Guidance for parameter interpretation General analysing of results – verification of validity Patient data plausible Pressure curve quality sufficient, no extra systoles or arrhythmias
Injected volume adequate
CVP value plausible
Pressure values & heart rate plausible
Thermodilution values reproducible and plausible ΔT sufficiently high (good signal) Tinj sufficiently low (cold injectate)
Page 3
Optimal thermodilution curves
PULSION is part of Maquet Getinge Group
Guidance for parameter interpretation General analysing of results – the parameters
-
Blood flow (CI) Preload volume (GEDI) Volume responsiveness (SVV), if applicable Afterload (SVRI) Contractility (CFI) Pulmonary Oedema (ELWI) Origin of pulmonary oedema (PVPI)
Trend important!
Page 4
Blood flow (Volume responsiveness)
Afterload
Contractility
Preload Pulmonary oedema
PULSION is part of Maquet Getinge Group
Guidance for parameter interpretation Example 1 Normal ranges CI
3 – 5 l/min/m2
GEDI SVV
680 – 800 ml/m2 0 – 10 %
SVRI
1700 – 2400 dyn·sec·cm-5·m2
CFI CPI
4.5 – 6.5 1/min 0.5 – 0.7 W/m2
ELWI PVPI
3 – 7 ml/kg 1 – 3 cardiogenic oedema > 3 permeability oedema
Conclusion: Low blood flow due to low preload and high vascular resistance: hypovolaemia Therapeutic recommendations: - Fluid loading - Vasodilator - Repeated thermodilution measurement after therapy
Page 5
PULSION is part of Maquet Getinge Group
Guidance for parameter interpretation Example 2 Normal ranges CI
3 – 5 l/min/m2
GEDI SVV
680 – 800 ml/m2 0 – 10 %
SVRI
1700 – 2400 dyn·sec·cm-5·m2
CFI CPI
4.5 – 6.5 1/min 0.5 – 0.7 W/m2
ELWI PVPI
3 – 7 ml/kg 1 – 3 cardiogenic oedema > 3 permeability oedema
Conclusion: High blood flow with low vascular resistance and high lung water due to high permeability: Hyperdynamic circulation, permeability pulmonary oedema (sepsis, ARDS)
Therapeutic recommendations: - Supportive therapy: ventilator settings, lung support (CARDIOHELP), kinetic therapy - Vasoconstrictor - Repeated thermodilution measurement after therapy Page 6
PULSION is part of Maquet Getinge Group
Guidance for parameter interpretation Example 3 Normal ranges CI
3 – 5 l/min/m2
GEDI SVV
680 – 800 ml/m2 0 – 10 %
SVRI
1700 – 2400 dyn·sec·cm-5·m2
CFI CPI
4.5 – 6.5 1/min 0.5 – 0.7 W/m2
ELWI PVPI
3 – 7 ml/kg 1 – 3 cardiogenic oedema > 3 permeability oedema
Conclusion: Low blood flow with high vascular resistance, severe pulmonary oedema due to high preload (fluid overload) and low contractility: cardiac failure, cardiogenic pulmonary oedema (cardiogenic shock) Therapeutic recommendations: - Fluid withdrawal - Inotropic drugs - Repeated thermodilution measurement after therapy
Page 7
PULSION is part of Maquet Getinge Group
Be aware of pros and cons! Possible consequences of therapeutic decisions
Page 8