pH Electrode Technical Education : Support - Technical Education :: Sensorex

/ Sensorex.com Home / Support / Technical Education

email page

Page 1

print page

pH Electrode Technical Education What Is pH? pH is a measure of acidity or alkalinity. The amount of hydrogen ions (H+) causes a liquid to be acidic (high concentration of hydrogen ions) or alkaline (low concentration of hydrogen ions). The pH range is measured from 0 to 14. Values below 0 or above 14 are possible but rare and cannot be measured with our electrodes. The pH scale is derived from the dissociation constant of water in the following equation: H2 O - > H+ + OH- = 1 x 10- 14 ( mol/ L) 2 = Kw ( Kw is the dissociation constant of water) .

H y droge n I on Con ce n t ra t ion in M ole s/ Lit e r a t 2 5 ° C pH

H+ conc.

OH - conc.

0

1.0

0.00000000000001

1

0.1

0.0000000000001

2

0.01

0.000000000001

3

0.001

0.00000000001

4

0.0001

0.0000000001

5

0.00001

0.000000001

6

0.000001

0.00000001

7

0.0000001

0.0000001

8

0.00000001

0.0000001

9

0.000000001

0.00001

10

0.0000000001

0.0001

11

0.00000000001

0.001

http://www.sensorex.com/support/education/pH_education.html

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

12

0.000000000001

0.01

13

0.0000000000001

0.1

14

0.00000000000001

1.0

Page 2

pH of Common Materials

Why is pH an Important Measurement? It determines product quality in: Sugar Refining Pulp and Paper Mills Latex Coagulation Photo Developing It enhances product efficiency of: Flue gas scrubbing Circuit Board Etching Fermentation processes It assures product safety Chromate and

cyanide

http://www.sensorex.com/support/education/pH_education.html

destruct

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

Page 3

Potable and waste waters Food, low pH to prevent botulism

How Can pH be Measured? Colorimetric Methods Reagent addition pH paper Electrochemical Methods (pH electrodes) A pH Measuring System Consists of: 1. A pH electrode, an electrode whose output voltage changes as the pH (hydrogen ion concentration) changes 2. A reference electrode is an electrode whose voltage output stays constant 3. A pH meter, a millivolt meter with a special high impedance input circuit and circuits to change the electrode's millivolts into pH unit readouts. 4. Optionally, an automatic temperature compensator, a device which senses temperature so that the meter can correct for the effects of temperature changes.

How Does A pH Electrode Work? Special composition glass senses H+ and a millivoltage is generated (59.2 mV per pH unit at 25C.) A filling solution picks up the signal from the special pH glass A pure silver wire dipped in silver chloride passes the signal from the solution whose pH is being measured to the electrode's cable or connector Sodium Ion error for solutions > pH 12.3

How Does A Reference Electrode Work?

http://www.sensorex.com/support/education/pH_education.html

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

Page 4

A porous reference junction separates the filling solution in the electrode from the solution whose pH is to be measured. The filling solution's constant chloride ion concentration generates a millivoltage at a pure silver wire with silver chloride on it. The silver wire passes the signal from the solution being measured to the electrode's cable or connector.

http://www.sensorex.com/support/education/pH_education.html

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

Page 5

Single vs. Double Junction Reference Chemicals that cause silver to precipitate at the reference junction will contaminate and plug single junctions. These may be such compounds as sulfides, mercaptans, cyanides, Iodides, and proteins. Other

http://www.sensorex.com/support/education/pH_education.html

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

Page 6

compounds such as silver, lead, mercury, and other heavy metal compounds will react with the chloride in the gel, causing a reduction in the reference output. Selection of the proper chemistry in the lower (double) junction will prevent or at least minimize the negative effects of these reactive compounds.

How Does A Combination Electrode Work? A combination electrode consists of a pH electrode and a reference electrode built into a single body or housing. A combination electrode therefore works like the pH and reference electrodes combined!

How does a pH meter Work? A pH meter takes the input from the pH glass (high impedance mV) and the input from the reference sensor and compares these 2 values to get a resulting millivolt reading. The reading in mV is converted to pH by the following guidleine: Zero mV = pH 7 59.2mV per pH unit change mV are + for pH <7 and mV are - for pH >7

Example of pH Circuit

Note that the pH and reference input go into an operational amplifier (op amp) due to the very high resistance of the pH glass. The meter will also adjust zero and span offsets and can do automatic temperature compensation for pH error (discussed below).

Temperature Compensation When measuring pH using a pH electrode the temperature error from the electrode varies based on the Nernst Equation as 0.03pH/10C/unit of pH away from pH7. As shown in the table below, the error due to

http://www.sensorex.com/support/education/pH_education.html

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

Page 7

temperature is a function of both temperature and the pH being measured. Note that there is no error at pH7 and 25 Deg C. Temperature compensation can be achieved manually or automatically. Manual temperature compensation is usually achieved by entering the temperature of the fluid being measured into the instruments menu and then the instrument will display a "Temperature Compensated" pH reading. This means that the temperature is corrected to the value expected at 25 Deg C. Automatic temperature compensation requires input from a temperature sensor and constantly sends a compensated pH signal to the display. Automatic temperature compensation is useful for measuring pH in systems with wide variations in temperature.

pH v s. Te m pe ra t u re Error Ch a rt pH 2

pH 3

pH 4

pH 5

pH 6

pH 7

pH 8

pH 9

pH 10

pH 11

pH 12



.30

.24

.18

.12

.06

0

.06

.12

.18

.24

.30

15°

.15

.12

.09

.06

.03

0

.03

.06

.09

.12

.15

25°

0

0

0

0

0

0

0

0

0

0

0

35°

.15

.12

.09

.06

.03

0

.03

.06

.09

.12

.15

45°

.30

.24

.18

.12

.06

0

.06

.12

.18

.24

.30

55°

.45

.36

.27

.18

.09

0

.09

.18

.27

.36

.45

65°

.60

.48

.36

.24

.12

0

.12

.24

.36

.48

.60

75°

.75

.60

.45

.30

.15

0

.15

.30

.45

.60

.75

85°

.90

.72

.54

.36

.18

0

.18

.36

.54

.72

.90

Note: Values in light blue are less than .1 error and may not require temperature compensation. Values in gray are temperature and pH in which there is no error in pH from temperature.

How are pH Systems Calibrated? The electrodes are placed in buffers. Buffers are solutions of know, stable pH value The commonly used buffers have pH values of 4.01, 7.00 and 10.00 The pH values of buffers change with changes of temperature (and so does the pH of solutions being measured) First the the system's (electrode and meter together) Zero Point is adjusted Zero point is usually determined with pH 7.00 buffer Ideally, a buffer is used whose value is close to that of the material or solution being measured A "calibrate "or "standardize" adjustment potentiometer knob or push button on the pH meter is used to set the system (electrode and meter together) to read the buffer's pH value Next, the system's span is checked and/or adjusted. The electrodes are rinsed and put into a second buffer The system should read close to the pH value of the second buffer. Most meters have controls labeled "SPAN" or "SLOPE" which compensates for electrodes with spans that are too short. The frequency of calibration is at the discretion of the user.

http://www.sensorex.com/support/education/pH_education.html

11/20/2009 10:57:06 PM

pH Electrode Technical Education : Support - Technical Education :: Sensorex

Page 8

NOTE: 2 - point calibration is necessary to ensure electrode works properly since a broken electrode can give acceptable pH7 output in calibration mode.

pH Applications Wastewater neutralization Electroplating Chemical Manufacturing Circuit board etching Flue gas scrubbers Boilers and cooling towers Pulp and paper mfg. Food and Beverage Pharmaceuticals inks, paints latex Water treatment (sewage) Aquariums, aquaculture Fermentation (wine, beer, alcohol)