S3

Stainless steel

Process Connection

Refer to note below for selection

-JPT

Rc1

-NPT

1 NPT female thread

-J10

JIS 10K 25 FF flange

-A15

ANSI Class 150 1 FF flange equivalent

(for polypropylene holder -PP) ANSI Class 150 1 RF flange with serration

(for Stainless steel holder -S3) pH Measuring System

Cleaning System

Cable Length for Ultrasonic Cleaning

Connector for Jet or Brush Cleaning Style Code Options

-T

Always -T

-NN

None

-S3

For ultrasonic cleaning (Transducer: SUS316) (*1)

-TN

For ultrasonic cleaning (Transducer: Titanium) (*2)

-HC

For ultrasonic cleaning (Transducer: Hastelloy C) (*3)

-JT

For jet cleaning. The solenoid valve must be specified separately

-BR

For brush cleaning. The solenoid valve must be specified separately

-NN

None

-C1

1m

-C3

3m

-C5

5m

-C6

7m

-C7

10m

-C8

15m

-C9

20m

-JP

Rc1/2

-NP

1/2 NPT Style A

*A O-ring

/MF1

Mounting bracket (stainless steel)

/PF

Daielperfrow (*4)

(3) For stainless steel, normally a 3 to 14 pH value is recommended.

Temperature

*1: General purpose (Normal pH 3 to 14) *2: For salt water *3: For acid (Normal pH 0 to 4) *4: Choose Daielperfrow when this holder is used in organic solvent, high alkali or high temperature alkali. Notes:(1) Mounting bracket is generally not required when the stainless steel holder is installed in-line in a pipe It is required where the holder is installed in a sampling rack (in which case the U-bolt included in /MF1 in not used). (2) Criteria for material selection (-PP or -S3) 105 In general, polypropylene is recommended from the viewpoint of chemical resistance. 100 However stainless steel is recommend in any of the following cases: * (a) The liquid contains organic reagent, oxidizing agents, etc., which can attack Stainless Steel 80 polypropylene. * (b) The temperature/pressure correlation of the process condition falls within the hatched 60 area of the diagram shown right. (8C) (c) The use of polypropylene is not reasonable from a viewpoint of strength or past 40 experience. 20

Polypropylene (or Stainless Steel) 0 100 200 300 400 500 Pressure (kPa) T16.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

11

<> <> d pH Holder for High Purity Water Suffix Code

Model

Option Code

5. Specifications

Model

pH holder for high purity water, wall-mount type

PH8HH Connection -JPT ports -NPT

Style A

*A

Pipe mounting bracket

Option Mounting Bracket /P

T17.EPS

3.

Interigent pH Converter

Power Supply -100 -110 voltage -200 -220 Power Supply -50 Frequency -60

Air Water 100V AC 110V AC 200V AC 220V AC 50Hz 60Hz

Style Code

Style D

*D

T20.EPS

d 4-Wire Type pH Converter Model

Description Solenoid valve

-A -W

Fluid

Always -H

Style

Suffix Code Option Code

PH8MV

Inlet; Rc1/4, Outlet; Rc1/2 Inlet; 1/4NPT, Outlet; 1/2NPT

-H

Solenoid Valve

Option Suffix Code Code

Description

PH400G

4-wire pH Converter -1 -2

Power Supply

100/110 V AC, 50/60 Hz 200/220 V AC, 50/60 Hz Japanese English

Language for -J warning, etc -E A

Always A

*B

Style B

Mounting hardware Hood

/U /PM /H3 /H4 Options /X1 Tag plate /SCT Conduit Adapter /AFTG /ANSI /SPS

Pipe, wall mounting bracket (stainless steel) Panel mounting bracket (stainless steel) Awning hood (carbon steel) Awning hood (stainless steel) Baked epoxy resin Stainless steel tag plate G1/2 1/2 NPT Teflon coated SUS steel screws T18.EPS

4.

Ultrasonic Oscillator (PUS400G) Model

Option Code

Suffix Code

-NN Application

Description Ultrasonic oscillator for pH meter

PUS400G

Always -NN General purpose

-NN -1

100V AC 50/60Hz

-2

110V AC 50/60Hz

Power Supply -3

115V AC 50/60Hz

-4

200V AC 50/60Hz

-5

220V AC 50/60Hz

-6

240V AC 50/60Hz

Language

-J

Japanese

-E

English /PS

Pipe mounting bracket (stainless steel)

/W

Wall mounting bracket (stainless steel)

/PA

Panel mounting bracket (stainless steel)

/H

Awning hood (carbon steel)

/H2

Awning hood (stainless steel)

Special Finish

/X1

Baked epoxy resin

Tag Plate

/SCT

Stainless steel tag plate

/AFTG

G1/2

/ANSI

1/2 NPT

/SPS

Teflon coated SUS steel screws

Mounting Bracket

Options

Hood

Conduit Adapter

T19.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.22-00

12

<> <>

6.

Cleaning Pump / Tank Model

Description

Suffix Code Option Code

PH8PU1

Cleaning tank / pump assembly

Power Supply -3

200V AC, 50/60 Hz

-5

100V AC, 50/60 Hz

Style code

*B

Style B

Options

/PT

With Rc 1/2 adapter

Cleaning Water Inlet/Outlet

/NP

With 1/2 NPT adapter

Anchor volts

/AN

4 L-shape M12 x 160 (SS41) bolts

Conduit adapter /APUG /ANSI

G 1/2 1/2NPT T21.EPS

7.

Terminal Box Model

Suffix Code

Option Code

PH8TBG

Specifications

Model

Terminal Box pA

Suffix Code

PH8TBS

Option Length of special cable (max. 20 m) Adapter for conduit construction

-A

/P /W

Pipe mounting hardware Wall mounting hardware

/ C uu

u; Standard 03 (3 m), 07 (7 m), 10 (10 m)

/ ATBG

Rc 1/2 screw thread

Specifications Terminal Box

Style A

Mounting hardware

Option Code

Always -A

Mounting hardware Option Length of special cable (max. 20 m) Adapter for conduit construction

/P /W

Pipe mounting hardware Wall mounting hardware

/ C uu

u; Standard 03 (3 m), 07 (7 m), 10 (10 m)

/ ATBG

Rc 1/2 screw thread T22-1.eps

T22.eps

8.

Accessories Model

Suffix Code

9. Option Code

PH8AX Calibration Reagents

Style Option (*2)

Specifications

Consumables Part Name

Accessories for pH meter (*1)

-L

Two bottles, each containing 250ml solution (pH7 and pH4) 24 bags, each bag containing powder for 500ml solution (pH7 X 12 bags and pH4 X 12 bags) and two 500ml polyethylene bottles.

-P

*A

Style A

/STD /KCLL /KCLP /TMP

Sensor stand (with mounting bracket for 50A 2-inch pipe) KCl solution (one 250ml polyethylene bottle) KCl powder (three bags, 250ml solution each) Thermometer (0 to 100 8C)

* 1: Including the following: Two 200ml polyethylene cups One cleaning bottle * 2: Either /KCLL or /KCLP is required for PH8EFP-u-u-TT2. T23.EPS

General purpose Glass electrode

High alkali

Part Number

Remarks

K9142TN One for PH8ERP, PH8EFP, PH8EHP K9319NA One for PH8ERP/PF, PH8EFP/PF K9142TU One for PH8EFP/HA K9319NC One for PH8EFP/HA, /PF

General K9142TH One for PH8ERP, PH8EFP purpose K9319QA One for PH8ERP, PH8EFP/PF High purity K9142TK One for PH8EHP water Junction K9142HW One for PH8ERP/TF, PH8EFP/TF Teflon

K9319QB One for PH8ERP/TF, /PF,PH8EFP/TF, /PF

KCl solution (3.3mol/l)

Buffer solution for calibration (pH 4) Buffer solution for calibration (pH 7) Buffer solution for calibration (pH 9) Powder for buffer solution (pH 4) Powder for buffer solution (pH 7) Powder for buffer solution (pH 9) KCl powder (for PH8EFP, PH8EHP) KCl powder (for PH8ERP)

K9084LP Six 250 ml polyethylene bottles K9084LL Six 250 ml polyethylene bottles K9084LM Six 250 ml polyethylene bottles K9084LN Six 250 ml polyethylene bottles K9020XA 12 bags,each for preparation

of 500 ml

K9020XB 12 bags, each for preparation

of 500 ml 12 bags, each for preparation

K9020XC of 500 ml

8 bags, each for preparation K9020XU of 250ml

2 bags, 1 bottle of 3.3 mol/l KCl,

K9142UT 1 syringe

(Note) The pH value of the calibrating buffer solution may vary depending on storage conditions. Prepare a new solution from powder for accurate instrument T24.EPS calibration

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

13

<> <>

WIRING DIAGRAMS (1) Wiring for pH conveter PH400G pH converter

*3

*5

+ mA + mV -

Output signal (4 to 20mA DC) Output signal (0 to 1V DC) Contact input PH8TBG *1 (cleaning start Terminal box command)

R1 R2

L1 L2

Power supply *5

S1

Contact output S1

S2

Contact output S2

High and low alarms

*5

PH8ERP PH8EFP PH8EHP pH sensor

S GE T1 T2 RE SE

*2

RE T1 T2 SE GE S

S3

Contact output S3 (cleaning or high and low alarms)

FAIL

Contact output FAIL (failure)

Note : External wiring connection terminal size is for M3 screw. *4

Grounding terminal (M4 screw) Grounding (100 or less)

*1: Terminal box is used only where pH converter is installed some distance from pH sensor (ordinary not needed). *2: This cable is specified in the option code for the PH8TBG. *3: Use only shielded cable with an outside diameter of 9 to 12 mm. *4: Be sure to ground pH converter case grounding terminal (grounding resistance; 100 or less). Only when the above grounding is impossible, ground at power cable side. But be sure to avoid two point-grounding. *5: Be sure to use cable with the outside diameter of 9 to 12 mm.

F04.EPS

Wiring to use S3 contact output for cleaning PH400G Converter 51 Power supply

S3

Solenoid valve

52

(2) Wiring for ultrasonic cleaning system pH Meter Holder Ultrasonic Transducer Cable

PUS400G Ultrasonic Oscillator U1 U2

S Ultrasonic Vibrator

L1 L2

Power Supply

Grounding (100V or less) F05.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

14

<> <>

WIRING USING CONDUIT INSTALLATION Notes for Conduit Installation pH converter PH400G Ultrasonic oscillator PUS400G

Terminal box PH8TBG (*1): Hanging (for drip)

(*2)

(*2): Bend (for water proofing) (*3): Conduit use is prohibited for sensor cable and ultrasonic (*3) cleaning cable

(*2)

(*2) (*3) (*3)

(*1)

(*1)

(*1)

(a) Recommended conduit installation (adapter not needed)

(*3): Conduit use is prohibited for sensor cable and ultrasonic cleaning cable

(*4) (*4) (*5)

(*4)

(*3)

(*6)

(*5)

(*3)

(*6)

(*5) (*6)

(*4): Conduit adapter (for installation) (*5): Flexible conduit (*6): Union

(*3) F06.EPS

(b) Conduit installation using a conduit adapter

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

15

<> <>

DIMENSIONS

Unit : mm KCl refillable pH/ sensor

KCl filling pH sensor

PH8ERP-u

PH8EFP-u

pH sensor for pure water

PH8EHP-u Medium pressure KCl reserve tank for KCl filling pH sensor (with mounting bracket)

(PH8EFP-u)-TT2

88

f27.5

;3m,5m,7m, 10m,15m,20m)

Tube length 64

JIS 2-inch stanchion

(Standard

;3m,5m,7m, 10m,15m,20m)

Cable length (Standard

263

f60.5

308

Tube length 64

10m,15m,20m)

60

Coupling with valve

115

103

(Standard;3m,5m,7m,

;3m,5m,7m, 10m,15m,20m)

Cable length (Standard 103

64

10m,15m,20m)

Cable length (Standard:3m,5m,7m,

Rc 1 / 4

f27.5

f38

f27.5

f38 f38

Guide-pipe for submersion type (with mounting hardware) PH8HG

General purpose KCl reserve tank for KCl filling pH sensor and pH sensor for high purity water (with mounting bracket)

Min :110 Max:515

(PH8EFP-u)-TT1 (PH8EHP-u)-TT1 F07.EPS

95

85

JIS 2-inch stanchion Approx.115

f48.6

600

f63

JIS 2-inch stanchion (vertical) Approx.75

Approx.45

Big volume tank KCl reserve tank for KCl filling pH sensor and pH sensor for high purity water. (with mounting bracket)

f48.6

(PH8EFP-u)-TT3 (PH8EHP-u)-TT3

54

2000

Approx.151

Sensor

JIS 2-inch Pipe (f60.5)

Hole requirements for wall mounting

25

Approx. 20

F08.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

2-f(M8 screw) holes

7060.2

60

Min :50 Max:1900

Approx.105 f105

60

GS 12B7C1-E

6th Edition 2012.11.10-00

16

<> <> Submersion holder — Polypropylene — (See separate drawing for mounting hardware)

Unit:: mm

PH8HS-PP

83 62

<With Jet Cleaner>

<With Ultrasonic Cleaner>

<Without Cleaner>

(L: Holder length; 1, 1.5, 2, 2.5, 3 m)

Cable for ultrasonic cleaning (Cable length: 3, 5, 7, 10, 15, 20 m)

u 1/2NPT 35

42 44

<With Brush Cleaner>

u 1/2NPT 35

(Code : -NP)

(Code : -NP)

64 105

u Rc1/2

105

f34

55

55

f22

u Rc1/2

105

(Code : -JP)

(Code : -JP)

55

f22

f22

L+31 L-323

L-40

L-323

L-40

L+145

f51

L-323

L-40

L+165

55

L+165

55

55

138.5

f50

Approx.155

Approx.155

Approx.155

75

75

75

F10.EPS

Submersion holder — Stainless steel — (See separate drawing for mounting hardware) PH8HS-S3 <With Jet Cleaner>

<With Ultrasonic Cleaner>

<Without Cleaner>

<With Brush Cleaner>

(L: Holder length; 1, 1.5, 2, 2.5, 3 m) Cable for ultrasonic cleaning (Cable length: 3, 5, 7, 10, 15, 20 m)

83 62

u 1/2NPT (Code : -NP)

35

42 44

64 176

u Rc1/2

176

(Code : -JP)

42

f 34

L+31

L+31

u Rc1/2

176

(Code : -JP)

42

f 22

42

f 22

L+31

L-323

f 22

L+31

L-323

L+56

u 1/2NPT (Code : -NP)

35

L-323

L+56

49

L+56

49

49

141.0

f 48.6

Approx. 200

Approx. 200

75 All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

Approx. 200

75

75 GS 12B7C1-E

F11.EPS

6th Edition 2012.11.10-00

17

<> <> Mounting hardware for submersion type holder UNIT: mm

PH8HS.../MS1 1 set PH8HS.../MS2 2 sets

68

JIS 2-inch (OD 60.5) Stanchion

64

94

90 54 Max : 515 Min : 110

Sensor Holder

600 Max : 515 Min : 110

f48.6



95

62

600 Cleaner Holder

JIS 2-inch (OD 60.5) Stanchion

90

130 54 100

F12.EPS

Stainless steel hardware for mounting submersion type holder PH8HS.../MS3

1 set

PH8HS.../MS4

2 sets

Max : 500 Min : 236

Holder with cleaner Holder without cleaner 51 44

48.5

28 55 22

44

44

44

44

44

44

44

22

65

55

600 Cleaner holder Sensor holder

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

JIS 2-inch (OD 60.5) Stanchion

GS 12B7C1-E

F13.EPS

6th Edition 2012.11.10-00

18

<> <> Mounting hardware for submersion type holder

Mounting hardware for submersion type holder

— Flange mounting hardware for holder without cleaner

— Flange mounting hardware for holder with cleaner

— Material: Polypropylene or Stainless steel

— Material: Polypropylene or Stainless steel

f210

Unit: mm

f330

JIS 10K 100 FF flange (Note)

JIS 10K 200 FF flange (Note) 12-f23 holes

8-f19 holes

L: Holder length; 1, 1.5, 2, 2.5, 3 m

74

f290

18 f175

78

L+31 22

Min : 150 Max : L-115

Min : 405 Max : L-130

Note: Only mating dimensions are according to flange standard.

Approx. 155

F14.EPS

Flow-through type holder with mounting hardware PH8HF fa

(pH Sensor)

4-fD holes

b c

e Outlet

O-ring

fB

fd

fA

165 h

11061 E Inlet

C

g f

15061 O-ring Model and Code PH8HF-PP-J10-T-NN-NN PH8HF-PP-A15-T-NN-NN PH8HF-S3-J10-T-NN-NN PH8HF-S3-A15-T-NN-NN

A

B

C

D

E

125 90 14 19 108 79.2 14.2 15.7 125 90 14 19 108 79.2 12 15.7 2

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

a

b

80 9 80 9 70 10 70 10

c

d

e

f

g

h (Approx.)

11 11 6 6

60 60 60.5 60.5

25 25 26 26

30 30 26 26

70 70 69 69

250 250 245 245

GS 12B7C1-E

F15.EPS

6th Edition 2012.11.10-00

19

<> <>

Unit : mm

Holder for high purity water PH8HH

Hole dimensions for Holder mounting

60

Sensor holding screw

180

171

20

1. Pipe mounting (2-inch pipe)

64

70

102

SAMPLE IN

70

14

Ground connection terminal M5

SAMPLE OUT

2. Wall mounting

6- f10 holes

27

Sample inlet 1/4NPT adapter

56 70 100 120

73 100 148

15

Rc 1/4

50

35 50 70

110

28Max

Rc 1/2 Sample outlet 1/2 NPT adapter

100

F16.EPS

Interigent pH Converter PH400G Hood (optional) Option code : /Hh

Four M6 screws, 8 deep 220

184

80

72

20

144

80

23

144

112

Cable inlet port (13.5 dia. holes) For JIS A8 cable gland

36

Adaptor for conduit work (option code : /AFTG, /ANSI)

A

B 36 C

D

36

36

E

A: B: C: D: E:

For output signal / contact input For sensor cable For power supply For contact output (S1 and S2) For contact output (S3 and FAIL)

Adaptor

49

38

Ground terminal (M4 screw)

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

G 1/2 female ( / AFTG) 1/2 NPT female ( / ANSI)

GS 12B7C1-E

Approx. 55

Cable inlet port (21 dia. holes) equivalent to JIS A15 cable gland

F17.eps

6th Edition 2012.11.10-00

20

<> <> Pipe Mounting Bracket / Wall Mounting Bracket (option code : /U) d

Example of bracket used for pipe mounting 188

Unit : mm

M6, 4 screws

174 200

50

Nominal 50A (O.D 60.5mm) mounting pipe d

100

Example of bracket used for wall mounting 135

13

M6, 4 screws

224

200

35 15 70 10 mm dia., 3 holes

100

F18.eps

Panel Mounting Bracket (option code : / PM) 23 12 max.(panel thickness)

Unit: mm

M6, 4 screws

137 +20

100

137 +20 178

Panel cutout dimensions F19.eps

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

21

<> <> Ultrasonic oscillator (PUS400G) Pipe mounting

Panel/, Wall mounting Unit : mm

151

30

30 288

Panel mounting bracket

Hood (optional)

38

68

96

8

Ultrasonic transducer cable inlet Cable gland DIN PG16 equivalent 100 2-f5.5 holes +1.1 (for hood mounting) 7060.2 156 0 5660.2 121

121 JIS 2-inch pipe stanchion (horizontal)

13

27

2-f6.6 holes (for oscillator mounting)

15

173 0 138

JIS 2-inch (OD f60.5mm) stanchion (vertical)

203

206

Power supply cable inlet Cable gland DIN PG16 equivalent

42

+1.1

2-f6.6 holes (for hood mounting)

2-f6.6 holes (for hood mounting)

185

27

100 7060.2 5660.2

3.9

185

42

Ultrasonic transducer cable inlet Cable gland DIN PG16 or equivalent

27

8

Ground terminal (M4 screw) Power supply cable inlet Cable gland DIN PG16 equivalent

80

138

96

68

Pipe mounting bracket (optional)

Ground terminal (M4 screw)

60

110

130 180

1 to 10

162

48

50

155 115

180

202 162

7060.2

3-f10 holes <Panel cutout>

<Wall mounting bracket>

48 F20.EPS



46

14.5

102

General Purpose Solenoid Valve PH8MV*D 103 72

71

2-Rc1/2 G1/2

Cautions on Installation of Solenoid Valve for Jet / Brush Cleaning 1. Do not allow a sample solution to flow backward into the solenoid valve or to be replaced with the driving fluid. For this take relevant measures; e.g. install a check valve to prevent inverse pressure between the inlet and outlet of the solenoid valve, or install the solenoid valve higher than the holder, especially when using the air jet/brush cleaning system. 2. Make sure to avoid the risk of corrosion of the solenoid body (bronze) and seal (nitrilel rubber) by vapor or gaseous components generated from a sample solution, especially when using the air jet/brush cleaning system.

F21.EPS

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

22

<> <> Terminal Box PH8TBG 65

126

Unit: mm

48

30

178 75 60.2

254

Waterproof gland

Pipe mounting hardware (example)

4-M5 screws, 9mm deep

7060.2

For conduit construction

(for mounting)

Adapter (option)

<Wall mounting hardware>

100 7060.2

4 - [ 5.5 hole

Nominal 50A pipe mounting

4 - [ 5.5 hole

100 7060.2

8 8 7560.2

75 60.2

13

185

203

27 15 7060.2

48

3 - [ 10 hole F22.eps

Sensor stand (PH8AX-u)/STD Sensor (Example)

115

(64)

JIS 50A 2-inch stanchion

108 F24.EPS

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

23

<> <> Cleaning Pump/Tank Assembly PH8PU1

Unit: mm (450)

(408)

(442)

4-f15 holes

T : Tank 50 V1 : Float valve 1/2 B V2 : Check valve 1/2 B V3 : Ball valve G : Pressure gauge P : Pump Cleaning water inlet

Cleaning water outlet G 1/2B female(std.)

G 1/2B male(std.)

1/2 PVC Pipe Teflon braided horse

Flow diagram

Motor

(800)

Pump

Cable connection

WATER IN

OVER FLOW

Overflow outlet (16 ATS type)

(804)

(290) Height to inlet

Tank

(504) Height to outlet

Cleaning water outlet Cleaning water inlet

Electromagnetic actuator

(138) F25e.eps

Notes on installation of cleaning pump / tank assembly (1) The cleaning pump / tank assemble may be installed outdoors. However, the place must be sheltered from rain, wind, and direct sunlight. (2) Since the noise level will increase to a maximum of 100 dB (A), utmost attention should be paid to the installation environment. (3) Connect a wired soft vinyl-chloride tube to the cleaning water outlet. Do not connect a rigid PVC pipe directly.

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

24

<> <> SELECTION CRITERIA FOR pH SENSORS AND HOLDERS

(1) When any of the two conditions listed below are applicable, select a KCl filling type pH sensor and either the submersion or flow-through type holder. d The solution is out of the range 2 < pH < 12. d The solution contains organic or oil in the order of a few percent. (2) When any of the two conditions listed below are applicable, consult our salesperson. d Strong oxidizing solutions such as aqua regia, chromic acid , hypochloric acid, perchloric acid. d The solution contains corrosive gases (ammonia, chlorine, hydrogen sulfide).

s: Can be used, n:Shortens useful life, x:Cannot be used

Phosphoric acid Boric acid Carbonic acid Chromic acid Sulfurous acid

0.5 0.05 0.4 0.04 0.6 0.06 1.0 0.6 0.6 1.2 0.8

1.0 2.0 1.0 2.0 1.0 2.0 1.5 5.0 3.6 0.8 1.4

Organic acid

Acetic acid Formic acid Oxalic acid Lactic acid Phenol acid Monochloroacetic acid

0.6 0.5 0.9 0.9 0.9 0.9

2.8 2.3 1.0 2.4 5.4 1.8

Alkali

Calcium hydroxide Potassium hydroxide Sodium hydroxide

0.2 0.5 0.4

12.4 12.7 12.9

Acid salts

Ammonium chloride Aluminous water Zinc chloride Ferric chloride Ferric nitrate

5 5 5 5 5

1.3

Basic salts

Sodium sulfite Sodium carbonate Sodium phosphate

5 5 5

Neutral salts

pH * (25 8C)

Potassium chloride Sodium sulfate Calcium chlorine Sodium nitrate Aluminum chloride

5 5 5 5 5

Oxdizing agents

Concentration W/V (%)

Hydrogen peroxide Sodium hypochlorite solution Chlorinated lime Potassium bichromate

1 1 1 5

Organic solvents

Chemical

Alcohol Organic solvent or oil (excluding alcohol)

10

Inorganic acid

Sulfuric acid Hydrochloric acid Nitric acid

Holder Flow-through, Submersion

Guide-pipe

11.8

8.2

12.5 4.5

Note: pH values in table are calculated from dissociation constant (including measured value). T25.EPS

CAUTION Select the material of wetted parts with careful consideration of process characteristics. Inappropriate selection may cause leakage of process fluids, which greatly affects facilities. Considerable care must be taken particularly in the case of strongly corrosive process fluid such as hydrochloric acid, sulfuric acid, hydrogen sulfide, and sodium hypochlorite. If you have any questions about the wetted part construction of the product, be sure to contact Yokogawa.

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

25

<> <>

Table of Corrosion-Resistant Materials (The data should be used for reference only) Note: This table shows corrosion resistance for each single substance alone. If a sample contains two or more substances, then the corrosion resistance may differ from that given in this table. : Excellent : Good : Not so good X : Unusable

Ultrasonic transducer material Sensor solution ground tip

Holder material Polypropylene Sulfurous acid

Inorganic acid

Hydrochloric acid

Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge

100 20 90 20 5 80 5

6

30

5

30

20 20 10 Hypochlorous Acid 10

20 40 20 40

Sulfuric acid Phosphoric acid Ammonia water

10 10 57 57 3 3 30 30 15 15

20 80 20 70 20 100 60 100 80 100

Alkali

Sodium hydroxide

20 20

Sodium hydroxide, Sodium hydroxide9 to 11% +Sodium chloride 15%

80 100 100

Potassium carbonate Sodium carbonate sat. 100 Zinc chloride

Chlorides

Aluminum chloride Ammonium chloride 35 Potassium chloride sat. sat. Calcium chloride sat. 20 Ferric chloride 20 Sodium chloride, 20% + Saturated Cl2 (Electrolysis solution)

14

30

6

30

5

30

5 5

30 b

40 60 80 100 40 60 100

X

10 10 10 10

20 90 20 90 90

5 35 25

b b 90

25 20 25

90 90 90

20

60

30 100

57

30

10

10

30

57

25

6 5 15 5 10 28 10 25 20 20

30 100 30 b b 65 b b 30 b

5 35 25 20 25 25 25 sat. 25

b b b b 25 25 b 60 b

30

b

X

90

X

X

X

5 5 5 5 10 20 10 25 20 20

30 70 30 b b 65 b b 30 b

5 5 5 5 10 20 10 25 20 20

30 100 30 60 b 65 b b 30 b 93

5 35 25 20

b b b b

5 35 25 20 10 25 25 sat. 25

b b b b b b b 60 b

30

b

X X 25 b sat. 60 25 b 30

b

X

90

X

X

Strong alkaliX Weak alkali

X

X

40 20 sat. 10 20 20

24 b b 30 b b b

10

24 80 90

10 10 10

90 90 90

50 20 10

b b 30

10

30

5

90

10

90

10

b

80

10 10 10

X

30

b b b 30

Sea water + Saturated Cl2

95

X

Bromine gas Hydrogen sulfide

95 30

20

b b 30 b b b

X

20

b b b 30

42 20 sat. 10 20 20

Strong acid Weak acid

90

10 10 10

Potassium bichromate

Sulfurous acid gas

30

30

40 10

30

20 40 60

15

5

15

Wet chlorine gas

20 90 90

40

60-90

60

90 30 85

20

2

Sodium sulfide Sodium bisulfate

X

43

60-90 X

Corrsion resistance is good for usual salts. 30 20

20 60

15

2

24 sat. 80 60 5

Hydrogen peroxide

Sodium sulfide

5 10

b

90 80

Sodium nitrate Sodium sulfite

X

10

b b 30

Ammonium nitrate

20 40

30

50 20 10

Potassium sulfate Sodium sulfate

30 60 90 20

20

24 b b 30 b b b

Ammonium sulfate

Remarks

Concent’n Temp Judge

5 37 37 20

42 20 sat. 10 20 20

Seawater, Magnesium chloride

Nitrates Sulfates

b

30

Viton

X

Potassium hydroxide

Others

10

X

6

Hydrobromic acid

Hydroiodic acid

Titanium

Concent’n Temp Judge

Chromic acid

Nitric acid

Gases

Hastelloy C

SUS 316

Sensor body material Ryton

Seal O-ring material

Polypropylene may sometimes be eroded by ammonium sulfate crystals

20

X

30

X

X

80 100

95 30 20 30-90

80 T26.EPS

Note: "b" refers to the boiling point.

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation

GS 12B7C1-E

6th Edition 2012.11.10-00

26

<> <>

Ultrasonic transducer material Sensor solution ground tip

Holder material Polypropylene Acetaldehyde Acetone Aniline Ether Ethylene glycol Ethyl alcohol Methyl chloride Glacial acetic acid Glycerin Chlorophenol Xylene Chlorobezene Chloroform

Organic substances

Dioxane Dichloroethare Ethyl nitrate Carbon tetrachloride

Trichloroethylene Toluene Benzophenone Benzaldehyde

SUS 316

Concent’n Temp Judge

Concent’n Temp Judge

20 100 20

100 30 50 25 100 110

100 100 100 100 100 100 96

20 70 100 20 70 100 70

100 100 100 100 100 100 100 100 100 100 100 100 100 100 100

20 70 100 70 100 20 70 100 20 20 100 20 20 70 100

X

100 100 100 100 100 100

70 20

X

20 20 20

X X X

100 20 100 70 100 100

Hastelloy C

Seal O-ring material

Titanium

Viton

Concent’n Temp Judge Concent’n Temp Judge Concent’n Temp Judge

Sensor body material Ryton 100

20

100

b

100 25

100

90

100 25 100 25

100

20

100

100

90

100

20

100

20

100

20

100

90

100

90

100

90

100 100

90 90

100 100

20 90

100

90

100

90

100 100

25 90

100

90

100

20

100 100

20 20

100 25

X

b

Remarks

Concent’n Temp Judge

100 25 100 24

X

100 25

X X X X X

100

b

100

b

100

b

X

100

105

90 100

b b

100

b

90 100

b b 145

100

24

X

Benzyl alcohol benzene

100 20

100

30

Fomaldehyde

10 10

37

b

70 100

37

b

100

30

37

b

100

25

X X

X

Methylnaphthelen Methyl ethyl ketone Methyl alcohol Nitrobenzene

Acetic acid

100 20 70 100 20 100 20 100 70 100 100 100 20 100 70 100 100 100 20

100 25

X

X 10

b

95

30

10

b

95

30

X 95

30

Phenol Benzonic acid Motor oil Petroleum ether Kerosene Tartaric acid Oil and fats Carbon sulfide

100 100 100 100 100 100 100 100 10 10 10 100

100 20 70 100 20 20 70 40 60 80 70

100 20

101 X 50

100

100 25 X

50

100

100 180

100 25

50

100

100 180 100 25

Note: "b" refers to the boiling point.

All Rights Reserved. Copyright © 1991, Yokogawa Electric Corporation

T27.EPS

GS 12B7C1-E

6th Edition 2012.11.10-00

27

<> <>

Enquiry Specifications Sheet for PH400G pH Converter System For enquires on the Yokogawa pH converter system, please tick (v) the appropriate box u and write down the relevant information in the blanks.

1 General Information Company name; Department; Contact Person; Plant name; Measurement location; Purpose of use; u Indication, u Recording, u Alarm, u Control V AC Hz Power supply;

2 Measurement Conditions (1) Process temperature; to Normally [8C] to Normally [kPa] (2) Process pressure; to Normally [l/min] (3) Flow rate; to Normally [m/s] (4) Flow speed; (5) Slurry or contaminants; u No, u Yes (6) Name of process fluid; (7) Components of process fluid; (8) Others;

3 Installation Site (1) Ambient temperature; (2) Location; u Outdoors, u I ndoors (3) Others;

4 Requirements (1) Measuring range; u pH 0 to 14 u (2) Transmission output; u 4 to 20 mA DC u (3) System configuration selection; u Sensor, u Holder, u pH Converter, u Cleaning system, u Terminal box, u Accessories m (4) Sensor cable length; u 3m, u 5m, u 7m, u 10m, u 15m, u 20m, u (5) Sensor operating pressure; u10 kPa or less, u Greater than 10kPa (6) Type of holder; u Guide pipe, u Submersion, u Flow-through, u Suspension, u Angled floating ball, u Vertical floating ball (7) Cleaning method; u No cleaning, u Ultrasonic cleaning, u Jet cleaning, u Brush cleaning (8) Sample temperature; u -5 to 1058C, u -5 to 1008C, u -5 to 808C (9) Others;

All Rights Reserved. Copyright ©1991, Yokogawa Electric Corporation Subject to change without notice