Thermocouple Thermometry
มว NIMT
Theory
The principle of The Thermocouple (Thermo+couple) is based on the Seebeck effect that Thomas Seebeck observed this effect in 1821 . Seebeck effect: A temperature gradient or A temperature difference between two points in a metal results in voltage difference between these. The thermoelectric voltage developed (dV) per unit temperature difference(dT) in a conductor is call Seebeck coefficient. S = dV/dT
The emf between the thermocouple wire is given by T
T
E = ∫ ( S − S )dt = ∫ S dt AB
A
T0
B
V
Metal B, SB
AB
T0
Where SAB = The Seebeck coefficient for the thermocouple pair A-B
Metal A, SA
Thermocouple Voltage,emf
Reference junction/ Cold junction
Measuring junction/ Hot junction
Thermocouple Circuit
Inhomogeneity Error In Thermocouple “The variation of the thermocouple output along the used length.” •TC As Use
1
TC emf,E 2
3
4
E=S12(T1-T2)=small E=S23(T2-T3)=largeE=S34(T3-T4)=small
Un-Aged wire
Transition wire
Aged wire
1
2 3
4
•TC in calibration furnace Causes of inhomogeneity in thermocouple wire: -Cold work (bending, twisting,etc.) -Chemical (environment and in-use) -Heat treatment (above 250oC for base-metal)
- Change in immersionÆMeaningless results.
Typical inhomogeneity for new Thermocouple; Pt thermocouple S R B ≈ ± 0.02%of E Base-metal thermocouple ≈ ± 0.1%of E
(test by Thermal Scan)
Thermocouple Types Standard Type (Letter-designed types) Noble (Rare) Metal Type B, R ,S
TC In Use Non Standard Type (Special designed)
Base Metal Type E,J,K,N,T
Pt/Pd, Au/Pt, W3-Re/W25-Re(Type C)
General Range use : •Type R or Type S < 1300 oC , Type B > 1300 oC. •Type T,E and K At low temp / Room temp. •Type K (< 200 oC), Type N (300 – 600) oC . • Type K, N > 600 oC.
NIMT Thermocouple Calibration System - Calibration MethodITS-90
มว N IM T
SPRT / HSPRT (-38.8344 oC to 961.78oC )
Nobel Metal TC
NIMT Fixed Point Cell Triple point of Mercury,Triple point of Water,Melting point of Galium, Freezing point of Tin,Freezing point of Zinc,Freezing point of Aluminum,Freezing point of Silver (-38.8344 oC to 961.78oC )
NPL(UK), NMIA(Aus)
Reference Standard Thermocouple (Pt/Pd ,Au/Pt,Type B,Type R ,Type S) (0 oC to 1100 oC )
1.Comparison with Fixed Point Range (0oC-962oC) Uncertainty 0.4 oC – 0.5 oC
Calibration Laboratory
SPRT
IPRT
Thermocouple (All Type)
DTM
LIG
USERS
NIMT Traceability Chart(2006)
2.Comparison with SPRT / Ref. TC Range (0oC-1100oC) Uncertainty 0.5 oC – 1.5 oC