PQ TechWatch A product of the EPRI Power Quality Knowledge program
Power Quality for
Healthcare Facilities December 2007
Philip Keebler, EPRI
CONTENTS Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . .1 The Healthcare Environment . . . . . . . . . . . .1 Power Quality in Healthcare Facilities . . . . .3 Recognizing Power Quality Problems . . . . . .4 Symptoms and Their Causes . . . . . . . . . . . .4 Sources of Electrical Disturbances . . . . . . .8 Improving Power Quality in the Healthcare Environment . . . . . . . . . . . . . . . .13 Meeting the Power Quality Challenges of the Healthcare Industry . . . . . . . . . . . . .13 Establishing Partnerships . . . . . . . . . . . . . .13 Creating a Power Quality Checklist for Procuring Equipment . . . . . . . . . . . . . . . . .14 Using Power-Conditioning Devices to Improve Equipment Compatibility . . . . . . .16 Understanding Facility Voltage Requirements, Grounding, and Dedicated Circuits . . . . . . . . . . . . . . . . . . .17 Conclusion . . . . . . . . . . . . . . . . . . . . . . . . . . .24 Bibliography . . . . . . . . . . . . . . . . . . . . . . . . . .24
EXECUTIVE SUMMARY
The healthcare environment is made up of perhaps the most unusual combination of electronic loads found in any facility. Healthcare facilities not only rely upon commercial loads (such as computers, servers, and lighting system) and industrial loads (such as food preparation equipment, laundry equipment, medical gas systems, but also rely on electronic medical loads (that is, medical equipment) to operate the facility and provide patient care services. As in other facilities, when an electrical disturbance such as a voltage sag, voltage transient, or voltage swell reaches the service entrance of the healthcare facility or medical location, computers in the accounting department may shut down, and motor starters and contactors providing power to the air-conditioning and ventilation system may change the environment within the facility. Unlike other places, however, a patient’s life could be threatened when an aortic balloon pump trips off-line during a cardiovascular surgery. The costs associated with downtime can be staggering, but no bounded cost can be placed on the irreversible result of loosing a patient. Building, electrical, and healthcare codes in the United States require that hospitals and other medical clinics have emergency power ready to activate upon the detection of a power quality problem and assume the load within 10 seconds of the detection. However, even though a generator may be used at a healthcare facility or medical location, it cannot be on-line to support critical medical equipment with an activated transfer switch in less than about 2 to 3 seconds at best. This duration of time might as well be forever in terms of the ability of electronic medical equipment to continue operating. In fact, an undervoltage as short as ¼ of a cycle (about 4 milliseconds) is often sufficient to confuse sensitive electronic devices. This PQ TechWatch will introduce the typical problems found in healthcare facilities, enlighten the reader on some new issues, and provide practical guidelines for avoiding those problems.
About the EPRI Power Quality Knowledge Program The EPRI Power Quality Knowledge program provides a wealth of resources in well-designed, readable, and accessible formats. Paramount among these resources are documents covering a wide range of PQ topics, written not only for use by busy PQ professionals, but also to be shared with important end-use customers and internal utility managers. The program’s website, www.mypq.net, is the most comprehensive electronic PQ resource available, providing 24-7 access to proven expertise via the PQ Hotline, hundreds of PQ case studies, over 200 PQ technical documents, PQ standards references, indexes, conference presentations, and a wealth of other resources. For more information, please visit www.mypq.net.
Copyright 2007, EPRI (www.epri.com). All rights reserved. Distribution of PQ TechWatch is subject to a license agreement with EPRI. No portion of this report may be reproduced or redistributed in any form, including electronic copies, without prior written consent from EPRI. Distribution outside the licensed organization is expressly forbidden. Product and company names mentioned in this document may be trademarks of their respective companies. Mention of third-party products is for informational purposes only and constitutes neither a recommendation nor an endorsement. EPRI and the EPRI logo are trademarks of EPRI.
ii
Power Quality for Healthcare Facilities
Healthcare providers have little time to be concerned with the quality of power or to find a reliable source of power to operate their equipment.
INTRODUCTION
Although the electricity provided to a healthcare facility or medical location is an absolute necessity for healthcare providers to operate their facilities, it is usually not given a lot of thought. The widespread growth of new and lingering illnesses and diseases, the call for increasingly critical emergency services, and the pressure to reduce healthcare costs force healthcare providers to keep their minds on their business—caring for their patients, enlisting the best possible healthcare professionals, and purchasing and installing the best medical equipment that money can buy. Turning on a heart-lung bypass machine prior to a six-hour openheart surgery where the operating room
problem to cripple the emergency medical staff. A second CT machine may not be an option, and the nearest machine may be many miles away in another hospital. This mission-critical imaging system could be taken off-line by a minor voltage sag to 80% of nominal (i.e., a 20% sag), lasting for only three 60-hertz cycles (50 milliseconds). The U.S. power quality community has estimated that $10 billion is lost yearly when automated control systems in industrial plants are upset by voltage sag events. Such numbers have not been estimated specifically for healthcare facilities or providers, but one can assume that the cost of downtime will also include possibly placing one or more patients at risk.
lights “are always on” has become as routine as activating a medical gas supply of oxygen
The Healthcare Environment
for a patient and then adjusting the flow rate
The healthcare environment in the United
so the patient receives the desired amount of
States is in continual transition in efforts to
oxygen. Healthcare providers have little time
improve patient care. Aside from the
to be concerned with the quality of power or
practice of medicine, nursing, and other
to find a reliable source of power to operate
medical-related fields, two areas key to the
their equipment. They need quality power 24
success of these transitions are (1)
hours per day, 365 days per year. Moreover,
improvements in the design, construction,
the time spent on power quality concerns is
and maintenance of healthcare facilities,
becoming shorter and shorter as bottom-line
and (2) the identification, selection,
pressures continue to be applied.
installation, and maintenance of medical equipment. Lessons learned in the area of
In most situations, instead of focusing on
power quality for healthcare demonstrate
the power quality, they have learned ways to
that efforts made beforehand to incorporate
“work around” malfunctioning and failed
power quality into these two areas usually
medical equipment. When one blood-
prevent significant interruptions in patient
pressure monitor is broken (possibly from a
care services and escalations in the costs of
voltage surge), a nurse or medical
medical equipment downtime.
technician goes and finds another monitor.
1
But, in smaller healthcare facilities where
The healthcare environment encompasses
equipment may be limited, providers may
everything associated with patient care and
find themselves with fewer pieces of
the healthcare facility from the time the
redundant medical equipment and without
patient enters the facility to the time the
resources including power to operate the
patient leaves the facility. This environment
facility. To healthcare providers, the
includes healthcare functions that occur
malfunction or failure of one key piece of
outside and inside the facility. Healthcare
medical equipment—a computed
facility designers, planners, architects, and
tomography (CT) scanner in an emergency
engineers and facility operating engineers
room, for example—would be enough of a
and maintenance support personnel should
Power Quality for Healthcare Facilities
Healthcare staff can contribute to improving patient care and the environment through increasing their level of awareness in recognizing equipment malfunctions that may be caused by power quality problems.
focus upon those parts of the environment that contribute to shaping the quality of power and depend upon the quality of the
Complex Electronic Medical Equipment Used in Patient Care Areas
power in providing patient care. Healthcare staff, including medical professionals, can also contribute to improving patient care and the environment through increasing their level of awareness in recognizing equipment malfunctions that may be caused by power quality problems. New electrotechnologies are continually introduced into this complex environment (see figure on right), placing new challenges upon the healthcare and facility staff, the quality of power delivered to the facility and to the equipment, and the electricity demand. These electrotechnologies may also consume additional floor space and weight load and place new burdens upon the facility infrastructure—electrical and mechanical systems. These new technologies include medical, functional, and facility equipment. Examples of new medical
electrotechnologies include diagnostic imaging systems capable of
New technologies, such as electronic machines in the
resolving more patient detail,
intensive care unit (top) and those used for laparoscopic
computer-based wireless clinical
imaging (bottom), are continually being introduced into the healthcare environment.
information systems, and advanced patient diagnostic and therapeutic equipment. Additionally, much of the medical equipment is mobile,
Today, the public and the government are
requiring reliable, well-regulated
making unprecedented demands upon the
electricity on tap throughout a facility.
healthcare industry to provide high-quality,
Examples of new functional
technologies include microprocessor-
restructuring and mergers are just two
based food preparation equipment
examples of how the healthcare industry is
and laundry equipment that use
meeting a financial challenge that leaves
adjustable speed drives.
little room for equipment malfunction.
Examples of facility equipment
include energy management systems, electronic controls for facility HVAC systems and equipment, and medical gas systems.
2
cost-effective patient care. Corporate
To ensure that the safe operation of medical equipment does not become a casualty of this new corporate mentality, the U.S. Congress passed the Safe Medical Device Act in 1990 (Public Law 101-69), which
Power Quality for Healthcare Facilities
For equipment with low immunity, electrical disturbances are a primary cause of damage and malfunctions.
establishes a partnership in safety between
Power Quality in Healthcare Facilities
the healthcare industry and manufacturers
Although inadequate and faulty wiring and
of medical equipment in the United States.
grounding systems and equipment interactions can exacerbate power quality
This act is required to track all implantable medical devices and life-supporting or lifesustaining devices listed in the act—such as
problems in healthcare facilities, electrical disturbances can damage low immunity
pacemakers, pulse generators, and
equipment or cause malfunction. In facilities
automatic defibrillators—that were distributed outside healthcare facilities after
where wiring and grounding systems are error free and equipment immunity is known,
August 29, 1993.
electrical disturbances are less likely to cause
The electrical environment in U.S.
power quality problems. Additional causes of power quality problems include the generation
National Electrical Code (NEC). The purpose
of disturbances from the normal operation of medical, functional, and facility equipment.
of this code is to provide minimum
For example, a contactor that controls power
standards to safeguard life or limb, health,
to part of the heating system in a facility can generate voltage transients that could impact
healthcare facilities is regulated by the
property, and public welfare by regulating and controlling the design, construction, installation, quality of materials, location, operation, and maintenance or use of electrical systems and equipment. This code regulates the design, construction, installation, alteration, repairs, relocation, replacement, addition to, use, or maintenance of electrical systems and
the operation and reliability of electronic medical equipment powered by the same panel that powers the heating system. In this situation, using a contactor that contains a snubber to limit the voltage transients and powering the heating system from a separate feeder circuit than the one powering the medical equipment will help resolve the problem.
equipment.
Microprocessor-Based Electronic Medical Equipment
Before the introduction of electronic medical equipment, common electrical disturbances were inconsequential to healthcare operations. Today, however, common electrical disturbances may cause high-tech medical equipment to malfunction, which is a problem given the intimate connection between this equipment and the patients that hospitals serve (see figure at left). Much of this equipment incorporates sensitive electronic power supplies and microprocessors (see figure on top of following page)—possibly resulting in extended patient discomfort, misdiagnoses, increased equipment downtime and service costs, and even life-threatening situations. Moreover, equipment damage and malfunctions can jeopardize patient safety and increase the cost of healthcare.
The healthcare environment is a unique one because of the intimate proximity of people to equipment.
3
Power Quality for Healthcare Facilities
Medical equipment used in the United
Circuit Boards from a Medical Imaging System
States, such as diagnostic imaging systems, that present dynamic loads to the facility electrical systems can cause power quality problems internal to the facility. The figure at lower left is an example of a nonlinear current waveform captured by a power quality monitor connected to the input of a CT scanner during imaging system operation. From the figure, one can see that the current is very nonlinear and is characteristic of a high inrush current when the system is placed into the scan mode. If the healthcare facility contains wiring and ground errors with its earthing system, then
Integrated circuits, sensitive to electrical and electromagnetic disturbances, are used in electronic medical equipment.
Although patient safety is the number one reason for reducing the potential for equipment malfunctions, healthcare administrators must also consider the bottom line. Electrical disturbances can result in repeated diagnostic tests, wasted medical supplies, and expensive service and repair calls. These unexpected events are not covered by any healthcare insurance provider. The increasing use of healthcare insurance and the increased coverage limitations therefore compel healthcare facilities to minimize all equipment malfunctions.
Current (50 amps/division)
Non-linear (Harmonic-Rich) Load Current from a CT System
Time (10 milliseconds/division)
This medical imaging system creates dynamic power quality problems in healthcare facilities with wiring and grounding errors.
4
dynamic loads such as those characteristic of diagnostic imaging system operation cause PQ disturbances that may impact other electronic devices in the hospital or even interfere with the operation of the dynamic load itself.
RECOGNIZING POWER QUALITY PROBLEMS
Symptoms and Their Causes Disturbances can enter healthcare equipment through any electrical port—the AC power input, telecommunications, or network—common in the facility’s electrical environment. Most disturbances will enter the AC power port and present themselves to equipment’s power distribution unit or power supply. Because most medical equipment in a healthcare facility is networked to other equipment, variations in the facility grounding system provide paths for disturbances to enter the equipment’s telecommunications and network ports. The effects of electrical disturbances upon healthcare equipment can be noticeable or unnoticeable. Disturbances entering AC power input, telecommunications, or network ports may not cause immediate damage to electrical and electronic components or cause equipment to fail suddenly. Depending upon the type of
Power Quality for Healthcare Facilities
The most common equipment malfunctions are caused by the inputs and outputs of microprocessors erroneously switching on and off because of voltage sags, swells, transients, andmomentary power interruptions.
disturbance—undervoltage or overvoltage, its duration, and the immunity of the equipment to that disturbance—gradual or fast occurring damage to electrical and electronic components may result. A disturbance such as a voltage surge entering the AC power input of medical equipment may not be sufficiently mitigated by internal overvoltage and overcurrent protection devices and may propagate through the power supply to other sensitive electronic subsystems and components. Voltage sags may cause post-sag inrush currents, which may cause permanent damage to overcurrent protection devices. A series of disturbances occurring over the period of a few hours or a few months, for example, may chip away at internal protection devices and electronic components, although damage to equipment may be virtually unnoticeable. Intermittent equipment malfunctions may be noticeable until eventual failure occurs. However, the most common equipment malfunctions are caused by the inputs and outputs of microprocessors switching between an on and off state resulting from voltage sags, voltage swells, voltage transients, and momentary power interruptions. For example, a voltage sag may cause the DC voltage (produced by the power supply) to the microprocessor of a blood-pressure monitor to decrease or suddenly change such that one or more of the microprocessor inputs or outputs drop from an on state to an off state. Or, a voltage transient incident upon the power supply may cause a change from an off state to an on state. In either case, data may be lost or scrambled, or the microprocessor may lock up or otherwise misoperate. Additionally, such changes in logical states can alter stored data, such as the control parameters of a defibrillator, ventilator, or an imaging system. Healthcare staffs have also reported power quality problems that are obviously not related to the malfunction of a microprocessor, such as 60-hertz artifacts on the signal recordings of biomedical equipment. The following are the most common symptoms of medical equipment
5
malfunction, including malfunctions not related to microprocessors. Distortion of Displayed Medical Information
Medical information displayed on cathode ray tubes (CRTs), liquid crystal displays (LCDs), printouts, and film may be distorted by disturbed DC voltages powering the display, a microprocessor malfunction, or faulty data from memory. For example, a waveform from an electrocardiogram printout may be disfigured, film from an X-ray may have a hot spot (a white area without any detail), or a video display on a physiological monitor may be distorted. Faulty data from memory or a microprocessor may also degrade the quality or resolution of an image captured by an imaging system such as a CT scanner (see figure below). Caregivers who encounter distorted information often report that they had to repeat tests or were unable to make timely, critical decisions because of the distortion.
Distorted Computed Tomography Image and Digital Readout
Variations in DC voltages can cause problems with the images and digital readouts from CT scanners.
Power Quality for Healthcare Facilities
Electrical disturbances can cause microprocessor -based medical equipment to malfunction.
Incorrect Diagnostic Results
unacceptable levels of 60-hertz current), and
Electrical disturbances can alter the control
miswired or damaged equipment that forces
parameters stored in electronic medical
supply current through ground conductors.
equipment and used to diagnose a patient’s
Electromagnetic fields from certain electrical
condition. For example, the status of a CT
distribution equipment, medical equipment,
system may be misreported via the digital
and facility equipment can also produce stray
readout as illustrated in the figure on the
magnetic fields that can cause these artifacts.
previous page. Moreover, biomedical
Artifacts in medical data may also be caused by
equipment such as blood-pressure monitors
current flowing in conductors that are not
may display diagnostic data, such as a digital
contained in conduits.
readout or level indicator, that disagrees with the patient’s prevailing condition. Incorrect diagnostic results may also be caused by 60-hertz noise coupled to the patient or to the leads of diagnostic equipment such as electrocardiographs (EKGs) (see figure below) and electroencephalographs (EEGs). Such noise is commonly associated with stray currents caused by faulty grounds (i.e., miswired ground conductors carrying
Equipment Lockup
Electrical disturbances can cause microprocessor-based equipment to lock up and fail to capture data used by caregivers to make critical medical decisions. Infusion equipment used to administer a patient treatment may fail to regulate or count the proper dosage. The lockup of a medical imaging system wastes the valuable time of patients, imaging technicians, and medical staff and may extend patient discomfort
Incorrect Diagnostic Results
when imaging scans must be repeated. Moreover, lockups of life-support equipment such as defibrillators pose lifethreatening risks to patients. Rebooting of medical equipment may take as long as two hours and in some cases cannot be accomplished if equipment software becomes damaged from electrical and electromagnetic disturbances.
Procedure Interruptions
Electrical disturbances may lock up microprocessor-based medical equipment, resulting in interrupted medical procedures. The consequences of these interruptions range from minor inconveniences to patient jeopardy. For example, if the video system fails during a routine laparoscopic surgery, the surgeon may have to incise the patient An artifact-infested electrocardiograph (top) appears to match a textbook example of
to complete the operation, an unplanned
arrhythmia (bottom) (reproduced from Capuano, 1993). The waveform on the top had a
procedure that significantly increases the
rate of 300 beats per minute or 5 hertz and was accepted and diagnosed as arrhythmia,
patient risk, recovery time, and the cost of
or atrial flutter (but actually was not).
6
patient care.
Power Quality for Healthcare Facilities
An electrical disturbance can damage an electronic component or circuit board in medical equipment causing a loss of data stored in memory or even destroying the memory altogether.
Loss of Stored Data
Control or Alarm Malfunctions
An electrical disturbance can damage an
The possible results of microprocessor
electronic component or circuit board in
malfunction include the loss of equipment
medical equipment causing a loss of data
control (see figure below) or the false
stored in memory or rendering the memory
sounding of an alarm. For example, the
inaccessible. Such losses can occur in data
keypad on an infusion pump may not
stored in the memories of biomedical
respond to finger touches of medical staff,
equipment and imaging systems, as well as
the pump may not remain in the desired
billing and patient records stored in
programmed state, or the equipment may
computer memory. If previously stored data
sound an alarm contrary to the condition of
suddenly becomes unavailable as a result of
the equipment or patient. Moreover, if an
a disturbance incident upon an electronic
unstable patient condition develops and an
data storage system, then patient tests may
equipment alarm does not sound, then the
need to be repeated, delaying patient
patient may be placed in a life-threatening
treatment. Power supply, mainframe,
situation. Some medical devices such as
memory, interface, and other types of circuit
infusion pumps have a built-in battery
boards may suffer damage from
backup that provides for internal backup
disturbances. Permanent damage to a power
power in the event of a sag or momentary
supply circuit board, like that shown in the
interruption. The use of a backup battery
figure below, may initiate the loss of stored
system in a medical device does not protect
data on a circuit board downstream of the
the device from malfunctions caused by
power supply board.
voltage transients and other disturbances.
Damage to a Power Supply Board
Nurse Checking on the Status of a Patient after Resetting a Medical Device
False alarms or, worse, alarm failures may result from any instrument malfunction, presenting a possible risk to patients and increased workload for healthcare professionals. A temporary overvoltage permanently damaged this power supply board from a medical instrument.
7
Power Quality for Healthcare Facilities
Equipment malfunctions can be avoided if the level of power quality is known and equipment selected or installed to be immune.
Sources of Electrical Disturbances
Faulty Facility Wiring and Grounding
The most common causes of electrical
In a fair number of cases, the cause of a
disturbances that lead to power quality
power quality problem in healthcare
problems in healthcare facilities and
facilities and medical clinics is simply a
medical clinics are
loose or corroded power or ground connection. Many medical equipment
low and unknown equipment
immunity; faulty facility wiring and grounding;
malfunctions attributed to poor power quality are caused by inadequate electrical wiring and grounding. Such problems frequently arise when
facility and equipment
modifications; high-wattage equipment; routine electric utility activities; accidents, weather, and animals; and a transfer to an emergency generator
or alternate feeder.
new electronic medical or office
equipment is connected to existing facility wiring; permanently installed medical
equipment is moved from one location to another; or underlying non-PQ-related
equipment malfunctions are not Low and Unknown Equipment Immunity
The immunity of most electronic medical equipment to electrical disturbances is low, unknown, or both. This is evidenced by the
resolved and changes to wiring and grounding are made in efforts to “enhance” the quality of power to the equipment.
number of cases of medical equipment malfunction and damage that are caused by
Wiring and grounding errors also enhance
power quality problems. Many power quality
the negative effects of neutral-to-ground
problems can be avoided if the quality of
transients, which disrupt electronic medical
power is known at the point of use within
equipment. Reversal of neutral and ground
the healthcare facility and if equipment
conductors; poor, missing, or redundant
immunity is known and high enough to
neutral-to-ground bonds; and poor, missing,
avoid equipment malfunction. When
or redundant equipment grounds are a few
immunity is unknown, healthcare providers
examples of faulty wiring and grounding
cannot determine if disturbances are likely
that can lead to medical equipment
to cause equipment malfunction and
malfunctions.
damage. As a result, healthcare providers cannot provide the utility with the data they
Many facility engineers and electricians in
need to warrant improvements to the power
healthcare facilities in the United States
system and cannot determine the degree of
used to mistakenly believe that if electrical
mitigation that can be provided by
systems are wired and grounded according
improving the operation of facility electrical
to Article 517 of the NEC (National Fire
systems (that is, identifying wiring and
Protection Association [NFPA] 70), there
grounding errors and resolving them) and by
should be no problems with the equipment.
utilizing power quality mitigation
By increasing the level of awareness of the
equipment.
impacts of power quality and compatibility on healthcare facilities and medical equipment through EPRI research, facility
8
Power Quality for Healthcare Facilities
The power supply equipment and wiring in a healthcare facility may fully comply with applicable standards, codes, and recommended practices and still be inadequate to prevent interruption of sensitive electronic equipment.
engineers, facility designers, and maintenance directors are realizing the importance of the integrity of their electrical
Power-Factor Correction Capacitors at a Substation Near a Healthcare Facility
systems in shaping the quality of power used for patient care. However, Article 517 focuses on electrical construction and installation criteria in healthcare facilities to reduce the risk of electrical shock and fire; it does not address power quality in the facility. The standard NFPA 99 entitled Handbook for Healthcare Facilities, also commonly used in the United States, focuses on the installation and performance of equipment in a healthcare facility, but also does not address power quality. The equipment and wiring in a healthcare
Switching capacitors in and out of service can create
facility may fully comply with applicable
transients that impact sensitive instrument.
standards, codes, and recommended practices and still be inadequate to support sensitive electronic equipment commonly found in a healthcare facility. (MRI) systems, CT scanners, and linear accelerators operate at high line voltages, Routine Electric Utility Activities
require high steady-state current, and
To correct the power factor of electricity,
present dynamic loading (see figure on
electric utilities routinely switch large
following page) to healthcare facility power
capacitors (see figure on top right) onto the
systems. During startup, this type of
power lines. These switching activities may
equipment draws very high inrush current—
generate transient overvoltages, called
as high as 70 times the normal operating
“capacitor-switching transients,” which may
current—which can cause voltage sags and
enter a healthcare facility or medical
other electrical disturbances on adjacent
location at the service entrance. These types
circuits not properly sized for these loads.
of electrical disturbances are more likely to
Problems occur when the circuits connected
occur in the morning and evening, when
to such disturbance-causing equipment
industrial facilities are powering up and
were not carefully planned for high-wattage
down. Other routine activities such as the
equipment. Such problems most often arise
operation of reclosures and breakers that
after a facility has recently undergone a
occur to maintain and stabilize the power
renovation or expansion or has recently
system and reduce the effects of electrical
moved existing medical equipment or
disturbances caused by natural events (e.g.,
installed new medical equipment. Also,
lightning) can result in some residual
installing high-wattage electronic
disturbances.
equipment without upgrading the existing facility power system (i.e., switchgear,
High-Wattage Medical Equipment with
transformers, and electrical wiring and
Dynamic Load
grounding) to accommodate the higher power consumption may result in overload,
9
Large medical equipment such as X-ray
undervoltage, and even overvoltage
machines, magnetic-resonance imaging
conditions.
Power Quality for Healthcare Facilities
Harmonic-Rich Current from an MRI System
facilities and medical clinics may find that equipment malfunctions are more prevalent on windy days when tree limbs may contact
Current (20 amps/division)
power lines. Voltage sags and interruptions may also be caused by lightning strikes, animals climbing atop the electrodes of a transformer or other utility equipment, and power-line conductor and insulator failures.
Downed Power Pole Adjacent to a Healthcare Facility
Time (25 milliseconds/division)
This distorted current waveform was captured with a power quality monitor during a PQ field investigation at a healthcare facility.
Voltage sags originating from outside a facility can be caused by downed, crossed, and contacted power lines.
Mechanical equipment containing loads that are inductive (e.g., motors) and resistive (e.g., heating elements)—such as heating, ventilation, air-conditioning, transportation, refrigeration, and pump equipment, which are controlled by starters and contactors—may also create electrical disturbances. The startup, normal operation, and shutdown of this equipment can cause voltage sags, transient overvoltages, and electrical noise.
Accidents, Weather, and Animals
Voltage sags originating from outside a
This toppled power pole caused a power outage at the
facility—which may account for more
healthcare facility nearby.
increased patient risk than any other single
10
type of disturbance—can be caused by
Facility electrical modifications
downed (like that shown in the figure on the
Renovating and annexing healthcare
right), crossed, and contacted power lines
facilities and medical clinics are common in
and are most likely to occur during
the global modern healthcare industry, as
inclement weather conditions and peak
are the addition of transformers, subpanels,
demand times. Cars crashing into utility
and circuits to an electrical system and the
poles and ice-laden, wind-blown, or
use of temporary circuits to power existing
overgrown limbs touching and landing on
equipment. The rerouting of feeder and
power lines may create a path from the
branch circuits can result in the
power line to ground, creating electrical
commingling of loads (powering sensitive
disturbances and power interruptions for
electronic medical equipment from the
some and voltage sags for many. Healthcare
same bus as disturbance-generating loads). Power Quality for Healthcare Facilities
To provide power to some construction
manual transfer switch. (Ideally, in facility
equipment, temporary electrical circuits
electrical designs where provisions for a
may be connected to the wiring of existing
second utility feed are included, the second
structures, or construction equipment may
feed should come from a different
be connected to the output of motor-
substation, but this is not always possible.)
generator sets. The operation of
If the transfer switch is not properly
construction equipment such as arc welders
installed, adjusted, and maintained to
(see figure below) and line-powered
ensure a smooth transfer of power, the
motorized rotary equipment on the center’s
transfer may produce electrical disturbances
wiring system may introduce electrical
that are severe enough to cause malfunction
disturbances into branch circuits powering
of electronic medical equipment. Inspection
sensitive electronic medical equipment.
of generator wiring (see figure below) will reveal important wiring and grounding characteristics that are vital to the
Construction of a Shielded Room for an MRI Suite Using an Arc Welder
emergency power system. Engineers in healthcare facilities and medical clinics may also find that malfunction and damage to medical equipment may occur during routine generator testing (if generator testing is required by local, state, and international codes and laws). Most master generator control centers include an adjustable time delay to ensure that the generators are placed online or offline without creating electrical disturbances.
Searching for a Neutral-to-Ground Bond in the Emergency Generator at a Healthcare Facility Arc welders can introduce electrical disturbances into the branch circuits on which medical equipment are operating.
Transfer to and from Emergency Generator or Alternate Feeder
To ensure that power is always provided to feeder circuits that power subpanels and branch circuits connected to critical-care equipment, some electrical codes require that healthcare facilities have ready access to emergency power. Whether the source of emergency power is an on-site generator or a second utility feed, transferring from the normal power source to the emergency source is accomplished with an automatic or
11
Generator wiring should be inspected and maintained to avoid producing electrical disturbances during a power transfer.
Power Quality for Healthcare Facilities
So You Think You Need Uninterruptible Power Supplies? In healthcare facilities where power quality
installation and maintenance of these
the facility. These devices are designed to
problems occur frequently, healthcare
systems, even when new large medical
operate on internal batteries, and thus
providers may be eager to purchase and
equipment is specified and purchased. Large
continuous operation of this equipment is
install power quality mitigation equipment to
medical equipment such as diagnostic
possible during a voltage sag or momentary
protect both small and large loads from
imaging systems can be fitted with a UPS at
interruption. One should note, however, that
electrical disturbances. In situations where
the installation site, but the barriers in doing so
electronic medical equipment with an onboard
small loads such as biomedical equipment do
are significant. Imaging suites are tight on
battery recharger and an internal rechargeable
not contain internal battery backup systems,
floor space, and the electrical system provided
battery may also malfunction during an electrical
installing an appropriately sized uninterruptible
for these spaces was not designed to
disturbance as the charger could be rendered
power supply (UPS) will increase the
accommodate the installation of power
inoperable as a result of a deep voltage sag;
immunity of these loads to common
mitigation equipment. Moreover, imaging
hence the need for characterizing this equipment
disturbances such as sags and momentary
system operators do not have time to routinely
for immunity to sags and interruptions.
interruptions. UPSs for large medical loads
test a UPS or maintain the UPS’s batteries. Even though electric utilities try to provide as
ranging from 10 kVA to a few hundred kilovoltamperes, which can cost as much as $1 million,
Unlike industrial and manufacturing facility
many nines of reliable power to a healthcare
may be installed on an individual medical
environments where industrial process
facility as possible, healthcare providers must
imaging system, can support multiple systems
systems can be made much more robust to
realize that their facilities are also fed from
in a medical imaging department, or can be
voltage sag phenomenon with proper
typical power distribution networks. Utilities
used for a group of critical equipment such as
electrical and software design techniques,
will make every effort to ensure that a direct
ventilators in an intensive care unit (ICU).
most medical equipment is not designed to
service feed (service entrance) to a hospital is
offer this option. Medical equipment is
properly maintained and that second feeds are
In many situations where large UPSs are
designed for individual use in an array of
provided from a second substation whenever
thought to be needed (and some are needed),
equipment and for compact use. For example,
possible. However, redesigning distribution
healthcare providers discover that common
the ten different types of medical equipment
systems or making other investments in the
disturbances are exacerbated by typical wiring
used in an ICU are not linked together with
utility’s power delivery infrastructure may also
and grounding errors within the healthcare
one downstream system depending upon the
be prohibitively costly. Given that the cost of
facility’s electrical system. Prior to the
results from an upstream system. Instead,
the events and facility-level solutions can be
decision to purchase and install a large UPS, a
each piece of medical equipment is designed
very expensive, electric utilities and their
well-developed power quality investigation
to carry out a specific task such as monitoring
healthcare customers search for ways to ease
should be done within the facility to determine
blood pressure, monitoring blood oxygen
the financial burden of increasing the immunity
the extent to which wiring and grounding
level, and providing breathing assistance to a
of their healthcare customers to common
errors contribute to the root cause of
patient. However, the typical solutions that
electrical disturbances such as voltage sags,
malfunctions with small and large medical
can be applied in manufacturing environments
momentary interruptions, and surges.
loads. In almost all situations, typical wiring
to solve power quality problems with
and grounding errors internal to the facility can
industrial equipment can also be applied to a
Healthcare providers are not willing to install a
be linked to the severity of common
healthcare facility.
power mitigation device on each piece of medical equipment. However, they can be
disturbances entering the facility from
12
everyday electrical events occurring on the
The types of portable electronic medical
persuaded to have their maintenance staff sift
utility power system and from events
equipment that can be fitted with a low- to
through the details of a facility’s power
generated by the operation of large loads in
mid-power UPS are limited (to some less than
distribution system through learning how to
neighboring customer facilities and/or
10 kVA machines. Because of the need to
conduct power quality investigations.
generated by the operation of large loads
provide safe patient environments, most
Moreover, healthcare providers may also be
within the healthcare facility.
typical power quality solutions, such as
persuaded to improve their medical
constant voltage transformers and sag-
equipment procurement process by learning
Purchasing and installing large UPS systems
reducing technologies, cannot be
how to specify an acceptable level of
to protect individual imaging systems or
implemented on medical equipment in the
immunity to voltage sags and momentary
several systems in a medical imaging suite
patient environment. Most medical
interruptions and voltage surges that is
can present additional problems for the
equipment are designed to be portable and
suitable to most healthcare facility electrical
healthcare provider. Healthcare facility
are placed on high-quality equipment carts
environments. But, before this concept can be
designers do not make accommodations for
without space provided for a UPS. Some
widely applied, medical equipment
such large pieces of power mitigation
devices such as blood-pressure monitors and
manufacturers must succumb to determining
equipment. Healthcare providers, operating on
infusion pumps must have power maintained
the full immunity capability of their equipment
extremely tight budgets, do not budget for the
to them as the patient is moved throughout
to these common disturbances.
Power Quality for Healthcare Facilities
Power quality in the healthcare environment can be improved through enhancing the level of awareness among utilities, healthcare facility designers, and medical equipment manufacturers.
IMPROVING POWER QUALITY IN THE HEALTHCARE ENVIRONMENT
fewer electrical disturbances, effectively use power-conditioning
Power quality in the healthcare environment
technologies for existing medical
can be improved through enhancing the
equipment in accordance with
level of awareness among the stakeholders:
standards and recommended
utilities, healthcare facility and medical
practices,
staff, healthcare facility designers, and medical equipment manufacturers. Power quality problems in this mission-critical environment present a series of challenges among stakeholders. Meeting these challenges helps to prevent these problems
carefully plan new construction or
renovation of existing healthcare facilities with regard to power quality concerns, maintain existing wiring and medical
before they become monumental to
equipment in healthcare facilities,
healthcare providers.
and learn from past power quality
Meeting the Power Quality Challenges of the Healthcare Industry Although healthcare staffs rely upon advanced medical procedures using advanced medical equipment to provide immediate patient care, they must sometimes plug equipment into antiquated and unreliable electrical systems. Moreover, some equipment manufacturers design equipment without fully considering and understanding the electrical environment of a healthcare facility. Because of its obligation to human care, the healthcare industry must demand high standards of performance from facility designers, equipment manufacturers, equipment service companies, facility and equipment support staff, and electric supply companies. To meet the challenges of the healthcare industry, these people must meet on common ground to establish new partnerships to
improve power quality in the healthcare environment, improve the procurement process for
new medical equipment, encourage equipment manufacturers
13
disturbances and that generates
problems.
Establishing Partnerships Preventing or resolving power quality problems should be a cooperative effort between healthcare facilities, equipment vendors, equipment manufacturers, and electric supply companies. Electric supply companies have always offered assistance to customers in emergencies and have sometimes promoted new energy-efficient technologies to improve productivity and reliability as well. As problems associated with new technologies were revealed, many electric supply companies established power quality programs that invested in power quality research to assist utility customers and manufacturers with equipmentcompatibility problems. Electric supply companies especially recognize the necessity of identifying or providing power quality engineering services to their healthcare customers. These services enable healthcare staff to learn how to identify wiring and grounding problems that exacerbate power quality problems, select the proper power-
to design medical equipment that is
conditioning equipment to mitigate these
more immune to electrical
problems, develop specifications (that
Power Quality for Healthcare Facilities
Building strong relationships between healthcare facilities, equipment vendors, equipment manufacturers, and electric supply companies offers many benefits.
include power quality specifications) for
and clinics where significant and
purchasing medical equipment problems,
costly power quality problems have
establish correct installation guidelines, and
occurred find it cost-effective to
plan center renovations or the construction
purchase a monitor and learn to use
of new healthcare facilities and medical
it. Consider tapping the expertise of
clinics to help avoid problems.
your local utility company or independent consultants. Determine
Building strong relationships between
the characteristics of your facility’s
healthcare facilities, equipment vendors,
electrical system: Can it tightly
equipment manufacturers, and electric
regulate equipment voltage? Is
supply companies offers many benefits.
voltage to equipment continuous?
These benefits include learning how to avoid
Does high-wattage equipment create
wiring and grounding errors, reducing or
electrical disturbances in the facility
eliminating controllable electrical
wiring? Your local utility company
disturbances, managing common
may also provide site-specific
uncontrollable electrical disturbances,
characteristics such as expected
encouraging equipment manufacturers to
voltage regulation and statistical
design and build robust equipment immune
analysis of electrical disturbances.
to most electrical disturbances, significantly reducing the potential for lawsuits by healthcare patients involved in events possibly initiated by equipment malfunctions, and avoiding citations and penalties from international regulatory agencies.
Evaluate the immunity performance
requirements of existing equipment. How susceptible is each type of medical equipment to common electrical disturbances such as voltage sags and transient overvoltages? Set your expectations for the
Creating a Power Quality Checklist for Procuring Equipment
performance of new equipment, and
Healthcare facilities and medical clinics
help in specifying design features
routinely procure and install medical,
that enhance compatibility between
functional, and facility equipment. To
the equipment and its intended
reduce power-quality-related problems
electrical environment.
between equipment and the intended electrical environment, equipmentprocurement procedures should include the following steps.
then ask your utility company for
Identify and repair all wiring and
grounding problems. Identify all areas where critical
electronic medical equipment may Planning for Additional Equipment Begin a sound in-house power
quality program with the purchase of a PQ monitor to conduct an on-site survey to identify potential power quality problems and diagnose problems with sensitive electronic
be used and the special power requirements of such equipment. With assistance from your local
utility company or independent consultants, identify appropriate power-conditioning devices for critical electronic equipment.
medical equipment. Some facilities
14
Power Quality for Healthcare Facilities
Purchasing Additional Equipment Disclose to equipment suppliers the
Installing Additional Equipment Use high-performance wiring and
power quality characteristics of the
proper grounding techniques
electricity and wiring where the new
specified in the International
equipment will be installed.
Electrical Code (IEC), the Institute of
Ask the manufacturer’s representative
about known power quality problems with the equipment and if the equipment has been tested for compatibility with the utility power system. If there is reason to believe that compatibility may be an issue, ask to see the power quality test report. For all new equipment, specify the
voltage range (required voltage regulation), frequency, and voltage sag immunity (i.e., ride-through) performance.
Electrical and Electronics Engineers (IEEE) Standard 602-1996 (White Book; Recommended Practice for Electric Systems in Healthcare Facilities), and the IEEE Standard 1100-1992 (Emerald Book; Powering and Grounding Sensitive Electronic Equipment). For circuits connected to sensitive
electronic equipment, use singlepoint grounding, locate equipment as electrically close to the source as possible, and make sure that the sizing of phase, neutral, and ground
Purchase equipment with an input
conductors follow international and
voltage rating matched to the voltage
local codes and manufacturer
at the installation site when possible.
installation requirements.
Purchase high-quality matching transformers with new equipment when the voltage ratings of the equipment do not match the available voltage at the installation site.
When adding grounding conductors
to an existing facility, run the grounding conductors parallel to the existing power conductors to reduce stray electromagnetic fields.
If a power-conditioning device is
needed, make sure that it is designed for compatibility with electronic medical equipment. Medical equipment such as imaging systems with dynamic load behavior may not function properly when connected to some power conditioners.
When installing high-wattage
medical equipment in an existing facility, monitor the input voltage at the proposed installation site for electrical disturbances for at least a 30-day period before completing the installation.
Make sure that all medical and
power-conditioning equipment complies with applicable
Regularly review equipment
international codes, standards, and
performance and continue the
recommended practices.
relationship between healthcare
To reduce susceptibility to common
15
Maintaining Equipment
facility staff, utility company
electrical disturbances, select the
representatives, equipment vendors,
highest input voltage rating for
equipment manufacturers, and
equipment known to be sensitive to
medical equipment service
common electrical disturbances.
companies.
Power Quality for Healthcare Facilities
Document all facility power outages,
noticeable disturbances (i.e., light flicker), and equipment problems.
Using Power-Conditioning Devices to Improve Equipment Compatibility Some power quality problems in healthcare
Include patient schedules, the location of equipment, the
facilities and medical clinics can be solved with appropriate power-conditioning
symptoms, suspected causes, time
devices. Some of these technologies are listed in the table on the left and include
and date of occurrence, and any other related events. Checking
isolation transformers, surge-protective
disturbance logs against utility company records and facility
devices, voltage regulators, and UPSs.
activities can help reveal the source
However, power-conditioning devices are
of electrical disturbances. These logs can also be used to specify future
not always the answer to a power quality
equipment purchases and determine correct installation methods.
quality mitigation equipment can worsen a
problem. In some cases, installing power medical equipment malfunction, especially in cases where medical equipment loads are
General Summary of Available Power-Conditioning Technologies
very dynamic in nature, like that of diagnostic medical imaging equipment. In addition, low-kilovolt-ampere powerconditioning devices and “ice-cube” relays, power supplies, and contactors routinely used in industrial facilities can be used in the physical plants (i.e., where HVAC, steam, air, vacuum, and other mechanical systems are located) but cannot be used with medical equipment to solve power quality problems. In other cases, installing such equipment is not necessary and can have no effect on the problem. For example, power-conditioning devices will not protect equipment against radiated emissions or electrostatic discharge, which has been reported as one of the electromagnetic-related causes of equipment malfunction. In some cases, the potential for this problem can be virtually eliminated by maintaining correct humidity levels or installing building materials that reduce the buildup of static charge. In other cases where wiring and grounding problems exacerbate equipment malfunctions caused by voltage transients, installation of a UPS can provide enhanced immunity to voltage sags and momentary interruptions and some mitigation of transients. However, if equipment damage is
16
Power Quality for Healthcare Facilities
Good grounding is essential for good power quality and safety at any healthcare facility.
caused by a wiring and grounding problem and voltage transients developed at the point of use (where the equipment is connected to the center electrical system), then installing an upstream UPS will not resolve the problem. Consult the equipment
medical equipment such as imaging
and radiology equipment and medical air pumps, and mechanical equipment such as
adjustable speed drives, chillers, fans, pumps, and HVAC equipment.
manufacturer and local utility company to determine whether a power-conditioning device can be used effectively.
Other support equipment, such as biomedical and laboratory equipment and low-power kitchen and laundry equipment,
Understanding Facility Voltage Requirements, Grounding, and Dedicated Circuits
are powered at 120 volts.
The voltage level provided to the service
tolerance levels for the electric supply
entrance of a healthcare facility will impact
voltage with a range of ±10% will again be
the voltage that is provided to all loads in
unified for European healthcare facilities
the facility, especially the medical
and medical clinics. Thus, European
equipment loads. Because the healthcare
manufacturers of medical equipment used
provider must provide healthcare services to
in the United States, who have integrated
patients in real-world power quality
design changes into their equipment to help
environments, grounding the facility
ensure reliable operation in Europe, may
infrastructure, the secondary of the utility
find that United States users file fewer
company’s transformer at the service
complaints regarding medical equipment
entrance, within the switchgear, throughout
malfunctions. Healthcare facilities and
the facility electrical system, and at the end-
medical clinics in the United States may
use level where the equipment is connected
experience fewer malfunctions caused by
and used is also critical. Moreover, many
long-term steady-state undervoltage
end-use loads in healthcare facilities require
conditions and possibly minor voltage sags.
the use of a dedicated feeder or branch
In areas such as medical laboratories where
circuit, which helps to maintain voltage and
microscopes are used and surgical suites
current quality to critical equipment.
where eye surgery and other surgical
However, effective January 1, 2008, the
procedures are performed, high power The voltage levels selected for new
quality lighting that is immune to more
equipment will depend upon the available
types of voltage fluctuations and other
utility voltage, the size of the healthcare
electrical disturbances may operate with
facility or medical location, voltage levels
less flicker to lamps, thus improving light-
used within, type of equipment, building
assisted and light-dependent medical
layout, voltage regulation requirements, and
procedures.
cost. Typically, power to a healthcare facility or medical location is supplied by the utility company at a medium voltage level for large
Voltage Matching
facilities and clinics and at 480 volts (with a
Once the nominal voltages of equipment
±5% range), three-phase. These voltages
have been selected, the voltage source for all
may be used to power:
medical equipment to be installed in the facility should be carefully checked to assure proper voltage levels. New equipment
17
Power Quality for Healthcare Facilities
should be ordered to match one of the
However, the focus of the discussion in this
planned voltage sources. Otherwise consider
section of this report is not on patient
using buck/boost transformers,
safety, but on wiring and grounding (i.e.,
autotransformers, or standard two-winding
earthing systems) as they relate to power
isolation transformers to match the voltage
quality in U.S. healthcare facilities and
requirement of the equipment to the voltage
medical clinics. Moreover, the compatibility
source. Variacs should never be used to
between medical equipment and the
match a source voltage to an equipment
electrical environment in these facilities and
voltage.
clinics is dependent upon the type of earthing system that powers and grounds
Regardless of the earthing system used, providing a solid lowimpedance ground to sensitive equipment will help minimize power quality problems.
Equipment from International Manufacturers
Equipment purchased from international sources originally designed to operate in countries with different nominal voltage levels requires careful consideration of the design of the facility distribution system so that the correct voltage can be supplied to the equipment. Equipment designed for nonstandard U.S. voltages may require matching transformers. The addition of a transformer may make equipment more sensitive to common electrical disturbances. Also, equipment designed for 60-hertz operation must be able to operate properly at 60 hertz. The voltage tolerance of overseas equipment may also be a concern and should be checked. Equipment purchased from European manufacturers not recognizing the standard U.S. nominal voltage may require a special transformer to be powered from U.S. voltage sources.
the medical equipment. Regardless of the earthing system used, providing a solid low-impedance ground to sensitive equipment—which is required by the NFPA NEC and healthcare facility codes and recommended by the IEEE Emerald, Green, and White Books—will help minimize power quality problems. Because patients are often moved from one location in the healthcare facility to another, grounded receptacles should be available at all possible equipment locations. Power cords should never be modified to accommodate an ungrounded receptacle by removing the grounding connector. Nor should grounding adapters be used on equipment requiring a ground. In some older healthcare facilities, grounding conductors may be present but may not be running parallel to the power conductors. In the course of enhancing the grounding system in these facilities, the grounding conductors should be run parallel to the circuit’s neutral and power
Ensuring Proper Grounding and Wiring
conductors, which will minimize stray
Power quality investigations carried out in
electromagnetic fields due to the presence
the United States are revealing that the
of any unwanted ground currents.
integrity of wiring and grounding systems in healthcare facilities and medical clinics has
Similar to the requirements of electrical
an even greater impact on the immunity of
systems for providing quality voltage and
medical equipment to common electrical
current to large loads such as chillers and
disturbances. Since the term leakage current
printing presses found in commercial and
was coined for the medical equipment
industrial facilities, large loads in healthcare
industry, much of the focus on the integrity
facilities must be circuited such that their
of grounding systems in healthcare facilities
operation does not affect other loads.
has been on patient safety.
Powering disturbance-generating loads such as HVAC equipment (e.g., motor contactors,
18
Power Quality for Healthcare Facilities
To avoid equipment malfunctions during renovation or new construction, healthcare facility engineers should coordinate with construction foremen before the electrical system is modified.
motor starters, chillers, heating systems,
The IEEE Standard 602 (White Book), and
etc.) from the same voltage bus that powers
IEEE Standard 1100 (Emerald Book) are also
critical medical loads (e.g., X-ray equipment
both excellent technical resources that
and medical imaging systems) is a
address power quality in healthcare facilities
prescription for incompatibility problems
and medical clinics and offer guidance on
between building and facility loads, and
powering and grounding sensitive electronic
critical medical loads. Large diagnostic
equipment during facility construction.
medical imaging systems, such as MRI systems, CT scanners, and various X-ray
Medical Equipment Power Supplies
machines require dedicated power, neutral, and ground conductors also, because they usually draw fluctuating dynamic currents. Providing dedicated conductors for power, neutral, and ground is not only concerned with individual circuits (i.e., the fact that the circuits are separate runs from switchgear and electrical panels) but also the size (i.e., wire gauge) of the conductors with respect to the required length and the allowable voltage drop from the supply to the load. Many power quality investigations result in findings that identify dedicated circuits to X-ray equipment and imaging systems that are sized too small in wire gauge. The size of the grounding conductor is also important and should be specified according to the requirements of the X-ray or medical imaging system manufacturer. When this equipment is installed, the facility electrician should also determine what other sensitive or disturbance-causing equipment may be powered by the common source. In some cases, the solution may require providing a dedicated circuit to certain sensitive medical equipment to isolate it from other disturbance-causing equipment.
In healthcare facilities and medical clinics, the failure of the facility power may pose life-threatening consequences to patients. Examples of these concerns are the failure of a power supply in a ventilator, a lighting system in an operating room, and the branch circuit to a life support system in an ICU. The restoration time for medical power supplies to restore power to the medical equipment is not specified in the United States for medical microprocessor-based equipment. Designers of medical power supplies must be conscious of the amount of leakage current they allow to flow out of the supply under certain conditions, and the allowed levels are governed by the Association for the Advancement of Medical Instrumentation (AAMI). Lower leakage currents equate to higher levels of conducted emissions, thus increasing the likelihood of a medical device creating an electromagnetic interference (EMI) problem. Careful balance between EMI filter design and leakage current helps to ensure success in both areas. However, as medical devices become more digital in the next 20 years, this balance will become more
To avoid equipment malfunctions during
difficult to achieve.
renovation or new construction, healthcare facility engineers should talk to the
Standards
designated construction contact before the
The healthcare and medical equipment
electrical system is modified. This
industries are heavily regulated to protect
precaution will help ensure that good power
patients. Both the United States and Europe
quality is maintained on circuits deemed
have developed and published standards,
essential to patient safety, critical care, and
recommended practices, and guidelines
other equipment necessary for the effective
related to power quality and
operation of the healthcare facility during
electromagnetic compatibility in the areas
the construction and renovation process.
19
Power Quality for Healthcare Facilities
of healthcare facility design, medical equipment design (i.e., product standards), and emergency preparedness. Standards, recommended practices, and guidelines have also been developed that define disturbances and test methods for power quality and electromagnetic compatibility. The United States has made significant
EN 61000-3-X: Emissions, limits EN 61000-4-X: Emissions,
measurement techniques EN 61000-5-X: Immunity, testing
techniques EN 61000-6-X: Installation and
mitigation guidelines
contributions in power quality standards and healthcare facility design standards.
Other healthcare codes, standards, and
Europe has made significant contributions
recommended practices are promulgated by
in the area of immunity standards (i.e.,
the NFPA, the IEEE, the American National
emissions and immunity) regarding product
Standards Institute (ANSI), the Federal
design and safety.
Communications Commission (FCC), the IEC, and the International Special
The top table on the following page presents
Committee on Radio Frequency (CISPR).
a summary of power-line and electromagnetic disturbances, power electronics technologies, emissions and
Healthcare Facility Standards
immunity standards, and equipment
Standards, recommended practices, and
performance standards relating to electronic
guidelines have also been developed in
medical equipment. Medical equipment
several areas related to the design of
designers and manufacturers in the United
healthcare facilities and medical clinics. The
States have become more cognizant of these
NFPA 99 (Standard for Healthcare Facilities),
standards. The emissions and immunity
the facility code standard developed and
standards listed in the bottom table on the
used in the United States; the NFPA
following page are the Basic
Standard 70 (The National Electric Code),
Electromagnetic Compatibility (EMC)
the electrical code standard developed in
standards prepared by the European-based
the United States; and the IEEE Standard
International Electrotechnical Committee
602-1996, White Book (Recommended
(IEC). In the past few years, they have been
Practice for Electric Systems in Healthcare
referred to as IEC 61000-X-X standards. After
Facilities), the electrical system design
the European Union (EU) recently adopted
practice developed and used in the United
them as European Norms (EN) standards,
States, provide guidance to designers of
they were referred to as EN 61000-X-X
healthcare facilities and medical clinics.
standards. The requirements listed in these
Facility designers in the United States also
standards serve as the basis for all present
commonly refer to the well-known IEEE
and future power quality and EMC
Standard 1100-2006, Emerald Book (Powering
requirements for all products traded
and Grounding Sensitive Electronic
internationally, including electronic medical
Equipment), for guidance on powering and
equipment. The Basic EMC standards
grounding electronic medical equipment.
consist of the following six parts: EN 61000-1-X: General EMC
standards EN 61000-2-X: Compatibility levels of
environments Continued on page 22 20
Power Quality for Healthcare Facilities
21
Power Quality for Healthcare Facilities
Cross-Reference of European Standards Applicable to Electromagnetic Compatibility of Electronic Medical Equipment
(Refer to table below)
Effects of Electromagnetic Disturbances on Power Electronics Technologies Used in Electronic Medical Equipment
Healthcare facility standards address important aspects of the electrical system in a healthcare facility from planning, voltage selection, loading, harmonics, disturbances, mitigation techniques, emergency power systems, renovation, telecommunica tions, and lighting.
Continued from page 20
strength, and isolation transformer
Healthcare facility standards address every
construction to enable its use in electronic
aspect of the electrical system in a
medical equipment. The “Y” capacitors
healthcare facility from planning; voltage
required in the input filter of a standard
selection; loading (e.g., historical load
switch-mode power supply for information
densities and profiles, demands, and
technology equipment would almost
factors); harmonics; disturbances;
certainly cause the power supply to fail on
mitigation techniques; emergency power
the grounds of excessive leakage current.
systems; renovation; telecommunications;
Briefly, the more-stringent requirements
and lighting. Guidance is given on how to
that are of particular relevance to power
avoid overloading, undervoltaging,
supplies used in electronic medical
overvoltaging, and equipment damage and
equipment are (1) service entrance to
shutdown caused by power problems.
secondary creepage and clearance distances for double or reinforced insulation for
Medical Equipment Safety Standards
equipment operating from 250 volts AC maximum must be 8 and 5 millimeters,
In the EU, technical safety problems of
respectively; (2) primary to secondary
electronic medical equipment are addressed
dielectric withstand test must be 4,000 volts
by the EN 60601 series of standards which
AC; (3) earth leakage current maximum is
follow IEC 601 (now referred to as IEC
0.5 milliamp for normal operation and 1
60601-1-2), Medical Electrical Equipment. In
milliamp maximum for a single fault
the United States, UL 544, Medical and
condition. These values are for type B, type
Dental Equipment, covers medical and
BF, and type CF equipment categories:
dental equipment, but in 1994 UL 2601-1, Medical Electrical Equipment—Part 1: General Requirements for Safety, came into effect. This standard is harmonized with IEC 60601-1-2, to be used at present in parallel with UL 544, Medical Equipment, and the U.S. safety standard for medical equipment, but it became the sole mandatory standard in 2004. In Canada, CSA 22.2-601.1, Medical
Type B—Non-patient-connected
equipment, or equipment with grounded patient connection. Type BF—Equipment with a floating
patient connection. Type CF—Equipment with a floating
connection for direct cardiac application.
Electrical Equipment—Part 1: General
22
Requirements for Safety, has been in use
Patient leakage current for the above
since 1990, again, alongside the existing
categories is 0.1 milliamp (0.5 milliamp for a
standard CSA 22.2-125, Electromedical
single fault condition) for type B and BF and
Equipment, and it became the sole
0.01 milliamp (0.05 milliamp for a single
applicable standard in the year 2000.
fault condition) for type CF.
The bulk of the electrical safety
In the EU, electronic medical equipment is
requirements detailed in IEC 60601-1-2 are
subject to the Medical Device Directives 93-
based on IEC 950, Safety of Information
42-EEC, which was implemented on January
Technology Equipment Including Electrical
1, 1995. These “New Approach” Directives
Business Equipment. However, an IEC 950
gave a three-year transitional period (up to
(EN 60950) approved power supply would
January 1, 1998) until CE marking (mandatory
need to pass the additional test and
marking to indicate conformity with the
inspection requirements of EN60601-1 for
health and safety requirements set out in the
separation, leakage current, dielectric
European Directives) was required.
Power Quality for Healthcare Facilities
Two New Approach Directives, 90/385/EEC Active Implantable Medical Devices (AIMD) and 93/42/EEC Medical Devices Directive (MDD) exempt those specific product categories from the EMC Directive. They contain their own specific EMC requirements. Probably only the MDD will be of interest to power supply designers and users. The EMC standards cited are IEC60061-1-2, adopted by European Committee for Electrotechnical Standardization (CENELEC) and published as EN60601-1-2. Emission standards required follow CISPR 11 (EN55011), Limits and Methods of Measurement of Electromagnetic Disturbance Characteristics of Industrial, Scientific, and Medical (ISM) Radio Frequency Equipment, normally class B, with a 12-dB relaxation for radiated emissions in X-ray rooms, for example. Immunity standards again rely heavily on IEC 801 as follows: IEC 801-2, Electrostatic Discharge:
3 kV contact, 8 kV air IEC 801-3, Radiated Radio-Frequency
Interference (RFI): 3 V/m from 26 to 1000 MHz, 80% amplitude modulation, 1 V/m in X-ray rooms
IEC 801-4, Electric Fast Transients:
1 kV at service entrance plug, 2 kV for hardwired service entrance, 0.5 kV on connecting leads greater than 3 m long IEC 801-5, Service Entrance Surges:
1 kV differential, 2 kV common mode After June 14, 2000, electronic medical equipment was allowed to be sold within the EU as compliant with either the EMC or MD Directives. An important point to note for all products subject to the AIMD, and many products under the MDD (except class I), is that they cannot be self-certified. Approvals must be carried out by Notified Test Organizations. Class I equipment is defined as equipment for which electric shock protection is achieved by basic insulation and protective earth. All conductive parts that could assume hazardous voltages in the event of failure of basic insulation must be connected to a valid protective earth conductor. The table below lists some additional medical equipment performance standards.
U.S. and European Electromagnetic Compatibility Standards Applicable to Healthcare Facilities and Electronic Medical Equipment
23
Power Quality for Healthcare Facilities
CONCLUSION
Important information has been provided here about how healthcare facilities and medical clinics view power quality problems, how such problems can be recognized by facility and medical staffs, definitions of the sources of electrical disturbances that can impact healthcare facilities and medical clinics, and how power quality challenges might be met in a complex environment where patient safety must prevail above power quality. Recognizing and correcting wiring and grounding errors and the commingling of loads are paramount in resolving power quality problems in healthcare facilities and medical clinics, and establishing partnerships between electric supply companies, facility designers, medical equipment manufacturers, and the facility and medical staffs is also critical. This approach is based on common practices employed in U.S. healthcare facilities to understand, identify, solve, and prevent
power quality problems. The information provided in the PQ TechWatch “Hardening Manufacturing Processes Against Voltage Sags” (EPRI, 200) can also be applied to the physical plant of healthcare facilities and medical clinics in efforts to harden mechanical equipment against voltage sags and momentary interruptions. Most voltage sag-sensitive components typically found in a healthcare facility or medical location cannot be placed on a power conditioner at the patient level. Diagnostic medical imaging systems are ultrasensitive to voltage disturbances, and many times these systems are not compatible with a UPS or cannot be placed on a power conditioner due to cost and space limitations in imaging suites. The cost of resolving underlying wiring and grounding errors and separating disturbance-causing loads from sensitive medical equipment is typically much less than the cost of placing an entire department or facility on conditioned power.
BIBLIOGRAPHY Capuano, Mike, Patrick Misale, and Dan Davidson, “Case Study: Patient-Coupled Device Interaction Produces Arrhythmia-Like Artifact on Electrocardiographs,” Biomedical Instrumentation & Technology, November/December 1993, pp. 475–483. Dorr, Douglas S., and Douglas C. Folts, “UPS Response to Power Disturbances,” Medical Electronics Magazine, December 1994, pp. 48–56. IEC 601-1-02, Medical Electrical Equipment, Part 1: General Requirements for Safety. 2. Collateral Standard: Electromagnetic Compatibility—
Requirements and Tests, 2nd edition (Geneva: International Electrotechnical Commission, June 1996). IEEE Standard 1602-1996, Recommended Practice for Electric Systems in Health Care Facilities—IEEE White Book (Piscataway, NJ: Institute of Electrical and Electronics Engineers, 1996). Keebler, Philip F., “Power Quality for Diagnostic Medical Imaging Systems,” EPRI, November 2006. Keebler, Philip F., “Power Quality for Healthcare,” BR-109172 (White Plains, NY: EPRI Healthcare Initiative, 1997). Keebler, Philip F., “Solving Power Quality Problems in Medical Imaging Systems,” PB-106393 (Knoxville, TN: EPRI Power Electronics Applications Center, 1996) Lamarre, Leslie, “Power Prescriptions for the Health Care Industry,” EPRI Journal, June 1994, pp. 14–21. Loznen, Steli P., “Product-Safety Requirements for Medical Electrical Equipment,” Compliance Engineering Magazine, March/April 1995, pp. 17–29. Russell, Michael J., “Cardiovascular Imaging Equipment Requires Emergency Power,” Power Quality Magazine, January–March 1992, pp. 8–19. Waterman, Craig, “Medical Facility Power Quality Problems Can Be Deadly,” Power Quality Magazine, Premier II 1990, pp. 82–90. Whitfield, John, The Electricians Guide to the 16th Edition of the IEE Wiring Regulations BS 7671 and Part P of the Building Regulations (Wendens Ambo, Essex: EPA Press, March 2005).
24
Power Quality for Healthcare Facilities