Recomendações Técnicas Câmara de Estabilidade NBR 5390 - Generalidades sobre os ensaios climáticos e mecânicos 1.DEFINIÇÃO Câmara ou espaço fechado no qual a temperatura e a umidade interna podem ser controladas dentro de limites específicos 2.N° DE SENSORES PARA QUALIFICAÇÃO1,2, 3 a)Para volume útil menor que 2000l, utilizar ao menos 9 sensores b)Dispor sensores em distâncias iguais a 1/10 de cada uma das dimensões do volume interno (LxAxP) c)Quando utilizar sensores higrômetros que medem umidade relativa, posicioná-los juntos aos sensores de temperatura. d)Quando utilizar um único sensor de temperatura de ponto de orvalho, localizá-lo no centro e calcular a distribuição espacial através da distribuição da temperatura do ar, considerando a umidade absoluta constante.

3.CIRCULAÇÃO DO AR a)Volume de ar deve circular completamente em 30s 4.SEGURANÇA a)Dotada de sistemas seguros quando utilizadas amostras que poder gerar vapores inflamáveis/combustíveis b)Sistema de alarme contra sobre-aquecimento com desligamento automático para evitar dano ao equipamento/produto/operador 5.ENSAIOS4,5 a)Determinação de correção da indicação ou diferença entre o local de medição de referencia e a indicação do Controlador Indicador de temperatura b)Determinação de não homogeneidade espacial no volume útil c)Determinação da instabilidade temporal no volume útil vazio d)Determinação do efeito de radiação e)Determinação do efeito da carga (comparação da medição no local com e sem carga).

f)Determinação da temperatura transitória excessiva g)Determinação da umidade transitória excessiva h)Determinação da taxa de mudança da temperatura i)Nos testes com carga devem ser ocupado ao menos 40% do volume útil j)Determinação do tempo de recuperação da temperatura k)Determinação do tempo de recuperação da unidade relativa 6.FATORES DE INFLUÊNCIA QUE DEVEM CUJOS LIMITES DEVEM SER DETERMINADO PELO FABRICANTE PARA GARANTIR O DESEMPENHO6 a)Poluição b)Luiz solar c)Vibrações acústicas d)Variação na tensão elétrica e)Variação na freqüência elétrica f)Qualidade, vazão e temperatura da água g)Ocorrência de condensação no interior 7.REGISTRO DAS MEDIÇÕES7 a)No máximo a cada 01 minuto. b)Após estabilização, no mínimo 30 medições devem ser tomadas por no mínimo 30 minutos. 8.MEDIÇÃO DA VELOCIDADE DO AR a)Segundo as normas, é meramente indicativo. 9.TIPOS DE SENSORES 9.1.Para a norma francesa NF X15-140 (AFNOR, 2002) a)Temperatura: Pt100 (platina 100 ohm a 0°C), termorpares tipo T, J ou K b)Umidade: Higrômetro à condensação (higrômetro ótico), psicrômetro e higrômetros à variação de impedância (capacitivo ou resistivo) 9.2.Para a norma francesa NF X15-140 (AFNOR, 2002) a)Não especificam os tipos. 10.CÁLCULOS a)Gradiente: diferença máxima, em um ambiente estável, entre os valores médios acrescidos da incerteza expandida. b)Gradiente máximo: diferença máxima, em um ambiente estável, entre os valores máximos e mínimos acrescidos da incerteza expandida.

c)Não homogeneidade espacial: maior diferença de umidade relativa e/ou temperatura entre um local de medição e o local de referência

11.LEMBRETES PARA A INSTALAÇÃO8 a)Move-in & Installation Customer Checklist b)All necessary permits and compliance to local and on site codes. c)Provide and coordinate rigging of equipment, uncrating and pallet removal, moving the chamber in the building and placing equipment at permanent location. d)Be sure to request a truck with a lift gate if there is not dock available. e)All chambers are shipped on pallets and a forklift will be needed to remove it from the pallet to move into your building. It is critical to know the exterior dimension of your chamber in order to move it into your building clearing doorways, etc. f)Installation of equipment, if not purchased from Cincinnati Sub-Zero. g)Proper power and voltage must be supplied for the chamber. The data plate label located in the back of the chamber describes the proper power information needed to operate the chamber. h)If the ambient environment of the area where the chamber will be installed exceeds 30°C (85°F), CSZ recommends a chamber with a water cooled condenser. Air-cooled chambers will lose capacity and damage can occur to the unit if the ambient room temperature is too high. i)A minimum of three feet clearance around the chamber is recommended for service and proper ventilation. j)If chamber is not on ground floor, ensure floor can handle the weight of the chamber and its components. k)Provide electrical disconnect, if needed, unless specified and included in quotation. l)Provide compressed air 80-110 psig, if needed, free of oil and entrained water. m)Provide water regulator (Humidity Chambers Only) and tap or de-ionized water. n)Connection of hose from chamber drain(s) to either floor drain or customer supplied condensate pump. o)A qualified professional must make final connections, in accordance with local and on site codes. p)Anchoring of any component to building floor where bolt down holes are present. q)If the chamber is and integral air-cooled unit, make sure the air-conditioning system in the area has enough capacity to remove that heat. The heat rejection information for the chamber will be listed in the chamber quotation and drawings located in the manual. r)For roof-mounted components, ensure that the building roof will support the added load of the components. Roof mounted components must meet local and on site codes. s)For roof-mounted components, provide all roof curbs and/or structure to support roofmounted components. t)Provide all penetrations and sealing of penetrations through the roof for refrigeration and electrical runs. u)Location of air-cooled condenser must allow adequate space for free air circulation and maintenance. v)Ground level components must be mounted on customer supplied concrete pad or equivalent.

1

Segundo IEC 60068-3-5 Environmental testing Part 3-5 International Electromechanical Commission NF X15-140 Measurement of air moisture (Association Françoise of Normalization - AFNOR, 2002) 3 Segundo Guia alemão DKD-R 5-7 (Deutscher Kalibrierdienst DKD ,2004) 4 NF X15-140 Measurement of air moisture (Association Françoise of Normalization - AFNOR, 2002) 5 Segundo Guia alemão DKD-R 5-7 (DKD,2004) 6 NF X15-140 (AFNOR, 2002) 7 NF X15-140 (AFNOR, 2002) 8 Points to Consider when Purchasing & Installing Environmental Chambers – Cincinnati Sub-zero company, http://www.cszindustrial.com/CHAMBER%20FAQ%20BOOK.pdf 2