UREA/BUN

Enzymatic-spectrophotometric ULTRAVIOLET

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the instrument. It is advisable to validate the use when there is any change in software or reagent versions.

Instrument:

AU 400 / 640

Reagent preparation

Working Reagent: Transfer the contents of one Reagent B vial into a Reagent A bottle. Stable for 2 months at 2-8 ºC.

Instrument settings Specific Test Parameters Sample: Volume 2 µL Reagents: R1 Volume 300 µL Wavelength: Pri. 340 Method: FIX Reaction slope: Measuring Point 1: First 2 Measuring Point 2: First 0 Linearity: % No Lag Time:

Min OD

First L -0.1 First H 0.5 Last L -0.1 Last H 0.5

Dynamic Range:

Last 5 Last 0

Correlation Factor:

L

0 H

300

A

1 B

0

On-board stability period:

7. None selected 8. Out of range L

Panic Value

Max OD H 2.5

L -0.1 Reagent OD limit:

Dilution 0 µL Dilution 0 µL Sec. 380

H

L

H

10

50

Unit: mg/dL

Decimal places: 0

Counts:

Process: CONC.

Calibration Specific Calibration Type: Point 1:

AB

Formula: Cal. Nº #

Y=AX+B OD

CONC *

2 Factor/OD-L -9999

Factor/OD-H 9999

* Enter the assigned standard value # User defined Vers. 0206

Related Documents

Urea Formaldehid
June 2020 34
Urea Modeling
November 2019 20
Pad Urea
November 2019 24
Tp-urea
April 2020 36
Urea Uv.pdf
April 2020 19
Urea Purification
May 2020 23

More Documents from "AFLAC ............"

Urea Uv.pdf
April 2020 19
June 2020 23
October 2019 25
October 2019 23
Excersice 2
December 2019 15
October 2019 28