How To Use 6211

  • May 2020
  • PDF

This document was uploaded by user and they confirmed that they have the permission to share it. If you are author or own the copyright of this book, please report to us by using this DMCA report form. Report DMCA


Overview

Download & View How To Use 6211 as PDF for free.

More details

  • Words: 641
  • Pages: 23
Connect NI USB-6211 • If this is the first time the DAQ has been plugged into the computer, you will need to follow the driver install procedure which is similar to any other USB device.

• Next, Open a Blank VI.

Select DAQ Assistant

Configure the DAQ Assistant

•This sub VI allows several operations to be configured. To read data into the computer, select Analog Input.

Configure the DAQ Assistant

Physical Channels You can select one channel or multiple channels here. This device allows for up to 16 single ended channels of input.

Select ai0 and click Finish.

Voltage Input Setup Signal Input Range defines the span of voltage the DAQ will measure. Narrowing the range allowing better resolution of measurements.

Voltage Input Setup Terminal Configuration can be made using Differential or Single Ended. Differential gives better common mode rejection and cleaner signals if the connections are properly wired.

Task Timing Several options exist for Acquisition Mode from 1 sample to continuous. We will look at the default selection, N Samples (Data points are putting them together like a moving picture show.

Task Timing • Samples to Read defines how many data points the DAQ will return when this VI is called. • Rate tells the DAQ how fast to acquire the data. • Divide the sample rate by the number of samples to determine the sample time. If we use 1000 samples at 1000Hz, that will be 1 second of data.

Getting Data out of the VI • Getting out of the VI • Create Graph Indicator

Data! • Click Run, wait for the second of data to be collected and… • But What is it? Shouldn’t it be zero? !!!!! Since the DAQ is not connected to anything, we don’t know what will result from the measurement. But we can see it is picking up some noise. If it were a flat line or just one point, that would indicate some other problem.

Wiring Use a cable with a BNC connector on one end and bare wire on the other end to connect the DAQ to the function Generator.

Analog In Terminals Connect the bare wires to the AI0 + and – terminals on the DAQ

Complete Circuit

Run the VI

• Now set the function generator on 10Hz

• Run the VI

Aliasing-Varying Sample Rate • Aliasing is when the DAQ is giving a result that is not what is really happing. • Filters at the Nyquist frequency are the only way to prevent aliasing. • Now set the function generator to 2 Vp-p, 100 Hz and Sine Wave. • Set the DAQ to 1000Hz sample rate, 50 Samples • What is the sample time? __(50mSec )_ • How many wave cycles should we see? ___(5)___

Result

Aliasing-Varying Sample Rate • Now try – 500Hz and 25 samples – 300Hz and 15 samples – 210Hz and 10 samples

Aliasing-Varying Sample Rate • Now sample rates below Nyquist Frequency

Sampling Below Nyquist • When the sample rate is less than half the frequency under measurement, aliasing or misinformation will result. • The apparent frequency will not be the actual frequency. • A perfect waveform can be observed that is likely some multiple of the actual frequency. • Hardware filtering is the only way to avoid aliasing when the sample rate cannot be set high enough to include all frequency content being measured.

Saving Data

• In order to keep the program from saving data every cycle of the while loop. • Set the action of the button to “Latch When Pressed”. This will allow the program to turn off the button once the desired action has been performed once. This is a kind of one-shot.

Examining Data File • Run the program and save some data. • Open the file with Excel and look at the data it recorded.

Related Documents

How To Use 6211
May 2020 0
6211
April 2020 1
How To Use Use Cases
May 2020 31
How To Use Netaxis
May 2020 16
How To Use Simulink
June 2020 19