Standard Operating Procedure (vario MACRO CHNS Analyzer) Sample Preparation •

Aggressive chemicals, acids, alkaline solutions, solvents, explosives or materials that can create explosive gas mixtures are explicitly forbidden.



Brief sample chemistry must be known, which the client will provide.



Samples (only solid sample matrix) should be well homogenized, ground and dried.



Sampling area and sample preparation tools (tweezers, spatula, etc.) must be contaminant free and swept with acetone.



Use tin foils or tin boats for sample packing.



Counterbalance tin boat or tin foil so that the balance scale shows “0.0000”.



Weigh sample (1 mg to 1g) and note down the required information in “Sample Weighing Log Book”.



Add sample additive (WO3 powder) to the sample in ratio of 1:1.



Close tin tightly above the sample with the help of tweezers and micro spatula and shape sample into a cube and press the air out.



Make sure that no bits of tin foil protrude from the fold.



Place prepared samples in appropriate sample trays.

Operation of the vario MACRO



Note down the temperature and humidity of the working area. Ambient temperature should be between 15 – 350C while relative humidity should be 30 – 80%.



Make sure that during operation of the equipment there will be no electric power failure. Generator and UPS must be ready in case of any power failure.



Only well-trained and authorized persons are allowed to work on CHNS elemental analyzer.



Check and note pressure from two stage regulators of oxygen and helium gas cylinders.



Switch on PC with screen and printer, and wait for booting procedure to be ended.



Remove carousel.

1



Switch on the equipment by pressing green button inside the right side front door.



The instrument will carry out a reference run of ball valve and carousel automatically.



Replace the carousel.



Load software by clicking “varioEL” icon.



Turn on operation gases o

For helium gas cylinder, set pre-pressure approximately 2.0 bar so that inlet pressure gauge shows 1.25 bar.

o

For oxygen gas cylinder, set pressure 2.5 bar.



Open file “template” and save as with new name.



Confirm that analysis mode is CHNS in “Mode” option.



Set CHNS parameters by clicking “Options” → “Parameters” (See page 5 – 12 of operation manual).



Observe in the status field, the furnace temperatures will rise up. Approximately 45 – 60 minutes are required for heaters to reach the set temperatures.



Perform leakage test by clicking “Options” → “Miscellaneous” → “Rough Leak Check”.



Position the pointer at the end sample number by clicking “Edit” option.



Input data by clicking “Edit” → “Input” for blank, run-in and sample in following sequence:

Sample Name

Sample Description

blank blank blank blank blank blank “run-in” “run-in” “run-in” “run-in” “run-in” “run-in” “run-in” “run-in” “run-in” “run-in”

Without anything Without anything Without anything Without anything Tin boat Tin + additive Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard Tin + additive + standard

Sample Weight (mg) 1 1 1 1 1 1 0.5 1.0 2.0 4.0 6.0 8.0 10.0 20 30 40

2

Method

Amount

glut 2 blank blind glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2 glut 2

6 – 12 2 1 4 2 2 1 1 1 1 1 1 1 1 1 1

Sample

Tin + additive + sample

blank

Without anything

According to sample nature 1

Glut 2

2

glut 2

6



Save the file.



Place the prepared samples in carousel according to sequence and note the carousel number in “Sample Weighing Log Book”.



Start analysis by clicking “System” → “Auto”.



Observe peak formation by clicking “Graphic View” button.



After completion of analysis compute daily factor by clicking “Math” → “Factor”.



Print out results by tagging the required samples.



Save file.

Shutdown Procedure



For short shutdowns (during analysis or breaks for one night or 2 –3 days, turn off He supply. O2 supply should remain open for flushing purpose. Click on “Sleep/Wake Up” option and select the appropriate shutdown conditions (see page 6-21 of operation manual).



For longer shutdowns (lasting several days or weeks), cool down all heaters and wait until the temperatures are lower than 1000C (approximately 2 hours). For this purpose, set nominal temperature to 00C by clicking “Parameters”. Exit software program via System – Offline option. Turn off gas supplies, vario MACRO equipment and peripheral instruments.