SIREGRAPH CF Service
Revision
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 6 TD SD 21
Siemens AG Medical Engineering
Contents
0-3 Page
1 _______General Information ____________________________________________ 1 - 1 Documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Measuring Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . Details Regarding Tolerances . . . . . . . . . . . . . . . . . . . . . . . . Safety Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Remarks Regarding Adjustment . . . . . . . . . . . . . . . . . . . . . . Potentiometers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Potentiometers, Remarks Regarding Replacement and Adjustment . . . . Allarmax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Lubricating Points / Bearings / Screws / Electrostatically-sensitive Devices Installation of Ball Bearings . . . . . . . . . . . . . . . . . . . . . . . . . Securing of Screws . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Protection of Electrostatically-sensitive Devices . . . . . . . . . . . . . . Manufacturer’s Note . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Overview I/O PC Board 24S2 . . . . . . . . . . . . . . . . . . . . . . . . PC Mode (Version Windows 3.11). . . . . . . . . . . . . . . . . . . . . . Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . PC Mode (Version Windows 95) . . . . . . . . . . . . . . . . . . . . . . Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . English Version of the Hardware Test and Settings . . . . . . . . . . . . . Error List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Replacing the 24S1 CPU Board and the 24S2 I/O Board . . . . . . . . . . 24S1 CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24S2 I/O Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Changes to the DW1 DIP Switches on the 24S2 Board. . . . . . . . . . . Table of DW1 Switch Positions (up to VC00A) . . . . . . . . . . . . . Table of DW1 Switch Positions (as from VC01A) . . . . . . . . . . . . Listing of all Programming Codes . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.1- 1 .1- 1 .1- 1 .1-2 .1-2 .1 - 2 .1-3 .1- 3 .1- 4 .1- 4 .1- 4 .1- 4 .1- 4 .1- 5 1 - 12 1 - 12 1 - 12 1 - 13 1 - 15 1 - 15 1 - 15 1 - 16 1 - 18 1 - 20 1 - 24 1 - 24 1 - 24 1 - 24 1 - 25 1 - 26 1 - 27
2 _______Adjustment Preparations ________________________________________ 2 - 1 Adjustment mode . . . . . . . . . . . Start adjustment mode. . . . . . . Switching off the adjustment mode Checking the voltages . . . . . . . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
.2-1 .2-1 .2- 2 .2- 3
3 _______Tilt Drive ______________________________________________________ 3 - 1 Control overview. . . . . . . . . . . Tilt drive overview . . . . . . . . . . Belt Tension, Adjustment Procedure Brake assembly . . . . . . . . . . .
Siemens AG Medical Engineering
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
RXD3-209.071.01 Rev. 05 09.00
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
Page 3 of 6 TD SD 21
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
.3 - 1 .3-2 .3- 3 .3- 4
SIREGRAPH CF Service
0-4
Contents Page Adjustment or replacement of potentiometer 3PT4. . . Set end positions . . . . . . . . . . . . . . . . . . Safety limit switch 3FC11 . . . . . . . . . . . . . . . . Movement times . . . . . . . . . . . . . . . . . . . . Safety limit switch 3FC16. . . . . . . . . . . . . . . . Checking the floor spacing with 0 degree table position
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. 3-5 . 3-6 . 3-8 . 3-8 . 3-9 .3 - 10
4 ______ Spotfilm Device Longitudinal Drive ________________________________4 - 1 Control overview . . . . . . . . . . . . . . . . . . Spotfilm device longitudinal drive overview . . . . Console . . . . . . . . . . . . . . . . . . . . . . . Adjustment or replacement of potentiometer 3PT2. Set end positions . . . . . . . . . . . . . . . . Safety limit switch 3FC14. . . . . . . . . . . . . . Checking the PC board SP 22Az4 . . . . . . . . . Spotfilm device drive, adjusting belt tension . . . . Adjustment procedure . . . . . . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
4-1 4-2 4-3 4-4 4-4 4-5 4-5 4-6 4-7
5 ______ Column Drive __________________________________________________5 - 1 Control overview . . . . . . . . . . . . . . . . . . . Column drive overview . . . . . . . . . . . . . . . . Adjustment and replacement of potentiometer 3PT1. Safety limit switch 3FC12. . . . . . . . . . . . . . . Blocking limit switch 3FC5 . . . . . . . . . . . . . . Safety limit switch 3FC15. . . . . . . . . . . . . . . Checking the PC board SP 22Az3 . . . . . . . . . . Room height switch 3FC13 . . . . . . . . . . . . . . Column drive, adjusting the belt tension . . . . . Adjustment procedure . . . . . . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
5-1 5-2 5-3 5-4 5-4 5-5 5-6 5-6 5-7 5-8
6 ______ Compression Drive _____________________________________________6 - 1 Control overview . . . . . . . Compression drive overview . Chain tension . . . . . . . . . Maximum compression force . Adjustment of switch 3FC4 . . Movement times . . . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
6-1 6-2 6-3 6-4 6-7 6-7
7 ______ SID Drive______________________________________________________7 - 1 Control overview . . . . . . . . . . . . . . . . . . SID drive overview . . . . . . . . . . . . . . . . . Adjustment or replacement of potentiometer 3PT7. Set end positions . . . . . . . . . . . . . . . .
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
Page 4 of 6 TD SD 21
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
7-1 7-2 7-3 7-4
Siemens AG Medical Engineering
Contents
0-5 Page
Adjustment of SID Limiting at 90° . Safety limit switch 3FC7 . . . . . . Adjustment of PC board 23Az7 . Movement times . . . . . . . . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
. . . .
.7- 4 .7- 5 .7- 6 .7- 6
8 _______Table Transverse Drive __________________________________________ 8 - 1 Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1 Table transverse drive overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2 Electrical Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 3
9 _______Cassette Carriage Drive _________________________________________ 9 - 1 Control overview. . . . . . . . . . . . . . . . . . Cassette carriage drive overview . . . . . . . . . Mechanical check . . . . . . . . . . . . . . . . . Adjust tension of the drive chain . . . . . . . . . Adjust belt tension of the cassette drive. . . . . . Adjusting the slip clutch . . . . . . . . . . . . . . Adjustment or replacement of potentiometer 3PT3 Set end positions . . . . . . . . . . . . . . . Movement times . . . . . . . . . . . . . . . . . . Maximum opening of the cassette jaws . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
. . . . . . . . . .
.9 - 1 .9- 2 .9-3 .9- 4 .9- 4 .9- 5 .9- 7 .9- 8 .9- 9 .9-9
10 ______Format Sensing _______________________________________________ 10 - 1 Control overview. . . . . . . . . . . . . . . . . . . . . . . . Overview of the format sensing drive . . . . . . . . . . . . . Adjustment or replacement of potentiometers 3PT10, 3PT11. Spacer Gauge . . . . . . . . . . . . . . . . . . . . . . . Adjusting the end positions and the format sensing . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
10 - 1 10 - 2 10 - 3 10 - 4 10 - 5
11 ______Diaphragm Plate Drive _________________________________________ 11 - 1 Control overview. . . . . . . . . . . . . . . . . . . Diaphragm plate drive overview . . . . . . . . . . . Mechanical check . . . . . . . . . . . . . . . . . . Symmetry and centering . . . . . . . . . . . . . . Chain tension . . . . . . . . . . . . . . . . . . . . Adjust the slip clutch . . . . . . . . . . . . . . Adjustment and replacement of potentiometer 3PT5 Adjustment of gap width, general remarks . . . . . Adjustment of gap width (subdivision mode) . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
. . . . . . . . .
11 - 1 11 - 2 11 - 3 11 - 3 11 - 4 11 - 4 11 - 6 11 - 7 11 - 8
12 ______Collimator____________________________________________________ 12 - 1 Control overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 1 Drive overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 - 2
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 6 TD SD 21
SIREGRAPH CF Service
0-6
Contents Page Adjustment after replacement of the potentiometers . . . . . . . . . . . . . . . . . .12 - 3 Adjusting the end positions . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 - 3 Correction Factors for Collimator . . . . . . . . . . . . . . . . . . . . . . . . . . . .12 - 4
13 _____ Grid _________________________________________________________13 - 1 Control overview . . . . . . . . . . . . . . . . . . . . . . . . Oscillation grid, general remarks . . . . . . . . . . . . . . . . Programming of oscillation grid / removable multi-line grid . Checking the grid centering . . . . . . . . . . . . . . . . . . Checking the grid movement . . . . . . . . . . . . . . . . . . Grid contact. . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
. . . . . .
.13 - 1 .13 - 2 .13 - 2 .13 - 3 .13 - 3 .13 - 3
14 _____ Tomography _________________________________________________14 - 1 Setting the tomographic height, beginning with firmware version VB00D Setting the column speed. . . . . . . . . . . . . . . . . . . . . . . Setting the spotfilm device speed. . . . . . . . . . . . . . . . . . . Software correction of the layer-height . . . . . . . . . . . . . . . . Setting the tomographic symmetry . . . . . . . . . . . . . . . . . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
. . . . .
.14 - 1 .14 - 1 .14 - 2 .14 - 3 .14 - 4
15 _____ Radiation Geometry____________________________________________15 - 1 Coincidence of light field/radiation field. Exposure position . . . . . . . . . . Evaluation. . . . . . . . . . . . . . Adjustment of the central beam . . . . Horizontal table direction . . . . . . Vertical table direction . . . . . . . Adjust 0 degree switch 3FC1 . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
. . . . . . .
.15 - 1 .15 - 1 .15 - 2 .15 - 3 .15 - 3 .15 - 4 .15 - 4
16 _____ Installing/Removing the Tabletop ________________________________16 - 1 Removing the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 1 Installing the tabletop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 - 1
17 _____ Installing/Remov. the Cover of the Spotfilm Dev. ____________________17 - 1 Removing the upper cover of the spotfilm device . . . . . . . . . . . . . . . . . . .17 - 1 Removing the lower cover of the spotfilm device. . . . . . . . . . . . . . . . . . . .17 - 1
18 _____ Changes to Previous Version ____________________________________18 - 1
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 6 TD SD 21
Siemens AG Medical Engineering
General Information
1-1
1
Documentation
1
• Wiring diagram SIREGRAPH CF
RHD 32
• Unit test certificate • Self-learning package on CD-Rom, RXD3-209.509.01.01.02
Measuring Equipment
NOTICE
1
All tools, measuring equipment and aids mentioned, with the exception of the “standard installation tools”, are listed and specified in the ARTD (Part 3)
• 2-channel memory oscilloscope
ARTD Part 3
• Digital multimeter Ri>10 MOhm
ARTD Part 3
• Fiber optic cable tester
84 28 039 GE999 New:K96 60 812 RE999
• Torque wrench >40 Nm
ARTD Part 3
• Spring balance 200 N
ARTD Part 3
• Spirit level
28 69 436 RE999
• Protective ground wire tester
85 73 628 RE999
Details Regarding Tolerances
1
General tolerances for measures of length according to ISO 2768 Limit sizes for nominal dimension range
over 3 mm to 6 mm
over 6 mm to 30 mm
Permissible ± 0.5 mm ± 1 mm tolerance
over 30 mm to 120 mm
over over over over 120 mm 400 mm 1000 mm 2000 mm to to to to 400 mm 1000 mm 2000 mm 4000 mm
± 1.5 mm ± 2.5 mm ± 4 mm
± 6 mm
± 8 mm
These tolerances are valid for all measurements indicated in these instructions as long as no other tolerance is explicitly indicated after the value. A tolerance of ±10% is permissible for torque values.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 30 TD SD 21
SIREGRAPH CF Service
1-2
General Information Safety Notes CAUTION
1
When carrying out the work steps and tests, the safety notes contained in the product-specific documentation as well as the general safety notes contained in the ARTD binder must be observed.
Checks and adjustments that must be carried out with radiation switched on are marked with the radiation symbol. When performing work procedures which are marked with this symbol, radiation protective clothing must be worn.
Remarks Regarding Adjustment
1
In the control panel, there are 7-segment display components which display the operating condition and, in the event of an error, an error number. The error numbers are listed (see appendix) and indicate the subassembly that is not functioning properly. If there should be a malfunction, check the drive mechanism first. In particular, check for easy movement of the moving parts Pay particular attention to the cleaning and lubrication of rails and guide parts. Check the chain tension and toothed rack guides for each drive. If the mechanism functions properly, check the electronics.
Potentiometers
1
Potentiometers P1 to P14 are located on CPU PC board 24S1. By means of these potentiometers, the voltages for test channel adjustment are set and coated with enamel. CAUTION
Do not turn these potentiometers.
Potentiometers which can be accessed from the front are also located on the Mini-Maestro (cassette carriage), Maestro (column/spotfilm device) and small voltage PC boards. These may be adjusted only if this is described in the corresponding chapter of these adjustment instructions. The same applies to the potentiometers on I/0 PC board 24S2. With replacement of a subassembly configured with potentiometers, the potentiometers have been adjusted at the factory and may not be changed, except for the potentiometers described in these adjustment instructions.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1-3
Potentiometers, Remarks Regarding Replacement and Adjustment
1
When replacing potentiometers for actual value sensing and for their adjustment, it must be made sure that there is always minimum play between the pinion gears. The potentiometers may not be subjected to mechanical load. Only the 3PT2 potentiometer (spotfilm device) is adjusted with a play of approx. 0.5 mm to achieve smooth movement. The 3PT10 and 3PT11 potentiometers, format sensing for height and width, are heat-sensitive!
Allarmax
1
This function monitors all drives and switches off if a drive is blocked. The movement must be blocked in some adjustment steps. This function must be switched off so that a shutdown does not take place.
NOTICE
Switching off the "Allarmax" software - Insert jumpers on board 24 S2 AMP3 (for service material set)
If this jumper is not configured, in the AMP 3 connector (24 S1 board) jumper Pin 16 with 1 and Pin 15 with 2. Overview of AMP 3
Siemens AG Medical Engineering
9
8
16
1
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 30 TD SD 21
SIREGRAPH CF Service
1-4
General Information
IMPORTANT
If one replaced subassembly does not function properly, do not turn the potentiometers; order a new subassembly.
Lubricating Points / Bearings / Screws / Electrostaticallysensitive Devices
1
Check the condition of all accessible chains, drives, slide and roller bearing guides prior to beginning each adjustment and if needed, apply bearing grease or a similar substance (lubricants are listed in Maintenance Instructions, RXD3-209.101.01...). If parts are worn or damaged, they may be replaced only with original parts.
Installation of Ball Bearings
1
Support rollers, rollers and ball bearings are adjusted so that the unstrained outer ring can still just be turned manually.
Securing of Screws
1
Secure all screws removed for adjustment purposes according to their original state. Secure screws by means of appropriate lock washers, counternuts or Loctite type 221, Part No.: 20 48 874. Securing by means of Loctite, if necessary, is listed in the text. When performing work steps with Loctite, please ensure that no securing compound will penetrate into any bearings. After some days, the gearing will have seized. Loctite may be removed by using heat (electric hair-dryer etc.)
Protection of Electrostatically-sensitive Devices CAUTION
1
The integrated circuits and PC boards used in the unit must be handled with particular care due to their electrically-sensitive structures and extremely high input resistance. The rules pertaining to this handling are listed in the ARTD. It is essential to use the protective devices listed there.
Manufacturer’s Note
1
This document was originally written in German.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1-5 F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L8
L97 L98
F3
F4
L99 L100
L41
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 1
Overview I/O PC Board 24S2 LED L01 L02 L03 L04
Color green green green green
L05 L06 L07 L08 L09 L10 L11 L12 L13 L14 L15 L16
green green green green green green green green green green green green
Siemens AG Medical Engineering
1
Function Control of unit tilting-up Control of unit tilting-down Control of lifting magnet SOL 1 with main contact Control of lifting magnet SOL 2 (Grid in initial position with precontact) no function no function no function no function Control of SID=115cm from tableside control panel Control of SID=135cm from tableside control panel Control of SID=150cm from tableside control panel Control of spotfilm device drive 3MT2 Control of column drive 3MT1 Control of cassette carriage drive 3MT3 no function Control of collimator plate drive3MT5
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 30 TD SD 21
SIREGRAPH CF Service
1-6
General Information
F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L41
L8
L97 L98
F3
F4
L99 L100
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 2
LED L17 L18 L19 L20 L21 L22 L23 L24 L25 L26 L27
Color green green green green green green green green green green green
L28 L29 L30 L31 L32
green green green green green
SIREGRAPH CF Service
Function Control of compression drive 3MT6 Control of SID drive 3MT7 Control of table transverse 3MT8 not used Control of collimator lamp ON Control of collimator drives (LED OFF) 3FC1 microswitch (tube unit centered to I.I.) open, LED OFF Control at kk-Interface IONTOMAT (central measuring field) Control at kk-Interface IONTOMAT (left measuring field) Control at kk-Interface IONTOMAT (right measuring field) Control at kk-Interface IONTOMAT (AC signal with collimator of height/width <12 cm) no function Control at kk-Interface (fluoroscopy ON) Control at kk-Interface (precontact ON) Control at kk-Interface (main contact ON) Control at kk-Interface (image inversion)
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1-7
F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L41
L8
L97 L98
F3
F4
L99 L100
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 3
LED L33 L34 L35 L36 L37 L38 L39 L40 L41 L42 L43 L44 L45
Color green green green green green green green green green green green green green
L46 L47 L48
green green green
Siemens AG Medical Engineering
Function Control at kk-Interface (zoom1) Control at kk-Interface (zoom2) Control at kk-Interface (zoom3) Control at kk-Interface (direct radiography) Control at kk-Interface (indirect radiography) DR Control at kk-Interface (room light) Control at kk-Interface (Fluoro 1/ADR 1) Control at kk-Interface (Fluoro 2/ADR 2) Control at kk-Interface (Fluoro stop/ADR stop) Control at kk-Interface (tomographic time A/0.6 sec.) Control at kk-Interface (tomographic time B/0.8 sec.) Control at kk-Interface (tomographic time C/1.2 sec.) Control at kk-Interface (Serial operation, direct or indirect technique with 2- to 4-subdivisions) no function no function no function
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 30 TD SD 21
SIREGRAPH CF Service
1-8
General Information
F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L41
L8
L97 L98
F3
F4
L99 L100
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 4
LED L49 L50 L51 L52 L53 L54 L55 L56 L57
Color green green green green green green green green green
L58 L59 L60 L61 L62 L63 L64
green green green green green green green
SIREGRAPH CF Service
Function Control at kk-Interface (voltage ON) Control at kk-Interface (exposure signal-as) Control at kk-Interface (blocking, e.g. door contact) Control at kk-Interface (blocking, e.g. 3D stand) no function no function no function Microswitch 3FC5 (blocking limit switch) actuated Proximity switch 3XS2 (or microswitch 3FC10) (grid contact) actuated, LED ON; removable multi-line grid inserted: LED ON Microswitch 3FC1 (tube centered onto I.I.) open, LED OFF Voltage applied to control PC board 22Az3 (column) Voltage applied to control PC board 22Az4 (spotfilm device) Microswitch 3FC8 (table transverse - wall side) actuated Microswitch 3FC9 (table transverse - doctor’s side) actuated no function no function
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1-9 F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L41
L8
L97 L98
F3
F4
L99 L100
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 5
LED L65
Color green
L66
green
L67 L68 L69 L70 L71 L72 L73 L74 L75 L76 L77 L78
green green green green green green green green green green green green
L79 L80
green green
Siemens AG Medical Engineering
Function Collimator control 24S3 ready for operation is OFF when tube assembly pivoted out 90/180° Microswitch 3FC2 (compression maximally decompressed) actuated Microswitch 3FC3 (cone inserted) actuated Microswitch 3FC4 (compression >20N - 30N) no function no function no function no function Programming switch position DW1/1 (room height, up to VC00A) Programming switch position DW1/2 (room height, up to VC00A) Programming switch position DW1/3 (I.I. size) Programming switch position DW1/4 (I.I. size) Programming switch position DW1/5 (number of zoom formats) Programming switch position DW1/6 (digital radiography/indirect radiography, up to VC00A) Programming switch position DW1/7 (inch/cm) Programming switch position DW1/8 (adjustment mode)
RXD3-209.071.01 Rev. 05 09.00
Page 9 of 30 TD SD 21
SIREGRAPH CF Service
1 - 10
General Information F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L41
L8
L97 L98
F3
F4
L99 L100
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 6
LED L81 L82 L83 L84 L85
Color green green green green green
L86 L87 L88 L89 L90 L91 L92 L93 L94 L95 L96
green green green green green green green green green green green
SIREGRAPH CF Service
Function no function no function no function no function Table-side control panel key P39A/B (cassette carriage Load-Park position) Table-side control panel key P44B (table-transv. doctor’s side) actuated Table-side control panel key P44D (table-transv. wall side) actuated Table-side control panel key P44A (long. carriage headward) actuated Table-side control panel key P44C (long. carriage footward) actuated Table-side control panel key P42A (tilt down unit) actuated Table-side control panel key P42B (tilt up unit) actuated Table-side control panel key P43A (pivot column to head end) actuated Table-side control panel key P43B (pivot column footward) actuated Table-side control panel key P21 (SID=115 cm) actuated Table-side control panel key P22 (SID=135 cm) actuated Table-side control panel key P23 (SID=150 cm) actuated
RXD3-209.071.01 Rev. 05 09.00
Page 10 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 11
F1
CN7
-
L1 PL1
F2 L16
-
L9
L24
-
L17
L32
-
L25
L40
-
L33
L48
-
L41
L8
L97 L98
F3
F4
L99 L100
L101
F5
L51 L54 L57 L60 L63 L66 L69 L72 L75 L78 L81 L84 L87 L90 L93 L96
DW 1-1 1-8
P1
-
L50 L53 L56 L59 L62 L65 L68 L71 L74 L77 L80 L83 L86 L89 L92 L95
L49 L52 L55 L58 L61 L64 L67 L70 L73 L75 L79 L82 L85 L88 L91 L94
P6
AMP5
Fig. 7
LED
Color
Function
L97
red
L98
red
L99
red
L100
red
L101
red
Short-circuit on CN7 OUT.9-16, protection for driver U1,actuate reset button PL1 Short-circuit on CN7 OUT.17-24, protection for driver U2, actuate reset button PL1 Short-circuit on CN6 OUT.25-32, protection for driver U3, actuate reset button PL1 Short-circuit on CN6 OUT.33-40, protection for driver U4, actuate reset button PL1 Short-circuit on CN6 OUT.41-48, protection for driver U5, actuate reset button PL1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 11 of 30 TD SD 21
SIREGRAPH CF Service
1 - 12
General Information PC Mode (Version Windows 3.11)
1
Tools required
1
• Laptop / notebook computer with Windows 3.11 • Connection cable (in the service bag), Part Number: 32 49 062 GE066 Preparations
1
• Connect cable to the plug CN 3 on the CPU board 24S1 and in the PC COM 1 • Switch on PC and system • Call up Windows • Click on "Accessories" • Click on "Terminal" • Click on "Settings" • Click on "Terminal Emulation" • Select "DEC VT52" and acknowledge with "OK" (Fig. 8)
Terminal Emulation (Allgemein/General)
OK Abbrechen CANCEL Cancel
Fig. 8
• Select "Data transmission" again under "Terminal" at "Settings" (Fig. 9)
Datenübertragung / Datatransfer Übertragungsrate / Transferrate (Baud) 110 300 600 1200 2400
4800
Datenbits / Databits 5 7 6 Parität / Parity Keine / None
9600
19200
OK Abbrechen / Cancel Stoppbits 1,5 1
8 Protokoll / Protocol Xon / Xoff
2
Anschluß / Connection Kein Kein Ke COM
/ None
Ungerade / Odd
Hardware
COM 1
Gerade / Even
Kein / None
COM 2
Markierung / Marker Leerzeichen / Space
Paritätsprüfung / Paritycheck Trägersignal entdecken / Discover carriersignal
Fig. 9
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 12 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 13
then select and acknowledge Connection: Baud Rate: Protocol: Parity:
COM 1 9600 none none
Procedure
1
• The menu appears with Shift-G: 0 3 8 9
= = = =
WARNING
Error memory / Messages file Hardware test and Settings, (Host us 1228) Keyboard test English / Italian changeover
Hardware test: Only the tests No 1. Battery check and No 2. Clock test / Adjust may be carried out. Carrying out the other tests may lead to damage of the unit.
English version
-------- Siregraph CF CARDS CONTROL -------*** Digit card number under test: *----> 0 = Messages file *----> 1 = *----> 2 = *----> 3 = Hardware test / Settings *----> 4 =
*----> 5= *----> 6= *----> 7= *----> 8= Keyboard Test *----> 9= Lingua italiana
Test on card No. (0..9):
NOTICE
Siemens AG Medical Engineering
With introduction of the firmware version VB00B, the text is in English/Italian< VB00B only in Italian.
RXD3-209.071.01 Rev. 05 09.00
Page 13 of 30 TD SD 21
SIREGRAPH CF Service
1 - 14
General Information Italian version
-------- Siregraph CF CARDS CONTROL -------*** Digita il numero della scheda sotto test: *----> 0 = Archivio messaggi *----> 1 = *----> 2 = *----> 3 = Host us 1228 *----> 4 =
*----> 5= *----> 6= *----> 7= *----> 8= Tastiera us 1272 *----> 9= English language
Test della scheda n. (0..9):
CAUTION
Select the entire screen area so that the text can be displayed and read completely.
In the error log mode, an inquiry is made about whether or not the contents of the error log are to be erased. The meaning of the errors is explained on the following pages (1 - 20). The function of the displays and control buttons on the remote control panel can be tested in the keyboard test. All steps are menu-prompted and are not described in greater detail here.
• Call up test No 1 (battery-voltage test) in the “Hardware test and Settings” (see also pages 1- 18 and 1 - 19) and check the voltage of the battery (CPU board 24S1). Call up test No 12 (Clock-Test /Adjust), check date and time and correct it if necessary. This must be done every time a startup or a CPU board replacement is carried out. Access to test No 12 is quick using the cursor keys. Observe the WARNING on page 1- 13.
• Exit the program with ESC.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 14 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 15
PC Mode (Version Windows 95)
1
Tools required
1
• Laptop / notebook computer with Windows 95 • Connection cable (in the service bag), Part Number: 32 49 062 GE066 Preparations
1
• Connect cable to the plug CN 3 on the CPU board 24S1 and in the PC COM 1 • Switch on PC and system • Call up Windows • “Start” • “Programs” • “Accessories” • “Hyperterminal” • “Hyperterminal.exe” • “Connection Description”: type in a suitable name (e.g. Siregraph CF) and select a symbol for the icon in the “Hyperterminal" menu or at the desktop
• Acknowledge with “OK” • “Phone Number” / “Connect using” / select “Direct to COM1” • Acknowledge with “OK” • “COM 1 Properties”: set the data as shown in the table below Bits per second
9600
Data bits
8
Parity
None
Stop bits
1
Protocol / Flow control
None
• Acknowledge with “OK” • “File” • “Properties” / “Settings” • “Emulation” / select “VT-52” • Acknowledge with “OK”
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 15 of 30 TD SD 21
SIREGRAPH CF Service
1 - 16
General Information Procedure
1
• The menu appears with Shift-G: 0 3 8 9
= = = =
WARNING
Error memory /Messages file Hardware test and Settings, (Host us 1228) Keyboard test English / Italian changeover
Hardware test: Only the tests No 1. Battery check and No 2. Clock test / Adjust may be carried out. Carrying out the other tests may lead to damage of the unit.
English version
-------- Siregraph CF CARDS CONTROL -------*** Digit card number under test: *----> 0 = Messages file *----> 1 = *----> 2 = *----> 3 = Hardware test / Settings *----> 4 =
*----> 5= *----> 6= *----> 7= *----> 8= Keyboard Test *----> 9= Lingua italiana
Test on card No. (0..9):
NOTICE
SIREGRAPH CF Service
With introduction of the firmware version VB00B, the text is in English/Italian, < VB00B only in Italian.
RXD3-209.071.01 Rev. 05 09.00
Page 16 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 17
Italian version
-------- Siregraph CF CARDS CONTROL -------*** Digita il numero della scheda sotto test: *----> 0 = Archivio messaggi *----> 1 = *----> 2 = *----> 3 = Host us 1228 *----> 4 =
*----> 5= *----> 6= *----> 7= *----> 8= Tastiera us 1272 *----> 9= English language
Test della scheda n. (0..9):
CAUTION
Select the entire screen area so that the text can be displayed and read completely.
In the error log mode, an inquiry is made about whether or not the contents of the error log are to be erased. The meaning of the errors is explained on the following pages (1 - 20). The function of the displays and control buttons on the remote control panel can be tested in the keyboard test. All steps are menu-prompted and are not described in greater detail here.
• Call up test No 1 (battery voltage test) in the “Hardware test and Settings” (see also pages 1- 18 and 1 - 19) and check the voltage of the battery (CPU board 24S1). Call up test No 12 (Clock-Test /Adjust), check date and time and correct it if necessary. This must be done every time a startup or a CPU board replacement is carried out. Access to test No 12 is quick using the cursor keys. Observe the WARNING on page 1- 16.
• Exit the program with ESC.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 17 of 30 TD SD 21
SIREGRAPH CF Service
1 - 18
General Information English Version of the Hardware Test and Settings
1
-------- CARD uS1228 CONTROL -------Controls in Consecutive mode or Test by test mode (C/T): ----- Battery voltage TEST. Test No 1 ------------------------------------------ Data RAM TEST. Test No 2. Not to be executed ! ----- RAM 1 BIT TEST. Test No 3. Not to be executed ! ----- Address RAM TEST. Test No 4. Not to be executed ! ----- SIGNALS *UDS, *LDS TEST. Test No 5. Not to be executed ! ----- TEST Checksum EPROM. Test No 6. Not to be executed ! ----- EEPROM TEST. Test No 7. Not to be executed ! ----- Time interrupt PIT1 TEST. Test No 8. Not to be executed ! ----- Led RUN and Relay OK TEST. Test No 9. Not to be executed ! ----- PIT 2 TEST. Test No 10. Not to be executed !
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 18 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 19
----- Serial line No. 2 TEST (OPTICAL FIBRE). Test No 11. Not to be executed ! ----- CLOCK TEST (RTC). Test No 12 ------------------------------------------ Digital INPUT/OUTPUT TEST. Test No 13. Not to be executed ! ----- Trimmer CALIBRATION uS1269. Test No 14. Not to be executed ! ----- Digital/analogical converter (DAC) TEST. Test No 15. Not to be executed ! ----- Analogical/digital converter (ADC) TEST. Test No 16. Not to be executed ! ----- TEST on precision of ADC reading. Test No 17. Not to be executed !
----- Press a key to continue:
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 19 of 30 TD SD 21
SIREGRAPH CF Service
1 - 20
General Information Error List
1
Drive
Error
State
Possible causes
Tilt
001
Rest
Potentiometer 3PT4 failed Potentiometer 3PT4 beyond end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
001
Movement
Potentiometer 3PT4 wrongly wired
002
Rest
Potentiometer 3PT1 failed Potentiometer 3PT1 beyond end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
012
Movement column + spotfilm device
Potentiometer 3PT1 mechanically not connected Potentiometer 3PT1 incorrectly wired Motor 3MT1 incorrectly wired Tacho generator 3DT1 failed / incorrectly wired
012
Actuated and no movement
DAC output on CPU board 24S1 failed Motor 3MT1 not connected Safety limit switch 3FC12 actuated Current limitation on 22AZ3 (Maestro) responded
022
Oblique projection
Potentiometer 3PT1 mechanically not connected Potentiometer 3PT1 incorrectly wired Motor 3MT1 wrongly wired Tacho generator 3DT1 failed / incorrectly wired
022
Actuated and no movement
DAC output on CPU board 24S1 failed Motor 3MT1 not connected Safety limit switch 3FC12 actuated Current limitation on 22AZ3 (Maestro) responded
032
Tomography
Too big difference between nominal and actual value
042
Movement
Too big difference between nominal and actual value
Rest
Potentiometer 3PT2 failed Potentiometer 3PT2 beyond end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
Movement Column + spotfilm device
Potentiometer 3PT2 failed Potentiometer 3PT2 beyond end position Motor 3MT2 incorrectly wired Tacho generator 3DT2 failed / incorrectly wired
Column
Spotfilm device 003
013
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 20 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
SID drive
"Width" collimator
Siemens AG Medical Engineering
1 - 21
013
Actuated and no movement
DAC output on CPU board 24S1 failed Motor 3MT2 not connected Safety limit switch 3FC14 actuated Current limitation on 22AZ4 (Maestro) responded
023
Oblique projection
Potentiometer 3PT2 mechanically not connected Potentiometer 3PT2 incorrectly wired Motor 3MT2 incorrectly wired Tacho generator 3DT2 failed / incorrectly wired
023
Actuated and no movement
DAC output on CPU board 24S1 failed Motor 3MT2 not connected Safety limit switch 3FC14 actuated Current limitation on 22AZ4 (Maestro) responded
033
Tomography
Too big difference between nominal and actual value
043
Movement
Too big difference between nominal and actual value
004
Rest
Potentiometer 3PT7 failed Potentiometer 3PT7 beyond the end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
004
Movement
Potentiometer 3PT7 mechanically not connected Potentiometer 3PT7 incorrectly wired Motor 3MT7 incorrectly wired
004
Actuated and no movement
DAC output on CPU board failed Motor 3MT7 not connected Safety limit switch 3FC7 actuated Board 23AZ7 has responded / is damaged
005
Rest
Potentiometer defective Potentiometer beyond the end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
005
Movement
Potentiometer mechanically not connected Potentiometer incorrectly wired
015
Movement
Motor incorrectly wired Deviation between collimator and diaphragm plates too large
015
Actuated and no movement
DAC output on CPU board 24S1 failed Motor not connected Board 24S3 defective
RXD3-209.071.01 Rev. 05 09.00
Page 21 of 30 TD SD 21
SIREGRAPH CF Service
1 - 22 "Height" collimator
General Information 006
Rest
Potentiometer defective Potentiometer beyond the end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
006
Movement
Potentiometer mechanically not connected Potentiometer incorrectly wired
016
Movement
Motor incorrectly wired Deviation between collimator and diaphragm plates too large
016
Actuated and no movement
DAC output on CPU board 24S1 failed Motor not connected Board 24S3 defective
007
Rest
Potentiometer 3PT5 defective Potentiometer 3PT5 over end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
007
Movement
Potentiometer 3PT5 mechanically not connected Potentiometer 3PT5 incorrectly wired
017
Movement
Motor 3MT5 incorrectly wired Deviation between collimator and diaphragm plates too large
017
Actuated and no movement
DAC output on CPU board 24S1 failed Motor 3MT5 not connected Board 23AZ5 has responded
"Width" 008 clamping jaws
Rest
Potentiometer 3PT10 defective Potentiometer 3PT10 beyond the end position Connection cable 3PT10 interrupted / not connected ADC input on CPU board 24S1 failed
"Height" 009 clamping jaws
Rest
Potentiometer 3PT11 defective Potentiometer 3PT11 beyond the end position Connection cable 3PT11 interrupted / not connected ADC input on CPU board 24S1 failed
Cassette carriage
Rest
Potentiometer 3PT3 defective Potentiometer 3PT3 beyond the end position Connection cable interrupted / not connected ADC input on CPU board 24S1 failed
Diaphragm plates
SIREGRAPH CF Service
010
RXD3-209.071.01 Rev. 05 09.00
Page 22 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
General
Siemens AG Medical Engineering
1 - 23
010
Movement
Potentiometer 3PT3 mechanically not connected Potentiometer 3PT3 incorrectly wired Motor 3MT3 incorrectly wired Tacho generator 3DT3 failed / incorrectly wired
010
Actuated and no movement
DAC output on CPU board 24S1 failed Motor 3MT3 not connected Board 22AZ2 (Mini Maestro) has responded
098
Console-control cabinet fiber optic cable defective Fault in the initialization phase Wrong data transmission
099
Control console / spotfilm device EMERGENCY STOP actuated A safety limit switch actuated
100
E2-Prom on CPU board 24S1 defective / not present
101
E2-Prom on CPU board data loss Incorrect checksum
RXD3-209.071.01 Rev. 05 09.00
Page 23 of 30 TD SD 21
SIREGRAPH CF Service
1 - 24
General Information Replacing the 24S1 CPU Board and the 24S2 I/O Board
1
24S1 CPU Board
1
After a replacement of the CPU board 24S1, the E-Prom1 and E-Prom 2 and EE-Prom U83 from the board that has been removed must be taken over. Turn the unit on and on the I/O board 24S2, set the DW 1/8 switch to ON. At the remote control console type in code 117 using the keys P31, P32, P34 and P35. Store with key P33 (see page 2 -1). On the I/O board 24S2 set the DW 1/8 switch to OFF. When the initialization that follows takes place, the switch positions of the DW1 programming switch are read in. Carry out Clock-Test/Adjust in the PC-Mode (Hardware test and Settings), see page 1 - 12 respectively 1 - 15 following.
24S2 I/O Board
1
After a replacement of the I/O board 24S2, the unit configurations from the DW1 programming switches must be taken over from the board that has been removed. Turn the unit on and on the I/O board 24S2, set the DW 1/8 switch to ON. At the remote control console type in code 117 using the keys P31, P32, P34 and P35. Store with key P33 (see page 2 -1). On the I/O board 24S2 set the DW 1/8 switch to OFF. When the initialization that follows takes place, the switch positions of the DW1 programming switch are read in.
Changes to the DW1 DIP Switches on the 24S2 Board
1
If changes are made to the DW1 DIP switches, proceed as follows:
• Select the adjustment mode. - 24S2 I/O board, DW 1/8 on ON. - Make the changes to the switch positions, e.g. I.I. size, room height, DFR, inch-cm, Zoom steps - Set code 117 and save (see Page 2-1).
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 on OFF. - When this is done, the changed switch positions are read in.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 24 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 25
Table of DW1 Switch Positions (up to VC00A)
Roomheight Raumhöhe > 3265 mm
Fullformat Vollformat
9" Imageintensfr. 23 cm BV ON
1
ON
ON
OFF
Programming OFF Programmierung AUS ON
OFF
OFF
OFF
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
Roomheight Raumhöhe max. 3000 mm
13" Imageintensfr. 33 cm BV
One Zoomstep Eine Zoomstufe
Programming ON Programmierung EIN ON
ON
ON
ON
OFF
OFF
OFF
OFF
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
Roomheight Raumhöhe max. 2850 mm
DFR ON DFR EIN
Two Zoomsteps Zwei Zoomstufen
cm-Cassettes cm-Kassetten
ON
ON
ON
OFF
ON
OFF
OFF
OFF
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
Roomheight Raumhöhe max. 2600 mm
DFR OFF DFR AUS
Three Zoomsteps Drei Zoomstufen
Inch-Cassettes Zoll-Kassetten
ON OFF 1 2 34 5 6 7 8
Siemens AG Medical Engineering
ON
ON
OFF
OFF 1 2 34 5 6 7 8
ON
1 2 34 5 6 7 8
RXD3-209.071.01 Rev. 05 09.00
Page 25 of 30 TD SD 21
OFF 1 2 34 5 6 7 8
SIREGRAPH CF Service
1 - 26
General Information Table of DW1 Switch Positions (as from VC01A)
Not used Nicht verwendet
Fullformat Vollformat
9" Imageintensfr. 23 cm BV ON
ON
ON
OFF
1
Programming OFF Programmierung AUS ON
OFF
OFF
OFF
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
Not used Nicht verwendet
13" Imageintensfr. 33 cm BV
One Zoomstep Eine Zoomstufe
Programming ON Programmierung EIN ON
ON
ON
ON
OFF
OFF
OFF
OFF
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
1 2 34 5 6 7 8
Not used Nicht verwendet
Not used Nicht verwendet
Two Zoomsteps Zwei Zoomstufen
cm-Cassettes cm-Kassetten
ON OFF 1 2 34 5 6 7 8
ON
ON
OFF
OFF 1 2 34 5 6 7 8
1 2 34 5 6 7 8
ON OFF 1 2 34 5 6 7 8
Code ss211ss mit / with DFR Codesiiiis212 s DFR Vorzugslage / Default ss Codesis213 s ohne / without DFR Not used Nicht verwendet ON
Code für Raumhöhe / Code for room height Code 201 ssss 340 cm (keine Begrenzung / no limitation) Code 202 ssss 330 cm Code 203 ssss 320 cm Code 204 ssss 310 cm Code 205 ssss 300 cm Code 206 ssss 290 cm Code 207 ssss 280 cm Code 208 ssss 270 cm Code 209 ssss 260 cm
SIREGRAPH CF Service
ON
OFF
OFF 1 2 34 5 6 7 8
Inch-Cassettes Zoll-Kassetten ON
ON
OFF 1 2 34 5 6 7 8
Three Zoomsteps Drei Zoomstufen
Not used Nicht verwendet
1 2 34 5 6 7 8
OFF 1 2 34 5 6 7 8
Siehe auch Blatt 2-1, Justagemodus See also page 2-1, adjustment mode
RXD3-209.071.01 Rev. 05 09.00
Page 26 of 30 TD SD 21
Siemens AG Medical Engineering
General Information
1 - 27
Listing of all Programming Codes Adjustment of:
1
DIP switch
Code
Remark
2600 mm
1 Off, 2 Off
117
up to VC00A
2850 mm
1 On, 2 Off
117
up to VC00A
3000 mm
1 Off, 2 On
117
up to VC00A
>3260 mm
1 On, 2 On
117
up to VC00A
33 cm
3 Off
117
23 cm
3 On
117
Full format
4 Off, 5 Off
117
1 Zoom
4 On, 5 Off
117
2 Zoom
4 Off, 5 On
117
3 Zoom
4 On, 5 On
117
ON
6 Off
117
up to VC00A
OFF
6 On
117
up to VC00A
cm
7 Off
117
Inch
7 On
117
Room height, up to VC00A
I.I. size
I.I. Zoom
DFR selection, up to VC00A
CM-inch format
Tilt drive -18°
11
+90°
12
Basic setting
Column drive Left side
21
30 mm to the end
Right side
22
30 mm to the end
Left side
31
105 mm to the end
Right side
32
105 mm to the end
Bottom position
41
set at 45°
Top position
42
set at 45°
Spotfilm device drive
SID drive
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 27 of 30 TD SD 21
SIREGRAPH CF Service
1 - 28
General Information Adjustment of:
DIP switch
Code
Remark
Collimator Width closed
51
Width open
52
Height closed
61
Height open
62
Collimator plates Closed
71
min = 41 mm
Open
72
max = 418 mm
Width, min
81
197 mm
Width, max
82
467 mm
Height, min
91
193 mm
Height, max
92
471 mm
Load position
101
198 mm
Park position
102
584 mm
For cassette width
121-129
min-max
For cassette height
131-139
min-max
For I.I. width
141-149
min-max
For I.I. height
151-159
min-max
90°
161
89°
162
90° basic setting must first be performed.
88°
163
87°
164
Format sensing
Cassette carriage drive
Collimator correction
Collimator correction
Tilt down 87°- 90°, beginning with VC01A
Tomo height correction, beginning with VB00D Reduction
171-174
No correction
175
Increase
176-179
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
1 mm steps
1 mm steps
Page 28 of 30 TD SD 21
Siemens AG Medical Engineering
General Information Adjustment of:
1 - 29 DIP switch
Code
Remark
Subdivision gap width, beginning with VB00D Gap smaller
181-184
No correction
185
Gap larger
186-189
0.2 mm steps
0.2 mm steps
SID blocking, beginning with VB00D No blocking
191
SID blocking >135
192
Effective beginning at 45°
SID blocking >115
193
Effective beginning at 45°
>3400 mm
201
beginning with VC01A
3300 mm
202
beginning with VC01A
3200 mm
203
beginning with VC01A
3100 mm
204
beginning with VC01A
3000 mm
205
beginning with VC01A
2900 mm
206
beginning with VC01A
2800 mm
207
beginning with VC01A
2700 mm
208
beginning with VC01A
2600 mm
209
beginning with VC01A
With DFR
211
beginning with VC01A
DFR default
212
beginning with VC01A
Without DFR
213
beginning with VC01A
Room height, beginning with VC01A
DFR selection, beginning with VC01A
Grid Classic oscillating grid
221
Removable multi-line grid
222
FL-TOP or FL-COMPACT
307
Should not be performed because the values are not exact. Afterwards, all drives must be readjusted.
Download EE-Prom values Default values
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 29 of 30 TD SD 21
SIREGRAPH CF Service
1 - 30
General Information This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 30 of 30 TD SD 21
Siemens AG Medical Engineering
Adjustment Preparations
2-1
2
Fig. 1
Adjustment mode
2
Several buttons and the 3 on the 7-segment display of the remote control console are the input elements and input display for the adjustment mode (Fig. 1).
Start adjustment mode
2
Set DIP switch DW 1/8 on the I/O board 24S2 to "ON". With selection of the adjustment mode, you cannot control any movements from the remote and tableside control consoles. 000 appears in the display. Before each adjustment, create the mechanical conditions and enter a 3-digit code. The following buttons are used for the adjustment mode: P31
Adjustment
Unit only at 1/ for > 120 steps of one downwards
P32
Adjustment
Unit only at 2/ for > 120 steps of one upwards
P34
Adjustment upwards
Steps of ten upwards
P35
Adjustment downwards Steps of ten downwards
P33
Storage
P48
Reset
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 4 TD SD 21
SIREGRAPH CF Service
2-2
Adjustment Preparations Switching off the adjustment mode
2
Set the DIP switch DW 1/8 on the 24S2 I/O board to the OFF position. When this is done, an automatic reset takes place and the data/values that have been set are read in. However the reset key on the 24S1 CPU board can also be pressed or the unit can be switched off and back on again.
NOTICE
Pressing the reset button PL1 on the CPU board 24S1 will only bring about a reset exclusively on this board; however, no reset will take place on the CPU board 1S2 in the remote control console. A comprehensive reset will be achieved only with unit OFF and ON again.
The following short version is used for the adjustment steps. Example:
• Mechanical initial position: • Select adjustment mode - I/O board 24S2, DW 1/8 to ON - Set 011 and store (see 2-1)
• Exit the adjustment mode. - I/O board 24S2, DW 1/8 to OFF. - This allows the adjustment data to be taken over.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 4 TD SD 21
Siemens AG Medical Engineering
Adjustment Preparations
2-3
Checking the voltages
2
PC board
Plug
Pin
Nominal value Tolerance
24S1
CN9
1/2 3/4 5/6 7/8
16 V ~ 10 V ~ 19 V ~ 19 V ~
+/- 2 V ~ +/- 2 V ~ +/- 2 V ~ +/- 2 V ~
24S2
SC1
1/4
21 V ~
+/- 2 V ~
24S3
SC1
1/2 3/4 5/6
16 V ~ 16 V ~ 24 V ~
+/- 2 V ~ +/- 2 V ~ +/- 2 V ~
23S10
SC2
1/2 3/4
16 V ~ 16 V ~
+/- 2 V ~ +/- 2 V ~
1S2
CN6
1/2
16 V ~ 16 V ~
+/- 2 V ~ +/- 2 V ~
22AZ3
21/22 21/23 21/24
105 V ~ 105 V ~ 105 V ~
+8V~/-0V~ +8V~/-0V~ +8V~/-0V~
22AZ4
21/22 21/23 21/24
105 V ~ 105 V ~ 105 V ~
+8V~/-0V~ +8V~/-0V~ +8V~/-0V~
23S10
SC1
1/2 3/4+
58 V =
+/- 2 V =
22S10
CN1
3/14 + 15/16 -
64 V =
+/- 2 V =
1/3 1/5 3/5
240 V ~ 240 V ~ 240 V ~
+12 V ~ / - 0V ~ +12 V ~ / - 0V ~ +12 V ~ / - 0V ~
22K1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 4 TD SD 21
SIREGRAPH CF Service
2-4
Adjustment Preparations This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 4 TD SD 21
Siemens AG Medical Engineering
Tilt Drive
3-1
3
Control overview
3
Tableside control panel 3S1
I/O PC board 24S2
CPU PC board 24S1
EMERGENCY STOP
Remote control console Contactor subassembly
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 10 TD SD 21
SIREGRAPH CF Service
3-2
Tilt Drive
Safety limitswitch 3FC11
Potentiometer 3PT4
Tension screw
Brake assembly
3MT4
Poly-V-belt Fig. 2
Tilt drive overview
3
Drive:
3MT4
Control:
24 S2 Breaker:
Potentiometer:
3PT4
Fuse:
F26 F27 F28
Safety limit switch:
SIREGRAPH CF Service
22k1 22k2
6A/T 6A/T 6A/T
22TR1
3FC11
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 10 TD SD 21
Siemens AG Medical Engineering
Tilt Drive
3-3
a
B
A
A
10
0,2
Fig. 3
1
2
Fig. 4
Fig. 5
Belt Tension, Adjustment Procedure
3
• If UI 003/96 has not yet been performed, i.e., the stop (1/Fig. 5) is not installed: move the unit into the +60° to +70° position and move the exposure unit to max. head end. In this position, the force on the tilt drive is low and a release on its own is not very likely.
• If UI 003/96 has been performed and the stop (1/Fig. 5) is installed: Raise unit to +90°. • Loosen the four mounting screws (1/Fig. 3) on the motor support plate. At the adjustment screw (Detail 2/Fig. 3), turn until the Poly-V belt is lightly tensioned. The motor longitudinal axis must be parallel to the gear longitudinal axis. Tighten the mounting screws (1/Fig. 3). Check: the outside surface of the belt pulley on the drive must be flush with the level of the diameter reduction of the belt pulley on the motor (Fig. 4). If necessary, correct the position of the belt pulley on the motor in the shaft long. direction.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 10 TD SD 21
SIREGRAPH CF Service
3-4
Tilt Drive
a
B
A
A
10
0,2
Fig. 6
Brake assembly
3
Auxiliary device:
Feeler gauge 0.2 mm
Adjustment conditions:
Move the unit into +60° to +70° position and move the exposure unit to max. head end. With the nuts (A/Fig. 6) set the distance of all springs to a = 10 mm. This setting corresponds to a brake momentum of approx. 50 Nm. With 60 Hz power supply the brake momentum must be approx. 60 Nm, which requires a stronger pretension of the springs. To do this, mark the nuts (A/Fig. 6) with a waterproof pen (for control purpose) and tighten them just one turn each. Evenly adjust the gap of 0.2 mm + 0.1 mm between magnet and brake disk using the nuts (B/Fig. 6).
Replacement:
If UI 003/96 has not yet been performed, i.e., the stop (1/Fig. 5) is not installed: move the unit into +60° to +70° position and move the exposure unit to max. head end. In this position, the force on the tilt drive is low and movement on its own is very unlikely. Replace the brake. If UI 003/96 has been performed and the stop (1/Fig. 5) is installed: raise unit to +90°. Turn the flywheel of the tilt motor 3MT4, against the resistance of the brake until the stop (1/Fig. 5) is securely against the unit base (2/Fig. 5). Replace the brake. Turn the flywheel o the tilt motor 3MT4 in the opposite direction as previously until the stop (1/Fig. 5) as previously is no longer against the unit base and the limit switch 3FC11 (Fig.2 + 11) is no longer actuated.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 10 TD SD 21
Siemens AG Medical Engineering
Tilt Drive
3-5
36°
120
506
18°
506mm
143
Fig. 7
Fig. 8
Adjustment or replacement of potentiometer 3PT4 Auxiliary device:
3
Spirit level
• Press EMERGENCY STOP: Mechanical initial position: Electronic initial position:
+36°, ± 1° (adjust manually at the flywheel of the motor 3MT4) +5V, ± 50 mV, 24S1 CPU board, TP2/1, TP2/13,14 (0V)
• Unlock EMERGENCY STOP
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 10 TD SD 21
SIREGRAPH CF Service
3-6
Tilt Drive
36°
120
506
18°
506mm
143
Fig. 9
Fig. 10
Set end positions Auxiliary device:
3
Spirit level
Trendelenburg position, -18°
• Move the spotfilm device all the way to the foot end. • Move the unit into the Trendelenburg position, approx. -18°, press EMERGENCY STOP. - Manually set the exact position at the rotation disk.
• Select the adjustment mode. - On the 24S2 I/O board, set DW 1/8 to ON (up) - Set Code 011 and save by pressing the P33 key.
• Disengage the EMERGENCY STOP switch. • Exit the adjustment mode. - On the 24S2 I/O board, set DW 1/8 to OFF (down).
• Move into the 90° vertical unit position, press the EMERGENCY STOP switch. - Manually set the exact 90° position on the rotation disk.
• Select the adjustment mode. - On the 24S2 I/O board, set DW 1/8 to ON (up). - Set Code 012 and save the value by pressing the P33 key.
• Stay in the adjustment mode.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 10 TD SD 21
Siemens AG Medical Engineering
Tilt Drive
3-7
• Find the tilt position in the table (see below) (per the customer’s request) at which the unit is supposed to stop later during operation in the upright position. - Set the appropriate code and save by pressing the P33 key.
• Disengage the EMERGENCY STOP switch. • Exit the adjustment mode. - On the 24S2 I/O board, set DW 1/8 to OFF (down).
• Switch off the system.
Code
Tilt Position
161
90°
162
89°
163
88°
164
87°
• Move to 0° position and measure with the angle spirit level. Required: 0°, ± 1° Correction: - Readjust end positions
• If the 0° position is not correct, readjust the end positions.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 10 TD SD 21
SIREGRAPH CF Service
3-8
Tilt Drive
13
– 18°
B
3FC11
22.5
90°
3FC11
13 A
Fig. 11
Safety limit switch 3FC11 Auxiliary device:
3
Tape measure
Preadjustment of the trip cam Adjustment condition
Adjustment
Unit position approx. +75° to 80°, spotfilm device maximum to head end
• Trip cam (A/Fig. 11) distance = 13, ± 1 mm. • Adjust safety limit switch 3FC11 to dimension 22.5 mm. • The safety limit switch must be fixed so that it will properly hit the trip cam. However, the switch must not collide with the fixing screws of the trip cam. • Trip cam (B/Fig. 11) distance = 13, ± 1 mm.
Visual inspection
Move to both end positions. Check the distance to the point where the switch should be activated: Vertical unit position, (exactly +90°) Trendelenburg position – 18°
Movement times
5 ±1 mm 5 ± 1 mm 3
Tilting speed = 6°/ sec. The speed is not adjustable
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 10 TD SD 21
Siemens AG Medical Engineering
Tilt Drive
3-9
3F16
Fig. 12
Safety limit switch 3FC16
3
When the bottom switch is pressed (roller), switch 3FC16 positively actuate. This can be adjusted by moving the switch strike plate (arrow/Fig. 12).
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 9 of 10 TD SD 21
SIREGRAPH CF Service
3 - 10
Tilt Drive
A
BV 9"
183 ±3
A
BV 13"
122 ±3
Checking the floor spacing with 0 degree table position Nominal:
3
Image intensifier format 9" (23 cm) 183 mm, ± 3 mm Image intensifier format 13" (33 cm) 122 mm, ± 3 mm
Correction:
SIREGRAPH CF Service
Shifting the subassembly (A / drawing above) in the oblong holes.
RXD3-209.071.01 Rev. 05 09.00
Page 10 of 10 TD SD 21
Siemens AG Medical Engineering
Spotfilm Device Longitudinal Drive
4-1
4
Control overview
4
Tableside control panel 3S1 I/O PC board 24S2
CPU PC board
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 8 TD SD 21
SIREGRAPH CF Service
4-2
Spotfilm Device Longitudinal Drive
3MT2 3MT2 3DT2
3FC14
3FC14 / 3FC5
3DT2
3PT2 3PT2
3FC15 3FC15
3FC16 3FC16
Fig. 2
Spotfilm device longitudinal drive overview Drive:
3MT2
Control:
22Az4
Potentiometer:
3PT2
Tacho generator:
3DT2
Fuses:
F23 F24 F25
Safety limit switch:
SIREGRAPH CF Service
10A/T 10A/T 10A/T
4
22TR1
3FC14 3FC15
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 8 TD SD 21
Siemens AG Medical Engineering
Spotfilm Device Longitudinal Drive
4-3
Console
4
Joystick setting P44 (spotfilm device)
• Connect DVM (+) 1S2 - U19 - 19 (-) 1S2 - U19 - 20
• Set P1 / 1S2 5V, ± 0.05 V joystick in rest position and LED 12 and 13 on 24S2 "OFF"
19
Siemens AG Medical Engineering
20
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 8 TD SD 21
SIREGRAPH CF Service
4-4
Spotfilm Device Longitudinal Drive
A
105 875
A
X
105
105
B B
105
Fig. 3
Adjustment or replacement of potentiometer 3PT2
4
Reference point: outside edge of spotfilm device housing and outside edge of table frame. Mechanical initial position:
Electronic initial position:
Potentiometer play approx. 0.5 mm. Use the motor to move into position 875 mm (X/Fig. 3), press EMERGENCY STOP and exactly adjust 3MT2 on the motor shaft manually. Disengage EMERGENCY STOP. +5V ± 50 mV 24S1 CPU board, TP2/4, TP2/13,14, 0V.
Set end positions
4
• Move to the left end position of approx. 105 mm, press EMERGENCY STOP. - Manually set the exact position (on the motor axis).
• Select adjustment mode (see Page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Set 031 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF. - Release EMERGENCY STOP.
• Move to right end position of approx. 105 mm, press EMERGENCY STOP. - Manually set the exact position (on the motor axis).
• Select adjustment mode (see above). - Set 032 and store.
• Exit the adjustment mode (see above), release EMERGENCY STOP.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 8 TD SD 21
Siemens AG Medical Engineering
Spotfilm Device Longitudinal Drive
4-5
3FC14 X
3FC5
26, ±1 mm
Fig. 4
Safety limit switch 3FC14
4
With safety limit switch 3FC14 actuated, the drive must be moved out of the trip cam by hand. For this, there is a shaft extension with Allen screw on motor 3MT2. Remove cover (X/Fig. 4) Adjustment:
Adjust the same height of 26, ± 1 mm for 3FC14 and 3FC5. Move the drive maximally towards head end. Screw down switch plate. Distance of both trip cams to the contact surface approx. 1 mm (visual inspection).
Checking the PC board SP 22Az4 CAUTION
4
Do not readjust any potentiometers on this PC board.
For correct speed control, a 5110 Ohm resistor has been soldered on plug-in piggy-back board "SP 22Az4", pin 9-12. After a replacement of the "Maestro" assembly this plug-in piggy-back board "SP 22Az4" from the "Maestro" assembly that has been removed must be taken over. If the piggy-back board itself is to be replaced it must be ensured that a 5110 Ohm resistor is soldered on the new board. Movement times: Speed continuously controllable up to approx. 10 cm/sec in the whole travel range. The maximum speed is not adjustable.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 8 TD SD 21
SIREGRAPH CF Service
4-6
Spotfilm Device Longitudinal Drive
X 1 1
1st ring 2nd ring 1
1 Fig. 5
Fig. 6
1
Fig. 7
Spotfilm device drive, adjusting belt tension
4
• An adjustment of the belt tension is required after a replacement of the transmission or drive motor.
• Adjustment of tension is carried out with the help of the supplied measurement aid (1/Fig. 5).
• Adjustment is possible only with the complete disassembly of the transmission. • A toothed belt is used as the belt for the spotfilm device drive.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 8 TD SD 21
Siemens AG Medical Engineering
Spotfilm Device Longitudinal Drive
4-7
Adjustment procedure
4
• Loosen the four screws (1/Fig. 6). • Place on the measurement aid and press on it (X/Fig. 6) Requirement: 1st ring must touch the top edge of the window. Correction: - By turning the adjustment screw (1/Fig. 7) - Turning counterclockwise: the belt is tightened - Turning clockwise: the belt is loosened
• After adjustment is completed, tighten screws (1/Fig. 6). • Install the drive.
NOTICE
Siemens AG Medical Engineering
The 2nd ring is used only for factory adjustment of the measurement aid.
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 8 TD SD 21
SIREGRAPH CF Service
4-8
Spotfilm Device Longitudinal Drive This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 8 TD SD 21
Siemens AG Medical Engineering
Column Drive
5-1
5
Control overview
5
Tableside control panel 3S1 I/O PC board 24S2
CPU PC board
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 8 TD SD 21
SIREGRAPH CF Service
5-2
Column Drive
back of column, 3FC12 3MT1, 3MT13DT1
3DT1
3PT1 3PT1
3FC5 3FC5 tripSchaltnocke cam
trip cam Schaltnocke
3FC12 3FC14
3FC5 3FC12 14
3FC5 14 3FC12
Fig. 2
Column drive overview
5
Drive:
3MT1
Control:
22Az3
Potentiometer:
3PT1
Tacho generator:
3DT1
Fuses:
F20 F21 F22
Safety limit switch:
SIREGRAPH CF Service
10A/T 10A/T 10A/T
22TR1
3FC12 3FC5
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 8 TD SD 21
Siemens AG Medical Engineering
Column Drive
5-3 2340
980
A
X
30,+3mm
B 30,+3mm
Fig. 3
Adjustment and replacement of potentiometer 3PT1
5
Reference point: outside edge of the table frame. Mechanical initial position:
Electronic initial position:
Use the motor to move into position 980 mm (X/Fig. 3), press EMERGENCY STOP and exactly adjust 3MT1 on the motor shaft manually. Disengage EMERGENCY STOP. +5V, ± 50 mV 24S1 CPU board, TP2/3-TP2/13,14: 0V
• Move to column end position A and oblique projection head-end, press EMERGENCY STOP, manually set exact position (on motor axis).
• Select adjustment mode (see Page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Set 021 and store.
• Release EMERGENCY STOP • Exit adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
• Move to column end position B and oblique projection foot-end, press EMERGENCY STOP, manually set exact position (on motor axis).
• Select adjustment mode (see Page 2-1.) - Set 022 and store.
• Release EMERGENCY STOP • Exit adjustment mode. - 24S2 I/O board, DW1/8 to OFF
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 8 TD SD 21
SIREGRAPH CF Service
5-4
Column Drive
3FC12 back of column, 3FC12 3MT1, 3DT1 3MT1
29, ±1mm
3DT1 3PT1 3PT1
3FC5 trip cam Schaltnocke 3FC12
trip cam Schaltnocke
3FC14
3FC5 3FC12 14
3FC5 3FC12 14
3FC14 3FC5
26 ± 1 mm
Fig. 4
Safety limit switch 3FC12
5
With safety limit switch actuated, the drive must be moved out of the trip cam by hand. For this, there is a shaft extension with Allen screw on motor 3MT1.
Blocking limit switch 3FC5
5
This is actuated if the tube is pivoted towards the head end or foot end and if the longitudinal carriage movement has nearly reached the end position. Upon actuation, the swivel movement is blocked in one direction. This is actuated also if the tube is pivoted towards the head end or foot end and if the longitudinal carriage movement is controlled in the direction of the end position. Upon actuation, the longitudinal carriage movement is blocked in one direction.
Adjustment:
SIREGRAPH CF Service
Adjust the same height of 26, ± 1 mm for 3FC14 and 3FC5. Move the drive towards head end. Screw down switch plate. Distance of both trip cams to the contact surface approx. 1 mm (visual inspection).
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 8 TD SD 21
Siemens AG Medical Engineering
Column Drive
5-5
C+ 1, ±0.5
B=47.5
3FC15
A=0.5mm
B=47,5
1, ±0.5 C+
Fig. 5
Safety limit switch 3FC15
5
Initial position:
Set unit to 90°. Column vertical (centered). Spotfilm device in the center of the travel range.
Adjustment:
Adjust switch roller to A = approx. 0.5 mm. Adjust trip cam to center B = 47.5.
Check:
Move column towards head end and foot end. Max. oblique projection and SID = 115 cm. Gap C between switch roller and trip cam = 1, ± 0.5 mm.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 8 TD SD 21
SIREGRAPH CF Service
5-6
Column Drive
3FC13
Fig. 6
Checking the PC board SP 22Az3 CAUTION
5
Do not adjust any potentiometers on this PC board.
For correct speed control, a 9090 Ohm resistor has been soldered on plug-in piggy-back board "SP 22Az3", pin 9-12. After a replacement of the "Maestro" assembly this plug-in piggy-back board "SP 22Az3" from the "Maestro" assembly that has been removed must be taken over. If the piggy-back board itself is to be replaced it must be ensured that a 9090 Ohm resistor is soldered on the new board. Movement times: Speed = 4°/sec. The speed is not adjustable.
Room height switch 3FC13
5
Check:
Press the "antenna" (Fig. 6) manually, switch has to switch as soon as possible.
Adjustment:
See (Detail/Fig. 6).
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 8 TD SD 21
Siemens AG Medical Engineering
Column Drive
5-7
X
1
1
1st ring 2nd ring 1
1 Fig. 7
Fig. 8
1
Fig. 9
Column drive, adjusting the belt tension
5
Adjustment of the belt tension is required after replacement of the transmission or drive motor. Adjustment of belt tension is carried out using the supplied measurement device (1/Fig. 7). Adjustment is possible only when the complete drive is disassembled. For the column drive, the belt is comprised of several individual or one Poly-V belt. The adjustment procedure is the same.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 8 TD SD 21
SIREGRAPH CF Service
5-8
Column Drive Adjustment procedure
5
• Loosen the four screws (1/Fig. 8). • Place on the measuring aid as shown in (X/Fig. 8) and press down on it. Requirement: 1st ring must touch the top edge of the window. Correction: - By turning the adjustment screw (1/Fig. 9) - Turning counterclockwise: the belt is tightened. - Turning clockwise: the belt is loosened. - After adjustment is completed, tighten the screws (1/Fig. 8).
• Install drive. • Adjust potentiometer 3PT1 as described on Page 3.
NOTICE
SIREGRAPH CF Service
The 2nd ring is used only for factory adjustment of the measuring aid.
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 8 TD SD 21
Siemens AG Medical Engineering
Compression Drive
6
6-1
Control overview
6
I/O PC board 24S2
CPU PC board 24S1
Remote control console
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 8 TD SD 21
SIREGRAPH CF Service
6-2
Compression Drive
Fig. 2
Compression drive overview
6
Drive:
3MT6
Actuation:
24S2
Control:
23Az6
Fuse:
F3
10A/T
Limit switch:
3FC2 3FC3 3FC4
Cone above Cone inserted Cone force > 20 - 30N
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
23Az6
Page 2 of 8 TD SD 21
Siemens AG Medical Engineering
Compression Drive
6-3
Tomographic rod
A 1
f = 15 mm
420
Chain
F
Fig. 3
Chain tension
6
Auxiliary device:
Spring balance
Adjustment conditions:
Set unit at approx. 45°, exposure system approx. in the middle of travel range. Oblique projection max. towards foot end, spotfilm device max. towards head end.
Adjustment:
Loosen two Allen screws at the end of the tomographic rod (1/Fig. 3). Adjust the screw (A/Fig. 3) so that the following values are obtained: F = 7, ± 0.7 N f = 15 mm Tighten the two Allen screws (1/Fig. 3).
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 8 TD SD 21
SIREGRAPH CF Service
6-4
Compression Drive
P2
M
23AZ 6
F
Fig. 4
Fig. 5
Maximum compression force
6
Auxiliary device:
Spring balance and two cable loops
Adjustment conditions:
Set unit horizontally Set column vertically Push Stop button
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 8 TD SD 21
Siemens AG Medical Engineering
Compression Drive
6-5
C E
B D
A
Spring balance
Fig. 6
• Remove cover "A". • Attach spring balance as shown.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 8 TD SD 21
SIREGRAPH CF Service
6-6
Compression Drive
Electrical preadjustment:
NOTICE
Adjust potentiometer (P2/Fig. 5) to maximum value, turn it to the left.
The compression force is measured via the armature current of the compression motor 3MT6 and is directly proportional.
Mechanical adjustment:
Release EMERGENCY STOP Move compression down and read spring balance. Requirement: 200 N ± 30 Correction: Adjust slip clutch (Detail/Fig. 6) Bend the safety clip (C) out of the adjustment nut (B) Use the hook wrench to turn the adjustment nut Clockwise turning: force is increased Counterclockwise turning: force is reduced Secure adjustment screw.
Electrical adjustment:
Move compression down. Adjust the potentiometer (P2/Fig. 5) (turn clockwise) until the controller switches off the drive at a force of F = 150, +10N.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 8 TD SD 21
Siemens AG Medical Engineering
Compression Drive
6-7
Fd.
150mm
Fc.
Fig. 7
Adjustment of switch 3FC4
6
Auxiliary device:
Spring balance
Adjustment conditions:
Set unit horizontally Set column vertically
Adjustment:
maximum cone force Fc = 20, ± 2N Attach the spring scale as indicated (Fig. 7) and pull it with a force of Fd = 50N, switch 3FC4 triggers LED L68 on I/O 24S2.
Correction:
Set initial tension of the spring with the two horizontal set screws, (arrows) see detail.
Movement times
6
• The times are not adjustable. • Whole excursion range of the compression carriage: s = 630, ± 5 mm • Movement time in both directions with each unit position: t = 12 sec, ±1 sec
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 8 TD SD 21
SIREGRAPH CF Service
6-8
Compression Drive This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 8 TD SD 21
Siemens AG Medical Engineering
SID Drive
7-1
7
Control overview
7
Tableside control console 3S1 I/O PC board 24 S2
CPU PC board 24S1
Remote control console
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 6 TD SD 21
SIREGRAPH CF Service
7-2
SID Drive
Fig. 2
SID drive overview
7
Drive:
3MT7
Actuation:
24S2
Control:
23Az7
Potentiometer:
3PT7
Fuse:
F3
Safety limit switch:
3FC7
SIREGRAPH CF Service
10A/T
RXD3-209.071.01 Rev. 05 09.00
23Az7
Page 2 of 6 TD SD 21
Siemens AG Medical Engineering
7-3
0
175
350
SID Drive
1
1
3FC7
3PT7 3TP1 Fig. 3
Fig. 4
Adjustment or replacement of potentiometer 3PT7
7
• Set column vertically.
Set unit position to approx. 45° (see Notice).
Mechanical initial position:
Electronic initial position:
NOTICE
Siemens AG Medical Engineering
Adjust 175 mm (see Fig. 3). Disengage the motor gears. (Loosen screws (1/Fig. 4)). +5V, ±50 mV. CPU PC board 24S1, TP2/5 - TP2/13,14 0V.
At a table position of approx. 45 - 60° there is an approximate balance between gas spring force and telescopic tube weight. The required distances can be adjusted easily and the potentiometer can be easily replaced and adjusted.
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 6 TD SD 21
SIREGRAPH CF Service
7-4
SID Drive Set end positions
7
• Disengage the motor gear. Loosen the screws (1/Fig. 4). • Manually set the end position at 350 mm (SID = 150 cm) and hold it. • Select adjustment mode (see Page 2-1) - 24S2 I/O board, DW 1/8 to ON. - Set 042 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
• Manually set end position at 0 mm (SID = 115 cm). • Select adjustment mode (see Page 2-1) - 24S2 I/O board, DW 1/8 to ON - Set 041 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
• Reengage the motor gear without play and tighten the screws.
Adjustment of SID Limiting at 90°
7
SID programming is possible beginning with software version VB00D NOTICE
When shipped, SID limiting is not programmed.
• Select SID 115 cm. • Tilt unit up to 90°. • Measuring the distance from the safety limit switch bracket to the wall. • Program the SID accordingly while maintaining the safety distance of 50 cm. • Select the adjustment mode, see Page 2-1, (I/O board 24S2, DW 1/8 set to ON). Select the following code with keys P34/35:
• 191, no SID limiting • 192, SID limiting at 135 cm • 193, SID limiting at 115 cm • Save the particular code with the P33 key. • Deselect the adjustment mode, (I/O board 24S2, DW 1/8 set to OFF). The particular limiting is effective beginning at a unit position of 45°.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 6 TD SD 21
Siemens AG Medical Engineering
SID Drive
7-5
Focus 72 plane
X
3FC7 68 Film plane Fig. 5
Fig. 6
Safety limit switch 3FC7
7
• With safety limit switch actuated, please proceed as follows: • Mark the position of the switch and loosen it. • Reset switch until this is actuated by the trip cam. • Move out of the range with motor force. • Move safety limit switch to its initial position and fasten. Distance measurement:
Siemens AG Medical Engineering
SID = 115 cm
X = 1010 mm, ±5 mm
SID = 150 cm
X = 1360 mm, ±5 mm
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 6 TD SD 21
SIREGRAPH CF Service
7-6
SID Drive CN1
F1
F3
J3 P6 F2 P5
P3
P4
J2 J1
P1 P2 P3
BC1
23AZ7
L2 L1 Fig. 7
Adjustment of PC board 23Az7 CAUTION
Oscilloscope: Time: Amplitude: Nominal:
7
On this PC board, only potentiometer P3 may be readjusted. Maximum current after run-up time, approx. 6A. The potentiometer is located on PC board 23Az8. When the board is replaced, approx. the same adjustment as the P3 should be taken over. Probe 23 Az7 board, Chip U1, Pin 6, Ground ⊥ to J2 0.2 sec/cm 1 V/cm Set brake force at P1 (Fig. 7) so that braking takes place softly, however, the telescope must be held securely. SID movement with table vertical and horizontal must run smoothly.
Movement times
7
from 115 cm to 150 cm: 5.8, +1 sec, with all unit positions. from 150 cm to 115 cm: 5.8, +1 sec, with all unit positions.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 6 TD SD 21
Siemens AG Medical Engineering
Table Transverse Drive
8-1
8
Control overview
8
Tableside control panel 3S1 I/O PC board 24 S2
CPU PC board 24 S1
Remote control console
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 4 TD SD 21
SIREGRAPH CF Service
8-2
Table Transverse Drive
Mechanical limit stop Mechanischer Endanschlag Drive Antrieb 3MT8 LimitEndschalter switch 3FC8 3FC9 Spotfilm device Zielgerät
Counterweight Carriage Steel Seilrolle cable Gegengewichtswagen Seil Steel Seilrolle cable steel ca- pulley pulley ble
Fig. 2
Table transverse drive overview
8
Drive:
3MT8
Actuation:
24S2
Potentiometer:
23Az8
Control:
23Az8
Fuse:
F3
10A/T 23Az8
Switch:
3FC8 3FC9
(tower side) (doctor’s side)
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 4 TD SD 21
Siemens AG Medical Engineering
Table Transverse Drive
8-3
CN1
F1
F3
J3 P6 F2 P5
P3
P4
J2 J1
P1 P2 P3
BC1
23AZ8
L2 L1 Fig. 3
Electrical Adjustment CAUTION
8
On this PC board, only potentiometer P3 may be readjusted. Maximum current after run-up time, approx. 6A. The potentiometer is located on PC board 23Az8.
• Tabletop horizontal. • Actuate movement on doctor’s side Nominal: 180 N, +100 N Correction: - Potentiometer P3 on 23Az8
NOTICE
Siemens AG Medical Engineering
This force is adjusted at the factory; a device to hook on the spring balance is available for service purposes. When the board is replaced, approx. the same adjustment as the P3 should be taken over.
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 4 TD SD 21
SIREGRAPH CF Service
8-4
Table Transverse Drive This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 4 TD SD 21
Siemens AG Medical Engineering
Cassette Carriage Drive
9
Control overview
9-1 9
Tableside control panel 3S1 I/O PC board 24 S2
CPU PC board 24 S1
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 10 TD SD 21
SIREGRAPH CF Service
9-2
Cassette Carriage Drive
Tableside control panel
Fig. 2
Cassette carriage drive overview Drive:
3MT3
Tacho generator:
3DT3
Actuation:
24S2
Control:
22Az2
Potentiometer:
3PT3
Fuses:
F14 F15 F16 F
SIREGRAPH CF Service
4A/T 4A/T 4A/T 12A/T
RXD3-209.071.01 Rev. 05 09.00
9
22TR 1 22S10
Page 2 of 10 TD SD 21
Siemens AG Medical Engineering
Cassette Carriage Drive
9-3
3
1
4
2 A Fig. 3
Mechanical check
9
• Set unit in horizontal position. • Remove table top (see chap. 16-1). • Unscrew guide rail (1/Fig. 3). • Pull out cassette carriage maximally at doctor’s side. • Disconnect the cassette carriage from the chain (2/Fig. 3). • Move cassette carriage over the entire range. Required: easy movement and free of play. Correction: - Adjustment of the bearings on cassette carriage (drawing).
• Set unit in vertical position, column approx. middle of travel range. ➪ Adjust the head end (upper) bearings (3/Fig. 3) ➪ Adjust the foot end (lower) bearings (4/Fig. 3). Access for the screwdriver is only possible through the hole at (A/Fig. 3).
• It must be possible to still turn one roller manually. • Set unit in horizontal position. • Attach cassette carriage to the chain (2/Fig. 3). • Attach guide rail (1/Fig. 3). The adjustment takes place after the adjustment of the parking and loading position (see chap. 9-8).
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 10 TD SD 21
SIREGRAPH CF Service
9-4
Cassette Carriage Drive
F f F
f
860
Fig. 4
Fig. 5
Adjust tension of the drive chain
9
Auxiliary device:
Spring balance.
Adjustment conditions:
Set unit in vertical position.
Adjustment:
see Fig. 4
F = 20, ± 2N f = 15 mm
Adjust belt tension of the cassette drive Auxiliary device:
Spring balance
Adjustment conditions:
Set unit in vertical position
Adjustment:
see Fig. 5
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
9
F = 5, ± 0.5N f = 5 mm
Page 4 of 10 TD SD 21
Siemens AG Medical Engineering
Cassette Carriage Drive
9-5
F
A Fig. 6
Adjusting the slip clutch
9
Aids:
Spring balance 200 N and attachment parts
Preparations:
• • • •
Measurement:
Correction:
Cassette carriage without cassette in parking position Switch "Allarmax" off (plug in AMP3 / 24S2 I/O board) Press Emergency Stop button Attach spring balance and attachment parts as shown in page 9 - 6
• Unlock Emergency Stop button • Read spring balance Nominal: 170 N ± 20 N - Adjust the slip clutch (A/Fig. 6) + Page 9-6 - Bend the safety clip (D) out of the adjustment nut (C).
• Use the hook wrench to turn the adjustment nut Turn clockwise: force is increased Turn counterclockwise: force is reduced
• Secure the adjustment nut. Finally:
Siemens AG Medical Engineering
• Switch on "Allarmax", (pull out AMP3 on 24S2 I/O board)
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 10 TD SD 21
SIREGRAPH CF Service
9-6
Cassette Carriage Drive
C
D
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 10 TD SD 21
Siemens AG Medical Engineering
Cassette Carriage Drive
9-7
Loading position 3PT3
Cassette carriage Exposure position
Park position
X
A 198
C 193
B 584
Spotfilm device
Fig. 7
Fig. 8
Spotfilm device with cassette carriage
Adjustment or replacement of potentiometer 3PT3
9
• Set unit in vertical position. • Press EMERGENCY STOP. • Mechanical initial position:
Pos. C = exposure position C=193 mm (Fig. 8) cassette carriage front edge ==> front edge of the bottom sheet of the spotfilm device (X/Fig. 8).
• Electrical initial position:
Siemens AG Medical Engineering
+5V, ± 50 mV CPU-PC board 24S1, TP2/12 - TP2/13,14: 0V.
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 10 TD SD 21
SIREGRAPH CF Service
9-8
Cassette Carriage Drive
Loading position
Cassette carriage Exposure position
Park position
X
A 198
C 193
B 584
Spotfilm device
Fig. 9
Spotfilm device with cassette carriage
Set end positions
9
Auxiliary device:
Tape measure.
• Set distance: Parking position (B/Fig. 9) - Distance cassette carriage front edge ==> front edge of the bottom sheet of the spotfilm device (X/Fig. 9). B = 584 mm
Code = 102
• Select adjustment mode (see page 2-1). - I/O PC board 24S2, DW 1/8 to ON. - Enter code 102 and store.
• Stay in the adjustment mode • Set distance: Loading position (A/Fig. 9) - Distance cassette carriage front edge ==> front edge of the spotfilm device (X/Fig. 9). A = 198 mm
Code = 101
• Enter code 101 and store • Exit the adjustment mode - 24S2 I/O board, DW 1/8 to OFF.
• Release EMERGENCY STOP.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 10 TD SD 21
Siemens AG Medical Engineering
Cassette Carriage Drive
9-9
1
Fig. 10
Movement times
9
The times are not adjustable. Movement time parking position - exposure position t ≤ 0.9sec. Fast series Required: t ≤ 2sec - Cassette 35 cm x 43 cm (transverse). - 4 on 1 subdivision. - Without object.
Maximum opening of the cassette jaws
9
• Adjust the guide rail (1/Fig. 10) so that a 43 cm cassette can be inserted. It must be possible for the cassette carriage to reliably reach the loading position. The maximum tension jaw height opening is 467 mm, ± 1 mm. If the tension jaws are opened too soon, error 010 is signaled.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 9 of 10 TD SD 21
SIREGRAPH CF Service
9 - 10
Cassette Carriage Drive This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 10 of 10 TD SD 21
Siemens AG Medical Engineering
Format Sensing
10
10 - 1
Control overview
10
Tableside control panel 3S1
I/O PC board 24 S2
CPU PC board 24 S1
CPU PC board 24S1 EMERGENCY STOP
"Width"
"Height" Remote control console
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 6 TD SD 21
SIREGRAPH CF Service
10 - 2
Format Sensing
3PT11 "Height" "Höhe"
Spring Feder
Scissors ik Scherenmechan mechanism
Loading Ladeposition
3PT10 "Breite" "Width"
position
Fig. 2
Overview of the format sensing drive
10
Color coding on Potentiometers 3PT10 and 3PT11
M (2) Potentiometers on the connection side: M (2) = middle L.D. (3) = left side down L.D. (3)
R.D. (1)
R.D. (1) = right side down
3PT10 Middle (2): Left side down (3): Right side down (1):
SIREGRAPH CF Service
3PT11 green white brown
Middle (2): Left side down (3): Right side down (1):
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 6 TD SD 21
green brown white
Siemens AG Medical Engineering
Format Sensing
10 - 3
Fig. 3
Adjustment or replacement of potentiometers 3PT10, 3PT11
10
Remove table top and cover of spotfilm device (see chap.16 and chap.17).
NOTICE
Potentiometers 3PT10 and 3PT11 are heat-sensitive!
Cassette jaws
• Press EMERGENCY STOP. Mechanical initial position (Fig. 3) - Cassette jaw width, potentiometer 3PT10, Spacing of the cassette jaws C = 332 mm (spacer gauge, see page 10 - 4). Electrical initial position = + 5 V, ± 50 mV. 24S1 CPU board TP2/10 - TP2/13,14: 0V. - Cassette jaw height, potentiometer 3PT11, Spacing of the cassette jaws C = 332 mm (space gauge, see page 10 - 4). Electrical initial position = + 5 V, ± 50 mV. 24S1 CPU board TP2/11 - TP2/13,14: 0V.
• Release EMERGENCY STOP.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 6 TD SD 21
SIREGRAPH CF Service
10 - 4
Format Sensing
Park position
B
Loading position
Park position
A
D C B
Spacer Gauge
Loading position
A C E
10
A = 471 mm
D = 197 mm C = 332 mm
The spacer gauge makes it easier to mechanically adjust the cassette jaw spacing. It must be placed between the cassette jaws so that it is squared respectively straight. If this is not done correctly, it can snap together very easily.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 6 TD SD 21
Siemens AG Medical Engineering
Format Sensing
10 - 5
Adjusting the end positions and the format sensing Auxiliary equipment:
10
Spacer gauge (see page 10 - 4)
• Move cassette carriage to loading position and adjust the cassette jaws as follows: Width A = 471, ±1 mm (spacer gauge)
• Select adjustment mode (see Page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Enter code 082 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
• Move cassette carriage to loading position and adjust the cassette jaws as follows: Height B = 467, ±1 mm (cassette carriage is empty, maximum opening of the jaws)
• Select adjustment mode (see Page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Enter code 092 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
• Move cassette carriage to parking position and adjust the cassette jaws as follows: Height D = 197, ±1.5 mm (spacer gauge)
• Select adjustment mode (see Page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Enter code 091 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
• Move cassette carriage to parking position and adjust the cassette jaws as follows: Width E = 193, ±1.5 mm (cassette carriage empty, mechanical stop in park position)
• Select adjustment mode (see Page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Enter code 081 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 6 TD SD 21
SIREGRAPH CF Service
10 - 6
Format Sensing This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 6 TD SD 21
Siemens AG Medical Engineering
Diaphragm Plate Drive
11
11 - 1
Control overview
11
I/O PC board 24S2
CPU PC board 24S1
CP9
PC board 22 S1
EMERGENCY STOP Collimator control
Collimator workstation
diaphragm plates
Remote control console
3PT5
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 8 TD SD 21
SIREGRAPH CF Service
11 - 2
Diaphragm Plate Drive
Diaphragm plates Blendenplatten ballKugellager bearing chain Kette
3MT5
chain Kette
Bedienfeld control panel
3PT5
chain Kette
Brass conMessingnecting rod Schubstange guide Führung
Fig. 2
Diaphragm plate drive overview
11
Drive:
3MT5
Actuation:
24S2
Control:
23Az5
Potentiometer:
3PT5
Fuse:
F3
10A/T
23Az5
F17 F18 F19
6A/T 6A/T 6A/T
22TR1
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 8 TD SD 21
Siemens AG Medical Engineering
Diaphragm Plate Drive
11 - 3
7
3
A
C
Head end
B
D
5
Foot end
4
2 6 1
7 Fig. 3
Fig. 4
Mechanical check
11
Remove table top and spotfilm device cover (see chap.16 and chap.17). Auxiliary device:
Tape measure.
Adjustment conditions:
Manually shift diaphragm plates evenly and free of play.
Adjustment:
Adjustment of the bearings. Have the rails been greased? (Longtime PD2) Have the chains been greased? (Longtime PD2)
Symmetry and centering
11
Auxiliary device:
Tape measure, flat spanner 8 mm (in the service kit), put up a wrap wire or thread in center of spotfilm device and tighten.
Adjustment conditions:
A = B = C = D (Fig. 3)
Adjustment:
Start to adjust one diaphragm plate (1/Fig. 4): loosen the screws (6/Fig. 4), open the eccenter (5/Fig. 4), adjust parallelism using the eccenter (4/Fig. 4). Then adjust the eccenter (5/Fig. 4), at least one bearing must be turnable. Tighten the screws (6/Fig. 4). Proceed with the second diaphragm plate in the same manner. The bearings (7/Fig. 4) are no eccenters.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 8 TD SD 21
SIREGRAPH CF Service
11 - 4
Diaphragm Plate Drive Chain tension
11
415
Auxiliary device:
A
A
Spring balance F
Adjustment conditions: Open diaphragm plates (A/Fig. 5) to dimension 415 mm, ±1 mm.
f
Adjustment:
560
Adjust chain tension. F = 15 ± 8 N f = 15 mm Fig. 5
Adjust the slip clutch
11
Aids:
Spring balance 200N and attachment parts
Preparations:
Cassette carriage without cassette in Park Position. Switch off "Allarmax" (plug in AMP3 on 24S2 I/O board) Press the Emergency Stop button
• Attach the spring balance and attachment parts at "B" and at "A" to the diaphragm plates as shown in the drawing on page 11-5 and (Fig. 6). - Unlock the Emergency Stop button. - Select Bucky mode, select full format - Read spring balance Nominal: 40, ± 10 N Correction: - Bend locking tab D out from adjusting nut C - Adjust adjusting nut C with hook wrench clockwise: force increases counterclockwise: force decreases - Press a locking tab D into adjusting nut C. Switch on "Allarmax" ! (pull out AMP3 on 24S2 I/O board !)
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 8 TD SD 21
Siemens AG Medical Engineering
Diaphragm Plate Drive
11 - 5
spring balance
section A-A C
D
Mount this device under diaphragm plate with open side facing toward foot end.
Chain
Diaphragm plate, wall side Foot end
Head end Diaphragm plate, Doctor’s side
Mount this device under diaphragm plate by clamping to bar attached to underside of plate. Spotfilm device Fig. 6
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 8 TD SD 21
SIREGRAPH CF Service
11 - 6
Diaphragm Plate Drive
Central beam center
423 418
A
215 C 38 B
Fig. 7
Adjustment and replacement of potentiometer 3PT5
11
• Press EMERGENCY STOP. Auxiliary device: Tape measure. Mechanical initial position: 215 mm, +/- 1 mm (C/Fig. 7). Electronic initial position: + 5 V, ± 50 mV, 24S1 CPU board, TP2/9 - TP2/13,14 0V. Set end positions Auxiliary device: Tape measure. Set minimum diaphragm opening B = 38 mm, +/- 1mm (B/Fig. 7)
• Select adjustment mode (see page 2-1). - 24S2 I/O board, DW 1/8 to ON. - Set 071 and store.
• Exit the adjustment mode. - I/O PC board 24S2, DW 1/8 to OFF. Set maximum diaphragm opening A = 418 mm, +/- 1 mm (A/Fig. 7)
• Select adjustment mode (see Page 2-1). - Set 072 and store.
• Exit the adjustment mode. - 24S2 I/O board, DW 1/8 to OFF
• Release EMERGENCY STOP.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 8 TD SD 21
Siemens AG Medical Engineering
Diaphragm Plate Drive
11 - 7
Adjustment of gap width, general remarks
11
The factory adjusted spacing width on subdivided exposures is approx. 3 mm. For the most part the spacing width depends on the positioning accuracy of the cassette carriage, for both single exposures as well as series exposures. If the customer wishes to have either a wider or narrower spacing, this can be adjusted electronically. The same spacing width on exposures with 3 or 4 subdivisions cannot be obtained due to insufficient positioning accuracy of the cassette carriage.
• Select adjustment mode I/O PC board 24S2, DW 1/8 to ON Code 184 to 181 (reduction in steps of 0.2 mm each) Code 185 (no software correction) Code 186 to 189 (increase in steps of 0.2 mm each)
• Enter the desired code and store. • Exit the adjustment mode I/O PC board 24S2, DW 1/8 to OFF
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 7 of 8 TD SD 21
SIREGRAPH CF Service
11 - 8
Diaphragm Plate Drive Adjustment of gap width (subdivision mode)
11
• Insert 18 cm x 24 cm cassette transversely • Select 3 on 1 subdivision Measure diaphragm opening (1st measurement)
• Select 4 on 1 subdivision and 3 on 1 subdivision again, press "Preparation" briefly Measure diaphragm opening (2nd measurement) Nominal: difference from 1st and 2nd measurement < 1 mm Correction: - Potentiometer P1 on 23AZ5 (/Fig. 8), brake current. CN1
F1
F3
J3 P6 F2 P5 P4
J2 J1
P1 P2 P3
BC1
L2 L1
Fig. 8
• Insert 35 cm x 43 cm cassette transversely • Select 3 on 1 subdivision Measure diaphragm opening (1st measurement)
• Select 4 on 1 subdivision and 3 on 1 subdivision again, press "Preparation" briefly Measure diaphragm opening (2nd measurement) Nominal: difference from 1st and 2nd measurement < 1 mm Repeat process with 18 cm x 24 cm cassette until the difference is < 1 mm (optimization).
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 8 of 8 TD SD 21
Siemens AG Medical Engineering
Collimator
12 - 1
12
Control overview
12
I/O PC board 24S2
CPU PC board 24S1
Remote control console Collimator control 24S3 Height
Width
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 4 TD SD 21
SIREGRAPH CF Service
12 - 2
Collimator
Fig. 2
Drive overview
12
Drive:
MT1 MT2
Actuation:
24S2
Control:
23S3
Fuse:
F1/F2
1A
Potentiometer:
POT1 POT2
(Height) (Width)
SIREGRAPH CF Service
(Height) (Width)
RXD3-209.071.01 Rev. 05 09.00
23S3
Page 2 of 4 TD SD 21
Siemens AG Medical Engineering
Collimator
12 - 3
Adjustment after replacement of the potentiometers
12
Mechanical initial position for the width: Width collimator plates closed all the way
• Before installing, set the potentiometer to 900 Ohm (Ralco), 100 Ohm (Gorla), between "CW" and "wiper"
• Take note of the direction of turn: when the width collimator opens the resistance value must increase.
• On the CPU board 24S1, measure between TP2/13,14 = 0V and TP2/7: 1,45 V (Ralco), 700 mV (Gorla). Mechanical initial position for the height: Height collimator closed completely
• Before installation, set the potentiometer to 900 Ohm (Ralco), 100 Ohm (Gorla), between "CW" and "wiper"
• Take note of the turn direction: when the height collimator opens the resistance value must increase.
• On CPU board 24S1, measure between TP2/13, 14 = 0V and TP2/8: 1,45 V (Ralco), 700 mV (Gorla).
Adjusting the end positions
12
The procedure described here is repeated with each code listed in the table below.
• Select adjustment mode (see Page 2-1) - I/O PC board 24S2, DW 1/8 to ON
• Enter the code and store. Width collimator closed completely: Height collimator closed completely: Width collimator opened completely: Height collimator opened completely:
Code = 051 Code = 061 Code = 052 Code = 062
• Exit the adjustment mode - 24S2 I/O board, DW 1/8 to OFF.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 4 TD SD 21
SIREGRAPH CF Service
12 - 4
Collimator Correction Factors for Collimator
12
With an increase of the values in a unit’s place the collimation is increased. Example: Correction of the cassette width The current value is 125. If the width is to be lowered, the value 125 is to be reduced to values between 121 and 124 with the help of the P31/32 keys. If the width is to be raised, the value 125 is to be increased to values between 126 and 129 with the help of the P31/32 keys.
• Select adjustment mode (see Page 2-1) - I/O PC board 24S2, DW 1/8 to ON - Enter the desired code (see table below) and store
Display
Adjustment
Storage
H
Z
E
H
Z
E
Cassette width
1
2
1-9
P34/35
P34/35
P31/32
P33
Cassette height
1
3
1-9
P34/35
P34/35
P31/32
P33
I.I. width
1
4
1-9
P34/35
P34/35
P31/32
P33
I.I. height
1
5
1-9
P34/35
P34/35
P31/32
P33
• After a storage “0” appears in the unit’s place. • Exit the adjustment mode - 24S2 I/O board, DW 1/8 to OFF.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 4 TD SD 21
Siemens AG Medical Engineering
Grid
13
Control overview
13 - 1 13
I/O PC board 24S2 CPU PC board 24S1
3FC10 or 3XS2
Collimator Collimator control
Remote control console
Collimator workstation
Fig. 1
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 4 TD SD 21
SIREGRAPH CF Service
13 - 2
Grid Fastening Befestigung Raster Grid
Frame Rahmen SOL2 Hubmagnet lift magnet
The oscillating grid is in the parking position. It is pressed there by lift magnet SOL2 against the spring force of the leaf springs.
SOL1 lift magnet 3FC10 Hubmagnet 3XS2
Leaf springs Blattfedern Fig. 2
Oscillation grid, general remarks
13
The grid is actuated by means of magnet SOL1. Magnet SOL2 keeps the grid in a central position for DFR exposures. The microswitch 3FC10 / or proximity switch 3XS2 releases the exposure. Grid movement range approx. 15, ± 1 mm, minimum oscillation time > 30 sec. Control: 24S2. The grid is centered mechanically via the tapped guide of the SOL 2 magnet.
NOTICE:
The ohmic resistance of the SOL 1 coil is less than that of SOL 2.
Programming of oscillation grid / removable multi-line grid
13
• Select the adjustment mode (see Page 2 - 1) I/O board 24S2, DW 1/8 set to ON Enter the code: 221 normal oscillation grid 222 removable multi-line grid
• Save the code that was entered. • Exit the adjustment mode I/O board 24S2, DW 1/8 set to OFF
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 4 TD SD 21
Siemens AG Medical Engineering
Grid
13 - 3 Checking the grid centering
13
• Table in 0° position • Tension thread / wire in center of spotfilm device Nominal: Grid center +/- 1 mm from the thread / wire Correction: - Correct the position of the push rod guide
Checking the grid movement
13
• Table in 0° position • Unlock grid by hand and keep it unlocked. • Press grid by hand into one of the end stops and release it. Nominal: at least five oscillations Correction: - none, check the condition of the springs (breaks / installation)
Grid contact
13
Mechanical grid contact 3FC10
• The contact must switch safely at the first oscillation, 24S2 / LED 57 (green) ON. Correction: - Correct position of the switch Electronic grid contact 3XS2
• Unlock the grid by hand and press grid against stop on grid contact side. Nominal: Grid frame - grid contact distance = 0.5 mm Correction: - Correct position of the switching element.
• The grid contact must switch safely at a distance < 5 mm, LED (red) at the grid contact or 24S2 / LED 57 (green) ON.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 4 TD SD 21
SIREGRAPH CF Service
13 - 4
Grid This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 4 TD SD 21
Siemens AG Medical Engineering
Tomography
14 - 1
14
Column (head - foot)
Tacho generator Column (foot - head)
1.3 sec.
Fig. 1
Setting the tomographic height, beginning with firmware version 14 VB00D Setting the column speed
14
• Connect and set the oscilloscope CH1 Column motor control 22AZ3
C1 11 (tacho signal) C1 12 (ground) Time base: 0.2 cm/sec, Trigger internal, Level 1, Slope + The speed of the column is shown on channel 1, the signal is positive or negative according to the movement direction. The radiation duration is shown on channel 2. Setting conditions:
• Tomography 40° • Tomographic height 200 mm • Exposure system in table center • Switch off "Allarmax" (plug in AMP3 on 24S2) • Release tomographic exposure without radiation • Measure travel time of column from head to foot (diagram) Required: 1.3 sec. (+ 0.05 sec/ - 0.0 sec) (Fig. 1) Correction: Potentiometer P2 on SP 22AZ3 (Pick-a-back board, max. Speed) clockwise: counterclockwise:
speed is increased speed is decreased
• Switch on "Allarmax", (remove AMP3 on 24S2) !
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 6 TD SD 21
SIREGRAPH CF Service
14 - 2
Tomography Setting the spotfilm device speed
14
It is possible to set a tomographic height correction digitally with this software version. The starting value is Code 175 and the correction can be made in 1 mm steps downwards with Code 174 to 171 and in 1 mm steps upwards with Code 176 to 179. A prerequisite is that the code 175 is programmed before the speed of the spotfilm device is adjusted. Otherwise the correct tomographic height can not be set.
• Select adjustment mode, I/O PC board 24S2, DW 1/8 to ON, (see page 2 - 1) • Verify if code 175 is programmed, search with keys P34 and P35 only • If not, enter code 175 and store • Exit adjustment mode, I/O PC board 24S2, DW 1/8 to OFF • Connect multimeter:
NOTICE
24S1, TP2 - 4 (+) 24S1, TP2 - 14 (0V)
The multimeter must be able to display three digits after the decimal point.
• At the generator, set "SS" to off • Switch off "Allarmax" (plug in AMP3 on 24S2) • Tomography 40° • Tomographic height 200 mm • Exposure system in table center • Press prep-button and hold until the column has reached its head-end start position and the multimeter displays a stable value. Read and note down that value (value A, wiper voltage of potentiometer 3PT2). Trigger a planigraphic exposure without radiation by pressing prep- and main-button. Column travels from head end to foot end. Release prep- and main-button and subsequently press both again thus triggering another planigraphic exposure allowing the column to travel back from foot end to head end. Attention: keep prep- and mainbutton depressed until the column has reached its head-end position and the multimeter displays a stable value (value B).
• Determine the difference from both measured values A and B Required: ± 5 mV Correction: Potentiometer P2 on SP 22AZ4 (Pick-a-back board, max. Speed) clockwise: counterclockwise:
if A > B if A < B
Repeat the process until the required ± 5 mV are reached.
• Switch on "Allarmax", (pull out AMP3 on 24S2) !
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 6 TD SD 21
Siemens AG Medical Engineering
Tomography
14 - 3
Software correction of the layer-height
14
• Perform tomographic exposure 40° and tomographic height 200 mm with tomographic test.
• Determine the difference of the tomographic height. • Select adjustment mode - I/O PC board 24S2, DW 1/8 to ON Code 174 to 171 (decreasing in 1 mm steps each) Code 175 (no software correction) Code 176 to 179 (increasing in 1 mm steps each)
• Enter the desired code from the above table and store • Exit the adjustment mode - I/O PC board 24S2, DW 1/8 to OFF
• Take new tomographic exposure under the same conditions. • If there is still a difference, make further corrections. • In systems with FLUOROSPOT take an indirect tomography control exposure.
NOTICE
The setting made at 40° / 200 mm applies for all tomographic angles, tomographic heights and operating modes.
• Make sure that "Allarmax" is switched back on, (jumper AMP3 on 24S2 removed) !
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 6 TD SD 21
SIREGRAPH CF Service
14 - 4
Tomography
kV- act t1
t2
Column (head - foot) Säule (KopfFuß)
Tacho generator Column (foot - head) Säule (Kopf- Fuß)
Radiation
Fig. 2
Setting the tomographic symmetry
14
• Connect and adjust the oscilloscope CH1 motor control column 22AZ3
C1 11 (tacho signal)
20V/div/40° 10V/div/20° 5V/div/ 8°
D100.X18 D100.X4 (ground)
500 mV/div
CH2 kV act generator POLYDOROS LX
D211 Σ kV-ACT D211 (ground) Time base: 0.2 cm/sec, Trigger internal, Level 1, Slope + POLYDOROS 50/80 SX
500 mV/div.
The column speed is displayed on channel 1; depending on the movement direction, the signal is positive or negative; the radiation time is displayed on channel 2. Adjustment conditions:
• Select 40 kV, minimum mAs on the generator. • Insert cassette, 4 on 1 subdivision, close collimator. • Select tomography and trigger exposure
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 6 TD SD 21
Siemens AG Medical Engineering
Tomography
14 - 5
Adjustment: (direct radiography) Required:
Radiation (CH2) symmetrical to column speed (CH1),i.e. t1 = t2 (Fig. 2).
Correction:
Potentiometer on I/O PC board 24S2. P2 for 40°, P4 for 20°, P6 for 8°.
NOTICE
P2 P4
turning these clockwise delays radiation release.
P6 Adjustment: (indirect radiography) Required:
Radiation (CH2) symmetrical to column speed (CH1),i.e. t1 = t2 (Fig. 2).
Correction:
Potentiometer on the I/O-PC board 24S2. P1 for 40°, P3 for 20°, P5 for 8°.
NOTICE
P1 P3
turning these clockwise delays radiation release.
P5
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 5 of 6 TD SD 21
SIREGRAPH CF Service
14 - 6
Tomography This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 6 of 6 TD SD 21
Siemens AG Medical Engineering
Radiation Geometry
15 - 1
15
Length Marker
Side marker
Fig. 1
Coincidence of light field/radiation field
15
• Move the unit into the 0° position. - SID = 115 cm. - Full format and Bucky
• At the generator, set - bed-side exposure - select large focus.
Exposure position
15
• Using the setting buttons on the collimator, starting from the maximum format, set a 24 cm x18 cm rectangular format on the collimator scale for SID = 115.
• Switch on light localizer and mark light field edges with radio-opaque material, e.g. washer or coins, attach a fifth marker as a side marking.
• Select approx. 50kV/2mAs and trigger an exposure.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 4 TD SD 21
SIREGRAPH CF Service
15 - 2
Radiation Geometry
Y1
Film Light field
X2
X1 Radiation field
Y2 Fig. 2
Evaluation
15
• On all four sides, measure the difference between the light field edges and the radiation field edges and calculate the total difference in the X- and Y-directions.
∑X=X1+X2 ∑Y=Y1+Y2
• If the tolerance is exceeded by ≥ 1.8%, adjust the collimator at the tube assembly flange so that light field and radiation field are within the tolerance.
• Repeat the exposure.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 4 TD SD 21
Siemens AG Medical Engineering
Radiation Geometry
15 - 3
1
∆ ∆
Fig. 3
Adjustment of the central beam
15
Horizontal table direction
15
• Set unit to 0° • Stretch wrap-wire over the notches (spotfilm device center). • SID = 135 cm. • Loosen the X-ray tube support cover (1/Fig. 3). • Switch on light localizer. • Turn the X-ray tube and adjust central beam center. • Deviation = 0 mm. • Tighten the X-ray tube support cover (1/Fig. 3). • Check: verify the difference at SID = 150 cm and SID= 115 cm. Maximum permissible difference = ± 2% from SID.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 3 of 4 TD SD 21
SIREGRAPH CF Service
15 - 4
Radiation Geometry
∆
1
1
Fig. 4
Fig. 5
Vertical table direction
15
Verify the difference in the vertical table direction by changing the SIDs and adjust by means of 2 Allen screws and 2 set screws (1/Fig. 4). Maximum permissible difference = ± 2% from SID. Check exposures:
• Insert a cassette that is marked so that the center and sides can be recognized. • Trigger exposure. • See table of contents, RXD3-209.034. . . .(SG CF Startup Instructions)
Adjust 0 degree switch 3FC1
15
• Set unit into 90° position. • Loosen Allen screw (1/Fig. 5) and adjust the disk so that the trip lever engages in the notch.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 4 of 4 TD SD 21
Siemens AG Medical Engineering
Installing/Removing the Tabletop
16
16 - 1
2
1
3
3TP1 Fig. 1
Removing the tabletop
16
• Disconnect the wires 5 and 7 on the terminal strip 3TB1 (back of column) (Fig. 1). NOTICE
If the 24S2.AMP1 connector is unplugged instead of leads 5 and 7, the compression carriage will move up after approx. 2 sec. and the slip clutch will respond.
• Remove the covers (2/Fig. 1). • Remove the ground wire and the end stop (3/Fig. 1). • Move the tabletop at doctor’s side until you can remove it.
Installing the tabletop
16
• Put the tabletop in position carefully and push it evenly into the guide until it reaches the mechanical stop. Pay special attention to the limit switches 3FC8 and 3FC9 (1/Fig. 1) to make sure they do not get bent or damaged.
• Now move the tabletop further in with motor force. • Attach covers (2/Fig. 1). • Restore the connections at the terminal strip 3TB1. • Reattach end stop and ground wire (3/Fig. 1).
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 2 TD SD 21
SIREGRAPH CF Service
16 - 2
Installing/Removing the Tabletop This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 2 TD SD 21
Siemens AG Medical Engineering
Installing/Remov. the Cover of the Spotfilm Dev.
17
17 - 1
a b1
1
b
b B A
b 1 a b1
Fig. 1
b
Fig. 2
Removing the upper cover of the spotfilm device
17
• Installation and removal with unit in 0° position • Loosen the 8 Allen connection screws and the cover panel ground wire (1/Fig. 1).
Removing the lower cover of the spotfilm device
17
• Installation and removal with unit in 90° position • During work on the cassette carriage motor and the potentiometer, remove the cover of the spotfilm device (A/Fig. 2). To do so, loosen screws a and b1.
• During adjustment of the cassette carriage bearings, remove the cover of the spotfilm device (B/Fig. 2). To do so, loosen screws b and b1.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 2 TD SD 21
SIREGRAPH CF Service
17 - 2
Installing/Remov. the Cover of the Spotfilm Dev. This page intentionally left blank.
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 2 TD SD 21
Siemens AG Medical Engineering
Changes to Previous Version
18
18 - 1
- Chapter 1-1, "Self-learn Package on CD-Rom" inserted. - Chapter 1-3, section regarding "Replacing and Adjustment Potentiometers" inserted. - Chapter 1-4, remark regarding "damaged or worn parts" as well as "Manufacturer’s Note" added - Chapter 1-6, Text corrected for "L23" - Chapter 1-7, Text corrected for "L37" - Chapter 1-8, Text corrected for "L51, L52, L57" - Chapter 1-12, Fig. 8 revised and Fig. 9 updated. - Chapter 1-20, designations of potentiometers, motors and boards corrected. - Chapter 1-23, designations of motors and boards corrected. - Chapter 1-24, section regarding "DIP switch DW1" inserted. - Chapter 1-26, remark regarding Code 201-209 and Code 211-213 inserted. - Chapter 1-27, 1-28, 1-29, "Listing of All Programming Codes" inserted. - Chapter 3-1, "Overview of Control" moved above Fig. 1. - Chapter 3-2, "Tightening screw" inserted into Fig. 2. - Chapter 3-3, Text rewritten. - Chapter 3-4, Text rewritten and section "Replacing the Brakes" inserted. - Chapter 3-5, Fig. 8 inserted - Chapter 3-6, Fig. 10 inserted. - Chapter 3-9, text rewritten and expanded. - Chapter 4-1, "Overview of Control" moved above Fig. 1. - Chapter 4-4, "Reference point....." and "Potentiometer play....." inserted. - Chapter 5-1, "Overview of Control" moved above Fig. 1. - Chapter 5-3, "Reference Point: Outside Edge of the Table Frame" inserted. - Chapter 6-1, "Overview of Control" moved above Fig. 1. - Chapter 6-4, Numbering of illustrations updated. - Chapter 7-1, "Overview of Control" moved above Fig. 1. - Chapter 7-4, section "Adjustment of the SID Limit at 90°" inserted. - Chapter 7-6, section "Adjustment of the 23Az7 Board" rewritten. - Chapter 8-1, "Overview of Control" moved above Fig. 1. - Chapter 8-3, section "Electrical Adjustment", text rewritten. - Chapter 9-1, "Overview of Control" moved above Fig. 1. - Chapter 9-5, section "Adjusting the Slip Clutch", text rewritten. - Chapter 9-6, illustration expanded. - Chapter 9-7, Fig. 8 revised. - Chapter 9-8, Fig. 9 revised. - Chapter 9-9, section "Maximum Tension Jaw Opening", text rewritten. - Chapter 10-1, "Overview of Control" above Fig. 1 moved. - Chapter 10-2, Fig. 2 revised, section "Overview of Drives for Format Sensing", text rewritten. - Chapter 10-3, remark regarding potentiometers inserted. - Chapter 10-4, illustrations revised. - Chapter 11-1, "Overview of Control" moved above Fig. 1, Fig. 1 revised. - Chapter 11-2, typographical error corrected from "3PT7" to "3PT5". - Chapter 11-3, section "Symmetry and Centering", "in the service kit" inserted. - Chapter 11-4, Fig. 5 and text arranged differently, section "Adjusting the Slip Clutch" from Chapter 11-5 rewritten and inserted.
Siemens AG Medical Engineering
RXD3-209.071.01 Rev. 05 09.00
Page 1 of 2 TD SD 21
SIREGRAPH CF Service
18 - 2
Changes to Previous Version - Chapter 11-5, Fig. 6 added, Section"Adjusting the Slip Clutch" moved to after Chapter 11-4. - Chapter 11-6, Fig. 6 revised, text updated. - Chapter 11-7, section "Adjustment of the Gap Width, General Remarks" from Chapter 11-8 inserted. - Chapter 11-8, section "Gap Width Correction with Subdivisions" and illustration from Chapter 11-7 inserted and rewritten. - Chapter 12-1, "Overview of Control" moved above Fig. 1. - Chapter 12-3, section "Adjustments Following Replacement of Potentiometers", text rewritten. - Chapter 12-4, Text arranged differently. - Chapter 13-1, "Overview of Control" moved above Fig. 1. - Chapter 13-2, section "Programming the Oscillation Grid" added. - Chapter 14-1, text rewritten - Chapter 16-1, text rewritten.
TD PS 21 / Schmidt, Tropia
SIREGRAPH CF Service
RXD3-209.071.01 Rev. 05 09.00
Page 2 of 2 TD SD 21