Disinfection

  • August 2019
  • PDF

This document was uploaded by user and they confirmed that they have the permission to share it. If you are author or own the copyright of this book, please report to us by using this DMCA report form. Report DMCA


Overview

Download & View Disinfection as PDF for free.

More details

  • Words: 561
  • Pages: 3
DISINFECTION & STERILISATION OF I.C.U. EQUIPMENTS DRAGGER VENTILATOR Unplug the machine before cleaning Use surface disinfectants For material compatibility, use disinfectants based on aldehydes/ alcohol/ quaternary ammonium compounds only. Material damage may occur if phenol based, halogen releasing, strong organic acids, oxygen-releasing compounds are used. •

Basic Machine Wiping with disinfectant For basic machine without Hose system, Aquapor without bowl & float, gas connecting hoses and temperature sensor.



Disconnected parts For ventilation hoses, watertraps, Aquapor with bowl & float, connector with Y piece, Nebuliser & their parts – Disinfect with moist heat at 930 C for 10 min in cleaning & disinfecting machine After disinfection, sterilize patient system and individual part at 1340 C Remove all water from the system If disinfecting machine not available – Disinfect with Sekuceft Rinse with clean soft water Remove all water & dry all parts well After rinsing, steam sterilize patient system and individual parts at 1340 C OR Rinse with clean water and dry Then steam sterilize at 1340 C

BEAR CUB VENTILATOR Unplug the machine before cleaning •

Ventilator Graphics Monitor (VGM) Front panel, Display shield and accessories may be wiped clean with water or isopropyl alcohol DO NOT USE – Phenols, Ammonium chloride, Quaternary compounds, Acrylate reactive solutions



Exterior of the Ventilator

Clean with an appropriate bactericidal/ germicidal agent like Isopropyl alcohol •



Ventilator tubings (except Bacterial filter) Clean with a warm detergent solution Then rinse with warm water Sterilize by autoclaving Flow Sensor Soak in enzymatic cleaner (Klenzyme/ Cidazyme) for 10 minutes. Periodic up & down agitation may assist in removal of protein particles from the device Remove & then rinse with distilled water Dry the device with a soft cloth Flow sensor may be sterilized using Ethylene oxide sterilization process.



Water traps Clean with a warm detergent solution & then rinse thoroughly with warm water.



Connectors, Adapters, Corrugated tubes May be sterilized by Autoclaving

CRITICARE MULTICHANNEL MONITOR Turn off and unplug the Monitor •

Exterior surface (except Display screen) Wipe clean with Isopropyl alcohol



Display Screen Wipe with a clean, soft, lint-free cloth sprayed with a common glass cleaner Do not spray glass cleaner directly on the screen



SpO2 Sensor Wipe clean with alcohol Disinfect by placing the paddles & cable (not connector) in 2% glutaraldehyde solution



B.P.Cuffs Clean by wiping it with a damp cloth or sponge Disinfect by wiping with 70% alcohol or a mild bleach solution Cloth cuff can be sterilized by Ethylene oxide or by Autoclaving

OXYPLETH PULSE OXIMETER Turn off and unplug the Monitor •

Monitor Can be cleaned or disinfected with solutions like 70% Isopropyl alcohol, 2% Glutaraldehyde or 10% bleach Do not IMMERSE or STERILISE the monitor



SpO2 Sensor Clean and disinfect with same solutions Wipe with a water dampened clean cloth Dry well before use

EMCO PULSE OXIMETER Turn off and unplug the Monitor •

Monitor Wipe the surface of monitor with Isopropyl alcohol/ Glutaraldehyde solution Then wipe with soft water moistened cloth



SpO2 Sensor Clean and disinfect with same solutions Wipe with a water dampened clean cloth Dry well before use

INFUSION PUMP Disconnect cable, Clean with a soft cloth with disinfectant like Ethyl alcohol or 3% Hydrogen peroxide, Allow to dry for 1 min. DO NOT OVERTURN the pump as the solution can enter inside the machine DO NOT AUTOCLAVE