Rosa-sl6 Folleto Rosa Sl-6

  • October 2019
  • PDF

This document was uploaded by user and they confirmed that they have the permission to share it. If you are author or own the copyright of this book, please report to us by using this DMCA report form. Report DMCA


Overview

Download & View Rosa-sl6 Folleto Rosa Sl-6 as PDF for free.

More details

  • Words: 516
  • Pages: 2
ROSA : SL-6 Beta-lactam Test for Raw Commingled Milk SL-6 AT A GLANCE H

Detects all six US required beta-lactam drugs in a single test (Table 1)

H

Closest test to Safe Level/Tolerance

H

Rapid One Step Assay (part of ROSA family of tests)

H

Approved by the NCIMS/FDA/AOAC-RI

H

Results in 8 minutes!

H

Save unnecessary rejection and disposal of 'safe' non-violative milk

H

First on farm test for Cloxacillin detection at Safe Level/Tolerance

Charm SL-6 Beta Lactam Kit

Table 1: SENSITIVITY LEVELS (ppb)*

ACCURATE AND RELIABLE

CHARM SL-6 7.1

SAFE LEVELS 10

Charm SL-6 is a simple and accurate

Ampicillin

9.6

10

lactam drugs within 30% of their safe

Cephapirin

18.7

20

level/tolerance in raw commingled milk.

Ceftiofur

37.5

50

This precision sets a new standard in

Cloxacillin

8.3

10

Penicillin G

4.2

5

BETA-LACTAMS Amoxicillin

* Calculated 90% positive at 95% confidence limit

TM

test that detects all six required US beta-

meeting regulatory levels of concern. Because SL-6 can be used by the farmer, hauler, fieldman, and processor, the dairy industry is assured a safer and more wholesome product, from farm to table.

OTHER AVAILABLE ROSA TESTS FOR: H

Aflatoxin M1

RAPID AND EASY

H

Enrofloxacin

SL-6 uses ROSA Technology based on

H

Beta-lactams (5 out of 6)

lateral flow diffusion where all test

H

Beta-lactams (MRL kit for E.U.)

ingredients are embedded in a single test

H

Tetracyclines

strip. SL-6 requires only one stepnjust

H

Sulfadimethoxine/Sulfamethazine

add milk to the strip, allowing for hands

H

Sulfamethazine

free, walk away convenience.

ROSA : SL-6 Beta-lactam

PROCEDURE ROSA READER PRINTOUT SAMPLE Unit Serial Number

Add milk sample to SL-6 strip. Close incubator lid

Date:

17 AUG 2003

Time:

10:17:51

Operator:

1

Channel:

SLBL6

Lot: Sample#:

After 8 minutes, incubator beeps and 'test complete' light will flash. Remove SL-6 Strip.

012 123-456

Limit:

< 0

Reading:

-1563 Lot Number

Result:

to start timer.

490

NEGATIVE

For unofficial testing: Visually compare Test line (T) Control line

C T

Test line

and Cloxacillin line (X) to the Control line (C). For official testing: Insert valid test in ROSA Reader.

X

Cloxacillin line

EQUIPMENT ROSA Reader

ROSA Reader

H

Records sample ID, test type,

H

Results in just 5 seconds

operator ID, date, time, lot

H

Download to printer or computer

number, reading and result

H

Stores 4500 data points

interpretation. It even sorts data by performance controls

ROSA Blue Incubator

ROSA Blue Incubator

H

Count down timer display

H

Quad and dual incubators

(on quad only)

H

Self timing with indicator lights/

H

Auto Start when lid closed

beeper

H

Auto Reset when lid opened

SL-6 KIT H

Available in 20, 100 or 500 test kits

H

Kit Codes (lab): LF-SL6-20 ; LF-SL6-100 ; LF-SL6-500

H

Kit Codes (farm): LF-SL6-20SUP ; LF-SL6-100SUP

H

One year refrigerated shelf life

H

5 ppb Penicillin G/10 ppb Cloxacillin positive control included with every lab kit

Charm Sciences, Inc. 659 Andover Street, Lawrence, Massachusetts 01843-1032, USA Copyright

Tel: 1.978.687.9200 Fax: 1.978.687.9216 Email: [email protected] www.charm.com

82003 Charm Sciences, Inc. All rights reserved. All trademarks are property of their respective owners. 200-2010-107-300-01

Related Documents

Rosa-sl6 Folleto Rosa Sl-6
October 2019 13
Rosa-slbl Folleto Rosa Slbl
October 2019 27
Folleto Rosa Mrlbl Europa
October 2019 12
Rosa
June 2020 19
Rosa
October 2019 37
Folleto
May 2020 28